K043046 · New Star Lasers, Inc. · GEX · Feb 1, 2005 · General, Plastic Surgery
Device Facts
Record ID
K043046
Device Name
COOLTOUCH, MODEL CT3S
Applicant
New Star Lasers, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 1, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the CoolTouch CT3S Nd:YAG Laser System is: a) for use in dermatology for incision, excision, ablation and vaporization with hemostasis of soft tissue, b) for treatment of fine lines and wrinkles, c) for treatment of back acne and atrophic acne scars, d) for treatment of reflux of the greater saphenous vein associated with varicose veins and varicosities.
Device Story
The CoolTouch CT3S is an Nd:YAG laser system emitting at 1320 nm. It comprises a cabinet housing a power supply, cooling system, and microcontroller, connected to a fiber optic and handpiece. Operated by clinicians in a medical setting, the device delivers laser energy to soft tissue for surgical or dermatological procedures. The system provides hemostasis during incision, excision, ablation, and vaporization. It is also used for cosmetic skin treatments and endovenous laser therapy for varicose veins. The clinician directs the laser output to the target tissue to achieve the desired therapeutic effect, benefiting patients by providing a minimally invasive method for tissue management and vein treatment.
Clinical Evidence
Clinical trials indicated effectiveness in treating reflux of the greater saphenous vein associated with varicose veins and varicosities. Results were referenced from previous 510(k) submissions. No new clinical performance data was generated for this specific notification.
Technological Characteristics
Nd:YAG laser system; 1320 nm wavelength; consists of cabinet (power supply, cooling system, microcontroller), fiber optic, and handpiece. Class II device.
Indications for Use
Indicated for dermatology patients requiring soft tissue incision, excision, ablation, or vaporization; treatment of fine lines, wrinkles, back acne, and atrophic acne scars; and treatment of greater saphenous vein reflux associated with varicose veins and varicosities.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
CoolTouch Corporation Model NS-160 (CTEV)
Submission Summary (Full Text)
{0}------------------------------------------------
## CoolTouch Corporation Model CT3S Nd:YAG Laser System 510(k) Premarket Notification 510(k) SUMMARY
人 043046
| Submitter: | CoolTouch Corporation |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 9085 Foothills Boulevard<br>Roseville, CA 95747 |
| Contact Person: | Donald V. Johnson<br>Vice-President of Operations |
| Telephone: | (916) 677-1912 |
| Facsimile: | (916) 677-1901 |
| Date Prepared: | |
| Device Trade Name: | CoolTouch Corporation Model CT3S Nd:YAG Laser<br>System |
| Common Name: | Nd: YAG Surgical Laser |
| Classification Name: | Laser Surgical Instrument.<br>21 C.F.R. § 878.4810 |
| Legally Marketed Predicate Device: | CoolTouch Corporation Model NS-160, CTEV. |
| Description of the CoolTouch<br>Nd:YAG Laser Systems: | The CoolTouch Nd:YAG Laser Systems are Nd:YAG<br>lasers producing laser emission at 1320 nm. The lasers<br>consist of three interconnected sections: The cabinet,<br>which houses the power supply, cooling system,<br>microcontroller and the laser, the fiber optic, and/or the<br>handpiece. |
| Intended use of CoolTouch<br>Nd:YAG Laser Systems: | The intended use of the CoolTouch CT3S Nd:YAG<br>Laser System is:<br>a) for use in dermatology for incision, excision, ablation<br>and vaporization with hemostasis of soft tissue,<br>b) for treatment of fine lines and wrinkles,<br>c) for treatment of back acne and atrophic acne scars,<br>d) for treatment of reflux of the greater saphenous vein<br>associated with varicose veins and varicosities. |
| Nonclinical Performance Data: | None |
| Clinical Performance Data: | Clinical trials also produced results that indicated that<br>the CoolTouch Nd:YAG Laser Systems are effective in<br>the treatment of reflux of the greater saphenous vein<br>associated with varicose veins and varicosities. See<br>previous related 510(k) submissions for clinical results. |
| Conclusion: | The CoolTouch CT3S Nd:YAG Laser System is<br>indicated:<br>a) for use in dermatology for incision, excision, ablation<br>and vaporization with hemostasis of soft tissue,<br>b) for the treatment of fine lines and wrinkles,<br>c) for treatment of back acne and atrophic acne scars,<br>d) for treatment of reflux of the greater saphenous vein<br>associated with varicose veins and varicosities. |
| Additional Information: | None requested at this time |
{1}------------------------------------------------
.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a seal for the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three overlapping lines, which represent the department's mission to protect the health of all Americans and provide essential human services.
FEB - 1 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Donald V. Johnson Vice President of Operations CoolTouch Corporation 9085 Foothills Boulevard Roseville, California 95747
Re: K043046
Trade/Device Name: CoolTouch CT3S Nd:YAG Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: January 21, 2005 Received: January 24, 2005
Dear Mr. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrease, 1976, the enactment date of the Medical Device Amendments, or to conniner of they 2011-11-11 in accordance with the provisions of the Federal Food, Drug, de vices that have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, mererors, mans of the Act include requirements for annual registration, listing of general controls profice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it rryour device is elassified (tools. Existing major regulations affecting your device can may be subject to suen adaterial evenlations. Title 21, Parts 800 to 898. In addition, FDA may be round in the over ments concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean ITCase oc advised that I DTC issuantes syour device complies with other requirements of the Act that I Dri has made a actor regulations administered by other Federal agencies. You must of ally it catal statutes und regalabents, including, but not limited to: registration and listing (2 h Comply with an the Hot 31equirements, and manufacturing practice requirements as set CFN in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Donald V. Johnson
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter will and in your and inding of substantial equivalence of your device to a legally premaince notificated in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please ir you dome office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other golferal micronal and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATION FOR USE STATEMENT
510(k) Number: __ K 04304b____________________________________________________________________________________________________________________________________________________
Device Name: CoolTouch CT3S Nd:YAG Laser System
Indications for Use:
The CoolTouch CT3S Nd:YAG Laser System is indicated:
a) for use in dermatology for incision, excision, ablation and vaporization with
hemostasis of soft tissue,
b) for use in the treatment of fine lines and wrinkles,
c) for treatment of back acne and atrophic acne scars,
d) for treatment of reflux of the greater saphenous vein associated with varicose veins and varicosities.
(Please do not write below this line - Continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE) Miriam C. Pror (Division Sign-Off) Division of General, Restorative, and Neurological Devices KO 4 38ver-the-Counter Use_ OR Prescription Use 510(k) Number 1 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.