K042419 · Draeger Medical, Inc. · BSZ · Oct 8, 2004 · Anesthesiology
Device Facts
Record ID
K042419
Device Name
FABIUS GS/FABIUS TIRO ANESTHESIA SYSTEM
Applicant
Draeger Medical, Inc.
Product Code
BSZ · Anesthesiology
Decision Date
Oct 8, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fabius GS/Fabius Tiro may be used for spontaneous, manually assisted, automatic, pressure support, or synchronized intermittent mandatory ventilation of patients during anesthesia, and delivery of gases and anesthetic vapor. The Fabius GS/Fabius Tiro can monitor inspired oxygen concentration, breathing pressure, and respiratory volume.
Device Story
Fabius GS/Fabius Tiro is a continuous flow anesthesia system for clinical use by physicians. Device delivers anesthetic gases/vapors and provides mechanical ventilation (spontaneous, manual, automatic, pressure support, SIMV/PS). System monitors inspired O2, breathing pressure, and respiratory volume. Software version 3.00 adds SIMV/PS mode, synchronizing mechanical ventilation with patient spontaneous breathing efforts; mandatory un-synchronized breaths delivered between intervals. Hardware includes SIMV/PS hard key on control panel. Clinicians adjust parameters (Tidal Volume, Frequency, Inspiratory Time, PEEP, Pressure Support level) via control panel to augment patient breathing. System provides respiratory support and monitoring during anesthesia, potentially improving patient ventilation management.
Clinical Evidence
Bench testing only. Qualification included hazard analysis, system-level qualification, and verification/validation tests to confirm software modification performance and safety.
Technological Characteristics
Continuous flow anesthesia system. Software-controlled ventilation modes (SIMV/PS). Hardware includes operator control panel with SIMV/PS hard key. Parameters: Pmax, Vt, Freq, TInsp, TIP:TI, PEEP, ΔPPS, Inspiratory Flow, Trigger Level. Pneumatic/mechanical subassemblies unchanged from predicate.
Indications for Use
Indicated for patients undergoing anesthesia requiring continuous flow anesthesia delivery and mechanical ventilation. Supports spontaneous, manually assisted, automatic, pressure support, or synchronized intermittent mandatory ventilation. Prescription use only.
Regulatory Classification
Identification
A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).
{0}------------------------------------------------
007 8 - 2004
K042418
### SUMMARY OF SAFETY AND EFFECTIVENESS DATA RELATING TO SUBSTANTIAL EQUIVALENCE
| Proprietary Name: | Fabius GS/Fabius Tiro Anesthesia System |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Classification Name: | Gas Machine, Anesthesia - 73 BSZ |
| Device Class: | Class II |
| Initial Distributor: | Draeger Medical, Inc.<br>3135 Quarry Road<br>Telford, Pennsylvania 18969 USA |
| Establishment Registration No.: | 2517967 |
| Devices to which substantial<br>equivalence is claimed: | Fabius GS/Fabius Tiro Anesthesia System – K042080<br>Evita 4 Continuous Ventilator - K961687<br>7900 Ventilator - K023366 |
#### Device Description:
The Fabius GS/Fabius Tiro is a continuous flow gas anesthesia system.
#### Intended Use:
The Fabius GS/Fabius Tiro may be used for spontaneous, manually assisted, automatic, pressure support, or synchronized intermittent mandatory ventilation of patients during anesthesia, and delivery of gases and anesthetic vapor. The Fabius GS/Fabius Tiro can monitor inspired oxygen concentration, breathing pressure, and respiratory volume.
#### Substantial Equivalence:
The current software used in the Fabius GS/Fabius Tiro Anesthesia System (K042086) is being modified to include Synchronized Intermittent Mandatory Ventilation with Pressure Support (SIMV/PS) as an optional ventilation mode. The addition of SIMV/PS is essentially a software change and is incorporated in Fabius GS/Fabius Tiro software version 3.00. The only hardware change was the addition of a SIMV/PS hard key on the operator control panel. The basic infrastructure, operating principle, alarm strategies, fault detection circuitry, and mechanical/pneumatic subassemblies within the Fabius GS/ Fabius Tiro remain unchanged.
SIMV/PS mode is a mechanical ventilation mode influenced by spontaneous breathing. In SIMV/PS mode, when the patient breathes spontaneously, the ventilator will synchronize mechanical ventilation to the patient efforts. Between these intervals, mandatory (automatically delivered) un-synchronized ventilation is delivered to ensure a minimum degree of ventilation. SIMV ventilation strokes are the same as those for Volume Controlled ventilation. The synchroneous trigger mechanism for SIMV/PS is the
> 510(k) Summary Page 1 of 2
{1}------------------------------------------------
same as that used in pure pressure support mode. Adjustable parameters are Tidal Volume (Vt), Ventilator Frequency (Freq), Inspiratory Time (Tinsp), Inspiratory Pause (TIP: TI), and Positive End Expiratory Pressure (PEEP).
Pressure Support ventilation can be added to augment the patient's spontaneous breathing efforts by adjusting the Inspiratory Pressure Setting (ΔPPS) level to a value other than "Off." When turned "On", the Pressure Support aspect of SIMV functions the same as the Pressure Support ventilation mode cleared for commercial distribution on the Fabius GS/Fabius Tiro under 510(k)s K030624 and K042086.
Adjustable ventilation settings available in SIMV/PS are:
- -Maximum Ventilation Pressure (Pmax)
- Tidal Volume (Vt) -
- -Ventilator Frequency (Freq)
- -SIMV Inspiratory Time (TInsp)
- Inspiratory Pause (TIP:TI) -
- Positive End Expiratory Pressure (PEEP) -
- Inspiratory Pressure Setting (APPS)* -
- Inspiratory Flow* -
- -Trigger Level*
* For Pressure Support aspect of SIMV if Inspiratory Pressure Setting (ΔPPS) is set to a value other than "Off."
The SIMV/PS ventilation mode for the Fabius GS/Fabius Tiro is substantially equivalent to the SIMV with Pressure Support ventilation mode in the Evita 4 Ventilator (K961687) and 7900 Ventilator (K023366). Similarities are:
- All provide a user settable number of volume controlled ventilator delivered breaths per minute.
- -All synchronize to spontaneous breaths.
- -All incorporate the option of adding pressure support to assist the patient's spontaneous breaths between ventilator breaths.
- All provide user selectable ventilation parameters during SIMV/PS for; Maximum । Ventilation Pressure (Pmax), Tidal Volume (Vt), Ventilator Frequency (Freq), Inspiratory Time (TInsp), and Positive End Expiratory Pressure (PEEP). Additionally, when SIMV is augmented with Pressure Support, all provide user selectable ventilation parameters for; Inspiratory Pressure Setting (APPS), Inspiratory Flow, and Trigger Level.
Qualification included hazard analysis, system level qualification, and verification/ validation tests.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, composed of three parallel lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus, indicating the department's name and national affiliation.
Public Health Service
OCT 8 - 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Michael A. Kelhart Regulatory Affairs Project Manager Dräeger Medical, Incorporated 3135 Quarry Road Telford, Pennsylvania 18969
Re: K042419
Trade/Device Name: Fabius GS/Fabius Tiro Anesthesia System Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: II Product Code: BSZ Dated: September 3, 2004 Received: September 7, 2004
Dear Mr. Kelhart:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Kelhart
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Quls
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): Koyz419_
Device Name: Fabius GS/Fabius Tiro Anesthesia System
Indications for Use:
The Fabius GS/Fabius Tiro is indicated as a continuous flow anesthesia system. The Fabius GS/Fabius Tiro can be used for spontaneous, manually assisted, automatic, pressure support, or synchronized mandatory intermittent ventilation, delivery of gases and anesthetic vapor, and monitoring oxygen concentration, breathing pressure and respiratory volume of patients during anesthesia. Federal law restricts this device to sale by or on the order of a physician.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Aun Sileom
Page 1_of 1
(Division Sign-OK) Division of Anesthesiology, General Hospital, Infection Control. Dental Devic
510(k) Number k042819
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.