← Product Code [QEY](/productcode/QEY) · K042456

# Lysus Infusion System (K042456)

_Ekos Corp. · QEY · Sep 23, 2004 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K042456

## Device Facts

- **Applicant:** Ekos Corp.
- **Product Code:** [QEY](/productcode/QEY.md)
- **Decision Date:** Sep 23, 2004
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.5150
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Lysus® Infusion System is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.

## Device Story

Lysus Infusion System consists of disposable infusion catheter with removable ultrasound core and control instrument. Catheter features multiple side holes along treatment zone for fluid delivery. Ultrasound core contains up to 30 evenly spaced elements; thermal sensors monitor transducer temperature. System delivers energy to treatment zone during infusion. Used in clinical settings by physicians for peripheral vasculature procedures. Device facilitates targeted delivery of thrombolytics or other fluids to treat vascular conditions.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

System comprises disposable infusion catheter and energy control instrument. Catheter includes multiple side holes; ultrasound core features up to 30 elements. Thermal sensors integrated for temperature monitoring. Energy-based infusion system.

## Regulatory Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

## Predicate Devices

- Lysus® Infusion System ([K041629](/device/K041629.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font below.

January 7, 2022

EKOS Corporation Jocelyn Kersten Regulatory Affairs Manager 22030 20th Avenue SE, Suite 101 Bothell, Washington 98021

Re: K042456

Trade/Device Name: Lysus Infusion System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEY, KRA

Dear Jocelyn Kersten:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated September 23, 2004. Specifically, FDA is updating this SE Letter as an administrative correction because FDA has created a new product code to better categorize your device technology.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075, Gregory.Oconnell(@FDA.HHS.gov.

Sincerely,

Digitally signed by Gregory W. Gregory W. O'connell -O'connell -S Date: 2022.01.07 13:32:41 -05'00'

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Public Health Service

Image /page/1/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling three stylized human figures or birds in flight, stacked vertically.

SEP 2 3 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

EKOS Corporation c/o Ms. Jocelyn Kersten Director, Regulatory Affairs 22030 20th Avenue, SE, Suite 101 Bothell, WA 98021

Re: K042456

Lysus Infusion System Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II (two) Product Code: KRA Dated: September 9, 2004 Received: September 10, 2004

Dear Ms. Kersten:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 – Ms. Jocelyn Kersten

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I four ove as note a determination that your device complies with other requirements of the Act that I Dr Has Interes and regulations administered by other Federal agencies. You must or uny 1 edetar station in the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of K Fur 0077, havemes (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product taxation of allow you to begin marketing your device as described in your Section 510(k) I mis letter with and in your le FDA finding of substantial equivalence of your device to a legally promated predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address

(2017119 097) 6. a. gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Dauna R. Vochner

( ) Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known): ¥04 乙456

Device Name: Lysus® Infusion System

Indications For Use: The Lysus® Infusion System is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Donna R. Jackson

(Division Sign-Off) Division of Cardiovascular Devices Page 1 of 1 __

510(K) Number K042456

{4}------------------------------------------------

SEP 2 3 2004

## Section 4. 510(k) Summary

General Provisions

| Submitter's Name and Address: | EKOS Corporation<br>22030 20th Ave. SE<br>Suite 101<br>Bothell, WA 98021       |
|-------------------------------|--------------------------------------------------------------------------------|
| Contact Person:               | Jocelyn Kersten<br>425-482-1108<br>425-482-1109 (fax)<br>jkersten@EKOSCORP.com |
| Classification Name:          | Catheter, Continuous Flush (KRA)                                               |
| Common or Usual Name:         | Continuous Flush Catheter                                                      |
| Proprietary Name:             | Lysus® Infusion System                                                         |
| Name of Predicate Device:     | Lysus® Infusion System                                                         |
| 510(k) Reference No.:         | K041629                                                                        |

#### Device Description

The system consists of a disposable infusion catheter with removable ultrasound core and an instrument that generates and controls the delivery of energy to the catheter. The infusion catheter contains multiple side holes distributed over the length of the treatment zone. The ultrasound core contains up to 30 ultrasound elements, evenly spaced over the treatment zone. Thermal sensors in the treatment zone monitor transducer temperature.

#### Intended Use

The Lysus® Infusion System is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.

#### Summary of Technological Characteristics

The proposed Lysus Infusion System is similar in construction to the previously cleared Lysus Infusion System.

### Test Summary

The proposed Lysus Infusion System is considered to be substantially equivalent to the currently marketed Lysus Infusion System based on a comparison of the intended uses and designs and results of the testing and evaluations performed.

---

**Source:** [https://fda-staging.innolitics.com/device/K042456](https://fda-staging.innolitics.com/device/K042456)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).