← Product Code [MSX](/productcode/MSX) · K042845

# THE PHILIPS MP20, MP30, MP40, MP50, MP60, MP70, AND MP90 INTELLIVUE PATIENT MONITORS (K042845)

_Philips Medizin Systeme Boblingen GmbH · MSX · Nov 12, 2004 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K042845

## Device Facts

- **Applicant:** Philips Medizin Systeme Boblingen GmbH
- **Product Code:** [MSX](/productcode/MSX.md)
- **Decision Date:** Nov 12, 2004
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.2300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters patients. Intended for monitoring, recording and alarming of of multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The MP20, MP30, MP40 and MP50 are additionally intended for use in transport situations within hospital environments. EASI 12-lead ECG is only for use on adult and pediatric patients. ST Seqment monitoring is restricted to adult patients only. The transcutaneous gas measurement (tcp02/tcpCO2) is restricted to neonatal patients only.

## Device Story

IntelliVue patient monitors (MP20-MP90) collect physiological data via various sensors/probes (ECG, blood pressure, gas, etc.). Release B.1.1 software enables integration with M1013A EGM Gas Analyzer. System processes inputs to provide real-time monitoring, recording, and alarming of patient status. Used in hospital settings and during intra-hospital transport by healthcare professionals. Output displayed on monitor for clinical review; supports immediate clinical decision-making regarding patient status. Benefits include continuous patient surveillance and timely intervention through automated alarms.

## Clinical Evidence

Bench testing only. No clinical data presented. Verification and validation activities included system-level performance tests and safety testing based on hazard analysis. Pass/Fail criteria derived from predicate device specifications.

## Technological Characteristics

Multi-parameter physiological monitor system. Software-based signal processing and alarm management. Connectivity via network/communication system. Integrates with external gas analyzer (M1013A).

## Regulatory Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

## Predicate Devices

- [K014159](/device/K014159.md)
- [K021778](/device/K021778.md)
- [K032858](/device/K032858.md)
- [K040304](/device/K040304.md)
- [K040183](/device/K040183.md)
- [K040259](/device/K040259.md)
- [K040357](/device/K040357.md)
- [K041235](/device/K041235.md)

## Reference Devices

- Gas Analyzer model M1013A EGM ([K041956](/device/K041956.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.

KO42845

1. The submitter of this premarket notification is: Egon Pfeil Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: ++49 7031 463-1926 Fax: ++49 7031 463-2442 e-mail: egon.pfeil@philips.com

This summary was prepared on October 13, 2004.

- 2. The names of the devices are the Philips MP20, MP30, MP40, MP50, MP60, MP70, and MP90 IntelliVue Patient Monitor. Classification names are as follows:

| Device Panel                             | Classification  | ProCode | Description                                                                 |
|------------------------------------------|-----------------|---------|-----------------------------------------------------------------------------|
| Circulatory<br>System Devices<br>(12625) | \$870.1025, III | DSI     | Detector and alarm, arrhythmia                                              |
|                                          | \$870.1025, III | MLD     | Monitor, ST Segment with Alarm                                              |
|                                          | \$870.1025, III | MHX     | Monitor, Physiological, Patient<br>(with arrhythmia detection or<br>alarms) |
|                                          | \$870.1100, II  | DSJ     | Alarm, Blood Pressure                                                       |
|                                          | \$870.1110, II  | DSK     | Computer, Blood Pressure                                                    |
|                                          | \$870.1130, II  | DXN     | System, Measurement, Blood-<br>Pressure, Non-Invasive                       |
|                                          | \$870.1435, II  | DXG     | Computer, Diagnostic, Pre-<br>Programmed, Single-Function                   |
|                                          | \$870.1915, II  | KRB     | Probe, Thermodilution                                                       |
|                                          | \$870.2060, II  | DRQ     | Amplifier and Signal<br>Conditioner, Transducer Signal                      |
|                                          | \$870.2300, II  | DRT     | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate Alarm)                   |
|                                          | \$870.2340, II  | DPS     | Electrocardiograph                                                          |
|                                          | \$870.2340, II  | MLC     | Monitor, ST Segment                                                         |
|                                          | \$870.2350, II  | DRW     | Electrocardiograph, Lead<br>Switching Adapter                               |
|                                          | \$870.2370, II  | KRC     | Tester, Electrode, Surface,<br>Electrocardiograph                           |
|                                          | \$870.2450, II  | DXJ     | Display, Cathode-Ray Tube,<br>Medical                                       |
|                                          | \$870.2600, I   | DRJ     | System, Signal Isolation                                                    |
|                                          | \$870.2700, II  | DQA     | Oximeter                                                                    |
|                                          | \$870.2770, II  | DSB     | Plethysmograph, Impedance                                                   |
|                                          | \$870.2800, II  | DSH     | Recorder, Magnetic tape,<br>Medical                                         |
|                                          | \$870.2810, I   | DSF     | Recorder, Paper Chart                                                       |
|                                          | \$870.2850, II  | DRS     | Extravascular Blood Pressure<br>Transducer                                  |
|                                          | \$870.2900, I   | DSA     | Cable, Transducer and<br>Electrode, incl. Patient<br>Connector              |

{1}------------------------------------------------

<0428-5
p.2/3

|                                                      |                | MSX | System, Network and<br>Communication, Physiological<br>Monitors              |
|------------------------------------------------------|----------------|-----|------------------------------------------------------------------------------|
| Anesthesiology<br>and Respiratory<br>Therapy (12624) | \$868.1400, II | CCK | Analyzer, Gas, Carbon Dioxide,<br>Gaseous-Phase                              |
|                                                      | \$868.1500, II | CBQ | Analyzer, Gas, Enflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)     |
|                                                      | \$868.1500, II | NHO | Analyzer, Gas, Desflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)    |
|                                                      | \$868.1500, II | NHP | Analyzer, Gas, Sevoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)   |
|                                                      | \$868.1500, II | NHQ | Analyzer, Gas, Isoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)    |
|                                                      | \$868.1620, II | CBS | Analyzer, Gas, Halothane,<br>Gaseous-Phase (Anesthetic<br>Concentration)     |
|                                                      | \$868.1700, II | CBR | Analyzer, Gas, Nitrous Oxide,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
|                                                      | \$868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-<br>Phase                                     |
|                                                      | \$868.2375, II | BZQ | Monitor, Breathing Frequency                                                 |
|                                                      | \$868.2480, II | LKD | Monitor, Carbon Dioxide,<br>Cutaneous                                        |
|                                                      | \$868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for<br>Infant not under Gas Anesthesia           |
| General Hospital<br>and Personal Use<br>(12520)      | \$880.2910, II | FLL | Thermometer, Electronic,<br>Clinical                                         |
| Neurological<br>(12513)                              | \$882.1400, II | GWR | Electroencephalograph                                                        |
|                                                      | \$882.1420, I  | GWS | Analyzer, Spectrum,<br>Electroencephalogram Signal                           |

- 3. The modified devices are substantially equivalent to previously cleared Philips devices marketed pursuant to K014159, K021778, K032858, K040304, K040183, K040259, K040357 and K041235.
- 4. The modification is the introduction of Release B.1.1 software for the IntelliVue patient monitor devices, MP20, MP30, MP40, MP50 interfacing the Gas Analyzer model M1013A EGM (K041956).
- 5. The modified devices have the same intended use as the leqally marketed predicate devices. They are intended for the monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in hospital environment and during transport within hospital environments.
- 6. The modified devices have the same technological characteristics as the legally marketed predicate devices.
- 7. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety

{2}------------------------------------------------

testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue Patient Monitor meets all reliability requirements and performance claims.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized graphic of what appears to be an abstract bird or wing-like design.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 12 2004

Philips Medizin Systeme Böblingen GmbH c/o Mr. Egon Pfeil Sr. Regulatory Affairs Engineer Cardiac and Monitoring Systems Hewlett-Packard-Str. 2 71034 Böblingen GERMANY

Re: K042845

K042643
Trade Name: Philips Medical System MP20, MP30, MP40, MP50, MP60, MP70 and MP90 Intellivue Patient Monitors, Release B.1.1 Regulation Number: 21 CFR 870.2300 Regulation Name: Physiological Monitors Network and Communication System Regulatory Class: Class II (two) Product Code: MSX Dated: October 13, 2004 Received: October 15, 2004

Dear Mr. Pfeil:

We have reviewed your Section 510(k) premarket notification of intent to market the device w & nave rowled your be determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use surfale in the encreated 776, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may a works provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Egon Pfeil

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Bhummar for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use

510(k) Number (if known): Koy28GS

The Philips IntelliVue MP20, MP30, MP40, MP50, MP60, Device Name: MP70 and MP90 Patient Monitors, Release B.1.1

Indicated for use by health care professionals Indications for Use: whenever there is a need for monitoring the physiological parameters patients. Intended for monitoring, recording and alarming of of multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The MP20, MP30, MP40 and MP50 are additionally intended for use in transport situations within hospital environments.

EASI 12-lead ECG is only for use on adult and pediatric patients.

ST Seqment monitoring is restricted to adult patients only.

The transcutaneous gas measurement (tcp02/tcpCO2) is restricted to neonatal patients only.

Prescription Use yes (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use __No___ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Blumensa

Page **_ of _**

---

**Source:** [https://fda-staging.innolitics.com/device/K042845](https://fda-staging.innolitics.com/device/K042845)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com