← Product Code [FAJ](/productcode/FAJ) · K043022

# STRYKER FLEXIBLE CYSTONEPHROSCOPE (K043022)

_Stryker Endoscopy · FAJ · Mar 15, 2005 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K043022

## Device Facts

- **Applicant:** Stryker Endoscopy
- **Product Code:** [FAJ](/productcode/FAJ.md)
- **Decision Date:** Mar 15, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Stryker Flexible CystoNephroscope is indicated for use during minimally invasive urological procedures through natural body orifices or through perculaneous access, and is intended for, but not limited to, examining the urinary tract and the interior of the kidney, and, using additional accessories, diagnostic and therapeutic procedures.

## Device Story

Flexible endoscope for urological procedures; used via natural body orifices or percutaneous access; enables visualization of urinary tract and kidney interior; facilitates diagnostic and therapeutic interventions when used with additional accessories; operated by clinicians in clinical settings; provides direct visualization to assist in clinical decision-making; benefits patients by allowing minimally invasive access to internal structures.

## Clinical Evidence

No clinical data. Bench testing only, including conformance to voluntary safety standards for endoscopic equipment and ISO 10993 for biological evaluation.

## Technological Characteristics

Flexible endoscope; features active deflection capability; conforms to ISO 10993 for biological evaluation and safety standards for endoscopic equipment.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- ACMI ACN-2 11001 DD Nephro-Fiberscope ([K970427](/device/K970427.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

K0430yy

Device Name

| Proprietary Name:      | Stryker Flexible CystoNephroscope  |
|------------------------|------------------------------------|
| Common and Usual Name: | Endoscope, Cystoscope, Nephroscope |
| Classification Name:   | Endoscope, Cystoscope, Nephroscope |

This 510(k) summary of safety and effectiveness is being submitted in accordance with requirements of the SMDA 1990.

The Stryker Flexible CystoNephroscope is substantially equivalent in terms of safety and The Stryker Flexible CystoNephroscope is Substantially Equivalin in Childer Collection Corporation of Corporation of Storz
effectiveness to the currently marketed ACMI ACN-2 11001 DD Nephro-Fiberscope (K970427).

The Stryker Flexible CystoNephroscope is a modification of the currently marketed Stryker The Stryker Flexible Cystolephroscope is a mounce, materials, and intended use. The Stryker 
Flexible Ureteroscope, with changes to labeling, performance, mad intended use. Flexible Ureteroscope, with changes to abeling, peromance, ma interess and and of the country of the country of the more in interest of the first Plexible CyStoricphroooops of active deflection capability.

The Stryker Flexible CystoNephroscope is indicated for use during minimally invasive urological The Stryker Flexible Cystonephilosope is through perculaneous access, and is intended for but not
procedures through natural body offices or through perculaneous accessories, procedures through natural body offices of the intent of the kidney, and, using additional accessories,
limited to, examining the uning tract and the intenseduces illulled to, examining the and therapeutic procedures.

The Stryker Flexible CystoNephroscope conforms to the following voluntary safety of Endossanio The Stryker Flexible CystonesticsCope Connemis to the Stocking Comments for the Safety of Endoscopic Equipment, ISO 10993 Biological Evaluation of Medical Devices.

There are no significant technological or performance differences between the Stryter Flexible There are no significant technological of penomance there any new questions raised
CystoNephroscope and the identified predicale devices , nor are there any new substantially CystoNephroscope and the Identified predicate devices, nor and any any any and equivalent to the identified predicate devices.

Contact:

11/1/04
Date:

Christopher L. Cook

Stryker Endoscopy 5900 Optical Court San Jose, CA 95138 Phone: 408-754-2288 408-754-2521 Fax: Email: chris.cook@stryker.com

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image shows a logo for the Department of Human Services. The logo features a stylized eagle with three lines representing its wings. The text "Department of Human Services" is arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 5 2005

Mr. Christopher L. Cook Regulatory Supervisor Stryker Endoscopy 5900 Optical Court SAN JOSE CA 95138

Re: K043022

Trade/Device Name: Stryker Flexible CystoNephroscope Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 FAJ and FGA Dated: February 16, 2005 Received: February 17, 2005

Dear Mr. Cook:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) I ms letter with anow you to ocgin maxing of substantial equivalence of your device to a legally premarket notification: "The PDA mianing of bassanian for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your de roo of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification the Also, please note the regulation entitled, formation on your responsibilities under the Act from the 807.77). " Ou may obtain other general and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Device Name: Stryker Fiexible CystoNephroscope

1 : 420 %

Indications for Use:

The Stryker Flexible CystoNephroscope is indicated for use during minimally invasive undigital The Stryker Flexible Cystonephosope is number on through percularies and is intended for procedures accessed through the united on the interior of the kidney, and, using additional 
but not limited to, examining the unities and the interprogrammand but not limited to, examining the ennaly and therapeutic procedures.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IE NEEDED) TE DELOW THOLE. Of CDRH, Office of Device Evaluation (ODE)

Kaila Lyons

(Division Sign-Off) Division of Reproductive, Abdominal and Badiological Devices 510(k; Number ________________________________________________________________________________________________________________________________________________________________

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter-Use_

---

**Source:** [https://fda-staging.innolitics.com/device/K043022](https://fda-staging.innolitics.com/device/K043022)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).