VITEK GRAM POSITIVE SUSCEPTIBILITY (GPS) CARD FOR RIFAMPIN

K042946 · bioMerieux, Inc. · LON · Dec 7, 2004 · Microbiology

Device Facts

Record IDK042946
Device NameVITEK GRAM POSITIVE SUSCEPTIBILITY (GPS) CARD FOR RIFAMPIN
ApplicantbioMerieux, Inc.
Product CodeLON · Microbiology
Decision DateDec 7, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2
AttributesAI/ML

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Antimicrobial susceptibility to rifampinRegression analysis and discriminate analysis98.9% Category Agreement (Combined)Clinical study at three sites including fresh clinical isolates, stock isolates, and a challenge set

Indications for Use

The VITEK® Gram Positive Susceptibility Test is intended to be used with the VITEK® System for the automated quantitative or qualitative susceptibility testing of isolated organisms for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae. VITEK® Gram Positive Rifampin is designed for antimicrobial susceptibility testing of Staphylococcus aureus and Staphylococcus epidermidis. VITEK Gram Positive Rifampin is intended for use with the VITEK System as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.

Device Story

VITEK® Gram Positive Susceptibility Card is a 45-well test card used with the VITEK® System for automated antimicrobial susceptibility testing. Clinical laboratory technicians inoculate the card with a saline-diluted isolate suspension via a vacuum filling process. The VITEK® Reader/Incubator monitors organism growth optically every hour for 6–15 hours. The system uses regression and discriminate analysis algorithms to interpret growth patterns against the antimicrobial agent. The device outputs a qualitative susceptibility interpretation (SIR). This automated process eliminates manual handling after inoculation, providing standardized susceptibility results to guide clinical antimicrobial therapy.

Clinical Evidence

Bench testing only. External evaluation conducted using fresh and stock clinical isolates and challenge strains. Performance compared to NCCLS agar dilution reference method. Results demonstrated 98.9% overall Category Agreement. Reproducibility and Quality Control testing yielded acceptable results.

Technological Characteristics

Miniaturized microdilution card for use with VITEK System. Automated optical growth monitoring. Measures MIC in range of 0.25 - 1.5 µg/ml (or ≤0.25 - ≥4 µg/ml per labeling). System-based incubation and detection. Software-driven automated calculation of MIC based on growth thresholds.

Indications for Use

Indicated for qualitative susceptibility testing (SIR) of Staphylococcus spp., Enterococcus spp., and Group B and Group D streptococci to rifampin at concentrations of 0.25, 0.5, and 1.5 µg/mL (calling range ≤ 0.25 – ≥ 4 µg/ml). For prescription use only.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k042946 B. Purpose for Submission: Addition of rifampin to the Vitek® Antimicrobial Susceptibility Test (AST) System C. Measurand: Rifampin ≤0.25 – ≥4 μg/ml D. Type of Test: Qualitative AST growth based detection E. Applicant: bioMerieux, Inc. F. Proprietary and Established Names: VITEK® Gram Positive Susceptibility Card G. Regulatory Information: 1. Regulation section: 21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System 2. Classification: II 3. Product code: LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation 4. Panel: 83 Microbiology H. Intended Use: 1. Intended use(s): The VITEK® Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® System for the automated quantitative or qualitative susceptibility testing of isolated colonies for most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae. The VITEK® Gram Positive Susceptibility Card is intended for use with the {1} Page 2 of 6 VITEK® system in clinical laboratories as an in vitro test to determine the susceptibility of staphylococci, enterococci and Group B and Group D streptococci to antimicrobial agents when used as instructed in the “pinsert” and operator’s manual. 2. **Indication(s) for use:** This submission is for the addition of the antibiotic rifampin at concentrations of 0.25, 0.5, and 1.5 µg/mL for a calling range of ≤ 0.25 – ≥ 4 µg/ml on the VITEK® Gram Positive Susceptibility Card to provide qualitative results (SIR). 3. **Special conditions for use statement(s):** For prescription use Qualitative SIR reading only, MIC results not available 4. **Special instrument requirements:** N/A I. **Device Description:** Each VITEK® test card contains 45 wells. The positive control well determines organism growth without antimicrobial inhibition. A suspension of the isolate to be tested is diluted with 0.45 – 0.5% sterile saline. The VITEK® Card is inoculated with the diluted suspension using a vacuum filling process in the VITEK® Filling Module. After the card is inoculated and placed inside the VITEK® Reader/Incubator, no further handling is required. Organism growth inside the card is optically monitored throughout the 6 – 15 hours incubation cycle. J. **Substantial Equivalence Information:** 1. **Predicate device name(s):** VITEK® Gram Positive Susceptibility Card for Gatifloxacin 2. **Predicate 510(k) number(s):** N50510/S143 3. **Comparison with predicate:** {2} | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Determine antimicrobial susceptibility to antimicrobial agents | Same | | Test Organism | Gram Positive Cocci | Same | | Test Card | VITEK® card format with base broth | Same | | Instrument | VITEK® System | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Antibiotic | Rifampin at specific concentrations | Gatifloxacin at specific concentrations | | Reading algorithm | Unique for rifampin | Unique for gatifloxacin | # K. Standard/Guidance Document Referenced (if applicable): "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA"; NCCLS M7 (M100-S14) "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard." # L. Test Principle: The VITEK® System determines when a well demonstrates growth (positive) based on the attenuation of light measured by optical scanner. Organism growth is expressed as increased turbidity in wells. Optical measurements are taken on an hourly basis. If during the 6 - 15 hour incubation cycle, bacterial growth occurs at levels equal to or greater than a predetermined threshold, regression analysis is utilized, along with the organism's identification, to determine the appropriate MIC value for the antimicrobial. The VITEK® Susceptibility Card test is based on the microdilution minimum inhibitory concentration technique with concentrations equivalent to standard method concentrations. Several parameters based on the growth characteristics observed are used to provide appropriate input for calculations. Discriminate analysis is used to develop the algorithm that determines the susceptibility result for all antimicrobials on the VITEK® system. An algorithm of using the 3 dilutions on the card allows for a calling range of $\leq 0.25 - \geq 4\mu \mathrm{g / mL}$ . The AST result must be linked to organism identification in order to determine a category interpretation. A category interpretation (SIR) will be reported. {3} Page 4 of 6 # M. Performance Characteristics (if/when applicable): ## 1. Analytical performance: a. Precision/Reproducibility: Ten gram-positive on-scale organisms were tested three times at three sites to determine within site and site to site reproducibility demonstrating >95% reproducibility. b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): The recommended QC isolates were tested on every test occasion with the reference method and the VITEK®. The reference method QC results were in range for every day tested. The VITEK® was tested a sufficient number of times to demonstrate that the system can produce QC results in the recommended range. Results demonstrated that the QC organisms had the same mode. Quality Control Table | ORGANISM | conc. | Reference | | | Vitek® | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | | | | | S. aureus | ≤0.25 | | 77 | | | 96 | | | ATCC 29213 | | | | | | | | | Expected Range: | | | | | | | | | 0.004 – 0.016 μg/ml | | | | | | | | | | | | | | | | | | E. faecalis | 0.50 | | | | | | | | ATCC 29212 | 1 | | 28 | | | | | | Expected Range: | 2 | | 48 | | | 78 | | | 0.5 – 4 μg/ml | 4 | | | | | 10 | | | | | | | | | | | Inoculum density control was monitored using a colorimeter and colony count. This was standardized weekly with all results recorded and in the expected range. Verification was performed during internal testing. A 0.5 McFarland is used to determine the correct inoculum. Colony counts were performed periodically at each site to demonstrate that the inoculum procedure results were in the expected CFU/ml. d. Detection limit: Not Applicable {4} Page 5 of 6 e. Analytical specificity: Not Applicable f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: A clinical study was conducted at three sites using the VITEK® cards with rifampin and the NCCLS reference agar dilution method prepared as recommended in NCCLS M7 approved standard. Inoculum was prepared with direct colony suspension. The testing included both fresh clinical isolates and stock isolates along with a challenge set with known results. The test device had a growth rate of >90.0%. A comparison was provided to the reference method with the following agreement. Essential agreement was not calculated because the Vitek card contained <5 dilutions of rifampin. Summary Table for Staphylococcus spp. | | CA Tot | CA N | CA % | #R | Min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | 299 | 298 | 99.7 | 7 | 1 | 0 | 0 | | Challenge | 77 | 74 | 96.1 | 5 | 3 | 0 | 0 | | Combined | 376 | 372 | 98.9 | 12 | 4 | 0 | 0 | CA-Category Agreement R-resistant isolates min- minor discrepancies maj-major discrepancies vmj-very major discrepancies CA is when the interpretation of the reference method agrees exactly with the interpretation of the VITEK® results. b. Matrix comparison: Not Applicable 3. Clinical studies: a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not Applicable c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable 4. Clinical cut-off: {5} Page 6 of 6 Not Applicable 5. Expected values/Reference range: Staphylococcus spp. ≤ 1 (S), 2 (I), ≥4 (R) The Interpretative criteria, QC isolates and the expected ranges are the same as recommended by the NCCLS and the FDA. All values will be included in the package insert. The ability of the VITEK® system to detect resistance to rifampin in S. aureus and S. epidermidis organisms is unknown because resistant organisms were not available at the time of comparative testing. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: 1. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...