USCOM 1A records and monitors beat-to-beat cardiac heamodynamics in paediatric and adult patients. Its recording and storage of patient information provides trending information of cardiovascular performance for quantitative evaluation. The USCOM 1A is for use by clinicians in the hospital environment.
Device Story
USCOM 1A is a portable, battery-powered, non-invasive continuous wave (CW) Doppler device; uses ultrasonic transducers to measure blood flow velocity transcutaneously at pulmonary or aortic valves; clinician-operated in hospital environments; device processes Doppler signals to calculate hemodynamic parameters including stroke volume and cardiac output; results displayed on touch screen; system stores patient data for trending and generates reports for clinical analysis; output assists clinicians in evaluating cardiovascular performance and hemodynamic status changes; benefits include non-invasive monitoring of cardiac function.
Clinical Evidence
No clinical data. Bench testing for acoustic output, electrical safety, thermal properties, EMC, biocompatibility, and mechanical/environmental characteristics demonstrated compliance with consensus standards.
Technological Characteristics
Portable CW Doppler device; <7kg; internal battery; touch screen interface; ultrasonic transducers (2.2MHz/10mm, 2.2MHz/15mm, 3.3MHz/10mm); conforms to 21 CFR 820 and ISO 13485; electrical safety/EMC compliant.
Indications for Use
Indicated for pediatric and adult patients requiring non-invasive monitoring of cardiac hemodynamics, including stroke volume and cardiac output, in a hospital setting.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
GE Vivid 3 Expert/Pro Diagnostic Ultrasound System (K020789)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for USCOM LTD, a company that specializes in ultrasonic cardiac output monitors. The logo features a circular graphic on the left, followed by the company name in bold, sans-serif font. Below the company name is the tagline "Ultrasonic Cardiac Output Monitors" in a smaller font size. The overall design is clean and professional, conveying a sense of expertise and reliability in the medical technology field.
# Ko43139
# 510(k) Summary - USCOM 1A
# 510(k) Summary
## Per 21 CFR Part 807.92
#### Section a):
FEB 1 5 2005
| Date Prepared | Tuesday, 9th Nov 2004 |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Mr Nick Schicht<br>USCOM Ltd<br>Level 7<br>10 Loftus Street<br>Sydney, Australia<br>2000 |
| Classification Reference | 90 IYN |
| Product Code | 21 CFR 892.1550 |
| Common/Usual Name | Pulsed-Doppler |
| Proprietary Name | USCOM 1A |
| Predicate Device(s) | 1): (K020789) GE Vivid 3 Expert/Pro Diagnostic Ultrasound System.<br>Ultrasonic Pulsed Doppler Imaging System, 21 CFR 892.1550,<br>90-IYN<br>2): (K011439) BioZ.com HemoDynamic Monitor, 21 CFR 870.2770,<br>DSB |
| Reason for submission | New Device |
| Indications for Use | USCOM 1A records and monitors beat-to-beat cardiac heamodynamics<br>in paediatric and adult patients. Its recording and storage of patient<br>information provides trending information of cardiovascular performance<br>for quantitative evaluation.<br>The USCOM 1A is for use by clinicians in the hospital environment. |
#### Device Description
The USCOM 1A is a non-invasive tool that measures and records changes in the heamodynamic status of a patient.
It is a portable Continuous Wave Doppler device using an Ultrasonic Transducer to measure blood flow. it is a portable of an easy to use touch screen display with monitoring, trending and storage The GOOM TH consisted on an odo, to a see navigates intuitive and simple screens to enter patient details and perform examinations.
Targeting the patient's pulmonary or aortic valves transcutaneously, the blood flow profile can be Targeting the palifically of assisted blood flow profile, the clinician can determine a number of parameters including Stroke Volume and Cardiac Output.
The USCOM 1A saves the measures in a patient file. Multiple saved measures over time are displayed The USOW TA Saves the model of the patient's change in heamodynamic status. Reports can be generated on patient measures which can be printed and exported for analysis.
The device is portable, weighing less than 7 kilograms and incorporates an internal battery.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for USCOM LTD, a company that specializes in ultrasonic cardiac output monitors. The logo consists of a circular emblem on the left, featuring a stylized image of sound waves emanating from a central point. To the right of the emblem, the company name "USCOM LTD" is displayed in bold, uppercase letters. Below the company name, the tagline "Ultrasonic Cardiac Output Monitors" is written in a smaller font size.
# Comparison with Predicate Device(s)
The USCOM 1A system is of comparable type and is substantially equivalent to the:
- GE Vivid 3 Expert/Pro system for CW Doppler performance having similar technological ﺍﻟﺘ characteristics for safety and effectiveness features, design, construction and materials with the similar Intended Use and clinical application, and;
- BioZ.com HemoDynamic Monitor for patient data storage, and retrieval functionality. ﻟﻠﺰ
#### Section b):
#### Non-clinical tests
The device has been tested for acoustic output, electrical safety including thermal properties, EMC, biocompatibility, mechanical and environmental characteristics and has been found to comply with applicable medical device safety and performance standards.
#### Clinical tests
None required. Testing to consensus performances standards and well proven test methods is sufficient to demonstrate that the Continuous Wave Doppler technology used in the USCOM 1A, can perform clinical requirements as intended by USCOM Ltd.
#### Conclusion
The intended use and key features are consistent with traditional clinical practice, FDA guidelines and established methods of patient examination. The product development and design of the device conforms to 21 CFR 820 Quality System Regulation (QSR) and ISO 13485 Quality Management System standards for medical device manufacturers. The product is designed to conform to FDA consensus standards such as electrical safety, EMC and performance, compliance has been verified through use of independent test houses. Therefore, it is concluded that the USCOM 1A system is substantially equivalent with respect to safety and effectiveness to the predicate devices, Vivid 3 from GE Medical Systems (K020789) and BioZ from CardioDynamics (K011439)
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 5 2005
USCOM Ltd. % Ms. Christina L. Kichula Manager, Regulatory Affairs PPD Medical Device 1700 Rockville Pike, Suite 400 ROCKVILLE MD 20852
Re: K043139
Trade Name: USCOM 1A Regulation Number: 21 CFR 870.2770 Regulation Name: Impedance plethysmograph Product Code: 74 DSB Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN and ITX Dated: January 27, 2005 Received: January 27, 2005
#### Dear Ms. Kichula:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your over determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device micrslate commerce pror to that have been reclassified in accordance with the provisions of the Amendinents, of to devices and Cosmetic Act (Act). You may, therefore, market the device, subject to I cleral i ood, Drug, and Oosment 110 et. The general controls provisions of the Act include the general conners provisions of and listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the USCOM 1A, as described in your premarket notification:
Transducer Model Number
2.2MHz 10mm diameter probe 2.2MHz 15mm diameter probe
{3}------------------------------------------------
#### 3.3MHz 10mm diameter probe
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device is exassined (sis a a) =>
it may be subject to such additional controls. Existing major regulations and control It may be subject to sach additions, Title 21, Parts 800 to 898. In addition, FDA can be found in the overouncements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Flease oe advised that I Dri of issualites vour device complies with other requirements of the Act that I DA has made a actuations administered by other Federal agencies. You must Of any I cocharates and regaranents, including, but not limited to: registration and listing (21 comply with an the 11et 31equirements)01); good manufacturing practice requirements as set CFK Part 807), labeling (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed noutication. The I D7X Imanig of succiantial vice and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific auriles in 1988. at (240) 276-0120. Also, please note the regulation entitled, Colliact the Office of Compunance as (21 t notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
If you have any questions regarding the content of this letter, please contact REVIEWER at (301) 594-1212.
Sincerely yours,
David A. Ingram
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for USCOM LTD, a company that specializes in ultrasonic cardiac output monitors. The logo consists of a circular graphic on the left, followed by the company name in bold, uppercase letters. Below the company name is the tagline "Ultrasonic Cardiac Output Monitors" in a smaller font.
#### Diagnostic Ultrasound Indications for Use Form USCOM 1A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(specify) | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | N | | | | | |
| Small Organ<br>(specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | N | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral<br>Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K043139
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for USCOM LTD, a company that specializes in Ultrasonic Cardiac Output Monitors. The logo consists of a circular graphic on the left, followed by the company name in bold, uppercase letters. Below the company name is the tagline "Ultrasonic Cardiac Output Monitors" in a smaller font. The logo is simple and professional, and it effectively communicates the company's focus on cardiac output monitoring technology.
## Diagnostic Ultrasound Indications for Use Form USCOM 1A with 2.2MHz 10mm diameter probe
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | Mode of Operation | | | | | | | | | |
|--|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| | Ophthalmic | | | | | | | | | | |
| | Fetal | | | | | | | | | | |
| | Abdominal | | | | | | | | | | |
| | Intraoperative<br>(specify) | | | | | | | | | | |
| | Intraoperative<br>Neurological | | | | | | | | | | |
| | Pediatric | | | | | N | | | | | |
| | Small Organ<br>(specify) | | | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | | | |
| | Adult Cephalic | | | | | | | | | | |
| | Cardiac | | | | | N | | | | | |
| | Transesophageal | | | | | | | | | | |
| | Transrectal | | | | | | | | | | |
| | Transvaginal | | | | | | | | | | |
| | Transurethral | | | | | | | | | | |
| | Intravascular | | | | | | | | | | |
| | Peripheral<br>Vascular | | | | | | | | | | |
| | Laparoscopic | | | | | | | | | | |
| | Musculo-skeletal<br>Conventional | | | | | | | | | | |
| | Musculo-skeletal<br>Superficial | | | | | | | | | | |
| | Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
(Please do not WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Yorick R. Lymm
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for USCOM LTD, a company that specializes in Ultrasonic Cardiac Output Monitors. The logo consists of a circular emblem on the left, followed by the company name in bold, uppercase letters. Below the company name, there is a tagline that describes the company's focus on ultrasonic cardiac output monitors. The logo is simple and professional, and it effectively communicates the company's brand identity.
# Diagnostic Ultrasound Indications for Use Form USCOM 1A with 2.2MHz 15mm diameter probe
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative<br>(specify) | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | N | | | | | | |
| Small Organ<br>(specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | N | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral<br>Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Lorida h. Lynn
(Division Sign Off)
(Division Sign-Off) Durision of Reproductive, Abdominal, ﺔ ﻓ Pornitogical Dence к Матрет 15
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for USCOM LTD, a company that specializes in Ultrasonic Cardiac Output Monitors. The logo features a circular emblem on the left, containing a stylized representation of sound waves or a fingerprint. To the right of the emblem, the company name "USCOM LTD" is displayed in bold, uppercase letters. Below the company name, the tagline "Ultrasonic Cardiac Output Monitors" is printed in a smaller font size.
#### Diagnostic Ultrasound Indications for Use Form USCOM 1A with 3.3MHz 10mm diameter probe
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(specify) | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | N | | | | | |
| Small Organ<br>(specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | N | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral<br>Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation-(ODE)
David A. Hyon
(Division Sign-Off)
Division of Reproductive, Abdominal മാർ വാർശിന്ത്രിയിരുന്നു. in 145 vumper ________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.