The Quest Diagnostics Hair Check-DT (Opiates) is a test system that utilizes the IDS One-Step ELISA Opiates Kit for the qualitative detection of opiates at or above 500 pg/mg in head hair samples from chronic heroin users. This test system has not been evaluated for use with other user populations or with hair specimens other than the head. It is an in vitro diagnostic device intended exclusively for in-house professional use only and is not intended for sale to anyone. The QUEST DIAGNOSTICS HairCheck-DT (Opiates) provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed result. Gas Chromatograph - Mass Spectrometry operating in the selected ion monitoring (SIM) mode or GC/MS/MS in selected reaction mode (SRM) is the preferred method with deuterated internal standards. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.
Device Story
The Quest Diagnostics HairCheck-DT (Opiates) is an in vitro diagnostic test system for professional laboratory use. It utilizes the IDS One-Step ELISA Opiates Kit to analyze head hair samples from chronic heroin users. The device performs a qualitative screening to detect opiates at or above a 500 pg/mg threshold. As a preliminary analytical test, it requires confirmation via a more specific chemical method, such as Gas Chromatography-Mass Spectrometry (GC/MS) in SIM mode or GC/MS/MS in SRM mode using deuterated internal standards. Healthcare professionals use the preliminary results in conjunction with clinical judgment to inform decision-making regarding drug use. The device benefits patients by providing a screening tool for identifying opiate presence in hair specimens.
Clinical Evidence
No clinical data provided; device relies on analytical performance characteristics of the IDS One-Step ELISA Opiates Kit for qualitative detection.
Indicated for qualitative detection of opiates at or above 500 pg/mg in head hair samples from chronic heroin users. Not evaluated for other populations or hair types.
Regulatory Classification
Identification
An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k042725
B. Purpose for Submission:
New submission
C. Measurand:
Opiates in hair
D. Type of Test:
Qualitative ELISA Immunoassay Test System, Home Brew
E. Applicant:
Quest Diagnostics, Inc.
F. Proprietary and Established Names:
Quest Diagnostics HairCheck-DT (Opiates)
G. Regulatory Information:
1. Regulation section:
CFR 862.3650, Enzyme Immunoassay, Opiates and Opiates Metabolites
2. Classification:
Class II
3. Product code:
DJG
4. Panel:
91 (Toxicology)
1
{1}
H. Intended Use:
1. Intended use(s):
Refer to Indications for use.
2. Indication(s) for use:
The Quest Diagnostics Hair Check-DT (Opiates) is a test system that utilizes the IDS One-Step ELISA Opiates Kit for the qualitative detection of opiates at or above 500 pg/mg in head hair samples from chronic heroin users. This test system has not been evaluated for use with other user populations or with hair specimens other than the head. It is an in vitro diagnostic device intended exclusively for in-house professional use only and is not intended for sale to anyone.
The QUEST DIAGNOSTICS HairCheck-DT (Opiates) provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed result. Gas Chromatograph - Mass Spectrometry operating in the selected ion monitoring (SIM) mode or GC/MS/MS in selected reaction mode (SRM) is the preferred method with deuterated internal standards. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.
Limitations: Evaluation of this assay was limited to head hair samples from a population of chronic heroin users. Interpretation of results must take into account that drug concentrations detected in hair from a single individual can vary extensively depending on the site of collection. Positive screening results only indicate the presumptive presence of opiates (codeine, morphine and/or 6-monoacetylmorphine), and require additional analysis by GC/MS or GC/MS/MS to confirm the result. A negative screening result does not necessarily rule out the possibility of opiate use, i.e., time of collection, frequency of use, mode of ingestion, dosage used, hair types and other factors may influence results.
It is not possible to document all possible effects due to treatments such as bleaching, straightening and dying.
There is a possibility that other substances and/or factors not evaluated in the interference studies may interfere with the test and cause false results that cannot be confirmed by mass spectrometry, e.g. technical or procedural errors.
3. Special conditions for use statement(s):
The assay is for Prescription In-House Use.
4. Special instrument requirements:
The device is for use with an automated microplate reader capable of measuring at 450 and 630 nm.
2
{2}
For confirmation testing, the sponsor uses an Agilent 5973N GC/MS in selected reaction monitoring (SRM) mode using deuterated internal standards.
I. Device Description:
The test obtaining clearance consists of two parts; a pre-analytical hair treatment procedure (to convert the solid matrix of hair to a measurable liquid matrix) and the screening assay. The screening assay, International Diagnostic Systems (IDS) Corporation One-Step ELISA (Enzyme-Linked ImmunoSorbent Assay) Opiates Kit, is purchased by Quest.
The screening portion of the test system consists of micro strip plates coated with rabbit anti-morphine polyclonal antibody, enzyme conjugate (horseradish peroxidase conjugated to morphine), substrate (containing tetramethylbenzidine), a proprietary diluent, and wash solution.
In-house prepared calibrators and controls are utilized. These are prepared solutions of morphine added to negative matrix tubes.
As the IDS kit is a component of the test system, the sponsor was asked to describe what specifications are provided to the component manufacturer, and what testing is performed when the kit is received to validate its performance. Details of the procedures used are described in Quest's 6/8/05 response. IDS performs a drift and precision procedure on selected plates, as well as visual inspection of all plates. Quest performs a linearity check on selected plates when they are received.
The procedures and reagents are briefly described in the "Test Principle" section, below. Trade secret information was not provided to FDA. Specific details regarding procedures and reagents provided by the sponsor have not been reproduced here because the product is not intended for sale to others.
The sponsor indicates there are no human source materials in their product, except for hair. Hair is not known to be a biological risk.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Dade Behring (Syva) EMIT II Opiates Assay
2. Predicate 510(k) number(s):
k011289
3. Comparison with predicate:
Both devices are qualitative assays for the detection of opiates. Both are immunoassays.
3
{3}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Method of measurement | Microplate reader | Spectrophotometer |
| Matrix | Head hair | Urine |
| Cutoff concentration | 500 pg opiates/mg hair | 2000 ng morphine/mL urine |
| Test Principle | ELISA | Competitive EIA |
## K. Standard/Guidance Document Referenced (if applicable):
No standards are referenced.
## L. Test Principle:
### Pre-Analytical:
The test utilizes a 3.9 cm sample of head hair. Approximately 120 strands taken from 2-3 different sites, and is cut as close as possible to the scalp, preferably from the back of the head at the crown. This amount of hair should weigh approximately 100–120 mg. In the laboratory the sample is cut from the root end, then cut into smaller lengths and mixed to ensure homogeneity.
Specimens are prepared by weighing out twenty milligram aliquots of the hair. In preparation for the screening test, an aliquot is washed with methanol for a brief period of time, then the methanol is discarded. This pre-wash is intended to rid the sample of external contamination. Methanol is added to the hair and it is heated for two hours. The methanol mixture is then transferred to a new tube and evaporated under nitrogen. The tubes are reconstituted with 0.6 mL of phosphate buffer prior to testing.
To minimize hair matrix effects calibrator and control stock solutions are added to a negative matrix tube prior to analysis. To prepare these tubes 10 grams of hair from non drug-users is weighed out and methanol is added. After soaking for a period of time the methanol is discarded. One liter of methanol is added to the methanol-washed hair and heated for 2 hours, then filtered. The collected methanol is diluted with methanol to 1 liter. One mL aliquots are pipetted into tubes and evaporated to dryness. Prior to analysis, 100 μL of prepared stock solutions of calibrator and control are pipetted into the negative hair matrix tubes, and 1.9 mL of phosphate buffer is added.
### Screening Assay:
Unknown samples, calibrators, and controls, as described above, are assayed using the IDS Opiates Kit. The kit is a solid-phase micro-titer plate immunoassay where labeled and unlabeled opiates bind to antibody. The two bind in proportion to their concentration.
Each sample is added to a well, followed by the enzyme conjugate. During this phase of
{4}
incubation the enzyme-labeled drug conjugate competes with drug in the sample for a limited number of binding sites on the antibody-coated microwells. A wash solution is then applied to remove any unbound materials. Enzyme substrate solution containing a chromagen is then added for the final color development process. The reaction is stopped with an acid and the absorbance is read using a plate reader at 450 nm. A background reading is also taken at 630 nm. Color intensity is inversely proportional to the amount of analyte present in the sample.
**Interpretation of Screening Results:**
Negative: Samples with an absorbance value higher than the Cutoff Calibrator are interpreted as negative. Either the sample does not contain opiates or opiates are present in concentrations below the cutoff level for the assay.
Presumptive Positive: Samples with an absorbance value equal to or lower than the Cutoff Calibrator are presumptively positive.
Other structurally similar compounds can produce positive results. Compounds that are not structurally similar to opiates have not been observed to produce positive results, however false positive screening results may occur because of non-specific binding or other technical problems.
**Confirmatory Testing:**
Negative hair matrix tubes are used in the confirmatory process. As in the screening procedure, control and calibrator solutions are added to the tubes prior to analysis. Negative hair matrix tubes are prepared in a similar manner to those prepared for the screening assay.
Confirmation testing is performed utilizing another aliquot from the original hair specimen. A 20 mg sample of each donor hair sample is washed four times prior to analysis. The first wash is performed with hexane. This wash is saved and analyzed along with the donor sample. (The concentration of drug in the hexane wash is multiplied by ten, then subtracted from the GC/MS result prior to applying the positive reporting criteria. This step is performed to mitigate the risk of external drug contamination.) The hexane wash is followed by 3 additional methanol washes. Each methanol wash is discarded.
Methanol is then added to the sample and it is heated for 2 hours. An acid/methanol mixture is then added, then the liquid is transferred to another tube. It is evaporated to dryness, and reconstituted with phosphate buffer.
A solid phase extraction is then performed on each standard, control, and unknown specimen, followed by GC/MS analysis. The hexane wash correction procedure is performed prior to determining the final test result.
**Interpretation of Confirmatory Testing Results:** Samples are considered positive for opiates use if morphine or codeine or 6-MAM are present at or above 500 pg/mg of hair.
For the run to be accepted, at least one control must be within 30% of the target value. (If only
5
{5}
one control passes, results are treated as qualitative results, rather than quantitative results.)
According to the sponsor, LOD’s for morphine, codeine and 6-MAM on their GC/MS system are 100, 200 and 100 pg/mg, respectively.
## M. Performance Characteristics (if/when applicable):
Two terms are utilized when describing study results.
- Normalized data is where the absorbance reading of the cutoff calibrator is considered. Formula for calculating the normalized absorbance value of unknown samples is: (1/absorbance value of cutoff calibrator) x absorbance value of unknown
- Non-normalized data refers to the raw absorbance readings observed in the run.
## 1. Analytical performance:
### a. Precision/Reproducibility:
A variety of specimen types were evaluated in precision studies, i.e., morphine solutions spiked into negative matrix tubes, pooled extracts of hair samples, and replicates of individual hair samples.
## Within-run and Between-run Precision Using Spiked Samples
Five solutions were prepared at 0, 0.5X, 1X, 1.5X and 2X of cutoff. These solutions were prepared by spiking test tubes containing negative hair matrix with the appropriate concentration of morphine to provide the percentages of cutoffs listed above. On day one, 15 replicates of each solution were pipetted into individual wells on a Microtiter plate and then analyzed by the ELISA screening method listed above. The data from day 1 was used to establish the Within-run precision for the ELISA screening method.
From these same solutions, 15 replicates were again analyzed in individual wells on days 2 and 3. Data from all 3 days was used to determine the Between-run precision, i.e., 45 replicates.
One experienced laboratory employee performed this study at Quest.
A total of 225 samples were assayed:
- 15X3 - 0.0 pg/mg Morphine
- 15X3 - 250 pg/mg Morphine
- 15X3 - 500 pg/mg Morphine
- 15X3 - 750 pg/mg Morphine
- 15X3 - 1000 pg/mg Morphine.
6
{6}
Calculations: Within-run precision was determined from Day 1 data for each concentration by calculating the mean, standard deviation and coefficient of variation for 15 samples. Between-run precision was determined by calculating the mean, standard deviation, and coefficient of variation for each concentration, 45 samples each, over three days. Precision is expressed as the percent coefficient of variation (%CV), where the standard deviation divided by the mean times 100% is equal to the percent coefficient of variation.
Within-Run Precision Study using Spiked Samples (non-normalized data)
| Opiates Spiked Sample | Negative | 50% | 100% | 150% | 200% |
| --- | --- | --- | --- | --- | --- |
| Mean | 2.217 | 1.257 | 0.400 | 0.253 | 0.211 |
| 95% CI Upper Limit | 2.267 | 1.297 | 0.413 | 0.26 | 0.218 |
| 95% CI Lower Limit | 2.167 | 1.218 | 0.386 | 0.246 | 0.214 |
| | | | | | |
| S.D. | 0.091 | 0.071 | 0.025 | 0.013 | 0.013 |
| 95% CI Upper Limit | 0.575 | 0.446 | 0.157 | 0.081 | 0.080 |
| 95% CI Lower Limit | 0.048 | 0.037 | 0.013 | 0.007 | 0.007 |
| | | | | | |
| CV% | 4.1% | 5.6% | 6.2% | 5.1% | 6.0% |
| 95% CI Upper Limit | 25.9% | 35.5% | 39.3% | 32.1% | 37.7% |
| 95% CI Lower Limit | 2.1% | 2.9% | 3.3% | 2.7% | 3.1% |
Between-Run Precision using Spiked Samples (non-normalized data)
| Opiates Spiked Sample | Negative | 50% | 100% | 150% | 200% |
| --- | --- | --- | --- | --- | --- |
| Mean | 2.008 | 1.148 | 0.354 | 0.220 | 0.181 |
| 95% CI Upper Limit | 2.041 | 1.169 | 0.361 | 0.224 | 0.186 |
| 95% CI Lower Limit | 1.976 | 1.126 | 0.347 | 0.215 | 0.177 |
| | | | | | |
| S.D. | 0.059 | 0.039 | 0.013 | 0.008 | 0.008 |
| 95% CI Upper Limit | 0.371 | 0.244 | 0.082 | 0.052 | 0.053 |
| 95% CI Lower Limit | 0.031 | 0.020 | 0.007 | 0.004 | 0.004 |
| | | | | | |
| CV% | 2.9% | 3.4% | 3.7% | 3.8% | 4.7% |
| 95% CI Upper Limit | 18.5% | 21.3% | 23.1% | 23.7% | 29.4% |
| 95% CI Lower Limit | 1.5% | 1.8% | 1.9% | 2.0% | 2.4% |
Within-run and Between-Run Precision using pooled extracts
Within-run precision was determined by analyzing fifteen replicate samples of pooled sample extracts. Results were similar to those of prepared solutions, above.
{7}
8
# Within-Run Precision using individual samples
Studies were done to characterize precision when replicate measurements of single hair samples were analyzed. Four hair specimens previously found to render absorbance readings close to the absorbance reading of the cutoff calibrator were analyzed. Each hair specimen was divided into 3 three aliquots of 20 mg each. Each 20 mg aliquot was taken through the entire ELISA screening process and measured in one run. The following table depicts the absorbance readings (not normalized) of the analysis.
Within-Run Precision of Opiates using individual hair (non-normalized data)
| Specimen # | 1 | 2 | 3 | 4 |
| --- | --- | --- | --- | --- |
| | 0.105 | 1.377 | 0.058 | 0.061 |
| | 0.094 | 1.289 | 0.062 | 0.047 |
| | 0.080 | 1.189 | 0.048 | 0.065 |
| Mean | 0.093 | 1.285 | 0.056 | 0.058 |
| 95% CI | | | | |
| Upper Limit | 0.124 | 1.519 | 0.074 | 0.081 |
| Lower Limit | 0.062 | 1.051 | 0.038 | 0.034 |
| Std Dev. | 0.013 | 0.094 | 0.007 | 0.009 |
| 95% CI* | | | | |
| Upper Limit | 0.079 | 0.591 | 0.045 | 0.059 |
| Lower Limit | 0.007 | 0.049 | 0.004 | 0.005 |
| %CV | 14.0% | 7.3% | 12.9% | 16.4% |
| 95% CI* | | | | |
| Upper Limit | 87.8% | 46.0% | 80.9% | 103.0% |
| Lower Limit | 7.3% | 3.8% | 6.7% | 8.5% |
Calibrator absorbance in this run was 0.577 with low and high control of 1.493 and 0.339 respectively. The absorbance value of the negative control was 2.004
b. Linearity/assay reportable range:
Not applicable. This is a qualitative assay. Representative absorbances are observable, however, in the precision section, above.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Commercially purchased materials consisting of morphine in methanol are used to prepare a working solution. Working solutions are then used to prepare calibrator and control solutions. (Calibrators and controls are prepared in a similar manner, however, they are made from different reference materials, each provided with a Certificate of Analysis.)
Assigned values of the gravimetrically prepared calibrator and control stock solutions
{8}
are verified by GC/MS analysis each time a new batch is prepared. The calibrator must fall within 20% of the targeted concentration. The sponsor indicates they have data on file to support the one year expiration date for these solutions.
At the time of analysis, the prepared calibrator and control stock solutions are pipetted into a negative matrix tube, and diluted with phosphate buffer. The final concentrations are as follows:
- Positive Calibrator containing 500 pg/mg hair of opiates
- Negative Blank Calibrator (negative matrix tube containing 0.0 pg/mg hair of opiates)
- Low Control containing 400 pg/mg hair of opiates
- High Control containing 600 pg/mg hair of opiates
Users are instructed to follow federal, state, and local regulatory guidelines regarding quality control procedures.
d. Detection limit:
The limit of detection (in pg/mg) was determined by calculating the mean negative calibrator absorbance (A₀) minus two times the SD (LOD = A₀ - 2SD).
The calculation of sensitivity was determined in hair matrix samples by calculating the mean absorbance value of each set of 18 zero calibrators (blanks) and adding two standard deviations for the corresponding group. To convert the value from absorbance units to pg/mg concentration, a regression line was constructed using the mean values of the zero standard, 225 pg/mg morphine standard, 300 pg/mg morphine standard, and 375 pg/mg morphine standard. Using the equation of the regression line, the absorbance value of the mean zero calibrator minus two standard deviations was converted to pg/mg of morphine.
One experience laboratory employee performed this study in one day.
The study took place at the Quest laboratory.
LOD Study Results
| | 0 pg/mg in buffer + matrix (Absorbance) |
| --- | --- |
| Mean (0 pg/mg) | 1.920 |
| Std Dev | 0.090 |
| Std Dev X 2 | 0.180 |
| Mean-2 Std Deviations | 1.740 |
| | |
| Mean (225 pg/mg) | 0.676 |
| Mean (300 pg/mg) | 0.572 |
| Mean (375 pg/mg) | 0.428 |
| R Square | 0.943 |
{9}
e. Analytical specificity:
Cross-Reactivity with structurally related compounds:
To determine cross-reactivity each compound was spiked into 46 mm phosphate buffer containing negative hair matrix.
Serial dilutions of each compound were prepared and analyzed. Resulting absorbance readings were plotted against the prepared concentration. The concentration of each compound that generated the same absorbance reading as the cutoff calibrator was extrapolated from the graph. The concentration of morphine in the cutoff calibrator was divided by the extrapolated concentration of the structurally similar compound and then multiplied by 100. (For example if it took 1200 pg/mg of a structurally similar compound to equal the absorbance value of 500 pg/mg of morphine then the cross reactivity would be 500/1200 x 100% = 50%.)
Percent Cross-reactivity of structurally related compounds
| Compound | Percent Cross-Reactivity | Amount (pg/mg) of Morphine Analog equivalent to produce a positive result at the cut-off of 500 pg/mg |
| --- | --- | --- |
| Ethylmorphine | 555.6 | 90 |
| Codeine | 285.7 | 175 |
| 6-Acetylcodeine | 133.3 | 375 |
| Heroin | 125 | 400 |
| Dihydrocodeine | 111.1 | 450 |
| Morphine | 100 | 500 |
| 6-Monoacetylmorphine | 100 | 500 |
| Morphine-3-beta-glucuronide | 100 | 500 |
| Thebain | 83.3 | 600 |
| Morphine-6-beta-glucuronide | 76.9 | 650 |
| Dihydromorphine | 71.4 | 700 |
| Hydrocodone | 33.3 | 1502 |
| Hydromorphone | 33.3 | 1502 |
| Nalorphine | 15.4 | 3247 |
{10}
| Compound | Percent Cross-Reactivity | Amount (pg/mg) of Morphine Analog equivalent to produce a positive result at the cut-off of 500 pg/mg |
| --- | --- | --- |
| Levorphanol | 14.3 | 3497 |
| Norcodeine | 5 | 10000 |
| Oxycodone | 1.1 | 45455 |
| Normorphine | 0.7 | 71429 |
| Diprenorphine | 0.6 | 83333 |
| Oxymorphone | 0.5 | 100000 |
| Apomorphine | 0.5 | 100000 |
Other compounds tested demonstrated less than 0.5% cross-reactivity.
## Cross-Reactivity with structurally unrelated compounds
Several (160) structurally unrelated compounds were added to 46 mm phosphate buffer to a concentration of 10,000 ng/mL then added to negative hair matrix tubes (equivalent to 300,000 pg/mg). Samples were analyzed along with replicates of blank negative hair matrix tubes. The mean absorbance readings from the samples were within 5% of the mean absorbance readings of the blank negative hair matrix tubes.
## Effect of Interfering Compounds
The same 160 structurally unrelated compounds were also tested for possible positive and negative interference with the Opiate ELISA assay. Two sets of negative hair matrix were prepared by adding morphine to achieve concentrations of at 333, 500 and 667 pg/mg hair. The second set of tubes was additionally spiked with the 160 structurally unrelated compounds to a concentration of 300,000 pg/mg hair. Absorbance readings of the tubes spiked with the structurally unrelated compound were within 5% of the absorbance readings of the negative hair matrix tube without the compound added.
There is a possibility that other substances and/or factors not listed above may interfere with the test and cause false results e.g., technical or procedural errors.
## ADDITIONAL ANALYTICAL STUDIES:
## Recovery studies/ Effectiveness of Screening Assay Extraction Method
The screening assay employs a process to extract drug from a hair sample, i.e., a 2-hour methanolic extraction of hair at 70°C. A study was done to demonstrate the effectiveness of this procedure.
In order to evaluate the effectiveness of the methanolic extraction procedure a definition of 100% recovery was established. According to the sponsor, treatment of hair with a strong base (NaOH) converts hair to liquid, which can be extracted by a solid phase extraction
{11}
technique for analysis. The strong base also converts a significant amount of 6-Acetylmorphine to morphine. Quest defines total recovery in this study as the sum of the codeine, morphine and 6-acetylmorphine as quantitated by GC/MS.
Two sample types were evaluated:
1) Hair spiked with morphine and 6-acetylmorphine:
Approximately 1 gram of drug-free hair was soaked in an aqueous solution containing 1000 ng/mL of morphine and 6-acetylmorphine for 24 hours. The hair was removed from the solution and washed with “Aloe Rid” shampoo once, water 5 times and allowed to air dry overnight at room temperature.
2) Hair samples which previously confirmed positive for morphine, codeine and/or 6-acetylmorphine.
Samples were aliquoted in duplicate; one aliquot was taken through the screening extraction and assay (described in Test Principle Section, above) and the other was taken through the 100% recovery extraction.
To perform extraction recovery determinations of each opiate compound, the screening extraction procedure was performed up to the point of evaporating the methanolic extract. At that point, internal standard was added and the solid phase extraction procedure used during confirmation testing was performed, followed by GC/MS analysis.
The method representing 100% recovery was accomplished by adding 0.5 mL of 1.0 N NaOH to 20 mg of hair and incubating at 70°C for 30 minutes. The sample is totally liquefied and the entire drug originally bound to the hair is now dissolved in the base solution. After base hydrolysis of the hair, the internal standard is added and the confirmation solid phase extraction procedure was performed.
The GC/MS results of the screening extraction were compared to the results of the confirmation extraction to determine the relative recovery of the drug using the 2-hour methanol incubation:
12
{12}
GC/MS Results and Recovery
| | Methanol | | | | Base Hydrolysis | | | | % Recovery |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | COD | MOR | 6-MAM | TOTAL | COD | MOR | 6-MAM | TOTAL | TOTAL |
| 1 | 168 | 3680 | 21 | 3869 | 173 | 4022 | 36 | 4231 | 91.4% |
| 2 | 0 | 96 | 7 | 103 | 0 | 65 | 69 | 134 | 76.9% |
| 3 | 102 | 25 | 0 | 127 | 164 | 157 | 67 | 388 | 32.7% |
| 4 | 1254 | 33 | 0 | 1287 | 2114 | 40 | 0 | 2154 | 59.7% |
| 5 | 211 | 13 | 0 | 224 | 213 | 88 | 0 | 301 | 74.4% |
| Spiked #1 | 0 | 2363 | 6487 | 8850 | 0 | 9303 | 65 | 9368 | 94.5% |
| Spiked #2 | 0 | 1980 | 5434 | 7414 | 0 | 8269 | 102 | 8371 | 88.6% |
| Spiked #3 | 0 | 1726 | 4834 | 6560 | 0 | 10761 | 78 | 10839 | 60.5% |
| | | | | | | | | | |
| | | | | | Average Recovery of Positive Donor Samples | | | | 67.0% |
| | | | | | Average Recovery of Spiked Samples | | | | 81.2% |
| | | | | | Average Recovery of All Samples | | | | 72.3% |
Hair Treatment Effects on Positive Hair Samples:
The effects of various hair treatments (i.e. bleaching, dyeing, shampooing) on the screening and confirmatory assay was examined. Forty-eight previously screened and confirmed opiate positive hair specimens were randomly assigned into one of three groups (16 in each group). Each group was subjected to one of the three treatments. Absorbance readings and GC/MS measurements were compared before and after treatment. Each of the three opiate compounds were examined separately during GC/MS analysis. Results appear below.
**Note: An increase in concentration correlates to a decrease in Absorbance and a decrease in concentration correlates with an increase in Absorbance.
{13}
Effects Observed in the Bleaching Study (Normalized data, n= 16)
| | | ELISA SCREENING DATA | | | |
| --- | --- | --- | --- | --- | --- |
| | Avg Abs/ Range of Abs* | # of samples that remained positive | Avg/ Range of Abs of all that had a decrease in Abs** | # of samples that became negative | Avg/ Range of Abs of all that had an increase in Abs** |
| Untreated | 0.135 (0.042 – 0.358) | | | | |
| Treated | 3.859 (2.721 – 4.247) | 0 | 0 | 16 | 3.859 (2.721 – 4.247) |
| | | GC/MS CONFIRMATION DATA | | | |
| | Avg / Range of sample concentrations (pg/mg) | # of samples that decreased in concentration | Avg/ Range of decrease in concentration | # of samples that increased in concentration | Avg/ Range of increase in concentration |
| Untreated | | | | | |
| Codeine | 1443 (297– 4432), n=7 | | | | |
| Morphine | 899 (159-4189), n=14 | | | | |
| 6 MAM | 4369 (1225-14116), n=14 | | | | |
| Treated | | | | | |
| Codeine | 143 (0-585) | 7 | 143 (0-585) | 0 | - |
| Morphine | None Detected | 14 | None Detected | 0 | - |
| 6 MAM | 95(0-370) | 14 | 95 (0-370) | 0 | - |
{14}
Effects Observed in the Dyeing Study (Normalized data, n= 16)
| | | ELISA SCREENING DATA | | | |
| --- | --- | --- | --- | --- | --- |
| | Avg Abs/ Range of Abs* | # of samples that remained positive | Avg/ Range of Abs of all that had a decrease in Abs** | # of samples that became negative | Avg/ Range of Abs of all that had an increase in Abs** |
| Untreated | 0.178 (0.025-0.410) | | | | |
| Treated | 2.300 (0.139-3.514) | 2 | 0 | 14 | 2.300 (0.139 -3.514) |
| | | | | | |
| | | GC/MS CONFIRMATION DATA | | | |
| | Avg / Range of sample concentrations (pg/mg) | # of samples that decreased in concentration | Avg/ Range of decrease in concentration | # of samples that increased in concentration | Avg/ Range of increase in concentration |
| Untreated | | | | | |
| Codeine | 586 (394-809), n=4 | | | | |
| Morphine | 815 (203-3894), n=13 | | | | |
| 6 MAM | 6911 (710-33032), n=14 | | | | |
| Treated | | | | | |
| Codeine | None detected | 4 | None detected | 0 | - |
| Morphine | 521 (0-4610) | 11 | 141 (0-1045) | 2 | 2613 (616-4610) |
| 6 MAM | 772 (0-6420) | 14 | 772 (0-6420) | 0 | - |
{15}
Effects Observed in the Shampoo Study (Normalized data, n= 16)
| | | ELISA SCREENING DATA | | | |
| --- | --- | --- | --- | --- | --- |
| | Avg Abs/ Range of Abs* | # of samples that remained positive | Avg/ Range of Abs of all that had a decrease in Abs** | # of samples that became negative | Avg/ Range of Abs of all that had an increase in Abs** |
| Untreated | 0.220 (0.062-0.522) | | | | |
| Treated | 0.450 (0.095-2.558) | 14 | 0.139 (0.139-0.139) | 2 | 0.471 (0.095-2.558) |
| | | | | | |
| | | GC/MS CONFIRMATION DATA | | | |
| | Avg / Range of sample concentrations (pg/mg) | # of samples that decreased in concentration | Avg/ Range of decrease in concentration | # of samples that increased in concentration | Avg/ Range of increase in concentration |
| Untreated | | | | | |
| Codeine | 896 (176-2788), n=5 | | | | |
| Morphine | 532 (124-1684), n=11 | | | | |
| 6 MAM | 3608 (401-15681), n=14 | | | | |
| Treated | | | | | |
| Codeine | 951 (0-2938) | 2 | 81 (0-162) | 3 | 1531 (216-2938) |
| Morphine | 360 (0-1007) | 8 | 393 (0-1007) | 3 | 270 (149-375) |
| 6 MAM | 3170 (486-15443) | 9 | 3258 (486-15443) | 5 | 3011 (838-7821) |
{16}
17
# Hair treatment Affects on Negative Hair:
In a separate study 52 previously screened negative specimens were randomly assigned to the same treatment groups. All 52 specimens remained negative after treatment. The percent difference between the mean normalized absorbance values of the treated and untreated groups was $-8.65\%$, $-9.82\%$ and $0.69\%$ for bleaching, dyeing and shampooing respectively.
Bleaching and dyeing had the greatest effect. Screening absorbance readings became more negative for the positive hair samples, and slightly more negative for the negative hair samples. (The decrease in absorbance reading is equal to an apparent increase in concentration.)
# Effectiveness of the Wash Procedure (Contamination Studies):
Two studies were done to determine whether confirmatory testing procedures (including the hexane wash correction) are able to distinguish between true analytically positive samples and those that have been externally exposed to drug. The focus of the studies was to demonstrate that the hexane wash correction procedure mitigates the risk of false positive results while not altering true analytical positive results.
1. The first study involved exposing drug-free hair to opiate compounds, performing confirmation testing on the samples and observing the final test result.
2. The second study involved performing confirmation testing on known positive samples and observing whether the hexane wash correction altered the final result.
# STUDY # 1
Five hair specimens that had previously screened negative for opiates were selected in order to represent five hair types.
Hair Types:
| Category | Hair Color | Hair Texture |
| --- | --- | --- |
| A | Black | Straight |
| B | Black | Curly |
| C | Brown | Thin |
| D | Brown | Thick |
| E | Blonde | All types |
{17}
One aliquot of each hair type was exposed to morphine and codeine (in separate experiments) according to the first three exposure modes listed below. Another aliquot of each hair type was exposed to heroin smoke. A twenty mg aliquot of all hair samples was then analyzed by GC/MS. Results are presented for each exposure mode, and according to the drug and category of hair types.
Exposure modes
| # | Type of Exposure | How performed |
| --- | --- | --- |
| 1 | Dry Contact | Ten (10) mg of morphine sulfate and ten (10) mg of codeine sulfate were weighed and combined with 10 grams of dextrose with maltodextrin, aspartame (Equal®). The powdered mixture was grinded to a fine powder and mixed thoroughly. Exactly 150 mg of powder was added to 150 mg of each hair type in individual plastic bags and the bags closed with a ziplock® seal. Bags were thoroughly shaken, manipulated and rubbed to ensure direct contact between the powder and hair. Hair specimens remained in the plastic bags for one hour. |
| 2 | Dry Contact plus water | After treatment by exposure mode #1, above, twenty mg aliquots of hair were weighed out and 2.0 mL of deionized water was added and then quickly removed to simulate rinsing the hair, as in a shower. An additional 2.0 mL of deionized water was then added and the tube was allowed to stand for 30 minutes at ambient temperature. The water was then removed and the hair dried overnight. |
| 3 | Dry Contact plus Saline | After treatment by exposure mode #1, above, twenty mg aliquots of hair were weighed out and 2.0 mL of 0.9% saline solution was added. The tube stood for 30 minutes at ambient temperature. Saline solution was removed and the hair dried overnight. |
| 4 | Smoke | One hundred mg of each hair type was exposed to heroin smoke by pyrolyzing 1.0 mg of diacetylmorphine (Heroin) in a glass device. The glass device was connected to carrier gas on one end and then connected to a tube containing the hair on the other end. The powder was burned and once smoke was produced, the gas was turned on which carried the smoke to the tubes containing the hair specimen. Smoke was held in the test tube for 5 minutes. |
{18}
Results from the study are presented in the tables below.
# Morphine Exposure:
# Exposure mode 1
| | Morphine (pg/mg) | | | |
| --- | --- | --- | --- | --- |
| | (A) Hexane Wash | (B) Final Extract | (D) x10 of Hexane Wash | (D-A) |
| A | 10501 | 28 | 105010 | -104982 |
| B | 16409 | 65 | 164090 | -164025 |
| C | 2640 | 73 | 26400 | -26327 |
| D | 22762 | 279 | 227620 | -227341 |
| E | 13587 | 90 | 135870 | -135780 |
# Codeine Exposure:
# Exposure mode 1
| | Codeine (pg/mg) | | | |
| --- | --- | --- | --- | --- |
| | (A) Hexane Wash | (B) Final Extract | (D) x10 of Hexane Wash | (D-A) |
| A | 2690 | 148 | 26900 | -26752 |
| B | 3642 | 151 | 36420 | -36269 |
| C | 3024 | 209 | 30240 | -30031 |
| D | 209797 | 1452 | 2097970 | -2096518 |
| E | 4176 | 157 | 41760 | -41603 |
# Morphine Exposure:
# Exposure mode
2
| | Morphine (pg/mg) | | | |
| --- | --- | --- | --- | --- |
| | (A) Hexane Wash | (B) Final Extract | (D) x10 of Hexane Wash | D-A) |
| A | 239 | 1546 | 2390 | -844 |
| B | 2539 | 5543 | 25390 | -19847 |
| C | 350 | 342 | 3500 | -3158 |
| D | 3248 | 9203 | 32480 | -23277 |
| E | 819 | 944 | 8190 | -7246 |
# Codeine Exposure:
# Exposure mode 2
| | Codeine (pg/mg) | | | |
| --- | --- | --- | --- | --- |
| | (A) Hexane Wash | (B) Final Extract | (D) x10 of Hexane Wash | (D-A) |
| A | 89 | 252 | 890 | -638 |
| B | 124 | 894 | 1240 | -346 |
| C | 83 | 342 | 830 | -488 |
| D | 581 | 2802 | 5810 | -3008 |
| E | 363 | 348 | 3630 | -3282 |
{19}
20
# Morphine Exposure:
## Exposure mode 3
| | Morphine (pg/mg) | | | |
| --- | --- | --- | --- | --- |
| | (A) Hexane Wash | (B) Final Extract | (D) x10 of Hexane Wash | D – A) |
| A | 428 | 1311 | 4280 | -2969 |
| B | 1978 | 4449 | 19780 | -15331 |
| C | 1885 | 832 | 18850 | -18018 |
| D | 829 | 3360 | 8290 | -4930 |
| E | 3398 | 2612 | 33980 | -31368 |
# Codeine Exposure:
## Exposure mode 3
| | Codeine (pg/mg) | | | |
| --- | --- | --- | --- | --- |
| | (A) Hexane Wash | (B) Final Extract | (D) x10 of Hexane Wash | (D – A) |
| A | 273 | 300 | 2730 | -2430 |
| B | 70 | 409 | 700 | -291 |
| C | 2250 | 769 | 22500 | -21731 |
| D | 1174 | 1147 | 11740 | -10593 |
| E | 511 | 464 | 5110 | -4646 |
# Heroin Exposure:
## Exposure mode 4
| | Morphine (pg/mg) | | | |
| --- | --- | --- | --- | --- |
| | (A) Hexane Wash | (B) Final Extract | (D) x10 of Hexane Wash | D – A) |
| A | 6056 | 2037 | 30280 | -28243 |
| B | 9534 | 244 | 47670 | -47426 |
| B1 | 23277 | 735 | 116385 | -115650 |
| C | 126 | 5 | 630 | -625 |
| D | 70 | 12 | 350 | -338 |
| E | 1054 | 132 | 5270 | -5138 |
# Heroin Exposure:
## Exposure mode 4
| | 6 MAM (pg/mg) | | | |
| --- | --- | --- | --- | --- |
| | (A) Hexane Wash | (B) Final Extract | (D) x10 of Hexane Wash | D – A) |
| A | 35850 | 4047 | 358500 | -354453 |
| B | 19432 | 1320 | 194320 | -193000 |
| B1 | 22902 | 14454 | 229020 | -214566 |
| C | 516 | 13 | 5160 | -5147 |
| D | 408 | 101 | 4080 | -3979 |
| E | 181 | 71 | 1810 | -1739 |
# STUDY #2
Nine clinically positive hair samples were selected for this study. All hair samples:
- were previously screened and confirmed positive
- came from subjects who admitted opiate or heroin use
- were accompanied by a urine specimen that screened positive
Confirmation testing procedures were performed on the 9 samples, i.e., 4 washes, extraction SEP, GG/MS, and hexane wash correction. Results are presented below:
{20}
CLINICAL POSITIVE SAMPLES
| | Morphine (pg/mg) | | | |
| --- | --- | --- | --- | --- |
| | (A) Hexane Wash | (B) Final Extract | (D) x10 of Hexane Wash | (E) = D minus A (pg/mg) Hair |
| #30 | 13 | 5998 | 130 | 5868 |
| #35 | 11 | 6184 | 110 | 6074 |
| #3 | 0 | 358 | 0 | 358 |
| #13 | 0 | 1762 | 0 | 1762 |
| #18 | 0 | 329 | 0 | 329 |
| #45 | 0 | 484 | 0 | 484 |
| #47 | 0 | 608 | 0 | 608 |
| #70 | 0 | 322 | 0 | 322 |
| #71 | 0 | 240 | 0 | 240 |
| | Codeine (pg/mg) | | | |
| --- | --- | --- | --- | --- |
| | (A) Hexane Wash | (B) Final Extract | (D) x10 of Hexane Wash | ((E) = D - A (pg/mg) Hair |
| #30 | 0 | 495 | 0 | 495 |
| #35 | 0 | 891 | 0 | 891 |
| #3 | 0 | 40 | 0 | 40 |
| #13 | 0 | 275 | 0 | 275 |
| #18 | 0 | 52 | 0 | 52 |
| #45 | 0 | 265 | 0 | 265 |
| #47 | 0 | 0 | 0 | 0 |
| #70 | 0 | 0 | 0 | 0 |
| #71 | 0 | 41 | 0 | 41 |
| | 6 MAM (pg/mg) | | | |
| --- | --- | --- | --- | --- |
| | (A) Hexane Wash | (B) Final Extract | (D) x10 of Hexane Wash | (E) = B minus D |
| #30 | 48 | 22401 | 480 | 21921 |
| #35 | 30 | 9458 | 300 | 9158 |
| #3 | 0 | 5300 | 0 | 5300 |
| #13 | 203 | 25847 | 2030 | 23817 |
| #18 | 0 | 4031 | 0 | 4031 |
| #45 | 0 | 3995 | 0 | 3995 |
| #47 | 134 | 9336 | 1340 | 7996 |
| #70 | 0 | 3692 | 0 | 3692 |
| #71 | 0 | 2469 | 0 | 2469 |
Summary of Study Results: Results from the analytically negative samples remained negative after being exposed to the three opiate compounds under the described conditions. All clinically positive samples remained positive.
## Stability Study:
Eleven samples previously screened positive for Opiates and analyzed by GC/MS for morphine, codeine and 6-monoacetylmorphine (6-MAM) were used in this
{21}
study. Samples were stored in a climate controlled space, then analyzed a second time approximately 3 years later.
The following table illustrates the mean concentration of morphine, codeine and 6-MAM in the samples. Based on this data there is an overall trend of drug loss.
Results From Stability Study on eleven Samples
| Study Observation | Morphine | Codeine | 6-MAM |
| --- | --- | --- | --- |
| Average Concentration, pg/mg hair, Baseline | 505 | 137 | 2045 |
| Mean Change in % | -29 % | -18 % | -31% |
| Range in concentration, pg/mg hair | 0-2055 | 0-786 | 0-12172 |
| % Maximum Decrease | -100% | -100% | -66% |
| % Maximum Increase | +81% | +100% | +33% |
| Number that increased in concentration | 3 | 3 | 1 |
| Number that decreased in concentration | 5 | 5 | 9 |
| Number that did not contain the specific drug at the beginning or ending of the study. | 3 | 3 | 1 |
Drugs rendering a 0 pg/mg hair result at initial testing were excluded from calculations involving a percent change.
Three of the eleven specimens analyzed changed from detectable to undetectable concentrations of both morphine and codeine. The original concentration of codeine in these three specimens was 116, 122 and 122 pg/mg. The original concentration of morphine in these three specimens was 424, 539 and 677 pg/mg.
10 of the 11 original specimens contained of 6-MAM. All ten contained measurable amounts of 6-MAM three years later. All eleven of the original specimens contained measurable amounts of one or more of the drugs three years later.
f. Assay cut-off:
The Substance Abuse and Mental Health Services Administration has not yet recognized hair testing in the Federal Workplace Drug Testing program. Preliminary recommendations, however, suggest the use of a 200 pg/mg cut-off level for opiates as the initial screening level. Confirmatory recommendations are
22
{22}
for there to be 200 pg/mg of either morphine, codeine, or 6-acetylmorphine to be reported as positive.
A screening cutoff of 500 pg/mg of opiates is used by Quest Diagnostics. The positive agreement studies appear to support that this is an effective cutoff for the claimed user population, i.e., chronic users.
Characterization of how the device performs analytically around the claimed cutoff concentration appears in the precision section, above.
2. Comparison studies:
a. Method comparison study: Agreement Studies
Clinical performance was evaluated with two studies, one involving chronic heroin users and one involving self-reported non-drug users. (Hexane wash correction procedures were not used at the time of these studies.)
Positive Agreement Studies
The study enrolled 105 subjects known to be chronic Heroin users and/or had a positive urine test for Opiates. Almost all participants reported last using heroin on a daily basis. Participants also indicated they had been using heroin for anywhere from 1 to 40 years. Each subject provided a urine and head hair sample.
Of the 105 volunteer subjects 88 were Caucasian, 16 Hispanic and one (1) did not note their race. The subjects ranged in ages from 21 to 70 years old. Of the 105 hair samples: 10 were light brown, 52 were medium brown, 33 were dark brown, 4 were gray, 3 were black and 3 were blonde. The curvature ranged from 68 straight and 34 curly. There was not sufficient hair to evaluate 3 of these samples as to curvature.
One hundred and five (105) were positive for Opiates in their urine using EMIT (300 ng/mL cutoff). Urine specimens were not confirmed by GC/MS. Ninety-seven (97) of the hair samples screened positive using ELISA (500 pg/mg hair cutoff). Ninety (90) of the hair samples confirmed positive for combinations of codeine, morphine or 6 monoacetyl morphine above the 500 pg/mg GC/MS cut-off. A positive was determined to be a value greater than 500 pg/mg hair for any of the three opiate analytes. The following table describes the findings.
23
{23}
Positive Agreement Study Results
| Subjects | Urine (EMIT) Results | Hair ELISA Results | Hair GC/MS Results | History on Survey |
| --- | --- | --- | --- | --- |
| 83 | + | + | + | + |
| 7 | + | + | + | - |
| 1 | + | + | - | - |
| 5 | + | + | - | + |
| 1 | + | + | QNS | + |
| 8 | + | - | -* | + |
* These eight specimens that screened positive in urine but screened negative in hair. These specimens were analyzed by GC/MS and were all found to contain at least two opiate compounds; all specimens contained Morphine (32-288 pg/mg hair) and 6MAM (40-225 pg/mg hair), three specimens contained Codeine (22- 51 pg/mg hair). In all cases, the concentrations were below the identified screening and confirmatory cutoff concentrations.
Positive Agreement Study ELISA absorbance value information:
Average Absorbance of the cutoff calibrator was 0.711
Normalized Absorbance range of 97 screened positive samples was 0.029 to 0.581
Average normalized absorbance value of all 105 enrollees is 0.305
QNS – specimen insufficient for GC/MS analysis
## Negative Agreement Study
Eighty-two (82) individuals who self-reported that they were non-drug users were enrolled in the study. Subjects provided urine and a hair sample.
Of the eighty-two (82) samples only thirty had race recorded. Twenty-three were Caucasian, 4 were African-American and 3 were Hispanic. No ages were collected on any of the volunteers. Of the eighty-two hair specimens 14 were black, 23 were dark brown, 20 were medium brown, 11 were light brown, 12 were blond and 2 were red. The curvature of the samples were estimated to be 45 straight, 32 curly, and 5 kinky.
All eighty-two (82) urine samples screened negative for Opiates using EMIT (300ng/mL cutoff). Urine samples were not confirmed. All 83 hair samples screened negative and contained no measurable amounts of Opiates when analyzed by GC/MS.
24
{24}
Negative Agreement Study Results
| Number of Subjects | Urine Results Screen | Hair Results Screen | Hair GC/MS Result |
| --- | --- | --- | --- |
| 82 | - | - | - |
Negative Agreement Study ELISA absorbance value information:
Normalized absorbance values for specimens in this negative agreement study ranged from 1.672 to 2.688 with a mean absorbance value of 2.213.
There were two hair runs included in the negative agreement study. The absorbance of the cutoff calibrator in the first run was 0.861. The range of absorbance values for the samples in that run was 1.478 to 2.075, with an average absorbance value of 1.875. The absorbance of the hair cutoff calibrator in the second run was 0.817. Samples in this run ranged from 1.366 to 2.196, with an average absorbance value of 1.825.
# Normalized Absorbance Readings from the two study populations
The following table displays the normalized absorbance readings from the negative and positive agreement studies.
Normalized data from Clinical Agreement Studies
| | Negative Agreement | Positive Agreement |
| --- | --- | --- |
| Mean Absorbance | 2.213 | 0.164 |
| Lowest Absorbance | 1.672 | 0.043 |
| Highest Absorbance | 2.688 | 0.866 |
| Standard Deviation | 0.188 | 0.154 |
| Mean Absorbance ± 2SDs | 2.213 ± 0.376 | 0.164 ± 0.308 |
b. Matrix comparison:
Not applicable. The assay is intended for only one sample matrix.
# 3. Clinical studies:
a. Clinical Sensitivity:
The clinical study conducted characterized the effectiveness of the Quest test system to identify chronic heroin users satisfying the inclusion criteria for the study. Performance in a general population of users is unknown.
To estimate confirmation rates of samples run in their lab that screened
{25}
positive, the sponsor analyzed 3 months of recent historical data. Of those samples approximately 37% of them confirmed positive. Some of the samples that failed to confirm contained drug, but the concentrations were below the positive reporting criteria.
b. Clinical specificity:
Clinical specificity of this drug is estimated in the negative agreement studies. However, the sample size is not adequate to rule out the possibility of a positive screening assay when no drug is present.
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
26
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.