The Neo Implant is intended for surgical placement into the bone of upper /lower jaw arches as a permanent anchorage for prosthetic devices, which can restore chewing function, speech and aesthetic appearance. This is accomplished using a single-stage and two-stage surgical procedure and cement or screw retained restorations. The Neo Implants - Neo Implant System are intended for immediate placement and function on single tooth and /or multiple tooth applications recognizing sufficient bone stability and appropriate occlusal loading, to restore chewing function. Multiple tooth applications may be splinted with a bar. The Neo Abutments includes a set of components that are intended to function on Neo implants or compatible external hex. Replace Select implants and Straumann implants, as a base prosthetic construction. Surgical drills, screwtaps, countersinks or burs are used in dental, oral/maxiofacial surgery for preparing the bone (osteotomy), to receive Neo implant(s) for restorative reconstruction.
Device Story
Neo Implant System comprises threaded, internal-connection, root-form titanium dental implants; associated titanium/PEEK cover screws and healing abutments; gold alloy or titanium abutments (Neolink); and stainless steel surgical instruments (round burs, twist drills, screwtaps, countersinks). Used by dental surgeons in clinical settings for osteotomy preparation and permanent prosthetic anchorage. Implants placed via single or two-stage surgical procedures. Abutments secured to implants via screws; pre-shaped plastic copings facilitate wax-up. Prepable abutments allow clinician/technician adjustment for cemented crowns/bridges. System restores chewing function, speech, and aesthetics. Performance relies on mechanical stability within jaw bone.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing and design comparison to legally marketed predicate devices.
Technological Characteristics
Root-form titanium dental implants; internal abutment connection. Abutments: gold alloy, titanium, or PEEK. Surgical instruments: medical grade stainless steel. Twist drills feature depth markings (7-17mm) and ISO 1797-1 handpiece connections. Sterilization: supplied sterile (implants/temporary parts) or non-sterile (abutments/instruments).
Indications for Use
Indicated for patients requiring dental restoration via single-stage or two-stage surgical placement of endosseous root-form implants in the upper or lower jaw. Supports single or multiple tooth applications, including immediate placement and function, provided there is sufficient bone stability and appropriate occlusal loading. Used for cement or screw-retained restorations to restore chewing function, speech, and aesthetics.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Astra Tech implant Dental System Additional components (K974738)
3i single use, disposable drills, taps, burs etc. (K962014)
Submission Summary (Full Text)
{0}------------------------------------------------
K04 3195
# SECTION 2.
MAR 7 - 2005
## A. 510(k) SUMMARY
## Summary of Safety and Effectiveness
In accordance with 21 CFR 807.92, the following information constitutes the Neoss Ltd summary for the Neo Implant System.
| SUBMITTER'S NAME: | Neoss Ltd |
|---------------------|------------------------------------------|
| ADDRESS: | Windsor House |
| | Cornwall Road |
| | Harrogate, HG1 2PW, UK |
| CONTACT PERSON: | Fredrik Engman |
| TELEPHONE NUMBER: | +44 (0) 1423 817733 , +46 (0)709 792 892 |
| FAX NUMBER: | +44(0) 1423 817744 |
| E-MAIL | fredrik@neossimplant.com |
| DATE OF SUBMISSION: | Nov 5, 2004 |
#### Identification of device 1.
| Classification name: | Implant endosseous root-form |
|------------------------|-----------------------------------------------------------------------------------------------------|
| Proprietary Name: | Neo Implant System |
| Common Name: | Dental implant, Dental abutment, Abutment screw<br>Round bur, Twist drill, Countersink and Screwtap |
| Classification Status: | Class II per regulations 872.3640 |
| Product Codes: | DZE |
#### 2. Equivalent devices
Neoss Ltd believes the Neo Implant is substantially equivalent to the following implant fixtures including temporary parts (cover screw and healing abutment);
- Astra Tech Implants Dental System- Immediate function, nr K041492 ﺳ
and the following abutments:
- Astra Tech Implants- Dental system, nr K931767 (Neo Titanium, Neo gold, ball, bar, abutment screw)
- Astra Tech implant- Dental System Additional components, nr K974738. (Ti prepable abutments).
Neoss Ltd also believes the bone cutting instruments are substantially equivalent to the following implant fixtures;
- 3i single use, disposable drills, taps, burs etc. - nr K962014 (non-sterile reusable components)
{1}------------------------------------------------
KC-4765
### Description of the Device 3.
### Implant fixture:
miplant is a threaded, internal abutment connection, root-form titanium dental implant. The Neo implant assortment consists of a number of implants with a diameter of mynant. The Neo miplangths between 7.0 -- 19.0 mm having the same internal abutment dimension independent of implant diameter. The internal connection being equiped with interlocking elements for an insertion tool and the non-rotational locking of the abutitient. Supplied sterile.
Temporary part related to the fixture - Cover screw; Made in titanium for implantation. Supplied sterile.
Temporary part related to the fixture - Healing abutment; Made in titanium or PFEK for implantation. Supplied sterile.
### Abutments:
The Neo Abutment system is a set of modifiable gold alloy or commercially pure titanium abutments (Neolink), which are secured directly to the implant using an abutment screw. For easy wax-up a set of pre-shaped plastic copings can be used to fit to the abutment. Supplied non-sterile.
Prepable Ti abutments: Prepable titanium abutments of straight or angle (20 degrees or less) design made to be adjusted by clinician or dental technician. Crown or bridge cemented on prepped abutment. Supplied non-sterile.
Bar abutment; straight titanium or gold abutment cylinder to which a bar can be bonded (i.e welded or soldered) or cast on to. Supplied non-sterile.
Ball abutment; conventional design (ball head diameter 2,25 mm) used with standard attachments. Supplied non-sterile.
Abutment screw made in gold alloy or titanium. Screw driver connection compatible with screwdrivers supplied by Neoss. Supplied non-sterile.
### Round bur;
The round bur with diameter Ø1,8-2 with is ISO 1797-1 hand piece connection and shaft.
### Twist drills:
Two flute twist drill with ISO 1797-1 hand piece connection and shaft. Depth marking corresponding to the implant lengths in the Neoss implant system, hence 7, 9, 11, 13, 15 and 17. Diameter from Ø2.2 to Ø5,1 in appropriate steps to match the Neoss implants.
### Screwtap:
Screwtap for the different implant diameters existing in the Neo Implant system (Ø3,5 to Ø5,5) with ISO 1797-1 hand piece connection and shaft.
{2}------------------------------------------------
Countersink:
Countersink, (05,5) with ISO 1797-1 hand piece connection and shaft. The tool has a centring tap in relation to the prepared hole.
Material of bone cutting instruments is medical grade stainless steel currently being used for similar components.
#### Intended use 4.
The Neo Implant is intended for surgical placement into the bone of upper /lower jaw arches as a permanent anchorage for prosthetic devices, which can restore chewing function, speech and aesthetic appearance. This is accomplished using a single-stage and two-stage surgical procedure and cement or screw retained restorations.
The Neo Implants - Neo Implant System are intended for immediate placement and function on single tooth and /or multiple tooth applications recognizing sufficient bone stability and appropriate occlusal loading, to restore chewing function. Multiple tooth applications may be splinted with a bar.
The Neo Abutments includes a set of components that are intended to function on Neo implants or compatible external hex. Replace Select implants and Straumann implants, as a base prosthetic construction.
Surgical drills, screwtaps, countersinks or burs are used in dental, oral/maxiofacial surgery for preparing the bone (osteotomy), to receive Neo implant(s) for restorative reconstruction.
#### Technological characteristics, comparison to predicate device. 5.
Substantial equivalence of the Neo Implant System is based on design similarities between the predicative device and the devices in this application, since the devices are very similar in terms of material, size and basic design.
Like the prodicate devices, the bone cutting instruments are intended to prepare the surgical site and the technique is identical to predicative devices in relation to handpiece connection and rotational specd. The material is stainless steel with machined electro polished surface with laser depth and type marking. The basic design is the same with twist drill design for the drills, countersink with centring tap and a screwtap with adjusted thread profile in relation to the Neo Implants.
### 6. Discussion of performance testing.
Mechanical testing requested for Screw-type Endosseous Implants are described in the Guideline "Information Necessary for Pre Market Notification Submission for Screw-Type Endosseous Implants", dated December 9, 1996.
Neo implant and abutment designs are within the scope of the Guideline "Information Necessary for Pre Market Notification Submission for Screw-Type Endosseous Implants", dated December 9, 1996, in terms of material, dimensions, and intended use why we have come to the conclusion that further testing will not race new issues of safety and efficacy.
{3}------------------------------------------------
Neo bone cutting instruments, are in terms of design, material, dimensions, and intended us in all relevant aspect identical to predicative devices why we have come to the aonclusion that further testing will not race new issues of safety and efficacy.
Please see section 5. Part B Performance Testing.
#### 7. Conclusion
Based on comparison, the Neo Implants System including bone cutting instruments, from Neoss Ltd, is substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
MAR 7 - 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NEOSS Limited C/O Dr. Russell P. Pagano Vice President M Squared Associates, Incorporated 719 A Street, NE Washington, DC 20002
Re: K043195
Trade/Device Name: Neo Implant System Regulation Number: 872.3640 Regulation Namc: Endosscous Implant Regulatory Class: II Product Code: DZE Dated: February 24, 2005 Received: February 28, 2005
Dear Dr. Pagano:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becares is ned the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mensiale commerce prior to frid , as , as , ac, accordance with the provisions of Amendinens, or to devrees that have e Act (Act) that do not require approval of a premarket the rederal I ood, Drug, und Commens , therefore, market the device, subject to the general approval appliednon (the Act. The general controls provisions of the Act include controls provisions of aregistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into cither class II (Special Controls) or class III If your device is classified (osuch additional controls. Existing major regulations affecting (FMA), it may of subject to bash adata and the end Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Lederal Register.
{5}------------------------------------------------
Page 2 - Dr. Pagano
Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that IDA S Issualled of a substantial vith other requirements
mean that FDA has made a determination that your device that hearther Federal assession mean that FDA has made a decemmanen that Journalisms administered by other Federal agencies.
of the Act or any Federal statutes and regulations administered by other Federal of the Act or any rederal statues and regulations, but not limited to: registration in mostical You must comply with an the Acc S requirements) in CFR Part 801); good manufacturing practice and listing (21 CFR Pat 607), laboling (21 CFR Pat 820), and if requirements as set form in the quality bjokens (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 5 10(k) I his letter will anow you to begin maileding of substantial equivalence of your device to a premarket nothleadon. The PDF mailing or classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), a If you desire specific advice for your de at (240) 276-0115. Also, please note the regulation
15 (240) 276-0116 at 12:40) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain other general miormation on consumer Assistance at its toll-free Division (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# B. INDICATIONS FOR USE
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Device Name: Neo Implant System
### Indications for Use:
The Neo Implants - Neo Implant System arc for single-stage and two-stage surgical The Neo implaine ment or screw retained restorations.
procedures and centent of screw reasing are intended for immediate placement and The Neo Implants - Neo Implane 5 Joical and Stations recognizing sufficient bone function on single tooth and rol many of munding, to restore chewing function. Multiple tooth applications may be splinted with a bar.
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
ાર
Over the Counter Use
Susan Rios
Trilogy General Hospital,
Lawn Center, Dental Devices
Number
8041395
9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.