K042436 · Smith & Nephew, Inc. · KTT · Oct 7, 2004 · Orthopedic
Device Facts
Record ID
K042436
Device Name
IIIZAROV PULLEY SYSTEM
Applicant
Smith & Nephew, Inc.
Product Code
KTT · Orthopedic
Decision Date
Oct 7, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
External fixation devices are used on adults or pediatric patients as required. External fixation systems consist of various components that are used to build fixator assemblies unique to the patient's need. These devices are modular, therefore, a multitude of different fixator frame configurations are possible. External fixation devices are used for the following indications: 1. Post-Traumatic joint contracture which has resulted in loss of range of motion 2. Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction 3. Open and closed fracture fixation 4. Pseudoarthrosis of long bones 5. Limb lengthening by epiphyseal or metaphyseal distraction (not applicable for COMPASS Universal Hinge or JET-X™ Fixator) 6. Correction of bony or soft tissue deformities (not applicable for COMPASS Universal Hinge) 7. Correction of segmental bony or soft tissue defects 8. Joint arthrodesis 9. Infected fractures or nonunions 10. Mini external fixator systems are indicated for the management of comminuted intra-articular fractures of the distal radius. 11. Calandruccio devices are indicated for arthrodesis of the ankle or subtalar joints, as well as some select fractures, nonunion, or osteotomy of the distal tibia; and acute transverse fractures or nonunion of the distal tibia. The indications for use listed above cover many of the external fixation systems marketed by Smith & Nephew. The indications for the Ilizarov Pulley System are pseudoarthrosis of long bones, included as number 4 above. These indications are similar to the indications of the predicate devices. The device is intended for single use.
Device Story
External fixation accessory; used in conjunction with Ilizarov External Fixation System. Components: stainless steel cables (1.0mm, 1.5mm, 1.8mm), pulleys, ratchet. Cables routed percutaneously through soft tissue and bone segments; secured via bone screws or eyelets. Ratchet mechanism attaches to external fixation half-ring construct; manual adjustment of ratchet tightens cable to move bone segments. Used by surgeons in clinical settings to manage pseudoarthrosis. Benefits: reduces required number of half-pins and wires; minimizes soft tissue displacement; decreases joint stiffness/movement limitation.
Clinical Evidence
No clinical data. Substantial equivalence based on design, material, and intended use comparisons to legally marketed predicate devices.
Indicated for adults and pediatric patients requiring treatment for pseudoarthrosis of long bones.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Smith & Nephew (formerly Richards) External Fixation System (Ilizarov) (K870961)
{0}------------------------------------------------
# 510(K) SUMMARY ILIZAROV PULLEY SYSTEM
K 042436
SUBMITTER'S NAME: SUBMITTER'S ADDRESS: SUBMITTER'S TELEPHONE NUMBER: CONTACT PERSON: DATE SUMMARY PREPARED: TRADE OR PROPRIETARY DEVICE NAME: COMMON OR USUAL NAME: CLASSIFICATION NAME:
0C7 7 - 2004
Smith & Nephew, Inc., Orthopaedic Division 1450 Brooks Road, Memphis, TN 38116 901-399-6670 John Reabe September 7, 2004 llizarov Pulley System External fixation system Single/multiple component metallic bone fixation appliances and accessories Class II Orthopedic/87
DEVICE CLASS: PANEL CODE:
## DEVICE INFORMATION:
## INTENDED USE:
External fixation devices are used on adults or pediatric patients as required. External fixation systems consist of various components that are used to build fixator assemblies unique to the patient's need. These devices are modular, therefore, a multitude of different fixator frame configurations are possible. External fixation devices are used for the following indications:
- 1. Post-Traumatic joint contracture which has resulted in loss of range of motion
- 2. Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction
- 3. Open and closed fracture fixation
- 4. Pseudoarthrosis of long bones
- 5. Limb lengthening by epiphyseal or metaphyseal distraction (not applicable for COMPASS Universal Hinge or JET-X™ Fixator)
- 6. Correction of bony or soft tissue deformities (not applicable for COMPASS Universal Hinge)
- 7. Correction of segmental bony or soft tissue defects
- 8. Joint arthrodesis
- Infected fractures or nonunions 9.
- 10. Mini external fixator systems are indicated for the management of comminuted intra-articular fractures of the distal radius.
- 11. Calandruccio devices are indicated for arthrodesis of the ankle or subtalar joints, as well as some select fractures, nonunion, or osteotomy of the distal tibia; and acute transverse fractures or nonunion of the distal tibia.
The indications for use listed above cover many of the external fixation systems marketed by Smith & Nephew. The indications for the Ilizarov Pulley System are pseudoarthrosis of long bones, included as number 4 above. These indications are similar to the indications of the predicate devices. The device is intended for single use.
## DEVICE DESCRIPTION:
The Ilizarov Pulley System consists of cables, pulleys and a ratchet. There are two styles of cable, with or without an eyelet on one end. The cables without an evelet are routed through the bone segment and perculaneously through the soft tissue. The cables with an eyelet are secured to the bone segment with a 3.5mm or 5.0mm diameter bone screw through the evelet. The cable is then routed percutaneously through the soft tissue. Cables are available in diameters of 1.0mm, 1.5mm and 1.8mm and a length of 1200mm. The 1.0mm cables feature an evelet on one end and the 1.5mm and 1.8mm cables are straight (without an eyelet).
The cable is routed through pulleys and connected to either a telescopic rod (predicate device) or a ratchet (new device). The telescopic rod or ratchet is attached to the external fixation half ring
{1}------------------------------------------------
construct. The telescopic rod or ratchet is turned to tighten the cable and move the bone seament.
The Ilizarov Pulley System is used in conjunction with the Ilizarov External Fixation System that consists of half rings, threaded rods, bolts, nuts, washers, pin clamps, wire fixation bolts, half pins and wires. The Ilizarov Pulley System will be attached to the Ilizarov External Fixation System for the indication for use of pseudoarthrosis of long bones.
The advantages of using the Ilizarov Pulley System include the ability to use fewer half pins and wires. Fewer half pins and wires will create less soft tissue displacement and less limitation to joint movement (joint stiffness).
## SUBSTANTIAL EQUIVALENCE INFORMATION:
The Ilizarov Pulley System is similar to the Smith & Nephew (formerly Richards) External Fixation System (Ilizarov) (K870961) and Smith & Nephew External Fixation System (K994143) in that all the devices are used for the same indications and consist of half pins, wires, rings and various components to construct an external fixation frame. The predicate devices use half pins, wires, telescopic rods, threaded rods and slotted threaded rods to adjust the external fixation frame and transport the bone segment. The Ilizarov Pulley System uses half pins, wires, cables, pulleys and a ratchet to adjust the external fixation frame and transport the bone segment.
The Ilizarov Pulley System includes a stainless steel cable. The stainless steel cable is similar to the Smith & Nephew Orthopaedic Cable System (K87516) in that both cables are made of stainless steel and the diameters are similar (1.5mm). The Ilizarov Pulley System cable is longer (1200mm) than the predicate cable (610mm). The indications for use are not the same, but the predicate cable includes indications for general orthopaedic repair procedures and any area in which monofilament wiring is used.
The Ilizarov Pulley System is substantially equivalent to the predicate devices. The differences between the lizarov Pulley System and predicate devices do not affect safety and effectiveness.
#### SUMMARY OF TECHNOLOGICAL COMPARISON:
The Ilizarov Pulley System is substantially equivalent to the predicate devices listed in the previous section in terms of material, indications for use and design.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized emblem featuring a symbol that resembles a bird or a stylized human figure, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
0CT 7 - 2004
Mr. John Reabe Director of Regulatory Affairs Smith & Nephew Inc. Orthopaedic Division 1450 Brooks Road Memphis, Tennessee 38116
Re: K042436
Trade/Device Name: Illizarov Pulley System Regulation Number: 21CFR 888. 3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: KTT Dated: September 7, 2004 Received: September 8, 2004
Dear: Mr. Reabe
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section10(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark N Milken
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Mark A. Milken
510(k) number (if known):
Device Name: Ilizarov Pulley System
## Indications for Use:
Division of General, Restorative. and Neurological Devices
510(k) Number K042436
External fixation devices are used on adults or pediatric patients as required. External fixation systems consist of various components that are used to build fixator assemblies unique to the patient's need. These devices are modular, therefore, a multitude of different fixator frame configurations are possible. External fixation devices are used for the following indications:
- 1. Post-Traumatic joint contracture which has resulted in loss of range of motion
- 2. Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction
- 3. Open and closed fracture fixation
- 4. Pseudoarthrosis of long bones
- 5. Limb lengthening by epiphyseal or metaphyseal distraction (not applicable for COMPASS Universal Hinge or JET-X™ Fixator)
- 6 Correction of bony or soft tissue deformities (not applicable for COMPASS Universal Hinge)
- 7. Correction of segmental bony or soft tissue defects
- 8. Joint arthrodesis
- 9. Infected fractures or nonunions
- 10. Mini external fixator systems are indicated for the management of comminuted intraarticular fractures of the distal radius.
- 11. Calandruccio devices are indicated for arthrodesis of the ankle or subtalar joints, as well as some select fractures, nonunion, or osteotomy of the distal tibia; and acule transverse fractures or nonunion of the distal tibia.
The indications for use listed above cover many of the external fixation systems marketed by Smith & Nephew The astions for the extential includes and the marketed
by Smith & Nephew. The indications for the lizarov Pulley System are pseudoarthrosis of long bones, included as number 4 above. These indications are similar to the indications of the predicate devices. The device is intended for single use.
Prescription Use × (Per 21 CFR 801.109)
OR
Over-the-Counter Use (Optional Format 1-2-96)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.