The Triage® TOX Drug Screen is a fluorescence immunoassay intended to be used with the Triage MeterPlus for the point-of-care qualitative determination of the presence of major metabolites above the threshold concentrations of up to 8 distinct drug classes, including assays for methamphetamines, acetaminophen/paracetamol, barbiturates, benzodiazepines, cocaine, opiates, phencyclidine, THC and tricyclic antidepressants in urine. The acetaminophen/paracetamol assay will yield positive results when acetaminophen/paracetamol is ingested at or above therapeutic doses. This test provides only a preliminary test result. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectroscopy (GC/MS) is the common confirmatory method. A quantitative serum acetaminophen/paracetamol measurement is the common confirmatory method for preliminary positive acetaminophen/paracetamol results.
Device Story
Triage TOX Drug Screen is a single-use competitive fluorescence immunoassay for qualitative drug detection in urine. Device uses murine monoclonal antibodies and fluorescently labeled drug metabolites. Urine sample added to device; moves via capillary action through filter to reaction chamber; reacts with fluorescent antibody conjugates. Presence of drug prevents binding of fluorescent conjugates to solid phase; unbound conjugate washed away. Fluorescence intensity on detection zone inversely proportional to drug concentration. Device inserted into Triage MeterPlus for automated reading. Used in point-of-care settings by healthcare professionals. Provides preliminary results; requires confirmatory testing (GC/MS) for clinical decision-making. Benefits include rapid, on-site screening for drug presence.
Clinical Evidence
Bench testing only. Method comparison study performed on 102 urine specimens comparing Triage TOX Drug Screen results against GC/MS. Overall agreement was 97.1%. Three discordant samples were positive on the Triage device and within 10% below the cutoff by GC/MS.
Technological Characteristics
Fluorescence immunoassay; point-of-care diagnostic test; used with Triage MeterPlus; qualitative detection of drug metabolites in urine; threshold-based detection.
Indications for Use
Indicated for point-of-care qualitative detection of major drug metabolites in urine for 8 drug classes (methamphetamines, acetaminophen/paracetamol, barbiturates, benzodiazepines, cocaine, opiates, phencyclidine, THC, tricyclic antidepressants) at specified threshold concentrations. Intended for prescription use.
Regulatory Classification
Identification
An acetaminophen test system is a device intended to measure acetaminophen, an analgestic and fever reducing drug, in serum. Measurements obtained by this device are used in the diagnosis and treatment of acetaminophen overdose.
Predicate Devices
Triage® TOX Drug Screen (k012745)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k043242
B. Purpose for Submission:
This is a new device for an Acetaminophen assay configured with other analytes which have been previously cleared (k012745) (see device description section).
C. Measurand:
Acetaminophen/paracetamol (APAP) Amphetamines (AMP), barbiturates (BAR), benzodiazepines (BZO), cocaine (COC), methamphetamines (mAMP), opiates (OPI), phencyclidine (PCP), THC and tricyclic antidepressants (TCA) in urine.
D. Type of Test:
Qualitative, Fluorescence Immunoassay
E. Applicant:
Biosite Incorporated
F. Proprietary and Established Names:
Triage® TOX Drug Screen (available in different configurations)
G. Regulatory Information:
1. Regulation section:
862.3030 - Acetaminophen Test System
862.3100 - Amphetamine Test System
862.3150 - Barbiturate Test System
862.3170 - Benzodiazepine Test System
862.3250 - Cocaine and Cocaine Metabolite Test System,
862.3610 - Methamphetamine Test System
862.3650 - Opiate Test System
862.3100 - Amphetamine test system (Phencyclidine)
862.3870 - Cannabinoid Test System and
862.3910 - Tricyclic Antidepressant Drug Test System
{1}
2. Classification:
Class II
3. Product code:
LDP, DKZ, DIS, JXM, JXO, LAF, DJG, LCM, LDJ and LFG; respectively
4. Panel:
91, Toxicology
## H. Intended Use:
1. Intended use(s):
See indications for use.
**Indication(s) for use:**
The Triage Tox Drug Screen is a Fluorescence immunoassay intended to be used with the Triage MeterPlus for the point-of-care qualitative determination of the presence of major metabolites above the threshold concentrations of up to 8 distinct drug classes, including assays for acetaminophen/paracetamol, amphetamines, barbiturates, benzodiazepines, cocaine, methamphetamines, opiates, phencyclidine, THC and tricyclic antidepressants in urine. The acetaminophen/paracetamol assay will yield positive results when acetaminophen/paracetamol is ingested at or above therapeutic doses.
The threshold concentrations are provided below:
| Acetaminophen/Paracetamol | APAP | 5 μg/mL |
| --- | --- | --- |
| Amphetamines | AMP | 1000 ng/mL |
| Methamphetamines | mAMP | 1000 ng/mL |
| Barbiturates | BAR | 300 ng/mL |
| Benzodiazepines | BZO | 300 ng/mL |
| Cocaine | COC | 300 ng/mL |
| Opiates | OPI | 300 ng/mL |
| Phencyclidine | PCP | 25 ng/mL |
| THC | THC | 50 ng/mL |
| Tricyclic Antidepressants | TCA | 1000 ng/mL |
{2}
2. Special conditions for use statement(s):
This test provides only preliminary test results. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectroscopy (GC/MS) is the preferred confirmatory method.
A quantitative serum acetaminophen/paracetamol measurement is the common confirmatory method for preliminary positive acetaminophen/paracetamol results.
For Prescription and Point-of-Care use. The Point-of-Care classification was address in a previously cleared submission (k012745)
4. Special instrument requirements:
Triage® MeterPlus; k973547
I. Device Description:
The Triage TOX Drug Screen is a competitive fluorescence immunoassay used for the qualitative determination of major metabolites of acetaminophen/paracetamol (APAP), amphetamines (AMP), methamphetamines (mAMP), barbiturates (BAR), benzodiazepines (BZO), cocaine (COC), opiates (OPI), phencyclidine (PCP), THC and tricyclic antidepressants (TCA). All analytes were previously cleared (k012745) except for the acetaminophen analyte. It is a single use test device and is used in conjunction with the Triage ® MeterPlus. The device contains murine monoclonal antibodies and drug metabolites labeled with a fluorescent dye or immobilized on the solid phase and stabilizers. The testing device is inserted into and read by the Triage MeterPlus. Threshold concentrations are used to separate a negative result from a presumptive positive result.
No human source materials are used during the manufacturing of this product.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Triage® TOX Drug Screen
2. Predicate 510(k) number(s):
k012745
{3}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | Qualitative determination of major metabolites in human urine. | Qualitative determination of major metabolites in human urine. |
| Matrix | Human Urine | HumanUrine |
| Test Principle | Fluorescence immunoassay | Fluorescence immunoassay |
| Differences | | |
| Item | Device | Predicate |
| Anaylets | Acetaminophen, Amphetamines, Methamphetamines, Barbiturates, Benzodiazepines, Cocaine, Opiates, Phencyclidine, THC and Tricyclic Antidepressants | Amphetamines, Methamphetamines, Barbiturates, Benzodiazepines, Cocaine, Opiates, Phencyclidine, THC and Tricyclic Antidepressants |
# K. Standard/Guidance Document Referenced (if applicable):
None stated.
# L. Test Principle:
The Triage Tox Drug Screen is a competitive fluorescence immunoassay used for the qualitative determination of parent compound and major metabolites of drugs of abuse in urine specimens. A urine sample is added to the device which then passes through a filter. The sample moves by capillary action into a reaction chamber and is allowed to react with murine monoclonal fluorescent antibody conjugates or mix fluorescent drug conjugates. After an incubation period, the reaction mixture flows through the device detection lane. The presence of drug or drug metabolite in the urine specimen prevents binding of the fluorescent conjugates to the solid phase on the detection zone. Unbound fluorescent conjugate is washed from the detection lane by excess urine. The concentration of the drug metabolite in the urine specimen is inversely related to the fluorescence bound to the detection zone. Threshold concentrations are used to separate a negative result from a presumptive positive result.
{4}
M. Performance Characteristics (if/when applicable):
See k012745 for amphetamines, barbiturates, benzodiazepines, cocaine, methamphetamines, opiates, phencyclidine, THC and tricyclic antidepressants. Sponsor states that performance studies have been conducted for the addition of acetaminophen through its internal Design Control process, the performance characteristics are exactly the same and data are on file at the Biosite.
1. Analytical performance:
a. Precision/Reproducibility:
The threshold for acetaminophen of 5 ng/ml was challenged by testing specimens containing the drug or drug metabolite spiked into drug-free urine at concentrations within 25% and 50% of the threshold. Each specimen was tested using the Triage® Tox Drug Screen. The data paralleled the expected agreement based on the coefficient of variation of the assay.
b. Linearity/assay reportable range:
Not applicable. The assay is intended for qualitative use.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Users are instructed to run external controls with each new lot or shipment of test materials, or every 30 days, and as otherwise required by federal, state or local guidelines. Commercial control materials are required but are not specifically identified in the labeling.
Each device has internal process controls which run with every sample. The controls monitor various procedural and functional parameters such as sufficient sample added to the test device and that the unbound fluorescent label has been washed sufficiently from the detection zone. During testing if the results are within the limits the results are reported. If the results are not within the limits the results are not reported and an error message will be displayed.
d. Detection limit:
Cutoff validation and sensitivity were evaluated. Four contrived samples were prepared by adding acetaminophen to drug-free urine at concentrations within 25% and 50% of the cutoff concentration. Samples were measured 10 times using the Triage TOX Drug Screen.
Table IV.1: APAP Cutoff Validation
| | 0.5x | 0.75x | 1.25x | 1.5x |
| --- | --- | --- | --- | --- |
| POS | 0 | 0 | 9 | 10 |
| NEG | 10 | 10 | 1 | 0 |
| Concordance | 100% | 100% | 90% | 100% |
{5}
# e. Analytical specificity:
Substances that may cause interference were added to samples containing acetaminophen within $25\%$ of the cutoff concentration. Cross-reactivity substances were added to drug-free urine. Samples were measured in duplicate. The substances did not interfere or have a cross-reaction with the acetaminophen assay at the concentrations listed below:
Table IV.2.1: Interfering Substances
| Substance | Concentration | Substance | Concentration |
| --- | --- | --- | --- |
| Acetone | 5 mg/mL | Hemoglobin | 1.2 mg/mL |
| Acetylsalicylic Acid | 1 mg/mL | Human Serum Albumin | 5 mg/mL |
| Ascorbic Acid | 15 mg/ml | Ibuprofen | 1 mg/mL |
| Bilirubin | 2.5μg/mL | Ketamine | 25 μg/mL |
| Caffeine | 0.125 mg/mL | Oxalic Acid | 10 mg/mL |
| Creatinine | 2.5 mg/mL | Riboflavin | 75μg/mL |
| Dextrose | 20 mg/mL | Scopalamine | 62.5 μg/mL |
| Ethanol | 5 mg mL | Sodium Chloride | 30 mg/mL |
| Fluoxetine | 0.5 mg/mL | Urea | 30 mg/mL |
| Gamma Globulin | 5 mg/mL | | |
Table IV.2.2: Cross-Reacting Substances
| Substance | Concentration | Substance | Concentration |
| --- | --- | --- | --- |
| Acetopromazine | 100 μg/mL | Methoxyphenamine | 100 μg/mL |
| Benzphentamine | 100 μg/mL | Methylphenidate | 100 μg/mL |
| Benztropine Methane | 100 μg/mL | Naloxone | 80 μg/mL |
| Bupropin | 100 μg/mL | Naproxen | 100 μg/mL |
| Butyrophenine | 100 μg/mL | Norpsuedoephedrine | 100 μg/mL |
| Cimetidine | 100 μg/mL | Phenethylamine | 100 μg/mL |
| Clonidine | 100 μg/mL | Phenmetrazine | 100 μg/mL |
| Cotinine | 100 μg/mL | Phenylephrine | 100 μg/mL |
| Dextromethorphan | 100 μg/mL | Phenylhydantoin, d/l-5-(p-hydroxyphenyl)-5- | 100 μg/mL |
| Dextrorphan | 100 μg/mL | Phenylpropanolamine | 100 μg/mL |
| Diphenhydramine | 100 μg/mL | Promrthazine | 100 μg/mL |
| Dopamine | 100 μg/mL | Propanolol, d/l | 100 μg/mL |
| Epinephrine, 1- | 100 μg/mL | Propoxyphene | 100 μg/mL |
| Fenfluramine | 20 μg/mL | Pseudoephedrine, d- | 100 μg/mL |
| Glutethimide | 100 μg/mL | Quinacrine | 100 μg/mL |
| KetorolacTromethane | 100 μg/mL | Ranitidine | 100 μg/mL |
| Levorphanol | 50 μg/mL | Thioridazine | 100 μg/mL |
| Meperidine | 100 μg/mL | Tramadol | 100 μg/mL |
| Mesoridazine | 100 μg/mL | Tyamine | 60 μg/mL |
| Methadone d/l | 100 μg/mL | Tranylcypromine | 100 μg/mL |
| Methaqualone | 100 μg/mL | Zolpidem | 100 μg/mL |
{6}
Acetaminophen was added at 75% and 125% of the cutoff concentration to urine samples with specific gravities ranging from 1.006 – 1.026 and to urine samples with the pH ranging from 4.5 – 7.5. Six (6) replicates of each sample were measured using the Triage TOX Drug Screen. There is no effect of specimen specific gravity or pH on the acetaminophen assay results.
f. Assay cut-off:
Characterization of how the device performs analytically around the claimed cutoff concentration appears in precision and sensitivity sections, above.
The cutoff concentration of this qualitative assay was arbitrarily chosen by the sponsor. There is no known clinical significance regarding the cutoff concentration. It indicates only that the analyte was or was not present at that level.
2. Comparison studies:
See k012745 for amphetamines, barbiturates, benzodiazepines, cocaine, methamphetamines, opiates, phencyclidine, THC and tricyclic antidepressants.
a. Method comparison with predicate device:
Because the candidate was compared to a reference method, GC/MS. It was not compared to a predicate device.
102 specimens with known acetaminophen concentration by GC/MS were measured using the Triage TOX Drug Screen. 8 specimens were within 25% below the cutoff concentration and 12 specimens were with 25% above the cutoff concentration. The cutoff concentration for both methods was 5 µg/mL.
The overall agreement of results from both methods is 97.1% (95% confidence interval 93.8%-100%).
b. Matrix comparison:
Not applicable. The assay is intended for only one sample matrix.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
{7}
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Not applicable.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.