K042904 · Draeger Medical Systems, Inc. · MHX · Feb 18, 2005 · Cardiovascular
Device Facts
Record ID
K042904
Device Name
INFINITY KAPPA XLT
Applicant
Draeger Medical Systems, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Feb 18, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Infinity Kappa XLT is capable of measuring heart rate, respiration rate, invasive pressure, I he inimity trapper HD rovingia, temperature, cardiac output, arterial oxygen saturation, pulse non invasite prossure, and 12-Lead ST Segment Analysis. The device produces rate, apped, of beginent if any of the physiological parameters monitored vary beyond preset Imits and timed or alarm recordings will be produced. The device will connect to an R50 millis and thised or thank rocereings with the evice is capable of displaying clinical data received over a hospital information system.
Device Story
Infinity Kappa XLT is a patient monitoring system combining features of Infinity Delta/DeltaXL/Kappa and Infinity Explorer. System components: Vital Signs Engine (signal acquisition), medical-grade display, and power supply. Inputs: physiological signals (ECG, pressure, temperature, SpO2, respiration, EEG via pod, muscle response via NMT pod, gas concentrations via SCIO module). Operation: continuous monitoring of vital signs; visual/audible alarms triggered when parameters exceed preset limits; data recording via R50 recorder; integration with hospital information systems. Used in clinical environments by healthcare professionals to assess patient condition and guide clinical decisions. Benefits: real-time patient status awareness and automated alarm notification.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing in accordance with applicable standards and internal design control procedures.
Technological Characteristics
Modular patient monitoring system. Components: Vital Signs Engine, display, power supply. Connectivity: Infinity Network, RS-232 interface for third-party devices, hospital information system integration. Standards: IEC 60601-1. Software-driven architecture combining monitoring and workstation functionality.
Indications for Use
Indicated for adult, pediatric, and neonatal patients requiring physiological monitoring. Parameters include heart rate, respiration, invasive/non-invasive pressure, arrhythmia, temperature, cardiac output, SpO2, pulse rate, apnea, and ST segment analysis. Cardiac output, ST segment, and arrhythmia monitoring are restricted to adult and pediatric populations. Intended for use by healthcare professionals (physicians, nurses, technicians) in clinical settings.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
K042904
#### 510(k) SUMMARY FEB | 8 2005
as required per 807.92(c)
Submitters Name, Address:
Draeger Medical Systems, Inc. Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879 Contact person for this submission: Penelope H. Greco October 20, 2004 Date submission was prepared:
Trade Name, Common Name and Classification Name:
A. Trade Name: Infinity Kappa XLT
# B. Common Name, Classification Name, Class and Regulation Number:
| Common Name | Classification Number | Class | Regulation Number |
|----------------------------------------------------------------------|-----------------------|-------|-------------------|
| Monitor, physiological, patient (with arrhythmia detection or alarms | MHX | 2 | 870.1025 |
| System, Network and Communication, Physiological Monitors | MSX | 2 | 870.2300 |
| Computers and Software, Medical | LNX | | |
Predicate Device Identification:
Infinity Explorer (K040945) Infinity Delta / DeltaXL / Kappa (K033957) Infinity SC 8000 (K012016)
Device Description:
The Infinity Kappa XLT is an addition to Draeger's INFINITY patient monitoring system that combines on one platform the patient monitoring features of the Infinity Delta/DeltaXL/Kappa and the Infinity Explorer, Draeger's software-driven critical care workstation.
The Kappa XLT System consists of three basic components:
- The Kappa XLT Vital Signs Engine that acquires all signals ●
- The Kappa XLT Display, a medical grade display unit .
- . The Kappa XLT Power Supply.
Page 1 of 2
### Draeger Medical Systems, Inc.
16 Electronics Avenue Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879
{1}------------------------------------------------
## Intended Use:
The Infinity Kappa XLT is capable of measuring heart rate, respiration rate, invasive pressure, I he inimity trapper HD rovingia, temperature, cardiac output, arterial oxygen saturation, pulse non invasite prossure, and 12-Lead ST Segment Analysis. The device produces rate, apped, of beginent if any of the physiological parameters monitored vary beyond preset Imits and timed or alarm recordings will be produced. The device will connect to an R50 millis and thised or thank rocereings with the evice is capable of displaying clinical data received over a hospital information system.
Assessment of non-clinical performance data for equivalence:
The Infinity Kappa XLT was tested in accordance with applicable standards and internal design control procedures and was determined to be as safe and effective for its intended use as the predicate devices.
Assessment of clinical performance data for equivalence:
Clinical performance evaluation indicates that the Infinity Kappa XLT is substantially equivalent to the Infinity Delta series monitors and the Infinity Explorer.
Biocompatability:
Not applicable
Sterilization:
Not applicable
Standards and Guidance: IEC 60601-1
Draeger Medical Systems, Inc. 16 Electronics Avenue Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
FEB 1 8 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Draeger Medical Systems c/o Ms. Penelope H. Greco Regulatory Submissions Manager 16 Electronics Avenue Danvers, MA 01923
K042904 Re: Trade Name: Infinity Kappa XLT Regulation Number: 21 CFR 870.1025 Regulation Name: Arrythmia Detector and Alarm Regulatory Class: Class II (two) Product Code: 74 MHX Dated: January 20, 2005 Received: January 21, 2005
Dear Ms. Greco:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becement on the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreativent of the enactment date of the Medical Device Amendments, or 10 eonimered prior to may 2017 11:11 in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Penelope H. Greco
Please be advised that FDA's issuance of a substantial equivalence determination does not mean i touse oc northou that I Dr bristian that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I oderated and the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (21 CFR Part 820); and if applicable, the electronic for at many a morely by so wisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) rms letter will anow yours ough finding of substantial equivalence of your device to a legally prestiated predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bhimma for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use K042904
510(k) Number (if known):_
Device Name: Infinity Kappa XLT
The Infinity Kappa XLT is capable of measuring heart rate, respiration rate, invasive pressure, non-invasive pressure, arthythmia, temperature, cardiac output, arterial oxygen saturation, pulse rate, apnea, ST Segment Analysis, and 12-Lead ST Segment Analysis.
The device produces visual and audible alarms if any of the physiological parameters monitored vary beyond preset limits and timed or alarm recordings will be produced. The device will connect to an R50 recorder via the Infinity Network and is capable of displaying clinical data received over a hospital information system.
Kappa XLT receives EEG signals when connected to an Infinity EEG pod.
The Infinity Trident NMT pod measures the muscle response to electrical stimulation of a peripheral nerve and when connected to a Kappa XLT transmits that information to the device.
The SCIO module sample breathing gases from adults and pediatrics. The gas module continuously measures the content of CO2, N2O, O2 and one of the anesthetic agents, halothane, isoflurane, Enflurane, Sevoflurane and Desflurane in any mixture and communicates real time and derived gas information to Kappa XLT.
With etCO2 the Kappa XLT can measure end tidal carbon dioxide, inspired carbon dioxide, and respiration rate in either mainstream or side-stream measurement mode; and with etCO2+Respiratory Mechanics, spirometry and carbon dioxide can be monitored. The device can interface with specific third party devices via an RS 232 connection.
The device is intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.
The devices are intended to be used on Adult, Pediatric and Neonatal populations, with the exception of the parameter Cardiac Output, ST Segment Analysis, and arrhythmia which are intended for use in the adult and pediatric populations.
### MRI Compatibility Statement:
The Infinity Kappa XLT is not compatible for use in a MRI magnetic field.
Prescription Use - -Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ OR (Per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blumuma
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.