TRIGEN HINDFOOT FUSION NAIL

K043052 · Smith & Nephew, Inc. · HSB · Nov 24, 2004 · Orthopedic

Device Facts

Record IDK043052
Device NameTRIGEN HINDFOOT FUSION NAIL
ApplicantSmith & Nephew, Inc.
Product CodeHSB · Orthopedic
Decision DateNov 24, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3020
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The TriGen Hindfoot Fusion Nail (HFN) is indicated for degeneration, deformity, or trauma of both the tibiotalar and talocalcaneal articulations in the hindfoot; tibiocalcaneal arthrodesis; combined arthrodesis of the ankle and sub-talar joints; avascular necrosis of the ankle and sub-talar joints; failed total ankle replacement with sub-talar intrusion; failed ankle arthrodesis with insufficient talar body; rheumatoid arthritis; severe deformity secondary to untreated talipes equinovarus or neuromuscular disease; and severe pilon fractures with trauma to the subtalar joints. The TriGen HFN is for single use only.

Device Story

Intramedullary fixation rod for ankle and hindfoot arthrodesis. Device implanted by orthopedic surgeons in clinical/OR settings to stabilize tibiotalar and talocalcaneal joints. Provides mechanical fixation to facilitate bone fusion. Used in cases of severe deformity, trauma, or failed prior surgeries. Benefits patient by restoring joint stability and reducing pain associated with degenerative or traumatic conditions.

Clinical Evidence

Bench testing only. Mechanical test data reviewed to confirm the device is equivalent to clinically used devices and capable of withstanding expected in vivo loading without failure.

Technological Characteristics

Intramedullary fixation rod. Material: Titanium alloy conforming to ASTM F 1472. Used with existing bone screws. Class II device.

Indications for Use

Indicated for patients with degeneration, deformity, or trauma of tibiotalar and talocalcaneal articulations, including rheumatoid arthritis, neuromuscular disease, talipes equinovarus, avascular necrosis, failed total ankle replacement, failed ankle arthrodesis, or severe pilon fractures with subtalar trauma.

Regulatory Classification

Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K043052 NOV 2 4 2004 ## Smith & Nephew, Inc. Summary of Safety and Effectiveness TriGen Hindfoot Fusion Nail Date of Summary: November 4, 2004 ## Contact Person and Address Janet Johnson Akil Director, Regulatory Affairs Smith & Nephew, Inc. Orthopaedic Division 1450 Brooks Road Memphis, TN 38116 1901) 399-5153 Name of Device: TriGen Hindfoot Fusion Nail Common Name: Intramedullary Fixation Rod ## Device Description The TriGen Hindfoot Fusion Nail (HFN) is an intramedullary nail used for fixation of the ankle and hind foot. The TriGen HFN nails are made from titanium alloy conforming to ASTM F 1472 and are used with existing bone screws. ## Device Classification 21 CFR 888.3020 Intramedullary fixation rod - Class II ## Mechanical and Clinical Data A review of the mechanical test data indicated that the TriGen HFN is equivalent to devices currently used clinically and is capable of withstanding expected in vivo loading without failure. #### Indications for Use The TriGen Hindfoot Fusion Nail (HFN) is indicated for degeneration, deformity, or trauma of both the tibiotalar and talocalcaneal articulations in the hindfoot; tibiocalcaneal arthrodesis; combined arthrodesis of the ankle and sub-talar joints; avascular necrosis of the ankle and sub-talar joints; failed total ankle replacement with sub-talar intrusion; failed ankle arthrodesis with insufficient talar body; rheumatoid arthritis; severe deformity secondary to untreated talipes equinovarus or neuromuscular disease; and severe pilon fractures with trauma to the subtalar joints. The TriGen HFN is for single use only. ## Substantial Equivalence Information The substantial equivalence of the TriGen HFN is based on its similarities in indications for use, design features, operational principles, and material composition to the following predicate devices - the Smith & Nephew, Inc. Tibia-Talus-Calcaneus Nail (K950394) and the Encore Orthopedics UltiMax Ankle Fusion Rod System (K991790). {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 2 4 2004 Ms. Janet Johnson Akil Director, Regulatory Affairs Smith & Nephew, Inc. Orthopaedic Division 1450 Brooks Road Memphis, Tennessee 38116 Re: K043052 Trade/Device Name: TriGen Hindfoot Fusion Nail Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: II Product Code: HSB Dated: October 4, 2004 Received: October 5, 2004 Dear Ms. Akil: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ # Page 2 - Ms. Janet Johnson Akil This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Mark A. Milburn Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Indications for Use 510(k) Number (if known): Device Name: TriGen Hindfoot Fusion Nail Indications for Use: The TriGen Hindfoot Fusion Nail (HFN) is indicated for degeneration, deformity, or rauma of both The Tricen Hindbot rusion Nak (11), is marbated for asso ; the tibiotalar and talocalcaneal articulations in the hindfoot; tibiocalcaneal arthrodest combined the tiblotalial and sub-talarons in the hillering in the nike and sub-talar joints; arthrodesis of the ankle and sub-talar joints; avascular necrosis of the anily incriff ris annrodesis of the ankle and sub-talar intrusion; failed ankle arthrodesis with insufficient lalled lotal anke replacentern with secondary to untreated talipes equinovarus or lalar body; medihatolu annitis; corors aufemily so the subtalar joints. The TriGen HFN is for single use only. | Prescription Use (Part 21 CFR 801 Subpart D) | X | |----------------------------------------------|---| |----------------------------------------------|---| AND/OR | Over-The-Counter Use (21 CFR 807 Subpart C) | | |---------------------------------------------|--| |---------------------------------------------|--| (Please do not WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Confurrence of CORH, 510(k) Number K043052 Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...