SUBMARINE PLUS PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER, MODEL SUVXXXXXXXXX
Applicant
Invatec Innovative Technologies
Product Code
DQY · Cardiovascular
Decision Date
Nov 8, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SUBMARINE PLUS Balloon Dilatation Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra popliteal, renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
Device Story
SUBMARINE PLUS is a coaxial PTA catheter with a semi-compliant inflatable balloon at the distal tip; designed for percutaneous transluminal angioplasty. The device is tracked over a 0.018-inch guidewire to the target vessel site. The balloon is inflated using a contrast medium/saline mixture to dilate stenoses or obstructive lesions in peripheral arteries and dialysis fistulae. Two radiopaque marker bands under the balloon facilitate fluoroscopic positioning. The distal shaft features a hydrophilic coating to improve trackability. Used by physicians in clinical settings (e.g., interventional radiology or vascular surgery suites). The device provides mechanical dilation of vascular obstructions, potentially restoring blood flow and improving patient outcomes in peripheral arterial disease or dialysis access maintenance.
Clinical Evidence
Bench testing only. Testing included balloon compliance, burst pressure, fatigue, shaft burst pressure, bond strength, catheter dimensions, and guidewire/introducer compatibility. No clinical data presented.
Technological Characteristics
Coaxial lumen catheter with semi-compliant balloon; hydrophilic coating on distal shaft; radiopaque marker bands; compatible with 0.018" guidewires; available in diameters 2.0-7.0mm and lengths 20-120mm; usable lengths 90cm or 130cm. Biocompatible materials.
Indications for Use
Indicated for patients requiring dilation of stenoses in iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Cordis SAVVY™ Balloon Dilatation Catheter
Submission Summary (Full Text)
{0}------------------------------------------------
K042537
page 1 of 2
## 510(k) Summary
## SUBMARINE PLUS Percutaneous Transluminal Angioplasty (PTA) Catheter
| 510(k) Summary | This summary of 510(k) safety and effectiveness information is<br>being submitted in accordance with the requirements of 21 C.F.R §<br>807.92. |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant<br>(Manufacturer) | Invatec Innovative Technologies<br>Via Martiri della Libertà, 7<br>25030 Roncadelle (BS) Italy<br>Tel: +39 030 258 93 11<br>Fax: +39 030 258 93 12<br>www.invatec.com<br>info@invatec.com |
| Submitter | ev3 Inc.<br>4600 Nathan Lane North<br>Plymouth, MN 55442<br>Tel: (763) 398 7000<br>Fax: (763) 398 7200 |
| Contact Person | Mike Winegar<br>Tel: (763) 398-7225<br>Fax: (763) 398-7200<br>E-mail: mwinegar@ev3.net |
| Date Prepared | September 17th, 2004 |
| Device Trade Name | SUBMARINE PLUS Percutaneous Transluminal Angioplasty<br>(PTA) Catheter |
| Device Common Name | Peripheral Transluminal Angioplasty (PTA) Catheter |
| Classification Name | 21 CFR 870.1250 Percutaneous Catheter |
| Device Classification | Regulatory Class: Class II<br>Product Code: LIT |
| Classification Panel | Cardiovascular |
| Predicate Device | Cordis SAVVY™ Balloon Dilatation Catheter |
| Intended use | The SUBMARINE PLUS Balloon Dilatation Catheter is intended to<br>dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra<br>popliteal, renal arteries and for the treatment of obstructive lesions<br>of native or synthetic arteriovenous dialysis fistulae. |
| Device Description | The SUBMARINE PLUS Balloon Dilatation Catheter is a catheter<br>with a semi-compliant inflatable balloon mounted at the distal tip.<br>It has a coaxial lumen. The central lumen of the catheter, which<br>terminates at the distal tip, is used to pass the catheter over a<br>guidewire with a maximum outer diameter of 0.018 inch (0.45 mm).<br>The other lumen is the balloon inflation lumen, which is used to<br>inflate and deflate the dilatation balloon with a mixture of contrast<br>medium and saline solution. The catheter tapers beneath the balloon<br>segment to achieve the lowest possible deflated profile. Two<br>radiopaque marker bands are placed under the balloon segment of<br>the catheter shaft to provide visual reference points for balloon |
| page dot d | |
| | positioning within the vessel. The distal catheter shaft is hydrophilic<br>coated (balloon excluded). The SUBMARINE PLUS Balloon<br>Dilatation Catheter is available in different balloon sizes (balloon<br>diameters of 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0, 5.5, 6.0 and 7.0mm and<br>balloon lengths of 20, 40, 60, 80 and 120mm). Nominal balloon<br>diameter and length are printed on the hub. The maximum<br>recommended guidewire diameter is 0.018" and the usable catheter<br>length is 130 or 90 cm. |
| Biocompatibility | All material used in SUBMARINE PLUS Percutaneous<br>Transluminal Angioplasty (PTA) Catheter are biocompatible based<br>on the biocompatibility testing. |
| Performance data | In vitro testing was conducted to demonstrate the safety and<br>effectiveness of SUBMARINE PLUS Percutaneous Transluminal<br>Angioplasty (PTA) Catheter. The testing included balloon<br>compliance, balloon burst pressure, balloon fatigue, shaft burst<br>pressure, bond strength, catheter dimensions and guidewire and<br>introducer compatibility. |
| Summary of Substantial<br>Equivalence | The SUBMARINEPLUS Balloon Dilatation Catheter is similar to<br>the predicate with respect to the intended use, and with respect to<br>their physical characteristics, such the catheter and balloon<br>dimensions and catheter design and material. SAILOR PLUS<br>Balloon Dilatation Catheter is therefore substantially equivalent to<br>the Predicate device. Further, the mechanical and biocompatibility<br>testing data is indicative of the safety and effectiveness of<br>SUBMARINE PLUS Balloon Dilatation Catheter. Therefore based<br>on its own performance characteristics and the technological<br>similarities, SUBMARINE PLUS Balloon Dilatation Catheter is<br>expected to perform similar to the predicate device and other<br>comparable, currently marketed PTA catheters. |
| Conclusion | SUBMARINE PLUS Balloon Dilatation Catheter is substantially<br>equivalent to the Predicate device and other currently marketed PTA<br>catheters. |
Submarine Plus OTW PTA Balloon Catheter
Traditional 510(k) 510(k) Summary
{1}------------------------------------------------
.
·
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image is a black and white circular seal. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 8 2004
Invatec Innovative Technologies c/o Mr. Mike Winegar Vice President, International RA ev3 Inc. 4600 Nathan Lane North Plymouth, MN 55442
Re: K042537
KU42337
SUBMARNE PLUSTM Percutaneous Transluminal Angioplasty (PTA) Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: DQY Dated: September 17, 2004 Received: September 20, 2004
Dear Mr. Winegar:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 310(x) premained in the tabstantially equivalent (for the indications
referenced above and have determined the device is substantial in intersta referenced above and nave declinined the arrased predicate devices marketed in interstate for use stated in the enclosure) to regally manced producal Device Amendments, or to
commerce prior to May 28, 1976, the enactment at the Federal Food. Dou commerce prior to May 28, 1970, the chactinent and of the Federal Food. Drug, devices that have been recults in accordance with as proval application (PMA).
and Cosmetic Act (Act) that do not require approval assesses approvisions of the Act. The and Cosmetic Act (Act) that do not require approval or a provisions of the Act. The
You may, therefore, market the device, subject to the general controls provisions of the You may, therefore, market the device, subject to the gentlation annual registration, listing of
general controls provisions of the Act include requirements misbranding and general controls provisions of the Act morate requirements.
devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) mo exisi crows in trepulations affecting your device can
may be subject to such additional controls. Tim may be subject to such additions. Laisung major regal to 898. In addition, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. be found in the Code of Pederal Regaraters, concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Mike Winegar
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease be deviloed alla i termination that your device complies with other requirements of the Act that I Dri has made a active and regulations administered by other Federal agencies. You must or uny I cather the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CFK in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in also quarty 25 cool provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product including in begin marketing your device as described in your Section 510(k) This icter will anow you to begin inding of substantial equivalence of your device to a legally promaticated predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you docure of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bhimmer for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number: K04 2537
Device Name: SUBMARINE™ PLUS PTA Balloon Dilatation Catheter
Indications For Use:
The SUBMARINE™ PLUS Balloon Dilatation Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blommor
Sivision Sign-Off) Division of Cardiovascular Devices 510(k) Number_
Page I of
Submarine Plus PTA Balloon Catheter Traditional 510(k) 510(k) Body
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.