TMD 0.5ML SAFETY SYRINGE (FA18 SERIES 0.5ML/FA58 SERIES U-100 INSULIN/FA78 SERIES TUBERCULIN)
K042426 · Taiject Medical Device Co., Ltd. · MEG · Sep 22, 2004 · General Hospital
Device Facts
Record ID
K042426
Device Name
TMD 0.5ML SAFETY SYRINGE (FA18 SERIES 0.5ML/FA58 SERIES U-100 INSULIN/FA78 SERIES TUBERCULIN)
Applicant
Taiject Medical Device Co., Ltd.
Product Code
MEG · General Hospital
Decision Date
Sep 22, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TMD™ 0.5ml Safety Syringe (FA18 Series 0.5ml) is intended for injection of fluids into the body after the aspiration of fluid. The TMD™ 0.5ml Insulin Safety Syringe (FA58 Series U-100 Insulin) is intended for use for subcutaneous injection of Insulin. The TMD™ 0.5ml Tuberculin Safety Syringe (FA78 Series Tuberculin) is intended intra-dermal of Tuberculin. All the three syringes above incorporate a safety feature that is designed to aid the reduction of needle stick injuries and the potential of syringe reuse.
Device Story
Single-use, sterile, disposable piston syringe with integrated active safety feature; manually activated by user to protect needle after administration. Input: fluid medication (insulin, tuberculin, or other). Output: delivered injection. Used in clinical or home settings by patients or healthcare providers. Safety mechanism reduces risk of sharps injuries and prevents syringe reuse. Device is a smaller volume (0.5ml) variant of existing predicate product lines.
Clinical Evidence
Bench testing only. Performance data generated in compliance with design control requirements demonstrated equivalence to predicate devices.
Indicated for patients requiring subcutaneous insulin injection, intradermal tuberculin injection, or general fluid injection. Syringes include safety features to reduce needle stick injuries and prevent reuse.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
FA11 Series 1ml/FA51 Series U-100 Insulin/FA71 Series Tuberculin (K031062)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 8 2004
Taiject Medical Device, Company Limited C/O Dr. Jim-Son Chou Vice President Achevé Technology, Incorporated P.O. Box 8853 Newport Beach, California 92658
Re: K042426
Trade/Device Name: TMD 0.5mL Safety Syringe (FA18 Series 0.5mL/FA58 Series U-100 Insulin/FA78 Series Tuberculin) Regulation Number: 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: September 3, 2004 Received: September 8, 2004
Dear Dr. Chou:
This letter corrects our substantially equivalent letter of September 22, 2004 regarding the trade name.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent [{for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
### Page 2 - Dr. Chou
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Carls
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
K042426
1 of 3
### SEP 2 2 2004
## 510(k) Summary of Safety and Effectiveness for the TMDTM 0.5ml Safety Syringe (FA18 Series 0.5ml/FA58 Series U-100 Insulin / FA78 Series Tuberculin) (per 21CFR807.92)
#### 1. SPONSOR
Taiject Medical Device Co., Ltd. 1F, No. 22, KeDung 3 rd Road , Science-Based Industrial Park Chu Nan, MiaoLi, Taiwan 350 Republic of China Tel: 886 37 585599 Fax: 886 37 585589 Contact person: Mr. David Huang Date Prepared: September 3rd, 2004
#### 2. DEVICE NAME
| Proprietary Name: | TMD TM 0.5ml Safety Syringe (FA18 Series 0.5ml/FA58<br>Series U-100 Insulin / FA78 Series Tuberculin) |
|----------------------|---------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Safety Syringe |
| Classification Name: | Piston syringe (FMF)<br>Anti-Stick Syringe (MEG) |
#### Predicate Device (Legally Marketed Device): 3.
Legally Marketed Device:
FA11 Series Iml/FA51 Series U-100 Insulin/FA71 Series Tuberculin) with 510K number K031062 and FA12 Series 3 ml/FA13 Series 5 ml with 510K number K022278
#### DEVICE DESCRIPTION 4.
The TMD 1 0.5ml Safety Syringe (FA18 Series 0.5ml/FA58 Series U-100 Insulin / FA78 Series Tuberculin) is a single use, sterile, disposable syringes that is designed to reduce the risk of sharps injuries. The insulin syringe has scale lines in insulin
Taiject Medical Device, Co., Ltd. Special 510(k) September 3, 2004 TMDTM 0.5ml Safety Syringe (FA18 Series 0.5ml/FA58 Series U-100 Insulin / FA78 Series Tuberculin) Page 18
{3}------------------------------------------------
K042426
2 of 3
units. The Tuberculin syringe has scale lines of Tuberculin.
#### INTENDED USE ડ.
--- The TMD™ 0.5ml Safety Syringe (FA18 Series 0.5ml) is intended for injection of fluids into the body after the aspiration of fluid.
--- The TMD™ 0.5ml Insulin Safety Syringe (FA58 Series U-100 Insulin) is intended for use for subcutaneous injection of Insulin.
--- The TMD™ 0.5ml Tuberculin Safety Syringe (FA78 Series Tuberculin) is intended intra-dermal of Tuberculin.
All the three syringes above incorporate a safety feature that is designed to aid the reduction of needle stick injuries and the potential of syringe reuse.
#### TECHNOLOGICAL CHARACTERISTICS 6.
The technological characteristics are the same as the legally market device, FA11 Series 1ml/FA51 Series U-100 Insulin/FA71 Series Tuberculin) with 510K number K031062 and FA12 Series 3 ml/FA13 Series 5 ml with 510K number K022278
#### 7. PERFORMANCE DATA
Performance data has been generated in compliance with the design control requirement and appropriate standards. The result demonstrated equivalent to the predicate devices.
#### COMPARISON INFORMATION 8.
Comparison of the TMD™ 0.5ml Safety Syringe (FA18 Series 0.5ml/FA58 Series U-100 Insulin / FA78 Series Tuberculin) with Legally Marketed Device TMD™ Safety Syringe (FA12 Series 3ml/FA13 Series 5ml)
| | Submission Device | Legally Market Device |
|--|-------------------|-----------------------|
|--|-------------------|-----------------------|
September 3, 2004 Taiject Medical Device, Co., Ltd. Special 510(k) TMDTM 0.5ml Safety Syringe (FA18 Series 0.5ml/FA58 Series U-100 Insulin / FA78 Series Tuberculin) Page 19
{4}------------------------------------------------
| | TMD™ 0.5ml Safety<br>Syringe (FA18 Series<br>0.5ml/FA58 Series U-100<br>Insulin / FA78 Series<br>Tuberculin) | TMD™ Safety Syringe<br>(FA11 Series 1ml/FA51 Series<br>U100 insulin/FA71 Series<br>Tuberculin ) |
|------------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | As a single use, hypodermic<br>syringe. Safety feature<br>protects after administration. | As a single use, hypodermic<br>syringe. Safety feature<br>protects after administration. |
| Volume (ml) | 0.5ml | 1ml |
| Needles | 25~31Garge | 25~31Garge |
| Gauge | 5/8" or Shorter | 5/8" or Shorter |
| Needle<br>Connection | Fixed needle | Fixed needle |
| Safety<br>Features | Active safety feature,<br>manually activated by users | Active safety feature, manually<br>activated by users |
| Syringe<br>Type | Plunger, Antistick with fixed<br>needles | Plunger, Antistick with<br>hypodermic needles |
| Material | Piston-Butyl Rubber<br>Barrel, plunger, Needle<br>holder - Polypropylene<br>Lubricant | Piston-Butyl Rubber<br>Barrel, plunger, Needle holder<br>- Polypropylene<br>Lubricant |
| Color | Parts- Clear<br>Printing-Black<br>Orange - Insulin<br>Gray - Tuberculin<br>Transparent - Safety syringe | Parts- Clear<br>Printing-Black<br>Orange - Insulin<br>Gray - Tuberculin<br>Transparent - Safety syringe |
In summary, TMD™ 0.5ml Safety Syringe (FA18 Series 0.5ml/FA58 Series U-100 Insulin / FA78 Series Tuberculin) is a smaller version of the legally marketed TMD™ Safety Syringe (FA11 Series 1ml/FA51 Series U-100 Insulin/FA71 Series Tuberclin) and FA12 Series 3 ml/FA13 Series 5 ml with 510K number K022278.
Taiject Medical Device, Co., Ltd. Special 510(k) September 3, 2004 TMD™ 0.5ml Safety Syringe (FA18 Series 0.5ml/FA58 Series U-100 Insulin / FA78 Series Tuberculin) Page 20
{5}------------------------------------------------
K042426
1 of 2.
## Indications for Use
K042426 510(k) Number (if known):
Device Name:TMDTM 0.5ml Safety Syringe (FA58 Series U-100 Insulin)
Indications For Use:
--- The TMD™ 0.5ml Insulin Safety Syringe (FA58 Series U-100 Insulin) is intended for use for subcutaneous injection of Insulin.
This syringe above incorporates a safety feature that is designed to aid the reduction of needle stick injuries and the potential of syringe reuse.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runne
ivision Sian-Off on of Anesthesiology, General Hospital. Infection Control, Dental Devices
Page i of ii
510(k) Number: K092896
Special 510(K) Tailect Medical Device, Co., Ltd. TMDTM 0.5ml Safety Syringe
September 3rd, 2004
{6}------------------------------------------------
K042426
2 of.2
# Indications for Use
K042426 510(k) Number (if known):
Device Name: TMD™ 0.5ml Safety Syringe (FA18 Series 0.5ml/ FA78 Series Tuberculin)
Indications For Use:
--- The TMD™ 0.5ml Safety Syringe (FA18 Series 0.5ml) is intended for injection of fluids into the body after the aspiration of fluid.
--- The TMD™ 0.5ml Tuberculin Safety Syringe (FA78 Series Tuberculin) is intended intra-dermal of Tuberculin.
All these two syringes above incorporate a safety feature that is designed to aid the reduction of needle stick injuries and the potential of syringe reuse.
AND/OR Prescription Use × (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runge
(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital Division of Anosthoental Devices
Page ii of iii
Koy2a4b
iect Medical Device, Co., Ltd. TMDTM 0.5ml Safety Syringe
Special 510(K)
September 3rd, 2004
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.