K042770 · Siemens Medical Solutions USA, Inc. · IYN · Oct 20, 2004 · Radiology
Device Facts
Record ID
K042770
Device Name
MODIFICATION TO ACUSON CV70 CARDIOVASCULAR SYSTEM
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
IYN · Radiology
Decision Date
Oct 20, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The CV70 ultrasound imaging system is intended for the following applications: Abdominal, Intraoperative, Small Parts, Transcranial, OB/GYN, Cardiac, Transesophageal, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
Device Story
The ACUSON CV70 is a mobile, software-controlled diagnostic ultrasound system. It acquires primary or secondary harmonic ultrasound echo data via various transducers (convex, linear, phased sector, laparoscopic, TEE). The system processes these inputs to display images in B-Mode, M-Mode, Pulsed (PW) Doppler, Continuous (CW) Doppler, Color Doppler, Amplitude Doppler, Harmonic Imaging, or 3D imaging on a CRT display. It is operated by healthcare professionals in clinical settings. The system provides measurement and analysis packages for anatomical structures and fluid flow, which clinicians use to support clinical diagnosis. It includes on-screen displays for thermal and mechanical indices to manage potential bioeffects.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing, including acoustic output measurements, biocompatibility (ISO 10993), and compliance with electrical safety and EMC standards (IEC 60601-1 series).
Technological Characteristics
Mobile ultrasound system; supports convex, linear, phased, and TEE transducers. Imaging modes: B, M, PW/CW/Color/Amplitude Doppler, Harmonic, 3D. Standards: UL 2601-1, CSA C22.2 No. 601-1, AIUM/NEMA UD-2/UD-3, ISO 10993, IEC 60601-1/1-1/1-2. Software-controlled architecture.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis of the human body, including abdominal, intraoperative, small parts, transcranial, OB/GYN, cardiac, transesophageal, pelvic, neonatal/adult cephalic, vascular, musculoskeletal, superficial musculoskeletal, and peripheral vascular applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
ACUSON Sequoia Diagnostic Ultrasound System (K041319)
Submission Summary (Full Text)
{0}------------------------------------------------
K042770
### 510(K) SUMMARY
# OCT 2 0 2004
### ACUSON CV70™ Cardiovascular system
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with the Safe Medical Device Act of 1990 revisions to 21 CFR, Part 807.92, Content and Format of a 510(k) Summary.
### Submitted By: 1.
Siemens Medical Solutions USA, Inc., Ultrasound Division 22010 S.E. 51st Street Issaquah, WA 98029
### Contact Person:
Patrick Lynch Regulatory Affairs
Phone: (425) 557-1825 FAX: (425) 391-9198
### Date Prepared:
September 3, 2004
### 2. Proprietary Name:
ACUSON CV70™ Cardiovascular System
### Common/ Usual Name:
Diagnostic Ultrasound System with Accessories
### Classification Name:
| 21 CFR 892.1550 | | |
|------------------------------------------|---------------|---------------------|
| Ultrasonic Pulsed Doppler Imaging System | FR # 892.1550 | Product Code 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | FR # 892.1560 | Product Code 90-IYO |
| Diagnostic Ultrasound Transducer | FR # 892.1570 | Product Code 90-ITX |
### 3. Predicate Device:
K032111, 7/18/2003, cleared as ACUSON CV70 Cardiovascular System, K041319, 6/7/2004, cleared as ACUSON Sequoia™ Diagnostic Ultrasound System.
### 4. Device Description:
The CV70 is a general purpose, mobile, software-controlled, diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bioeffect mechanisms. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in: B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, or Harmonic Imaging, or 3D imaging, on a CRT display.
The CV70 has been designed to meet the following product safety standards:
- UL 2601-1, Safety Requirements for Medical Equipment
- 트 CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment
- 용 AIUM/NEMA UD-3, 1998 Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- AIUM/NEMA UD-2, 1998 Acoustic Output Measurement Standard for Diagnostic Ultrasound
- 동 93/42/EEC Medical Devices Directive
{1}------------------------------------------------
- 트 Safety and EMC Requirements for Medical Equipment
- 트 EN/IEC 60601-1
- . EN/IEC 60601-1-1
- 포 EN/IEC 60601-1-2
- IEC 1157 Declaration of Acoustic Power
- 트 ISO 10993 Biocompatibility
### 5. Intended Uses:
The CV70 ultrasound imaging system is intended for the following applications: Abdominal, Intraoperative, Small Parts, Transcranial, OB/GYN, Cardiac, Transesophageal, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
### 6. Technological Comparison to Predicate Device:
The CV70 is substantially equivalent to the ACUSON CV70, cleared via K032111; and some features of the ACUSON Sequoia, cleared via K041319. All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. All systems allow for specialized measurements of structures and flow, and calculations.
### End of 510(k) Summary
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or bird.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 2 0 2004
Siemens Medical Solutions USA, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC · 1394 25th Street NW BUFFALO MN 55313
Re: K042770
Trade Name: Acuson CV70 Cardiovascular System
Regulation Number: 21 CFR 892.1550
- Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560
Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570
- Regulation Name: Diagnostic ultrasonic transducer
- Regulatory Class: II
- Product Code: 90 IYN, IYO, and ITX
- Dated: September 30, 2004 Received: October 5, 2004
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Acuson CV70 Cardiovascular System, as described in your premarket notification:
### Transducer Model Number
C5-2 Convex Array C6-2 Convex Array C8-5 Convex Array
5.0C50+ Convex Array 5.0L45 Linear Array 7.5L70 Linear Array
{3}------------------------------------------------
LB5-2 Linear Array L10-5 Linear Array VF13-5 Linear Array VF13-5SP Linear Array 7.5L50I Linear Array 7.5L50Q Linear Array LAP8-4 Laparoscopic
P4-2 Phased Sector Array 5.0P10 Phased Sector Array V5Ms Phased Sector Array TEE CW2 Continuus Wave Doppler CW5 Continuous Wave Doppler P9-4 Phased Sector Array
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), in it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the
{4}------------------------------------------------
Page 2 - Mr. Job
promotion and advertising of your device, please contact the Office of Compliance at (301) 594promotion and advertising of your ac-roos, please and the by reference to premarket 4039. Also, please note the regulation one and information on your responsibilities under the notification (21 CFA for 80777). Only Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address > "http://www.fda.gov/cdrh/dsmamain.html".
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
Nancy C brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
510(k) Number (if known):
### ACUSON CV70 Cardiovascular System
Device Name:
Intended Use:
Diagnostic imaging or fluid flow analysis of the human body as follows: Intended Use: Mode of Operation Color Clinical Application Color Amplitude Combined Other Velocity PWD CWD в M A (Specify) (Specify) Doppler Doppler Imaging Ophthalmic Note 2,3,4,5 P BMDC P P P P P Fetal BMDC Note 2,3,4,5 P P P P P P Abdominal Note 3 Intraoperative P BMDC P P P P (Note 6) Intraoperative P P BMOC Note 3 P P P Neurological Note 2,3,4,5 P P BMDC P P P P Pediatric Small Organ P P BMDC Note 2,3,4,5 P P P P (Note 1) P p BMDC Note 2,3,4,5 p P P P Neonatal Cephalic P P P BMDC Note 2,3 P P P Adult Cephalic Note 2,3,4,5,7 P P BMDC P P P P Cardiac Note 2,3 BMDC p P P P P P Transesophagea! Transrectal Transvaginal Transurethral Intravascular BMDC Note 2,3,4,5 P P P P P P Peripheral vessel BMDC Note 3 P P P P P Laparoscopic Musculo-skeletal Note 2,3,4,5 P BMDC P P P P P Conventional Musculo-skeletal Note 2,3,4,5 P P P BMDC P P P Superficial Other (specify)
N = new indication; P = previously cleared by FDA; E = added under Appendix E
For example: breast, testes, thyroid, penis, prostate, etc. Note 1
Ensemble tissue harmonic imaging Note 2
3D imaging Note 3
B&W SieScape panoramic imaging Note 4
Power SieScape panoramic imaging Note 5
For example: abdominal, vascular Note 6
Note 7 Contrast agent imaging
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
> > Prescription Use (Per 21 C
Nancy Creador
(Division Sign-Off)
yngdo
Division of Reproductive, Abdominal,
and Radiological Devices K042770
510(k) Number
{6}------------------------------------------------
510(k) Number (if known):
Device Name:
Intended Use:
### C5-2 Convex Array Transducer for use with: ACUSON CV70 Cardiovascular System Diagnostic imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Abdominal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Small Organ | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethrai | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K032111); E = added under Appendix E
Note 2 Ensemble tissue harmonic imaging
Note 3 3D imaging
Note 4 B&W SieScape panoramic imaging
Power SieScape panoramic imaging Note 5
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Dana C. Bergsten
(Division Sigi Division of Renrodi ive, Abdominal, and Radiological Devices 510(k) Number
{7}------------------------------------------------
510(k) Number (if known):
Device Name:
Intended Use:
C6-2 Convex Array Transducer for use with: ACUSON CV70 Cardiovascular System Ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation Color Clinical Application Combined Other Amplitude Color Velocity CWD A B M PWD (Specify) Doppler Doppler (Specify) Imaging Ophthalmic P P P BMDC Note 2,3,4,5 P P Fetal P BMDC Note 2,3,4,5 P P P P Abdominal Intraoperative Abdominal Intraoperative Neurological P Note 2,3,4,5 p P BMDC Pediatric P P Small Organ Neonatal Cephalic Adult Cephalic Cardiac Trans-esophageal Transrectal Transvaginal Transurethral Intravascular P P P P ြ BMDC Note 2,3,4,5 Peripheral vessel Laparoscopic Musculo-skeletai Conventional Musculo-skeletai Superficial Other (specify)
N = new indication; P = previously cleared by FDA (K032111); E = added under Appendix E
Ensemble tissue harmonic imaging Note 2
Note 3 3D imaging
B&W SieScape panoramic imaging Note 4
Power SieScape panoramic imaging Note 5
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Thury C. Burdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K042770
{8}------------------------------------------------
510(k) Number (if known):
Device Name:
Intended Use:
C8-5 Convex Array Transducer for use with: ACUSON CV70 Cardiovascular System ACUSON CV70 Cardiovascular System
Diagnostic imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Abdominal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Small Organ<br>(Note 1) | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Neonatal Cephalic | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Adult Cephalic | | E | E | E | | E | E | | BMDC | Note 2,3,4,5 |
| Cardiac | | E | E | E | | E | E | | BMDC | Note 2,3,4,5 |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Musculo-skeletal<br>Superficial | | E | E | E | | E | E | | BMDC | Note 2,3,4,5 |
| | | | | | | | | | | |
| Other (specify) (Society)
N = new indication; P = previously cleared by FDA (K032111); E = added under Appendix E
For example: breast, testes, thyroid, penis, prostate, etc. Note 1
Ensemble tissue harmonic imaging Note 2
3D imaging Note 3
B&W SieScape panoramic imaging Note 4
Power SieScape panoramic imaging Note 5
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> THE WRITE BELOW THIS LINE OF CONTINE Eveluation (ODE) I WRITE BELOW THIS EINE SONNINGE Evaluation (ODE)
Prescription Use (Per 21)CFR 801.109 ance (Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________
{9}------------------------------------------------
510(k) Number (if known):
Device Name:
Intended Use:
5.0C50+ Convex Array Transducer for use with: ACUSON CV70 Cardiovascular System ACOSON OVT o Gardlovacious. System of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5 |
| Abdominal | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5 |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5 |
| Small Organ<br>(Note 1) | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5 |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | E | E | E | E | E | E | | BMDC | Note 2,3,4,5 |
| Musculo-skeletal<br>Superficial | | E | E | E | E | E | E | | BMDC | Note 2,3,4,5 |
| Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K032111); E = added under Appendix E
For example: breast, testes, thyroid, penis, prostate, etc. Note 1
Ensemble tissue harmonic imaging Note 2
3D imaging Note 3
B&W SieScape panoramic imaging Note 4
Power SieScape panoramic imaging Note 5
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Hogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
Radiological Devices
K Number K042770
{10}------------------------------------------------
510(k) Number (if known):
Device Name:
Intended Use:
5.0L45 Linear Array Transducer for use with: ACUSON CV70 Cardiovascular System ACUSON CV70 Cardiovascular Oystern
Diagnostic imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5 |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Note 1) | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5 |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5 |
| Musculo-skeletaf<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K032111); E = added under Appendix E
For example: breast, testes, thyroid, penis, prostate, etc. Note 1
Ensemble tissue harmonic imaging Note 2
3D imaging Note 3
B&W SieScape panoramic imaging Note 4
Power SieScape panoramic imaging Note 5
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Darrey C. Brandon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K042770
{11}------------------------------------------------
510(k) Number (if known):
Device Name:
### 7.5L70 Linear Array Transducer for use with: ACUSON CV70 Cardiovascular System
Intended Use:
Diagnostic imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | P | | BMDC | Note 3,4,5 |
| Small Organ<br>(Note 1) | | P | P | P | P | P | P | | BMDC | Note 3,4,5 |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | E | E | E | | E | E | | BMDC | Note 3,4,5 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC | Note 3,4,5 |
| Musculo-skeletal<br>Superficial | | P | P | P | | P | P | | BMDC | Note 3,4,5 |
| Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K032111); E = added under Appendix E
For example: breast, testes, thyroid, penis, prostate, etc. Note 1
Note 3 3D imaging
B&W SieScape panoramic imaging Note 4
Power SieScape panoramic imaging Note 5
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Broadon
(Division Sign=Off) Division of Reproduct ve. Abdon and Radiological Devices 510(k) Number
{12}------------------------------------------------
510(k) Number (if known):
Device Name:
Intended Use:
### LB5-2 Linear Array Transducer for use with: ACUSON CV70 Cardiovascular System ACUSON OVT o Garanovaooan. Of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC | Note 4,5 |
| Abdominal | | P | P | P | | P | P | | BMDC | Note 4,5 |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | |
| Conventional<br>Musculo-skeletal | | | | | | | | | | |
| Superficial<br>Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K032111); E = added under Appendix E
B&W SieScape panoramic imaging Note 4
Power SieScape panoramic imaging Note 5
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices K042770
510(k) Number
{13}------------------------------------------------
510(k) Number (if known):
Device Name:
# L10-5 Linear Array Transducer for use with:
Intended Use:
ACUSON CV70 Cardiovascular System Ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Small Organ<br>(Note 1) | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Musculo-skeletal<br>Superficial | | P | P | P | | P | P | | BMDC | Note 2,3,4,5 |
| Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K032111); E = added under Appendix E
For example: breast, testes, thyroid, penis, prostate, etc. Note 1
Ensemble tissue harmonic imaging Note 2
Note 3 3D imaging
B&W SieScape panoramic imaging Note 4
Note 5 Power SieScape panoramic imaging
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Brogdon
(Division Sign-Off)
Division of Reproductive and Radiological Device 510(k) Number
{14}------------------------------------------------
510(k) Number (if known):
Device Name:
Intended Use:
VF13-5 Linear Array Transducer for use with: ACUSON CV70 Cardiovascular System ACUSON CV70 Gardiovusounar Oyetshof the human body as follows:
| | | Mode of Operation | | | | | | | | | |
|----------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative<br>Abdominal | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | P | | BMDC | Note 3,4,5 | |
| Small Organ<br>(Note 1) | | P | P | P | P | P | P | | BMDC | Note 3,4,5 | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral vessel | | P | P | P | P | P | P | | BMDC | Note 3,4,5 | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | P | P | P | | BMDC | Note 3,4,5 | |
| Musculo-skeletal<br>Superficial | | P | P | P | P | P | P | | BMDC | Note 3,4,5 | |
| Other (specify) | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K032111); E = added under Appendix E
For example: breast, testes, thyroid, penis, prostate, etc. Note 1
Note 3 3D imaging
B&W SieScape panoramic imaging Note 4
Power SieScape panoramic imaging Note 5
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
> > Prescription Use (Per 21 CFR 801.109)
Nancy C. Brogdon
(Division Sign Off) Division of Reproductive, A and Radiological Devices 510(k) Number _
{15}------------------------------------------------
510(k) Number (if known):
Device Name:
VF13-5SP Linear Array Transducer for use with: ACUSON CV70 Cardiovascular System
Intended Use:
# Diagnostic imaging or fluid flow analysis of the human body as follows:
| | | Mode of Operation | | | | | | | | |
|----------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(note 6) | | P | P | P | | P | P | | BMDC | Note 3 |
| Intraoperative<br>Neurological | | P | P | P | | P | P | | BMDC | Note 3 |
| Pediatric | | P | P | P | | P | P | | BMDC | Note 3 |
| Small Organ<br>(Note 1) | | P | P | P | | P | P | | BMDC | Note 3 |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC | Note 3 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC | Note 3 |
| Musculo-skeletal<br>Superficial | | P | P | P | | P | P | | BMDC | Note 3 |
| Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K032111); E = added under Appendix E
For example: breast, testes, thyroid, penis, prostate, etc. Note 1
Note 3 3D imaging
For example: abdominal, vascular Note 6
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Thornton
(Division Sign-Off)
Division of Reproductive, Ab and Radiological Devices 510(k) Number _
{16}------------------------------------------------
510(k) Number (if known):
Device Name:
### 7.5L501 Linear Array Transducer for use with: ACUSON CV70 Cardiovascular System
Intended Use:
Diagnostic imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | P | P | P | | P | P | | BMDC | Note 3,4,5 |
| Intraoperative<br>(Note 6) | | P | P | P | | P | P | | BMDC | Note 3,4,5 |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Note 1) | | P | P | P | | P | P | | BMDC | Note 3,4,5 |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC | Note 3,4,5 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC | Note 3,4,5 |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K032111); E = added under Appendix E
For example: breast, testes, thyroid, penis, prostate, etc. Note 1
- 3D imaging Note 3
Note 4 B&W SieScape panoramic imaging
Power SieScape panoramic imaging Note 5
For example: abdominal, vascular Note 6
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Vance Eric Brandon
(Division Sign Division of Reproductive, Abdor and Radiological Devices 510(k) Number _
{17}------------------------------------------------
510(k) Number (if known):
Device Name:
### 7.5L50Q Linear Array Transducer for use with: ACUSON CV70 Cardiovascular System AGUSON OVTO Ourorovation of the human body as follows:
Intended Use:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | P | P | P | | P | P | | BMDC | Note 3,4,5 |
| Intraoperative<br>(Note 6) | | P | P | P | | P | P | | BMDC | Note 3,4,5 |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Note 1) | | P | P | P | | P | P | | BMDC | Note 3,4,5 |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC | Note 3,4,5 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC | Note 3,4,5 |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K032111); E = added under Appendix E
For example: breast, testes, thyroid, penis, prostate, etc. Note 1
3D imaging Note 3
B&W SieScape panoramic imaging Note 4
Power SieScape panoramic imaging Note 5
For example: abdominal, vascular Note 6
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Brogdon
Division S Division of Reproductive, Abdominal Badiological Devices 's ' )(k) Number
{18}------------------------------------------------
510(k) Number (if known):
Device Name:
# LAP8-4 Laparoscopic Transducer for use with:
Intended Use:
ACUSON CV70 Cardiovascular System Diagnostic imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(Note 6) | | P | P | P | | P | P | | BMDC | Note 3,4,5 |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | P | P | P | | P | P | | BMDC | Note 3,4,5 |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K032111); E = added under Appendix E
3D imaging Note 3
B&W SieScape panoramic imaging Note 4
Note 5 Power SieScape panoramic imaging
For example: abdominal, vascular Note 6
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Houghton
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K042770
{19}------------------------------------------------
510(k) Number (if known):
Device Name:
### P4-2 Phased Sector Array Transducer for use with: ACUSON CV70 Cardiovascular System
Intended Use:
Diagnostic imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | P | P | P | P | P | P | | BMDC | Note 2,3 |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | P | P | P | P | P | P | | BMDC | Note 2,3 |
| Cardiac | | P | P | P | P | P | P | | BMDC | Note 2,3,7 |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.