THE TRABECULAR METAL GLENOID - BIGLIANI/FLATOW THE COMPLETE SHOULDER SOLUTION, MODEL 4316-XX-XX
Applicant
Zimmer Trabecular
Product Code
KWS · Orthopedic
Decision Date
Nov 23, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Prosthetic replacement with this device may be indicated for the treatment of severe pain or significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint; un-united humeral head fractures of long duration; irreducible 3- and 4-part proximal humeral fractures; avascular necrosis of the humeral head; or other difficult clinical management problems where arthrodesis or resectional arthroplasty is not acceptable. The assembled humeral component may be used alone for hemiarthroplasty or combined with the glenoid component for total shoulder arthroplasty. Humeral heads with heights greater than 27 mm may be used for difficult clinical management problems involving rotator cuff deficiency where arthrodesis or conventional non-constrained arthroplasty is not acceptable. The Trabecular Metal Glenoid must be fully or partially cemented in place in the USA.
Device Story
Monoblock glenoid component; Trabecular Metal base; direct compression molded polyethylene articular surface. Interfaces with Zimmer B/F humeral components. Used in total shoulder arthroplasty; implanted by orthopedic surgeons in clinical settings. Provides articular surface for humeral head; restores joint function; reduces pain. Manufacturing change increases interfacial area between polyethylene and metal base to improve attachment strength.
Clinical Evidence
Bench testing only. Engineering analysis performed to evaluate the effect of increased polyethylene infiltration area on device strength and attachment area.
Technological Characteristics
Monoblock design; Trabecular Metal base; direct compression molded UHMWPE articular surface. 5 mm thickness. Cemented fixation. No software or electronic components.
Indications for Use
Indicated for patients with severe pain or disability due to degenerative, rheumatoid, or traumatic glenohumeral joint disease; un-united humeral head fractures; irreducible 3- and 4-part proximal humeral fractures; avascular necrosis; or conditions requiring arthroplasty over arthrodesis/resection. Used for total shoulder arthroplasty or hemiarthroplasty. Must be cemented.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
# The Trabecular Metal Glenoid
Bigliani/Flatow® (B/F) The Complete Shoulder Solution
| Submitter Name<br>And Address: | Zimmer Trabecular Metal Technology, Inc.<br>80 Commerce Drive<br>Allendale, New Jersey 07401-1600 |
|------------------------------------|-------------------------------------------------------------------------------------------------------|
| Contact Person: | Marci Halevi |
| Phone Number: | (201) 818-1800 X 507 |
| Fax Number: | (973) 829-0825 |
| Date Prepared: | November 2, 2004 |
| Device Trade Name: | The Trabecular Metal Glenoid, the B/F Complete Shoulder Solutio |
| Device Common Name: | Glenoid Component |
| Classification Number<br>and Name: | 21 CFR 888.3660 & 888.3650; Prosthesis, shoulder, semi & non-<br>constrained, metal/polymer cemented. |
Substantial The term "substantial equivalence" as used in this 510(k) notification is Equivalence: limited to the definition of substantial equivalence found in the Federal Food, Drug and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without premarket approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts.
Device The device is a monoblock glenoid component comprised of a Description: Trabecular Metal base with an articular surface comprised of direct compression molded polyethylene. The TM Glenoid is designed to interface & articulate with Zimmer B/F humeral components and is available in one thickness option of 5 mm, and the same outer geometry and dimensions as the B/F glenoids cleared in K022377 and K031449.
{1}------------------------------------------------
Kod3061
TM Glenoid Bl Special 510(k) Premarket Notific
## 510(k) Summary (Continued)
- Indications for use: Prosthetic replacement with this device may be indicated for the treatment of severe pain or significant disability in degenerative. rheumatoid, or traumatic disease of the glenohumeral joint; un-united humeral head fractures of long duration; irreducible 3- and 4-part proximal humeral fractures; avascular necrosis of the humeral head; or other difficult clinical management problems where arthrodesis or resectional arthroplasty is not acceptable. The assembled humeral component may be used alone for hemiarthroplasty or combined with the glenoid component for total shoulder arthroplasty. Humeral heads with heights greater than 27 mm may be used for difficult clinical management problems involving rotator cuff deficiency where arthrodesis or conventional non-constrained arthroplasty is not acceptable. The Trabecular Metal Glenoid must be fully or partially cemented in place in the USA.
Device The subject device possesses the same external geometry and Technological dimensions, the same articulation geometry, the same minimum polyethylene thickness, and same size options as the predicate devices. Characteristics & The Trabecular Metal, direct compression molded polyethylene and Comparison to monoblock design is similar to numerous cleared Zimmer Trabecular Predicate Device: Metal devices. This 510(k) clearance provides a change in manufacturing technique to increase the interfacial area between the UHMWPE and the Trabecular Metal glenoid base.
- An engineering analysis showed that the increased area of polyethylene Performance infiltration does not adversely affect device strength and increases the Data: area of attachment between the UHMWPE and the Trabecular Metal glenoid base.
Conclusion: The Trabecular Metal Glenoid is substantially equivalent to the identified predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 23 2004
Ms. Marci Halevi Manager of Regulatory Affairs Zimmer Trabecular Metal Technology 80 Commerce Drive Allendale, New Jersey 07401-1600
K043061 Re:
R042001
Trade/Device Name: Trabecular Metal Glenoid – B/F Complete Shoulder Solution Regulation Numbers: 21 CFR 888.3660, 21 CFR 888.3650 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis, shoulder joint metal/polymer non-constrained cemented prosthesis Regulatory Class: II Product Codes: KWS, KWT Dated: November 3, 2004 Received: November 4, 2004
Dear Ms. Halevi:
We have reviewed your Section 510(k) premarket notification of intent to market the device wt nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreations of the enactment date of the Medical Device Amendments, or 10 conniner of price to may 20, 20, 20, 20, 2017 11:45 provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetier For (10) inter the device, subject to the general controls provisions of the Act. The r ou may, mereleve, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
# Page 2 -- Ms. Marci Halevi
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Special 510(k) Premarket N
510(k) Number (if known) :
K043061
Device Name:
The Trabecular Metal Glenoid - The B/F Complete Shoulder Solution
Indications For Use:
Prosthetic replacement with this device may be indicated for the treatment of severe pain or significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint; un-united humeral head fractures of long duration; irreducible 3- and 4-part proximal humeral fractures; avascular necrosis of the humeral head; or other difficult clinical management problems where arthrodesis or resectional arthroplasty is not acceptable. The assembled humeral component may be used alone for hemiarthroplasty or combined with the glenoid component for total shoulder arthroplasty. Humeral heads with heights greater than 27 mm may be used for difficult clinical management problems involving rotator cuff deficiency where arthrodesis or conventional non-constrained arthroplasty is not acceptable. The Trabecular Metal Glenoid must be fully or partially cemented in place in the USA.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH; Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|------------------------------------------------------------|
| | (Division Sign-Off) |
| | Division of General, Restorative, and Neurological Devices |
| Prescription Use (Per 21 CFR 801.109) | <div style="text-align:center;">X</div> |
|-------------------------------------------------------------|-----------------------------------------|
| OR... | |
| Over-The-Counter Use 510(k) Number (Optional Format 1-2-96) | K043061 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.