K043117 · Best Medical International, Inc. · KXK · Dec 9, 2004 · Radiology
Device Facts
Record ID
K043117
Device Name
BEST RADIOPAQUE STRAND; BEST RADIOPAQUE MARKER
Applicant
Best Medical International, Inc.
Product Code
KXK · Radiology
Decision Date
Dec 9, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5730
Device Class
Class 2
Indications for Use
The Best Radiopaque Strand (Marker Strand) is made of radiopaque xxx markers (i.e, seeds) and absorbable spacers, inside an absorbable strand, is indicated for use in brachytherapy, soft tissues, or organ tissue, but should not be used during cardiovascular or neurological procedures. The Best Radiopaque Marker (Marker) is made of a radiopaque xxx solid seed, is indicated for use in brachytherapy, soft tissues, or organ tissue, but should not be used during cardiovascular or neurological procedures.
Device Story
Best Radiopaque Strand and Marker are implantable devices used to improve dosimetry coverage in brachytherapy. The Strand consists of two radiopaque markers separated by absorbable spacers within an absorbable strand; the Marker is a single radiopaque seed (gold, platinum, stainless steel, titanium, or tungsten). Devices are placed in soft tissue or organ tissue prior to therapeutic procedures to provide clear identification of anatomic regions for radiation targeting. Used by clinicians in a clinical setting. Benefits include enhanced visualization of target sites for improved radiation dose delivery.
Clinical Evidence
Bench testing only.
Technological Characteristics
Materials: gold, platinum, stainless steel, titanium, or tungsten markers; absorbable spacers and strand. Form factor: implantable seeds and strands. Non-active, non-electronic, non-software device.
Indications for Use
Indicated for use in brachytherapy, soft tissues, or organ tissue. Contraindicated for cardiovascular or neurological procedures.
Regulatory Classification
Identification
A radionuclide brachytherapy source is a device that consists of a radionuclide which may be enclosed in a sealed container made of gold, titanium, stainless steel, or platinum and intended for medical purposes to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy.
Special Controls
*Classification.* Class II (special controls). A prostate seeding kit intended for use with a radionuclide brachytherapy source only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Submission Summary (Full Text)
{0}------------------------------------------------
## BEST MEDICAL INTERNATIONAL, INC.
### DEC - 9 2004
## Proprietary
K043117
#### 510(k) Summary of Safety and Effectiveness VII.
- 1. Manufacturer and Contact Information:
| Manufacturer: | Best Medical International, Inc.<br>7643 Fullerton Road<br>Springfield, VA 22153 |
|----------------------|-------------------------------------------------------------------------------------------------------------------|
| Contact Information: | Sankara Ramaswamy<br>Director, Regulatory Affairs<br>Best Medical International, Inc.<br>(703) 451-2378, Ext. 107 |
- 2. Device Classification Name:
Pursuant to 21 CFR 892.5730, both the Best Radiopaque Strand and the Best Radiopaque Marker are class II devices.
| Proprietary Name: | Best Radiopaque Strand<br>Best Radiopaque Marker |
|---------------------|-----------------------------------------------------------|
| Common/Usual Name: | Best Marker Strand<br>Best Marker |
| ClassificationName: | Best Implantable Marker Strand<br>Best Implantable Strand |
- 3. Intended Use:
Please see Appendix 2 below titled, "Indications for Use Form".
- 4. Device Description and Characteristics:
The Best Radiopaque Strand (marker strand) is composed of two radiopaque markers, separated by absorbable spacers, and enclosed in an absorbable strand. It is supplied sterile.
The Best Radiopaque Marker (marker) is composed of a radiopaque nonradioactive seed (gold, platinum, stainless steel, titanium, or tungsten) which is supplied non-sterile.
The placement of the markers and marker strand in soft tissue, prior to therapeutic procedures, allows better dosimety coverage of the targeted site due to clearer identification of the anatomic regions.
{1}------------------------------------------------
# BEST MEDICAL INTERNATIONAL, INC. Proprietary
- 5. Summary of Safety and Effectiveness:
This 510(k) Summary of Safety and Effectiveness is in accordance with the requirements of the Safe Medical Devices Act (SMDA) 1990, and is provided in Appendix 1 below.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a stylized symbol. The symbol consists of three curved lines that resemble a human figure or a stylized representation of a person. The logo is printed in black ink on a white background.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### DEC - 9 2004
Dr. Joseph C. Whang Official Correspondent Best Medical International, Inc. 7643 Fullerton Road SPRINGFIELD VA 22153
Re: K043117
Trade/Device Name: Best Radiopaque Strand Regulation Number: 21 CFR 892.5730 Regulation Name: Radionuclide brachytherapy source
Regulatory Class: II Product Code: 90 KXK Dated: November 11, 2004 Received: November 10, 2004
Dear Dr. Whang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced work and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reasy 20, 1110, accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the do not require approvisions of the Act. The Act. The general controls provisions of the Act include reguirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/dsma/dsmamain.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# BEST MEDICAL INTERNATIONAL, INC. Proprietary
### Appendix 2
Indications for Use Form
K043117
510 (k) Number: Reference xxx
(1) Device Name: Best Radiopaque Strand
Indications For Use:
The Best Radiopaque Strand (Marker Strand) is made of radiopaque xxx markers (i.e, seeds) and absorbable spacers, inside an absorbable strand, is indicated for use in brachytherapy, soft tissues, or organ tissue, but should not be used during cardiovascular or neurological procedures.
(2) Devise Name: Best Radiopaque Marker
Indications For Use;
The Best Radiopaque Marker (Marker) is made of a radiopaque xxx solid seed, is indicated for use in brachytherapy, soft tissues, or organ tissue, but should not be used during cardiovascular or neurological procedures.
gold, platinum, stainless steel, titanium, or tungsten a
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
David A. Norman
(Division Sian-Q 510kl Num
1 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.