K042491 · Becton, Dickinson & CO · LJS · Nov 22, 2004 · General Hospital
Device Facts
Record ID
K042491
Device Name
BD ONECATH PERIPHERALLY INSERTED CENTRAL CATHETER
Applicant
Becton, Dickinson & CO
Product Code
LJS · General Hospital
Decision Date
Nov 22, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The BD OneCath™ peripherally inserted central catheters are used to sample blood and administer fluids into the central vascular system such as medications, fluids, nutrients, blood and blood products.
Device Story
BD OneCath is a peripherally inserted central catheter (PICC) for intermediate-term central venous access. Device consists of radiopaque polyurethane tubing inserted into peripheral veins and advanced to the superior or inferior vena cava. Used for blood sampling and infusion of medications, fluids, nutrients, and blood products. Intended for clinical settings where IV therapy is required for up to 6-8 weeks. Operated by healthcare professionals. Provides central venous access to facilitate therapy delivery and blood collection, benefiting patients by reducing the need for repeated peripheral venipuncture.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility assessments and physical performance testing (pull strength and junction integrity).
Technological Characteristics
Radiopaque polyurethane tubing; improved material formulation for enhanced radiopacity and alcohol resistance compared to predicate. Form factor is a percutaneous catheter for central venous access. Sterilization method not specified.
Indications for Use
Indicated for patients requiring intermediate-term central venous access for blood sampling and administration of medications, fluids, nutrients, and blood products. Suitable for adults (via basilic, cephalic, or median cubital veins), neonates (via scalp veins), and pediatric patients (via saphenous, popliteal, or femoral veins).
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
Predicate Devices
BD L-Cath PICC
Submission Summary (Full Text)
{0}------------------------------------------------
K\u03b142491 left 2
BD Medical Systems 9450 South State Street andv Utah 84070 tel: 801.565.2300 fax: 801.565.2740 www.bd.com
NOV 2 2 2004
# Attachment D
Image /page/0/Picture/4 description: The image shows the BD logo. The logo consists of a circular emblem on the left and the letters "BD" on the right. The emblem features a stylized image of a person with arms raised, with rays emanating from above. The letters "BD" are in a bold, sans-serif font.
Helping all people live healthy lives
#### Premarket Notification [510(k)] Summary
| Submitter: | Becton Dickinson Infusion Therapy Systems Inc. |
|------------------------|------------------------------------------------|
| Address: | 9450 South State Street<br>Sandy, UT 84070 |
| Contact Person: | Leslie Wood<br>Manager, Regulatory Affairs |
| Telephone Number: | (801) 565-2504 |
| FAX Number: | (801) 565-2749 |
| Date Summary Prepared: | August 30, 2004 |
| Trade Name: | BD OneCath™™ |
| Common Name: | Peripherally Inserted Central Catheter or PICC |
| Classification Name: | Intravascular, therapeutic, long-term catheter |
| Classification: | Class II<br>880.5970 - LJS |
| Predicate Device: | BD L-Cath PICC |
# Description of the BD OneCath Peripherally Inserted Central Catheter:
Device selection is an important component of intravenous therapy. Some of the patient considerations that are included in this decision are the: (1) length and type of therapy, (3) adequacy of venous access, (4) lifestyle and activity, and (5) setting in which therapy will be administered. Peripherally inserted central catheters (PICC) are selected for intermediate length therapy, while tunneled percutaneous catheters and subcutaneous ports are selected for long-term central venous access.
A PICC is a central venous catheter that is inserted into a peripheral vein and advanced to the superior vena cava. The basilic, cephalic and median cubital veins of the arm are the preferred locations for PICC insertion in adults; scalp veins such as the posterior auricular and temporal veins may be used for neonatal patients. For pediatric patients, the saphenous, popliteal, and femoral veins of the leg may be used, with the tip location in the inferior vena cava. The PICC is generally selected for IV therapy that is prescribed for less than 6 to 8 weeks. Although a PICC may indwell for a shorter period, the average indwelling time for a PICC is 2-6 months.
{1}------------------------------------------------
The BD OneCath PICC products are similar in design to other PICC products. The catheter tubing is a radiopaque, polyurethane material that is inserted into the vascular system via a catheter introducer.
7.47
### Intended Use of the BD OneCath Peripherally Inserted Central Catheter:
The BD OneCath™ peripherally inserted central catheters are used to sample blood and administer fluids into the central vascular system such as medications, fluids, nutrients, blood and blood products.
#### Technological Characteristics Comparison:
The polyurethane used for the BD OneCath product has improved radiopacity and resistance to alcohol than the polyurethane used for the current L-Cath product. These improved characteristics will make the BD OneCath PICC more competitive in this market.
#### Nonclinical Tests Support Substantial Equivalence:
Biocompatibility testing of the new polyurethane tubing and physical testing on the areas of potential failure (such as pull strength and the security of junctions between the extension tubing and molded parts) were conducted.
## Conclusions from Nonclinical Tests:
The BD OneCath PICC is substantially equivalent to the BD L-Cath PICC.
The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence found in the Federal Food, Drug, and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without premarket approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The seal is black and white.
NOV 2 2 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Leslie Wood Manager, Regulatory Affairs Becton Dickinson Infusion Therapy Systems, Incorporated 9450 South State Street Sandy, Utah 84070
Re: K042491
Trade/Device Name: BD OneCath™ Peripherally Inserted Central Catheter Regulation Number: 880.5970 Regulation Name: Percutaneous Implanted Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: September 13, 2004 Received: September 14, 2004
Dear Ms. Wood:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Wood
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications For Use
510(k) Number (if known): K042491
BD OneCath™ Peripherally Inserted Central Catheter Device Name:
Indications For Use:
The BD OneCath™ peripherally inserted central catheters are used to sample blood and administer fluids into the central vascular system such as medication, fluids, nutrients, blood and blood products.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
1 of 1
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Tuan Dima
(Division Sign-Off) (Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: _ K P Y 2 40)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.