K042439 · Diamond Age Systems, Inc. · GEX · Oct 7, 2004 · General, Plastic Surgery
Device Facts
Record ID
K042439
Device Name
AZURYT MODEL CLT 1401, CO2 SURGICAL LASER SYSTEM
Applicant
Diamond Age Systems, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 7, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Azuryt Model CTL 1401, CO2 Surgical Laser System is indicated for the incision, excision, ablation, vaporization and coaqulation of body soft tissues, including intraoral tissues. These systems can be used in a variety of medical specialties, including general and plastic surgery, dentistry, oral/maxillofacial surgery, dermatology, gastroenterology, gynecology, otorhinolaryngology, aesthetical surgery, neurosurgery, oculoplastic, orthopedic, pulmonary/thoracic surgery, and urology.
Device Story
CO2 surgical laser system; utilizes 10.6 µm wavelength laser for soft tissue surgery. Components include power supply, cooling system, laser head, articulated arm, handpiece, microcomputer, and control panel. Operated by clinicians in various surgical settings. Laser energy delivered via handpiece to target tissue for incision, excision, ablation, vaporization, or coagulation. Microcomputer manages system parameters via control panel. Benefits include precise tissue interaction for diverse surgical applications.
Clinical Evidence
Bench testing only. Compliance with international safety standards (IEC 60825-1, IEC 60601-2-22, IEC 60601-1, IEC 60601-1-2) and US federal performance standards (21 CFR 1040.10, 21 CFR 1040.11) for Class IV laser products.
Technological Characteristics
CO2 laser system; 10.6 µm wavelength. Components: power supply, cooling system, sealed CO2 laser head, articulated arm, handpiece, microcomputer. Class IV laser product. Complies with IEC 60825-1, IEC 60601-2-22, IEC 60601-1, IEC 60601-1-2. Standalone system.
Indications for Use
Indicated for incision, excision, ablation, vaporization, and coagulation of soft tissues (including intraoral) across multiple surgical specialties including general, plastic, dental, oral/maxillofacial, dermatological, gastrointestinal, gynecological, ENT, aesthetic, neurosurgical, oculoplastic, orthopedic, pulmonary/thoracic, and urological procedures.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
## Premarket Notification [510(k)] Summary (per 21 CFR 807.92)
#### 1. Submitted by:
K042439
Diamond Age Systems, Inc. 3775 South Laurel Way Chandler, AZ 85249 Phone: (866) 546-5444
Contact Person: Geoffrey D. Swank Vice-President of Laser Operations Phone: (303) 263-3307 Fax: (303) 985-7957
Date Prepared: 29 August 2004
#### 2. Device Name
Trade/Proprietary Name: Common/Usual Name: Classification Name:
Azurvt Model CTL 1401. CO2 Surgical Laser System General Surgical Laser System Laser Instrument, Surgical, Powered (per 21 CFR 878.4810)
#### 3. Predicate Device:
The Azuryt Model CTL 1401, CO2 Surgical Laser System is substantially equivalent to other laser systems on the market, such as the Deka Smart CO2 laser, cleared under K031224 and K031440; and the Lumenis LX-20 CO2 Surgical Laser System, cleared under K896478, K953074, and K960475.
#### 4. Intended use of the device
The Azuryt Model CTL 1401, CO2 Surgical Laser System is indicated for the incision, excision, ablation, vaporization and coaqulation of body soft tissues, including intraoral tissues. These systems can be used in a variety of medical specialties, including general and plastic surgery, dentistry, oral/maxillofacial surgery, dermatology, gastroenterology, gynecology, otorhinolaryngology, aesthetical surgery, neurosurgery, oculoplastic, orthopedic, pulmonary/thoracic surgery, and urology.
{1}------------------------------------------------
#### Description of the Device 5.
The, Azuryt Model CTL 1401 CO₂ Surgical Laser System is a CO₂ laser consisting of a power supply, cooling system, laser head, articulated arm, handpiece, microcomputer and control panel.
### Summary of the technological characteristics of the device 6. compared to the predicate device.
Like the predicate devices, the Azuryt Model CTL 1401 is a CO2 surgical laser system, utilizing a laser with a wavelength of 10.6 µm. Each of these devices contains a power supply, a cooling system, a sealed CO2 laser head delivering a beam with a wavelength of 10.6 um, an arm to deliver the laser beam, a handpiece and a variety of accessories to address specific surgical needs.
#### 7. Testing
The manufacturer of the Azuryt Model CTL 1401certifies that its device complies with the following international standards:
EN60825-1. IEC 60825-1. Safety of laser products - Part 1: Equipment classification, requirements and user's guide.
EN60601-2-22, IEC 60601-2-22, Medical Electrical Equipment -- Part 2: Particular requirements for the safety of diagnostic and therapeutic laser equipment.
EN60601-1. IEC 60601-1. Medical Electrical Equipment - Part 1: General requirements for safety, as amended by amendment 1 (1991) and amendment 2 (1995).
EN60601-1-2, IEC 60601-1-2, Medical Electrical Equipment - Part 1: General requirements for safety - 2. Collateral Standard: Electromagnetic compatibility - Requirements and tests
The device also complies with the European Medical Device Directive 92/42/EEC and the US Federal Performance Standards 21 CFR 1040.10 and 21 CFR 1040.11 for Class IV Laser Products, except for deviations pursuant to Laser Notice No. 50, dated 26 July 2001.
#### 8. Conclusions
Based upon the testing and comparison to the predicate devices, the Azuryt Model CTL 1401 has the same intended uses, with similar technological characteristics as the predicate devices. The system performs as intended and raises no new safety or effectiveness issues.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Diamond Age Systems, Inc. c/o Mr. Ned E. Devine Entela. Inc. 3033 Madison Avenue, SE Grand Rapids, Michigan 49548
Re: K042439
Trade/Device Name: Azuryt Model CTL 1401, CO2 Surgical Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: September 21, 2004 Received: September 22, 2004
0C1 7 - 2004
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Ned Devine
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will and in your le FDA finding of substantial equivalence of your device to a legally promated nevice results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you attire of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general micronal and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Sincerely yours,
Mark A. Millerson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K042439
Device Name: Azuryt Model CTL 1401, CO2 Surgical Laser System
Indications For Use:
The Azuryt Model CTL 1401, CO2 Surgical Laser System is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues, including intraoral tissues. These systems can be used in a variety of medical specialties, including general and plastic surgery, dentistry, oral/maxillofacial surgery, dermatology, gastroenterology, gynecology, otorhinolaryngology, aesthetical surgery, neurosurgery, oculoplastic, orthopedic, pulmonary/thoracic surgery, and urology.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number K042439
Page | of |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.