The AN6250 is a digital X-ray general radiography system intended for use by qualified/trained doctor or technician and is designed to perform radiographic exposures of the whole body including skull, spinal column, chest, abdomen, extremities and other body parts excluding mammography. Radiographic exposures may be taken with the patient in the sitting, standing, or lying in the prone or supine position. Images from the AN6250 Digital Radiology System are available for preview by the doctor / x-ray technician on the operator's workstation within seconds of the x-ray exposure. After acceptance by the operator, the digital (DICOM) images can be stored on electronic media, or exported to a (DICOM/PACS) network, clinical review station or to a film printer.
Device Story
AN6250 is a stationary, general-purpose digital X-ray system. It captures radiographic images of the whole body (excluding mammography) using X-ray exposure. Operated by physicians or technicians in clinical settings, the system processes raw X-ray data into digital images. These images are displayed on an operator workstation for immediate preview. Following operator acceptance, images are stored locally or transmitted via DICOM/PACS network to clinical review stations or film printers. The device facilitates diagnostic imaging, enabling healthcare providers to visualize internal anatomy for clinical decision-making. Benefits include rapid image availability and digital workflow integration.
Clinical Evidence
Bench testing only. The device was validated at the unit and system level against Requirements Specifications and recognized consensus standards, including IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and radiation protection standards. No clinical data was required or provided.
Technological Characteristics
Stationary digital X-ray system. Components include X-ray source and digital detector. Connectivity via DICOM/PACS network. Complies with IEC 60601-1, 60601-1-1, 60601-1-3, 60601-2-7, 60601-2-28, 60601-2-32, and 60601-1-2. Tested for mechanical shock, vibration, and line dropout susceptibility.
Indications for Use
Indicated for general radiographic imaging of the whole body, including skull, spine, chest, abdomen, and extremities, in patients of all ages. Excludes mammography. Intended for use by qualified clinicians or technicians with patients in sitting, standing, or recumbent positions.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Analogic AN6150 Digital Radiology System (K040995)
Submission Summary (Full Text)
{0}------------------------------------------------
NOV 1 9 2004
Image /page/0/Picture/1 description: The image shows the word "ANALOGIC" in bold, black letters. To the right of the word is a registered trademark symbol. A black square is located to the right of the trademark symbol. The letters are sans-serif and appear to be a company logo.
K043825
## 510(k) Summary For Analogic Corporation AN6250 Digital Radiology System
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92
#### 1. Submitter's Name and Address:
Analogic Corporation 8 Centennial Drive Peabody, MA 01960
#### Date this Summary was Prepared: 2.
November 2, 2004
#### 3. Contact Person:
Donald J Sherratt, Regulatory Affairs Manager (978) 977-3000 extension 3049 Tel: (978) 977-6808 Fax:
#### Device Name: 4.
| Proprietary or Trade Name: | AN6250 Digital Radiology System |
|----------------------------|-------------------------------------------|
| Common Name: | Digital Radiology Systems |
| Classification Name: | Digital Radiology Systems and Accessories |
| Device Classification: | Class II |
#### 5. Predicate Device:
The legally marketed device to which equivalence is being claimed is the Analogic AN6150 Digital Radiology System. The predicate system was cleared under Premarket Notification K040995.
{1}------------------------------------------------
#### Device Description: 6.
The AN6250 is a Stationary, General Radiology X-ray System.
#### Intended Use 7.
The AN6250 is a digital X-ray general radiography system intended for use by qualified/trained doctor or technician and is designed to perform radiographic exposures of the whole body including skull, spinal column, chest, abdomen, extremities and other body parts excluding mammography. Radiographic exposures may be taken with the patient in the sitting, standing, or lying in the prone or supine position.
Images from the AN6250 Digital Radiology System are available for preview by the doctor / x-ray technician on the operator's workstation within seconds of the x-ray exposure. After acceptance by the operator, the digital (DICOM) images can be stored on electronic media, or exported to a (DICOM/PACS) network, clinical review station or to a film printer.
#### Comparison of Technological Characteristics: 8.
The design of the new AN6250 Digital Radiology System is derived from the design of the AN6150 Digital Radiology System and uses the same technological principles.
#### Compliance with Voluntary Standards 9.
Before final market release, the AN6250 Digital Radiology System will be thoroughly validated at the unit and system level to meets all elements of its Requirements Specification. This includes the following non-clinical tests:
- . IEC 60601-1:1988, +A1:1991, +A2:1995, an FDA recognized consensus standard for safety of medical electrical equipment.
- IEC 60601-1-1: .
- IEC 60601-1-3:1994. Medical electrical equipment Part 1: General . requirements for safety - 3. Collateral standard: General requirements for radiation protection in diagnostic X-ray equipment
- . IEC 60601-2-7:
- IEC 60601-2-28
- . IEC 60601-2-32:
- IEC 60601-1-2:2001, an FDA recognized consensus standard for electromagnetic . compatibility.
- Line Dropout and Variation Susceptibility were tested according to the FDA . Reviewer Guidance for premarket Notification Submissions, November 1993
- Mechanical Shock and Vibration Tests .
- . Shipping Container Transportation Test
#### 10. Conclusion
Based on the similarity in construction to the predicate device, same technological characteristics and the same intended use, the device is substantially equivalent.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows a circular logo for the Department of Health & Human Services - USA. The logo features the department's name in a circular arrangement around an abstract symbol. The symbol resembles an eagle or a bird in flight, represented by three curved lines.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# NOV 1 9 2004
*Mr. Donald J. Sherratt Regulatory Affairs Manager Analogic Corporation 8 Centennial Drive Centennial Industrial Park PEABODY MA 01960
Re: K043025
Trade/Device Name: AN6250 Digital Radiology System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: 90 KRP Dated: November 2, 2004 Received: November 3, 2004
Dear Mr. Sherratt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including. but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter with and by the FDA finding of substantial equivalence of your device to a legally prematice notification. "Thesalts in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specifice of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the 60 : 77). I ou may or man curers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Nancy C. brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "ANALOGIC" in bold, black letters. The letters are sans-serif and appear to be a custom font. To the right of the word is a registered trademark symbol and a black square. The background is white.
510(k) Number: Ko43025 25 ---
Device Name: AN6250 Digital Radiology System
Indications For Use:
The AN6250 is a digital X-ray general radiography system intended for use by qualified/trained The IIN0250 Is a creatived to perform radiographic exposures of the whole body including skull, spinal column, chest, abdomen, extremities and other body parts excluding mammography. Radiographic exposures may be taken with the patient in the sitting, standing, or lying in the prone or supine position.
Images from the AN6250 are available for preview by the doctor/X-ray technician on the operator's workstation within seconds of the X-ray exposure. Digital (DICOM) images can be stored on electronic media, or exported to a (DICOM/PACS) network, clinical review station or to a film printer.
Prescription Use ਮ (21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K043025
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.