ESTEEM BLUE POLYISOPRENE SURGICAL GLOVES WITH COATING (CONTAINING CHITOSAN, PROVITAMIN B, GLUCONOLACTONE AND GLYCEROL)

K042574 · Cardinalhealth · KGO · Dec 8, 2004 · General, Plastic Surgery

Device Facts

Record IDK042574
Device NameESTEEM BLUE POLYISOPRENE SURGICAL GLOVES WITH COATING (CONTAINING CHITOSAN, PROVITAMIN B, GLUCONOLACTONE AND GLYCEROL)
ApplicantCardinalhealth
Product CodeKGO · General, Plastic Surgery
Decision DateDec 8, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4460
Device ClassClass 1
Attributes3rd-Party Reviewed

Indications for Use

Esteem Blue Polyisoprene Surgical Gloves with coating are intended for use in environments within hospitals and other healthcare facilities. The gloves are appropriate for use during invasive and non-invasive medical procedures requiring sterility. They are intended to be worn by operating room personnel to protect a surgical wound from contamination.

Device Story

Esteem Blue Polyisoprene Surgical Gloves are sterile, powder-free gloves manufactured from synthetic rubber latex. The gloves feature a coating containing chitosan, provitamin B, gluconolactone, and glycerol. Designed for use by operating room personnel in hospitals and healthcare facilities, the gloves serve as a protective barrier during invasive and non-invasive medical procedures to prevent contamination of surgical wounds. The device is a physical barrier; it does not involve electronic inputs, algorithms, or software.

Clinical Evidence

No clinical data. Bench testing only: primary skin irritation (non-irritating), guinea pig maximization (no sensitization), ultimate elongation and tensile strength (exceeded ASTM D3577-01a2), barrier defects (exceeded 21 CFR 800.20 and ASTM D3577-01a2, AQL 1.5), and residual powder (below 2 mg/glove per ASTM D6124-01).

Technological Characteristics

Synthetic rubber latex surgical glove with coating (chitosan, provitamin B, gluconolactone, glycerol). Powder-free. Sterile. Performance meets ASTM D3577-01a2 for tensile strength/elongation and barrier defects. Residual powder meets ASTM D6124-01.

Indications for Use

Indicated for use by operating room personnel in hospitals and healthcare facilities to protect surgical wounds from contamination during invasive and non-invasive medical procedures requiring sterility.

Regulatory Classification

Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Cardinal Health. The logo consists of a stylized graphic above the company name. The graphic is composed of several curved lines, and the text "CardinalHealth" is written in a simple, sans-serif font. K042574 # DEC - 8 2004 # XIII. SUMMARY OF SAFETY AND EFFECTIVENESS ### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS ESTEEM BLUE POLYISOPRENE SURGICAL GLOVES WITH COATING (CONTAINING CHITOSAN, PROVITAMIN B, GLUCONOLACTONE AND GLYCEROL) | Regulatory Affairs Contact: | Erica Sethi<br>Cardinal Health<br>1500 Waukegan Road, MP-WM<br>McGaw Park, IL 60085 | |-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Telephone: | (847) 785-3337 | | Fax: | (847)785-2461 | | Date Summary Prepared: | 9/13/04 | | Product Trade Name: | Undetermined | | Common Name: | Surgical Glove | | Classification: | Glove, Surgeon's | | Predicate Devices: | Cardinal Health's Esteem Sterile Polyisoprene Surgical<br>Gloves (K011721) | | Description: | Esteem Blue Polyisoprene Surgical Gloves with coating are<br>formulated using Synthetic Rubber Latex. These are offered<br>powder-free and sterile. | | Intended Use: | Esteem Blue Polyisoprene Surgical Gloves with coating are<br>intended for use in environments within hospitals and other<br>healthcare facilities. The gloves are appropriate for use during<br>invasive and non-invasive medical procedures requiring<br>sterility. They are intended to be worn by operating room<br>personnel to protect a surgical wound from contamination. | {1}------------------------------------------------ | Substantial Equivalence: | Esteem Blue Polyisoprene Surgical Gloves with coating are<br>substantially equivalent to Esteem Sterile Polyisoprene Surgical<br>Gloves in that they provide the following characteristics:<br>- same intended use<br>- same sizes, product features, packaging<br>- both made of Synthetic Rubber Latex using similar manufacturing<br>process | |-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Summary of Testing: | | | Test | Result | | Primary Skin Irritation | Gloves are non-irritating. | | Guinea Pig Maximization | Gloves do not display any potential for sensitization. | | Ultimate Elongation<br>& Tensile Strength | Gloves exceed requirements for rubber surgical gloves<br>per ASTM D3577-01a2. | | Barrier Defects | Gloves exceed requirements per 21 CFR<br>§800.20 and ASTM D3577-01a2, AQL = 1.5. | | Data/Test Method | Gloves meet powder level requirements for "Powder Free"<br>designation using ASTM Standard D6124-01-Standard test method<br>for residual powder on medical gloves. Results generated values<br>below 2 mg of residual powder per glove. | {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES - USA Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC - 8 2004 Cardinal Health C/O Mr. Ned Devine Responsible Third Party Official Entela, Incorporated 3033 Madison Avenue, SE Grand Rapids, Michigan 49548-1289 Re: K042574 Trade/Device Name: Esteem Blue Polyisoprene Surgical Glove with Coating (Containing Chitosan, ProvitaminB, Gluconolactone and Glycerol) Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: I Product Code: KGO Dated: November 24, 2004 Received: November 26, 2004 Dear Mr. Devine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 - Mr. Devine Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, f Susan Tuono Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K042574 Esteem Blue Polyisoprene Surgical Glove with coating (containing Device Name: Chitosan, ProvitaminB, Gluconolactone and Glycerol). Indications For Use: These gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination in environments within hospitals and other healthcare facilities. These gloves are appropriate for use during invasive as well as non-invasive medical procedures requiring sterility. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use_X (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Suptie Michie Cnd. (Division Sign-Off) (Division Sign-On) Division of Anesthesiology, General Hospital, Division Control, Dental Devices Page 1 of _ _ _ _ _ 510(k) Number. K042574
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...