← Product Code [FRN](/productcode/FRN) · K043204

# SGARLATO LABS PAINFREE PUMP AND PAIN CONTROL INFUSION PUMP (K043204)

_Sgarlato Laboratories, Inc. · FRN · Mar 30, 2005 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K043204

## Device Facts

- **Applicant:** Sgarlato Laboratories, Inc.
- **Product Code:** [FRN](/productcode/FRN.md)
- **Decision Date:** Mar 30, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Sgarlato PainFree Pump is a single use, general purpose infusion device intended to provide continuous infusion of medication (such as local anesthetics or corticosteroids) directly into a target site for the purpose of pain relief. Infusion may be carried out via percutaneous, subcutaneous, and epidural routes. The PainFree Pump may be used to provide continuous infusion of a local anesthetic directly into an intraoperative site for postoperative management of pain following surgery. It is suitable for use as an ambulatory device. It is not intended for vascular drug delivery, hyperalimentation, or the delivery of blood or blood products.

## Device Story

Mechanically-driven, single-use ambulatory infusion pump; provides continuous delivery of liquid medication (local anesthetics/corticosteroids) to target sites. Device utilizes spring-powered syringe plunger to pressurize reservoir; micro-bore flow restrictors meter medication flow to infusion site. Operated by clinicians; used in surgical or clinical settings for postoperative pain management. Output is continuous infusion via catheter; allows for y-adapter use to split volume to multiple sites. Benefits patient by providing extended, portable pain relief without electrical power. Does not deliver blood, lipids, or TPN.

## Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence established through mechanical comparison, material compliance (ISO 10993-1), and comparison of flow rates, reservoir capacities, and intended infusion sites against predicate devices.

## Technological Characteristics

Mechanical, non-electric, spring-powered infusion pump. Materials conform to ISO 10993-1. Flow control via micro-bore tubing and 5-micron medication filter. Single-use, sterile. Dimensions/form factor support ambulatory use. No software or electronic components.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Predicate Devices

- Paragon Infusion Kit ([K984146](/device/K984146.md))
- Sgarlato Pain Control Infusion Pump ([K990101](/device/K990101.md))

## Reference Devices

- Sgarlato PCIP ([K896422](/device/K896422.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAR 3 0 2005

## SECTION E - 510(k) Summary for the Sgarlato PainFree pump

#### General

The Sgarlato PainFree pump is mechanically the same device currently legally marketed by Sgarlato Labs for continuous infusion of liquid medication into an intraoperative site for postoperative pain management. Similar mechanicallydriven, infusion pumps with the same flow and reservoir capacity include percutaneous, subcutaneous, intramuscular and epidural infusion sites as alternative routes for administration of medication. Because the Sgarlato PainFree pump offers equivalent flow rates, similar reservoir volumes, and similar materials in its construction, Sqarlato laboratories believes that its use with alternative administration routes should be considered substantially equivalent to currently cleared predicate devices. Predicate Devices cited are the Paragon Infusion Kit K984146 and the Sgarlato Pain Control Infusion Pump K990101.

### Similarities

This infusion pump is one of a multiplicity of similar general purpose devices marketed to provide continuous infusion of moderate amounts of liquid medication (100-200 ml) over an extended period of time without use of electrical power. These mechanically-driven, single-use or reusable infusion pumps are therefore suitable for ambulatory use and can provide medication infusion via catheters for extended periods ranging from a few hours to several days.

The Predicate Devices, the I-Flow Paragon Infusion Kit (K984146) and the Sgarlato Pain Control Infusion Pump (K990101) offer similar flow rates and reservoir capacities. They both are mechanical devices that use spring energy to provide pressurized medication. They use micro-bore flow restrictors to meter the flow from a pressure vessel to the desired at the infusion site. They both include epidural-style medication catheters for delivery of medication to the site and offer the ability to easily exchange flow restrictors in a given setting. They offer y adapters to split the reservoir volume to multiple sites in a given surgical setting. They provide convenience kits with a filling syringe, needle introducer for the catheter, and various accessories to affix the catheter to the patient, label and carry the pump, etc. Neither device is intended for use in delivering blood, blood product, lipids, or for other intravenous or TPN use.

### Differences

The primary difference between the original Sqarlato Pain Control Infusion Pump K990101) and the Sgarlato PainFree pump is expansion of the sites indicated for infusion. The original device listed percutaneous infusion via a catheter to an intraoperative site. The extended indication adds subcutaneous, and epidural sites.

{1}------------------------------------------------

The PainFree pump is intended to infuse liquid medications such as corticosteroids or other anti-inflammatory compounds of the same approximate density of local anesthetics. Compounds such as blood or lipids that may clog the medication filter included in the set are not intended to be used. The original the modion Control Infusion pump was primarily used for infusing local anesthetics.

The primary difference between the I-Flow pump and the PainFree pump is the method of pressurizing the reservoir. The I-Flow device uses a spring/scissors mechanism to apply pressure to a pliable drug bag within a hard shell. The PainFree pump uses a syringe-like reservoir that compresses the plunger against a self-contained spring to apply pressure to the liquid volume in the syringe. The Sgarlato device is distributed as a single-use sterile device. The I-Flow device may be reused and/or sterilized by the user.

#### Conclusion

The PainFree pump is substantially equivalent mechanically to its predecessor Pain Control Infusion Pump from Sgarlato. The PainFree pump is substantially equivalent in terms of identified infusion sites to the I-Flow Paragon device. See Table 1 for more specific information.

{2}------------------------------------------------

|                             | Paragon Infusion Kit<br>(K984146)                                                                                                                                     | Sgarlato PCIP<br>(K896422)                                                                                                                                       | Sgarlato PainFree<br>Pump (review item)                                                                                                                                         |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item                        |                                                                                                                                                                       |                                                                                                                                                                  |                                                                                                                                                                                 |
| Intended Use                | Continuous infusion of<br>medication into an<br>intraoperative site for<br>general surgery and<br>pain management.<br>The pump itself is for<br>general infusion use. | Continuous infusion of<br>local anesthetic into a<br>surgical site for<br>postoperative pain<br>management                                                       | Continuous infusion of<br>medication into a<br>target site for pain<br>management                                                                                               |
| Routes of<br>Administration | Percutaneous,<br>subcutaneous,<br>intramuscular, and<br>epidural                                                                                                      | Percutaneous                                                                                                                                                     | Percutaneous,<br>subcutaneous, and<br>epidural                                                                                                                                  |
| Contraindications           | Not for intravenous<br>delivery. Not intended<br>for blood, blood<br>products, lipids, or fat<br>emulsions                                                            | Not for epidural,<br>subcutaneous, or<br>vascular drug<br>delivery. Not for<br>delivery of blood,<br>blood products, or<br>TPN                                   | Not for intravenous<br>delivery. Not intended<br>for blood, blood<br>products, lipids, or fat<br>emulsions                                                                      |
| Reuse                       | Re-usable pump,<br>single use<br>administration sets                                                                                                                  | Single Use                                                                                                                                                       | Single Use                                                                                                                                                                      |
| Fill Volumes                | 100 ml                                                                                                                                                                | 100 or 200 ml                                                                                                                                                    | 100 or 200 ml                                                                                                                                                                   |
| Flow Rates                  | 0.5,1,2,4, or 10 ml/hr                                                                                                                                                | 0.5,1,or 2 ml/hr                                                                                                                                                 | 1,2, or 4 ml/hr                                                                                                                                                                 |
| Flow control                | Micro-bore tubing with<br>medication filter                                                                                                                           | Micro-bore tubing with<br>5 micron medication<br>filter                                                                                                          | Micro-bore tubing with<br>5 micron medication<br>filter                                                                                                                         |
| Pump Mechanism              | Spring-scissors plate<br>compressing PVC bag                                                                                                                          | Spring-powered<br>syringe plunger                                                                                                                                | Spring-powered<br>syringe plunger                                                                                                                                               |
| Power                       | Compressed spring                                                                                                                                                     | Compressed spring                                                                                                                                                | Compressed spring                                                                                                                                                               |
| Components and<br>Materials | Conform with ISO<br>10993 part 1 for fluid<br>path                                                                                                                    | Conform with ISO<br>10993 part 1 for fluid<br>path                                                                                                               | Conform with ISO<br>10993 part 1 for fluid<br>path                                                                                                                              |
| Safety/Alarms               | None except Fluid<br>level indicator to<br>monitor flow                                                                                                               | None except plunger<br>movement monitors<br>flow                                                                                                                 | None except plunger<br>movement monitors<br>flow                                                                                                                                |
| Packaging                   | Tyvek-sealed tray<br>containing pump,<br>administration sets,<br>and accessories.<br>Flow restrictors, and<br>y- connector sets,<br>may be separately<br>packaged     | Tyvek-sealed tray<br>containing pump,<br>administration sets,<br>and accessories.<br>Flow restrictors, and y<br>connector sets, may<br>be separately<br>packaged | Tyvek-sealed tray<br>containing pump,<br>administration sets,<br>and accessories.<br>Flow restrictors, y-<br>connector sets, and<br>catheter sets may be<br>separately packaged |

## Comparison to legally marketed Devices

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. Inside the circle is a stylized representation of a human figure, composed of three overlapping profiles, with the arms raised.

MAR 3 0 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Sgarlato Laboratories, Incorporated C/O Michael McDougall, Ph.D. Regulatory Affairs Michann Partners LLC 2315 South Bascom Avenue Suite #200 Campbell, California 95008

Re: K043204

Ro4920 |
Trade/Device Name: Sgarlato PainFree Pump; Sgarlato Pain Control Infusion Pump Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: February 24, 2005 Reccived: March 16, 2005

Dear Dr. McDougall:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 Mr.), it hay be doged in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be advised that I Dri 3 issualled or our device complies with other requirements of the Act that I DA has made a decemmations administered by other Federal agencies. You must of any I cactar stututes and regaranents, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI'N Fatt 607), adoning (21 OFR Part 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation conta of pro neeting your device as described in your Section 510(k) This letter will anow you to ogin mailing of substantial equivalence of your device to a legally premarket notication: The PDF Intelling sification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific arrive ior Jose (240) 276-0115. Also, please note the regulation entitled, Contact the Office of Coursemarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general miorinational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sulfite. Michael O. md.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K043204

Sgarlato Laboratories, Inc.

510(k) PainFree Pump - Additional Indication for Use, Section D, Page 4.1

# Indications for Use

K043204 510(k) Number (if known):

Sgarlato PainFree Pump; Sgarlato Pain Control Infusion Pump Device Name:

Indications For Use:

The Sgarlato PainFree Pump is a single use, general purpose infusion device intended to provide continuous infusion of medication (such as local anesthetics or corticosteroids) directly into a target site for the purpose of pain relief. Infusion may be carried out via percutaneous, subcutaneous, and epidural routes. The PainFree Pump may be used to provide continuous infusion of a local anesthetic directly into an intraoperative site for postoperative management of pain following surgery. It is suitable for use as an ambulatory device. It is not intended for vascular drug delivery, hyperalimentation, or the delivery of blood or blood products.

Prescription Use Yes (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

i.e. 10 m

ision Sign-Off) Chision of Anesthesiology, General Hospital, in setion Control, Dental Devices

1 (k) Number _ N 443244

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K043204](https://fda-staging.innolitics.com/device/K043204)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
