STANSION MATRIX

K042600 · Spineworks, LLC · MQP · Feb 18, 2005 · Orthopedic

Device Facts

Record IDK042600
Device NameSTANSION MATRIX
ApplicantSpineworks, LLC
Product CodeMQP · Orthopedic
Decision DateFeb 18, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Stansion Matrix is indicated for use in the thoracolumbar spine (i.e., T1 to L5) to replace a portion of a diseased vertebral body that is resected or excised for the treatment of tumors, where the defect is contained within a single vertebral body, in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The device is also indicated for treating fractures of the thoracic and lumbar spine. The Stansion Matrix is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period of time. The device is intended for use with supplemental rigid posterior pedicle screw fixation.

Device Story

Stansion Matrix is a single-piece titanium vertebral body replacement (VBR) device; open-sided design facilitates bone ingrowth. Implanted via posterior approach using provided insertion tool; intended to restore spinal column biomechanical integrity and height following vertebral body resection or fracture. Used in surgical settings by physicians; requires supplemental rigid posterior pedicle screw fixation. Device provides structural support to anterior, middle, and posterior spinal columns; aids in anterior decompression of spinal cord and neural tissues.

Clinical Evidence

No clinical data provided; substantial equivalence supported by performance testing.

Technological Characteristics

Titanium alloy VBR; open-sided geometry for bone ingrowth; available in three sizes; implanted via posterior approach; non-powered, mechanical device.

Indications for Use

Indicated for patients with thoracolumbar (T1-L5) vertebral body defects due to tumor resection/excision or thoracic/lumbar fractures. Requires use with supplemental rigid posterior pedicle screw fixation.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # FEB 1 8 2005 # KO42600 # 510(k) SUMMARY ### SpineWorks LLC's Stansion Matrix™ # Submitter's Name, Address, Telephone Number, And Contact Person SpineWorks, LLC 2802 Florida Street Huntington Beach, CA 92648 (805) 797-3003 Phone: Phone: Facsimile: (714) 960-5250 Douglas W. Neary Contact Person: #### Date Prepared August 2004 #### Name of the Device Stansion Matrix™ Vertebral Body Replacement #### Common or Usual Name Vertebral Body Replacement (MQP) #### Classification Name Spinal Vertebral Body Replacement #### Predicate Devices - Interpore Cross International GEO™ Structure . - Depuy AcroMed, Inc. Stackable Cage™ System . - Osteotech, Inc. VBR . - Orthovita Endoskeleton TA VBR . #### Intended Use The Stansion Matrix is indicated for use in the thoracolumbar spine (i.e., T1 to L5) to replace a portion of a diseased vertebral body that is resected or excised for the treatment of tumors, where the defect is contained within a single {1}------------------------------------------------ K042600 Page Zof Z vertebral body, in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The device is also indicated for treating fractures of the thoracic and lumbar spine. The Stansion Matrix is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period of time. The device is intended for use with supplemental rigid posterior pedicle screw fixation. #### Principles of Operation The device is a single piece titanium VBR that is open on all sides to maximize bone ingrowth. It is available in three sizes. The device is implanted via the posterior approach, according to similar methods as the predicates. An insertion tool is provided with the device to facilitate its implantation. #### Technological Characteristics The technological characteristics of the Stansion Matrix are very similar to the predicates with respect to shape, sizes, materials, and method of use. Like several of the predicates, the Stansion Matrix is made of titanium alloy. The minor differences in the design and the range of available sizes compared to the predicates do not raise any new questions of safety or effectiveness. #### Summary Basis for the Finding of Substantial Equivalence The Stansion Matrix has the same intended use and very similar indications for use to the predicate devices. The Stansion Matrix also is similar to the predicates with respect to its shape, size range, materials, and method of use. Any minor differences in technological features do not raise new issues of safety or effectiveness, as confirmed by performance testing. Therefore, the Stansion Matrix is substantially equivalent to the predicates. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. FEB 1 8 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Public Health Service Spine Works, L.L.C. C/o Jonathan S. Kahan, Esq. Hogan & Harston L.L.P. 555 13" Street N.W. Washington, DC 20004 Re: K042600 Trade/Device Name: Stansion MatrixTM Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: MQP Dated: January 21, 2005 Received: January 21, 2005 Dear Mr. Kahan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Codc of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 -- Mr. Kahan This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincercly vours. D. Mark A. Wilkerson Celia M. Witten, Ph.D., M.D. Director, Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use Statement 510(k) Number (if known): K042600 Device Name: Stansion Matrix™ Indications for Use: The Stansion Matrix™ ("Stansion Matrix") is indicated for use in the thoracolumbar spine (i.e., T1 to L5) to replace a portion of a diseased vertebral body that is resected or excised for the treatment of tumors, where the defect is contained within a single vertebral body, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The device is also indicated for treating fractures of the thoracic and lumbar spine. The Stansion Matrix is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period of time. The device is intended for use with supplemental rigid posterior pedicle screw fixation. Prescription Use _ ど (Part 21 C.F.R. 801 Subpart D) AND/OR Over-The-Counter Use_ (21 C.F.R. 807 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mike A. Mulkerson Division of General, Restorative, and Neurological Devices Page l_of_l 510(k) Number K042600
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...