EBI ARRAY SPINAL FIXATION SYSTEM AND SPINELINK II SPINAL FIXATION DEVICE
Applicant
Ebi, L.P.
Product Code
NKB · Orthopedic
Decision Date
Dec 13, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EBI® Array™ Spinal Fixation System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformities or curvatures (i.e.. scoliosis, kyphosis, and lordosis), tumor, stenosis, psuedoarthrosis, and failed previous fusion. The EBI® VuePASS™ Portal Access System when used with the EBI® Array™ Spinal Fixation System is indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery.
Device Story
System comprises EBI® Array™ or SpineLink® II Spinal Fixation implants (rods, screws, hooks, links) and EBI® VuePASS™ Portal Access Surgical System (retractor system). VuePASS includes stainless steel sequential dilators, radiolucent oval cannulas, and circular tubes. Used in posterior spinal surgery; provides minimally invasive access. Surgeon operates system in clinical setting. VuePASS use limited to specific implant lengths (50mm or less for Array; 36mm or less for SpineLink) and excludes cross connectors/links. Implants provide structural stabilization of spine. Benefits include reduced surgical invasiveness compared to traditional open procedures.
Clinical Evidence
Bench testing only. Verification testing demonstrated successful implantation of the spinal fixation systems utilizing the VuePASS Portal Access Surgical System.
Technological Characteristics
Implants: Titanium alloy (Ti-6Al-4V ELI, ASTM F136) and unalloyed titanium (ASTM F67-00). VuePASS components: Aluminum alloy (ASTM B 221), titanium alloy (ASTM F136), stainless steel (ASTM F899), Radel (ASTM D4181), and silicone. System is a mechanical spinal fixation and retractor set. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients requiring non-cervical spinal fixation for degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), deformities (scoliosis, kyphosis, lordosis), tumor, stenosis, pseudoarthrosis, or failed previous fusion. Used via minimally invasive approach for posterior spinal surgery.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
EBI® SpineLink® II Spinal Fixation System (K012516)
EBI® SpineLink® II Spinal Fixation System (K031355)
Submission Summary (Full Text)
{0}------------------------------------------------
# DEC 1 3 2004
#### 510(k) Summary
This 510(k) Summary for the EBI® Array™ Spinal Fixation System with the EBI® VuePASS™ Portal Access Surgical System is provided as required per Section 513(3) of the Food, Drug and Cosmetic Act.
| 1. Submitter: EBI, L.P.<br>100 Interpace Parkway<br>Parsippany, NJ 07054 | Contact Person: Frederic Testa, RAC<br>Telephone: (973) 299-9300, ext.2208 |
|--------------------------------------------------------------------------|----------------------------------------------------------------------------|
|--------------------------------------------------------------------------|----------------------------------------------------------------------------|
December 2, 2004 Date prepared:
EBI® Array™ Spinal Fixation System with the EBI® 2. Proprietary Name:
| VuePASS™ Portal Access Surgical System | | | | |
|----------------------------------------|--|--|--|--|
|----------------------------------------|--|--|--|--|
| Common Name: | Spinal Fixation Device |
|-----------------------|-------------------------------------------------------------------------------------|
| Classification Names: | Pedicle Screw Spinal System<br>Spinal Interlaminal Fixation Orthosis<br>Arthroscope |
- 3. Predicate or legally marketed devices that are substantially equivalent:
EBI® Array™ Spinal Fixation System (K033312) .
- 4. Description of the device: The EBI® Array™ Spinal Fixation System is a single use spinal fixation device that utilizes rods, screws, and hooks. The EBT® VuePASS™ Portal Access Surgical System is a retractor system for a minimally invasive approach
{1}------------------------------------------------
for posterior spinal surgery. The system includes a series of stainless steel sequential dilators and various sized reusable radiolucent oval cannulas and circular tubes. When used with the EBI® Array™ Spinal Fixation System, the use of the EBI® VuePASS™ Portal Access Surgical System is limited to the implant of a rod length of 50mm or less and excludes the use of system cross connectors or hooks.
- 5. Intended Use: The EBI® Array™ Spinal Fixation System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformities or curvatures (i.e.. scoliosis, kyphosis, and lordosis), tumor, stenosis, psuedoarthrosis, and failed previous fusion.
The EBI® VuePASS™ Portal Access System when used with the EBI® Array™ Spinal Fixation System is indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery.
- 6. Materials: The components of the EBI® Array™ Spinal Fixation System are manufactured from titanium alloy (Ti-6A1-4V ELI) per ASTM F136 and unalloyed titanium per ASTM F 67-00. The components of the EBI® VuePASS™ Portal Access System are manufactured from aluminum alloy as per ASTM B 221, titanium alloy as per ASTM F 136, stainless steel as per ASTM F 899, radel as per ASTM D 4181, and silicone.
{2}------------------------------------------------
- 7. Comparison of the technological characteristics of the device to predicate devices: There are no significant differences between the proposed EBI® Array™ Spinal Fixation System with the EBI® VuePASS™ Portal Access Surgical System and the currently marketed EBI® Array™ Spinal Fixation System. Verification testing demonstrated successful implantation of the EBI® Array™ Spinal Fixation System utilizing the EBI® VuePASS™ Portal Access Surgical System. It is substantially equivalent* to the predicate device in regards to design, intended use, and materials.
<sup>*</sup> Any statement made in conjunction with this submission regarding a determination of substantial equivalence to any * Any statement made in confuler the product can be lawfully marketed without pre-market approval or reclassification and is not intended to be interpreted as an admission or any other type of evidence in patent reclassitication and is not meetised to be interpreted as an actives on any environment final Regulation, Preamble, August 23, 1977, FR 42520 (Docket No. 76N-0355.)]
{3}------------------------------------------------
### 510(k) Summary
This 510(k) Summary for the EB1® SpineLink® II Spinal Fixation System with the EB1® VuePASS™ Portal Access Surgical System is provided as required per Section 513(3) of the Food, Drug and Cosmetic Act.
| 1. Submitter: | EBI, L.P.<br>100 Interpace Parkway<br>Parsippany, NJ 07054 | Contact Person: | Frederic Testa, RAC |
|---------------|------------------------------------------------------------|-----------------|--------------------------|
| | | Telephone: | (973) 299-9300, ext.2208 |
December 2, 2004 Date prepared:
EBI® SpineLink® II Spinal Fixation System with 2. Proprietary Name: the EBI® VuePASS™ Portal Access Surgical System
| Common Name: | Spinal Fixation Device |
|-----------------------|-------------------------------------------------------------------------------------|
| Classification Names: | Pedicle Screw Spinal System<br>Spinal Interlaminal Fixation Orthosis<br>Arthroscope |
## 3. Predicate or legally marketed devices that are substantially equivalent:
| • EBI® SpineLink® II Spinal Fixation System | (K012516) |
|---------------------------------------------|-----------|
|---------------------------------------------|-----------|
- EBI® SpineLink® II Spinal Fixation System (K031355) .
{4}------------------------------------------------
- 4. Description of the device: The EBI® SpineLink® II Spinal Fixation System is a fixation device that utilizes interconnecting links, screws (fixed and polydirectional designs), transverse connectors, hooks, endcaps, locknuts, and link ties. The EBI® VuePASS™ Portal Access Surgical System is a retractor system for a minimally invasive approach for posterior spinal surgery. The system includes a series of stainless steel sequential dilators and various sized reusable radiolucent oval cannulas and circular tubes. When used with the EBI® SpineLink® II Spinal Fixation System, the use of the EBI® VuePASS™ Portal Access Surgical System is limited to the implant of link lengths of 36mm or less and excludes the use of system cross links.
- 4. Intended Use: The EBI® SpineLink® II Spinal Fixation System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformities or curvatures (i.e.. scoliosis, kyphosis, and lordosis), tumor, stenosis, psuedoarthrosis, and failed previous fusion.
The EBI® VuePASS™ Portal Access System when used with the EBI® SpineLink® II Spinal Fixation System is indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery.
- 6. Materials: The components of the EBI® SpineLink® II Spinal Fixation System are manufactured from titanium alloy (Ti-6A1-4V ELI) per ASTM F136. The components of the VuePASS™ Portal Access Surgical System are manufactured from aluminum alloy as per ASTM B 221, titanium alloy as per ASTM F 136, stainless steel as per ASTM F 899, radcl as per ASTM D 4181, and silicone.
{5}------------------------------------------------
- 7. Comparison of the technological characteristics of the device to predicate devices: There are no significant differences between the proposed EBI® SpineLink® II Spinal Fixation System with the EBI® VuePASS™ Portal Access Surgical System and the currently marketed EBI® SpineLink® II Spinal Fixation System. Verification testing demonstrated successful implantation of the EBI® SpineLink® II Spinal Fixation System utilizing the EBI® VuePASS™ Portal Access Surgical System. It is substantially equivalent* to the predicate device in regards to design, intended use, and materials.
<sup>*</sup> Any statement made in conjunction with this submission regarding a determination of substantial equivalence to any other product is intended only to relate the product can be lawfully marketed without pre-market approval or only produce is mended only to readly interpreted as an admission or any other type of evidence in patent infringement litigation. [Establishment Registration and Premarket Notification Procedures, Final Regulation, Preamble, August 23, 1977, FR 42520 (Docket No. 76N-0355.)]
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized eagle with three lines representing its wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# DEC 1 3 2004
Mr. Frederic Testa, RAC Regulatory Affairs Project Manager EBI, L.P. 100 Interpace Parkway Parsippany, New Jersey 07054
Re: K042772
KU42 / /2
Trade/Device Name: EBI® Array Spinal Fixation System with VuePASS™ Portal Access
Crade/Device Name: EBI® Array Spinal Fixation System Fixetion System with EDI Array Spinal Pinaux SpineLink® Spinal Fixation System with VuePASS™ Portal Access Surgical System
Regulation Number: 21 CFR 888.3070, 21 CFR 888.3050 Regulation Name: Pedicle screw spinal interlaminal fixation orthosis, Regulatory Class: III Product Code: NKB, MNH, MNI, KWP, NBH Dated: November 18, 2004 Received: November 19, 2004
Dear Mr. Testa:
We have reviewed your Section 510(k) premarket notification of intent to market the device indication we nave reviewed your becaon 510(x) premainted is substantially equivalent (for the indications ferenced above and nave decembined by marketed predicate devices marketed in interstate for use stated in the encrosule for regars managed posice Americal Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Red Ford Ford. Ford, Day commerce proof to May 20, 1976, the encordance with the provisions of the Federal Food, Drug, devices that have been reclassinod in assess approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I bu may, increrore, market the device, odyde requirements for annual registration, listing of general controls provisions of ractice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (soc aborto) als. Existing major regulations affecting your device can
may be subject to such additional controls. Existing major regulations of may be subject to back acceleral Regulations, Title 21, Parts 800 to 898. In addition, FDA may be found in the Obaouncements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean 1 Toase oc advised that I Drivesantes syour device complies with other requirements of the Act that I Dri has intates and regulations administered by other Federal agencies. You must of any I cueral statutes and regulations and limited to: registration and listing (21 Compy with an the 110 - 11 CFR Part 801); good manufacturing practice requirements as se CI K Fart 6077, naomig (21 CFR Part 820); and if applicable, the electronic form in the quality byections (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2 - Frederic Testa, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) rins letter will and in yourse FDA finding of substantial equivalence of your device to a legally promation notification of results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you attent office of Compliance at (240) 276-__. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small general mistmation on your Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark M. Milkersen
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Statement of Indications for Use
Page 1 of 1
510(k) Number (if known):
Device Name:
·
Device Rainer
EBI® Array Spinal Fixation System with VuePASS™ Portal Access Surgical System
Indications For Use:
The EBI® Array™ Spinal Fixation System is a non-cervical spinal fixation device intended for use as an anterolateral pedicle screw fixation system, a posterior hook and sacral/iliac screw fixations ystem, or as an anterplated for pedicle screw fixation system, a posterior november patients. The device is indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographie studiosis, spondylolisthesis, trauma, (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, spondylorisations)
and lordosis), tumor, stenosis, psuedoarthrosis, and failed previous fusion.
The EBI® VuePASS™ Portal Access System when used with the EBI® Array™ Spinal Firation System is I he EDI Vuel ABS
Prescription Use X (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional Format 1-2-96)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark N. Millhuser
vision of Gen and Neurological Devices
510(k) Number K042772
{9}------------------------------------------------
Page
510(k) Number (if known):
Device Name:
EBI® SpineLink® II Spinal Fixation System with VuePASS™ Portal Access Surgical System
Indications For Use:
The EBI® SpineLink II Spinal Fixation System is a non-cervical spinal fixation device intended for use as an anagerolateral The EBT Spinellilik II Spillar I Ration System.
Pedicle screw fixation system, a posterior hook and sacrabiliates. The device is indicated for pedicle screw fixation system, a posterior nosition. In the patientive disactive is indicated for all of the following indications regardess of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studiosis discogenic back pain with defencialism or me
spondylolisthesis, trauma, (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, spondylonstrations) in the many psuedoarthrosis, and failed previous fusion.
The EBI® VuePASS™ Portal Access System when used with the EBI® Spinal Fixation The EBI * VuePASS * * Foral 766655 byers.
System is indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery.
Prescription Use _ X (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional Format 1-2-96)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for Murle N. Millerson
Neurological Devic
510(k) Number K042772
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.