HEIDELBERG RETINA TOMOGRAPH II / ROSTOCK CORNEA MODULE (HRT II / RCM)
Applicant
Heidelberg Engineering
Product Code
MXK · Ophthalmic
Decision Date
Oct 13, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1850
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Imaging and observation of the cellular layers of the anterior segment of the human eye, for the diagnosis and management of anterior segment diseases.
Device Story
The HRT II / RCM is a confocal laser scanning microscope used for imaging the anterior segment of the eye. It utilizes a 670 nm Class I diode laser to scan the cornea; light reflected or backscattered from the cornea is detected via a confocal optical arrangement. The device captures two-dimensional optical sections; by shifting the focal plane, it generates a sequence of images to create a three-dimensional representation of corneal layers. The system includes a laser scanning camera, Rostock Cornea Module optics, a mount with headrest, and a PC. A sterile, single-use PMMA plate provides corneal contact during examination. The device is operated by a clinician in a clinical setting. Images are displayed on a monitor, stored on a hard drive, and analyzed using Heidelberg Eye Explorer software, which includes cell density measurement tools. The output assists clinicians in diagnosing and managing anterior segment diseases by providing high-resolution, layer-by-layer visualization of corneal structures.
Clinical Evidence
Bench testing only. The device was tested for compliance with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (electromagnetic compatibility). Laser safety was verified per 21 CFR 1040.10 and IEC 60825-1:1993+A2:2001. No clinical trial data was required for this 510(k) clearance.
Technological Characteristics
Confocal laser scanning microscope; 670 nm Class I diode laser source; resonant and galvanometric scanning motors; avalanche photodiode detector; 63x water immersion lens; 2-micron lateral resolution; 4-micron depth resolution; sterile single-use PMMA contact plate; PC-based image acquisition and storage; Heidelberg Eye Explorer software; connectivity via PC hard drive/RAM.
Indications for Use
Indicated for imaging and observation of cellular layers of the anterior segment of the human eye for diagnosis and management of anterior segment diseases in patients requiring such examination.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
JE 1 3 21114
## 510(k) Summary of Safety and Effectiveness
## Manufacturer and Submitter
| Company Name: | Heidelberg Engineering GmbH |
|------------------------|--------------------------------------------------------|
| Company Address: | Gerhart-Hauptmann-Strasse 30 |
| | 69221 Dossenheim, Germany |
| | phone: +49 / 6221 / 64 643 0 |
| | fax: +49 / 6221 / 64 63 62 |
| Contact Person: | Dr. Gerhard Zinser |
| Date Summary Prepared: | August 30, 2004 |
| Device | |
| Trade/Device Name: | Heidelberg Retina Tomograph II / Rostock Cornea Module |
| | (HRT II / RCM) |
| Common/Usual Name: | Corneal Microscope |
| Classification Name: | AC-powered Slit-Lamp Biomicroscope |
| Regulation Number: | 21 CR 886.1850 |
| Product Code: | MXK |
| Classification Panel: | Ophthalmic |
### Substantial Equivalence
Classification Panel:
Classification:
The HRT II / RCM is substantially equivalent to a slit-lamp biomicroscope, a preamend-The FIXT 11 / RONA is buondinations of Secon 2" device, a Class II device that received 510(k) market clearance on October 26, 2001, document control number K012416,
Class II device
#### Device Description
The system consists of the HRT II laser scanning camera, the Rostock Cornea Module optics, a mount with headrest, a power supply, and a personal computer. For an examinaoptivs, a modily with nead sollule optics is placed on the laser scaming camera, positioned in front of the patient's eye, in contact with the cornea. The contact surface is a sterile PMMA plate. Images of the anterior segment are acquired, displayed on the comsterner monitor and stored on the hard drive of the computer from where the images can be retrieved at a later time.
Iceric ved at a faror thire. the cornea of the eye under examination and detecting the light reflected or backscattered from the cornea. The light detection system has a confocal optical arrangement, scatched nom all depth of focus. In consequence a two-dimensional image acquired windness in sincer as an optical section through the corneal layer located at the current Can be considered as an option result in addition, by moving the focal plane step by step, a sequence of section images at different locations of the focal plane can be acquired. This sequence of seculor mages a layer-by-layer three-dimensional image. The acquired im-Sequence of mages provides a the computer monitor and stored on the computer's hard disk for later use.
{1}------------------------------------------------
## Intended Use
The intended use of the Heidelberg Retina Tomograph II / Rostock Cornea Module is to The intended use of the netotoorg results + cannogior ======================================================================================================================== of the anterior segment.
| Comparison of similarities and differences | | | |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| | HRT II / RCM | Nidek Confoscan 2 * | Slit-Lamp Biomicro-<br>scope |
| Indications for use | Imaging and observation<br>of the cellular layers of<br>the anterior segment of<br>the human eye, for the<br>diagnosis and manage-<br>ment of anterior segment<br>diseases. | As a diagnostic tool for<br>observation of the cell<br>layers of the anterior<br>parts of the eye. | Imaging and observation<br>of the anterior segment<br>of the eye for diagnostic<br>purposes. |
| Corneal contact | Yes. Polymethyl-<br>methacrylate (PMMA)<br>plate. | Not in normal use; how-<br>ever, corneal contact<br>with the microscope lens<br>front surface can occur. | No |
| Working distance cornea<br>to objective | 0 (in contact) | 1.98 mm | ca. 50 mm |
| Corneal contact sensing<br>and warning feature | Not applicable. | Yes | Not applicable. |
| Pre-sterilized contact<br>surface | Yes. Single use PMMA<br>plate. | No. Microscope lens<br>disinfected according to<br>instructions for use. | Not applicable. |
| Front surface area | 75 mm2 (PMMA plate) | 16.6 mm2 (microscope<br>lens) | Not applicable. |
| Focus | Manually variable. | Fixed | Not applicable. |
| Focus adjustment range | 1 mm | Not applicable. | Not applicable. |
| Adjustment direction | Device is adjusted hori-<br>zontally while the patient<br>is sitting straight in front<br>of the device. | Device is adjusted hori-<br>zontally while the patient<br>is sitting straight in front<br>of the device. | Device is adjusted hori-<br>zontally while the patient<br>is sitting straight in front<br>of the device. |
| Working position | Horizontal | Horizontal | Horizontal |
| Optical setup | Confocal microscope | Slit-confocal microscope | Conventional micro-<br>scope. |
| Type of scanning aper-<br>ture | Point | Slit | Not applicable. |
| Scanning means | Resonant and galvano-<br>metric scanning motors. | Stepper motor. | Not applicable. |
| Light source | Class I diode laser, 670<br>nm wavelength. | Halogen lamp. | Tungsten lamp (typical). |
| Microscope lens | 63x water immersion | 40x water immersion | Not applicable. |
| Lateral optical resolution | 2 microns | 1 micron | ca. 10 microns |
| Optical depth resolution | 4 microns | 10 microns | ca. 10 microns |
| Detector | Avalanche photodiode. | CCD camera, mono-<br>chrome. | Examiner observes im-<br>age through an ocular, or<br>photographic film, or<br>CCD camera |
| Lateral field of view | 0.4 mm x 0.4 mm | Not known; approxi-<br>mately 0.4 mm x 0.4 mm | Up to 8 mm. |
| Lateral digital resolution | 1 micron/pixel | (Not known.) | Not applicable. |
| Digital depth resolution | 2 micron/pixel | Not applicable. | Not applicable. |
| Image acquisition time | 24 msec | (Not known.) | Not applicable. |
| Acquisition of three-<br>dimensional images | Yes | No | No |
| Microscope lens magni-<br>fication | 63x | 40x | ca. 16x |
| Magnification on screen<br>(15", 1024x768 pixels) | 300x | 500x | ca. 100x |
| Image storage | Directly into PC RAM,<br>then to PC hard drive. | Directly into PC RAM,<br>then to PC hard drive. | Not applicable. |
| Image compression<br>method | JPG, loss-free | (Not known.) | Not applicable. |
| Cell density measure-<br>ment software | Yes | Yes | No |
| Corneal profile meas-<br>urement | Optical sectioning, by<br>shifting the focal plane<br>of the confocal imaging<br>system. | Optical sectioning, by<br>recording of average<br>scatter from different<br>depth locations. | No |
| Operating and image<br>management software | Heidelberg Eye Explorer | NAVIS | Not applicable. |
| Physical layout | Lift table, PC, mount<br>with headrest, optical<br>head are separate com-<br>ponents. | Lift table, PC, and opti-<br>cal head integrated into<br>one piece. | Lift table, mount with<br>headrest, optical head and<br>separate components. |
# Technological Characteristics Compared to Predicate Device
{2}------------------------------------------------
K042742
# 510(k) Summary of Safety and Effectiveness
# Conclusions from Performance Testing
The HRT II / RCM has been tested according to IEC 60601-1 and EC 60601-1-2, and I he HKT 117 KCM has becirements. The system is a laser product of Class 1 according to
was found to meet all requirements. The system is a laser product of Class 1 according was found to moot an requent 1040.10 and IEC 60825-1:1993+A2:2001.
21 CFR Parl 1040 Section 1040.10 and ESC 80023 11491 images resulted in substantial I he evaluation of the devices with respect to intended use, technological characteristics, and safety and effectiveness.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines representing its wings or feathers. The bird is positioned above a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 3 2004
Heidelberg Engineering GmbH c/o TUV America, Inc. Mr. Stefan Preiss 1775 Old Highway 8 New Brighton, MN 55112-1891
Re: K042742
: K042742
Trade/Device Name: Heidelberg Retina Tomograph II / Rostock Cornea Module (HRT II / RCM) Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered Slit-Lamp Biomicroscope Regulatory Class: Class II Product Code: MXK Dated: September 30, 2004 Received: October 4, 2004
Dear Mr. Preiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 310(x) premainter is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the Chelosure) to regars actiment date of the Medical Device Amendments, or to commerce prior to May 26, 1776, the encordance with the provisions of the Federal Food, Drug, devices that have been reclassined in accessfire approval of a premarket approval application (PMA). and Cosment AC. (Act) that do not require appro the general controls provisions of the Act. The You may, merciole, market the devices, courses, coursements for annual registration, listing of general controls provisions of the 110 110 110 1100 1100 bitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) ins. Existing major regulations affecting your device can may be subject to such additional controller Linestigations, Title 21, Parts 800 to 898. In addition, FDA may be found in the Code of Peachar Ing your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I Dri- 3 issuates or wouldevice complies with other requirements of the Act that TDA has made a determinations administered by other Federal agencies. You must of any Federal statutes and regularions and limited to: registration and listing (21 Comply with an the 71ct 31equirements, need 801); good manufacturing practice requirements as set CTK I at 807); aboning (21 CFR Part 820); and if applicable, the electronic form in the quality systems (QD) regulations (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Palgi Korenthol
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
U.S. Food and Drug Administration
E
Image /page/5/Picture/2 description: The image shows the logo for the Department of Health and Human Services. The logo consists of an emblem on the left and the text "Department of Health and Human Services" on the right. The emblem appears to be an eagle or similar bird-like figure. The text is in a sans-serif font and is stacked in three lines.
FDA Home Page | CDRH Home Page | Search | CDRH A-Z Index | Contact CDRH
# Indications for Use
510(k) Number (if known): KO4Z742
Device Name:
Heidelberg Retina Tomograph "" / Rostock Cornea Module
Indications for Use: Imaging and observation of the cellular layers of the anterior segment of the human eye, for the diagnosis and management of anterior segment diseases.
Evenette T. Beem
510(k) Number K042742
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
ア
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF
NEWS (PLE NEW NEEDED) NEEDED)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.