← Product Code [FBK](/productcode/FBK) · K042615

# CARBON MEDICAL TECHNOLOGIES ENDOSCOPIC INJECTION NEEDLE (K042615)

_Carbon Medical Technologies, Inc. · FBK · Nov 5, 2004 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K042615

## Device Facts

- **Applicant:** Carbon Medical Technologies, Inc.
- **Product Code:** [FBK](/productcode/FBK.md)
- **Decision Date:** Nov 5, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The CMT Endoscopic Injection Needle is intended for use as an accessory for currently marketed endoscopes and provides delivery of injectable materials during an endoscopic procedure. The type of material to be injected is dependent on the nature of the procedure, but may include delivery of injectable bulking agent during endoscopic procedures, sclerosing agents during esophagoscopic and gastroscopic procedures, local anesthetics during cystoscopic or laryngoscopic procedures, or saline or contrast media during colonoscopic procedures.

## Device Story

Device is a sterile, single-use endoscopic injection needle; consists of stainless steel needle attached to plastic cannula with proximal luer lock hub. Operated by clinician during endoscopic procedures; connects to standard syringe for manual delivery of fluids or agents into tissue. Needle length and gauge vary to accommodate specific endoscope channels. Device facilitates targeted delivery of therapeutic or diagnostic substances; enhances procedural utility of existing endoscopes.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Stainless steel needle; plastic cannula; plastic luer lock hub. Manual operation via standard syringe. Sterile, single-use. Dimensions vary by model to fit endoscope channels.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Scerotherapy Needle Medi-Glove ([K955558](/device/K955558.md))
- CMT Endoscopic Injection Needle Carbon Medical Technologies, Inc. ([K982890](/device/K982890.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

| 510(k) SUMMARY      | NOV - 5 2004                                                                                 | K042615    |
|---------------------|----------------------------------------------------------------------------------------------|------------|
| General Information |                                                                                              | PG: 1 OF 2 |
| Date Prepared       | September 22, 2004                                                                           |            |
| Classification      | Class II                                                                                     |            |
| Trade Name          | Carbon Medical Technologies Endoscopic Injection Needle                                      |            |
| Common Name         | Endoscopic Needle                                                                            |            |
| Submitter           | Carbon Medical Technologies, Inc.<br>1290 Hammond Road<br>St. Paul, MN 55110<br>651-653-8512 |            |
| Contact             | Robert W. Johnson<br>Vice President of Regulatory Affairs and Quality Assurance              |            |
| Predicate Device    | Scerotherapy Needle<br>Medi-Glove (K955558)                                                  |            |
|                     | CMT Endoscopic Injection Needle<br>Carbon Medical Technologies, Inc. (K982890)               |            |

.

·

{1}------------------------------------------------

#### Device Description

The Carbon Medical Technologies Endoscopic Injection Needle consists of a stainless steel needle attached to a plastic insertion tube. The needle length and gauge will be identified on the label. The luer lock hub, which is molded onto the proximal end of the device, is designed to accommodate a standard syringe.

The CMT Endoscopic Injection Needle is intended for use as an accessory for currently marketed endoscopes and provides delivery of injectable materials during an endoscopic procedure. The type of material to be injected is dependent on the nature of the procedure, but may include delivery of injectable bulking agent during endoscopic procedures, sclerosing agents during esophagoscopic and gastroscopic procedures, local anesthetics during cystoscopic or laryngoscopic procedures, or saline or contrast media during colonoscopic procedures.

The CMT Endoscopic Injection Needle is provided sterile and is intended for single use only.

#### Indication

The CMT Endoscopic Injection Needle is an accessory for currently marketed endoscopes to allow delivery of injectable materials into tissues during an endoscopic procedure.

### Technological Characteristics

The CMT Injection Needle consists of a stainless steel needle attached to a plastic cannula. A plastic luer lock hub is molded onto the proximal end where a standard syringe can be attached for injection of materials through the lumen of the needle into tissue. Multiple needle lengths are available to accommodate the length of the endoscope channel.

## Summary

In summary, CMT believes the above listed predicate devices and the CMT Needle are substantially equivalent based on design, materials, methods of fabrication and indications for use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo is a circle with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is an abstract image of an eagle. The eagle is facing to the right and has three lines that represent its wings.

NOV - 5 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Robert W. Johnson Vice President, Regulatory and Clinical Affairs and Ouality Assurance Carbon Medical Technologies, Inc. 1290 Hammond Road ST PAUL MN 55110-5867

Re: K042615

Trade/Device Name: Carbon Medical Technologies Endoscopic Injection Needle Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 FBK Dated: September 22, 2004 Received: September 24, 2004

Dear Mr. Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we have it wear your bection by equivalent (for the indications for use stated in above and nave determined and onedicate devices marketed in interstate commerce prior to the cherosate) to regally mantowe of the Medical Device Amendments, or to devices that have been May 20, 1770, the onlinement the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that ICclassified in accordanted with the per al application (PMA). You may, therefore, market the do not require approval of a provisions of the Act. The general controls provisions of the Act de nec, subject to the general estistion, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket it your de riot is elabories (300 acch additional controls. Existing major regulations affecting your Apployal), It thay be subject to racin aderal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I lease be advised that i 27 i in that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the I cation statues and regulations as amontimited to registration and listing (21 CFR Part 807); labeling Act 3 requirements, mending out noing practice requirements as set forth in the quality systems (QS) (21 CFR Part 807), and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/dsmadsmam.html

Sincerely vours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K042615

Carbon Medical Technologies Endoscopic Injection Needle Device Name:

Indications For Use:

The Carbon Medical Technologies Injection Needle is an accessory for currently marketed endoscopes to allow delivery of injectable materials into tissue during an endoscopic procedure

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use ___ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C. Brogdon
(Division Sign-Off)

Division of Reproduc and Radiological Devi 510(k) Number

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda-staging.innolitics.com/device/K042615](https://fda-staging.innolitics.com/device/K042615)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
