QC-BEADS

K042750 · Bioscreen, Inc. · NRF · Oct 29, 2004 · Hematology

Device Facts

Record IDK042750
Device NameQC-BEADS
ApplicantBioscreen, Inc.
Product CodeNRF · Hematology
Decision DateOct 29, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.8625
Device ClassClass 2

Indications for Use

QC-Beads™ are intended for use as controls in counting sperm manually using a microscope or automatically using computer assisted semen analysis instruments.

Device Story

QC-Beads™ are quality control materials for sperm concentration analysis; consist of 3.7-μm polystyrene latex beads suspended in glycerol/detergent/water solution with 0.1% sodium azide preservative. Supplied in two concentration levels (Lo: 16-24 million/ml; Hi: 34-46 million/ml). Used by laboratory technicians with manual microscopy (hemacytometers/Makler chambers) or computer-assisted semen analysis (CASA) instruments to verify counting accuracy. Provides a stable, non-biological alternative to human semen samples for daily instrument/method validation.

Clinical Evidence

No clinical data. Bench testing only. Precision/reproducibility evaluated by counting duplicate aliquots of Hi and Lo beads 10 times using a Neubauer hemacytometer and bright-field microscope; results showed CV of 9-11%. Stability testing confirmed at least one-year shelf life at room temperature.

Technological Characteristics

Polystyrene latex beads (3.7-μm diameter) in glycerol/detergent/water solution; 0.1% sodium azide preservative. HDPE bottle packaging. Manual or automated optical counting principle. Room temperature storage.

Indications for Use

Indicated for use as quality control material for sperm counting procedures, including manual microscopic counting and automated computer-assisted semen analysis (CASA) systems.

Regulatory Classification

Identification

A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).

Special Controls

*Classification.* Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K042750 B. Purpose for Submission: New Device C. Measurand: Quality control material for sperm counts D. Type of Test: Quantitative E. Applicant: Bioscreen, Inc. F. Proprietary and Established Names: QC-Bead™ G. Regulatory Information: 1. Regulation section: 21 CFR 864.8625, Hematology Quality Control Mixture 2. Classification: Class II 3. Product code: NRF, Material, Quality Control, Semen Analysis 4. Panel: Hematology (81) {1} H. Intended Use: 1. Intended use(s): QC-Beads™ are intended for use as controls in counting sperm manually using a microscope or automatically using computer assisted semen analysis instruments. 2. Indication(s) for use: QC-Beads™ are intended for use as controls in counting sperm manually using a microscope or automatically using computer assisted semen analysis instruments. 3. Special conditions for use statement(s): Not applicable. 4. Special instrument requirements: QC-Beads™ are used in conjunction with single-use cell-counting chambers, the Makler Chamber, hemacytometers or similar cell-counting devices for sperm concentration. I. Device Description: QC-Beads™ consists of polystyrene latex beads of 3.7-μm diameter suspended in a solution of glycerol and detergent in water with 0.1% sodium azide added as a preservative. QC-Beads™ is supplied in two levels, one 4.0-ml bottle per level per box. The concentration of beads at each level is in the physiological range concentration of sperm in semen. According to the World Health Organization (WHO), the Lo QC-Beads™ fall in the range around the reference value of sperm concentration which is 20 million or more sperm/ml. The Hi QC-Beads™ fall in the range above the reference range value of sperm concentration. QC-Beads™ are packaged in 4-ml narrow-mouth bottles composed of high density polyethylene (HDPE) that are manufactured by Nalgene. J. Substantial Equivalence Information: 1. Predicate device name(s): QwickCheck-beads™ AQC™ Sperm Count 2. Predicate 510(k) number(s): {2} K051600, K041240 3. Comparison with predicate: | Similarities | | | | | --- | --- | --- | --- | | Item | Device | Predicate | Predicate | | | QC-Beads™ | QwickCheck-beads™ | AQCTM Sperm Count | | Intended Use | Used as quality control for sperm count | Same | Same | | Design | Two levels | Same | Same | | Differences | | | | | --- | --- | --- | --- | | Item | Device | Predicate | Predicate | | Materials | Liquid suspension containing latex particles. | Same | Liquid human semen matrix and stabilized human sperm cells. | | Storage | Room temperature | Same | 2-8°C | | Expected value | Hi QC-Beads™ = 34-46 million beads/ml Lo QC-Beads™ = 16-24 million beads/ml | High Level @ 40 million beads/ml Low Level @ 20 million beads/ml | Level One = 0-30 million/ ml Level Two = 30-75 million/ml | K. Standard/Guidance Document Referenced (if applicable): World Health Organization, 1992, 1999. WHO laboratory Manual for the examination of human semen and sperm-cervical mucus interaction, Cambridge University Press. L. Test Principle: Not Applicable. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: {3} Duplicate aliquots of the Hi and Lo QC Beads were counted 10 times using a Neubauer hemacytometer and a bright-field microscope. The results of the counts were within the expected range with a CV of 9-11% b. Linearity/assay reportable range: Not Applicable. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Stability: Duplicate bottles of the Hi and Lo QC Beads were stored at room temperature for 14 months and demonstrated a stability of at least one year when stored at room temperature. Expected Range: Lo QC Beads- 16-24 million beads/ml Hi QC Beads- 34-46 million beads/ml d. Detection limit: Not Applicable. e. Analytical specificity: Not Applicable. f. Assay cut-off: Not Applicable. 2. Comparison studies: a. Method comparison with predicate device: Not Applicable. b. Matrix comparison: Not Applicable. 3. Clinical studies: a. Clinical Sensitivity: Not Applicable. {4} b. Clinical specificity: Not Applicable. c. Other clinical supportive data (when a. and b. are not applicable): 4. Clinical cut-off: Not Applicable. 5. Expected values/Reference range: The concentration of beads at each level is in the physiological range of concentration of sperm to semen. According to the World Health Organization (WHO), the Lo QC-Beads™ fall in the range around the reference value of sperm concentration which is 20 million or more sperm/ml. The Hi QC-Beads™ fall in the range above the reference value of sperm concentration. Lo QC-Beads™ = 16-24 million beads/ml Hi QC-Beads™ = 34-46 million beads/ml N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...