K042394 · Kavo America Corporation · NBL · Aug 2, 2005 · Dental
Device Facts
Record ID
K042394
Device Name
DIAGNODENT PERIO TIP (ACCESSORY)
Applicant
Kavo America Corporation
Product Code
NBL · Dental
Decision Date
Aug 2, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1745
Device Class
Class 2
Indications for Use
For use as an aid in the detection and localization of subgingival dental calculus.
Device Story
The KaVo DIAGNOdent® Perio Tip is an accessory probe for the existing DIAGNOdent® system. It utilizes a 655 nm diode laser to induce fluorescence in subgingival dental calculus. The device is operated by dental professionals in a clinical setting. The probe tip is inserted subgingivally to scan the root surface; the system detects fluorescence emitted by calculus, which differs from the signal of healthy cementum. This signal is processed by the existing DIAGNOdent® electronics to provide an objective, reproducible indication of calculus presence. The output assists the clinician in identifying and localizing subgingival deposits, facilitating more accurate periodontal treatment and potentially improving patient outcomes by ensuring thorough cleaning of root surfaces.
Clinical Evidence
Evidence includes in-vitro and preclinical studies. Research (Folwaczny et al., 2002; Krause et al., 2003) demonstrated that 655 nm laser fluorescence is correlated with materials in calculi and concrements. Studies confirmed the method is reproducible and objective for assessing root surfaces, with highly significant differences in fluorescence intensity between cementum and calculus in air, electrolytic solution, and blood.
Indicated for use as an aid in the detection and localization of subgingival dental calculus in patients requiring periodontal assessment.
Regulatory Classification
Identification
A laser fluorescence caries detection device is a laser, a fluorescence detector housed in a dental handpiece, and a control console that performs device calibration, as well as variable tone emitting and fluorescence measurement functions. The intended use of the device is to aid in the detection of tooth decay by measuring increased laser induced fluorescence.
Special Controls
*Classification.* Class II, subject to the following special controls:(1) Sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter;
(2) Premarket notifications must include clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
(3) The labeling must include detailed use instructions with precautions that urge users to:
(i) Read and understand all directions before using the device,
(ii) Store probe tips under proper conditions,
(iii) Properly sterilize the emitter-detector handpick before each use, and
(iv) Properly maintain and handle the instrument in the specified manner and condition.
1. restrictions on the sale, distribution, and use of the device to prescription use in accordance with 21 CFR 801.109;
2. clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
3. labeling that includes detailed use instructions with precautions that urge users to (i) read and understand all directions before using the device, (ii) store probe tips under proper conditions, (iii) properly sterilize the emitter-detector handpiece before each use, and (iv) properly maintain and handle the instrument in the specified manner and condition.
{0}------------------------------------------------
P-1
AUG 2 - 2005
EXHIBIT 2 510(k) Summary KaVo America Corporation 340 East Main Street Lake Zurich. Illinois 60047 Toll Free: 800 323 8029 Tel: 847 / 550 - 6800 Fax: 847 / 550 - 6825 e-mail: info@kavousa.com Contact: John Franz, President August 30, 2004
- 1. Identification of the Device: Proprietary-Trade Name: KaVo DIAGNOdent® Perio Tip Classification Name: Dental Hand Instrument, Laser Fluorescence Caries Detection Device, Product Codes EJB, NBL Common/Usual Name: Periodontal Probe
- Equivalent legally marketed device: DIAGNOdent® K983658 and DETECTAR, K023367 . 2.
- Indications for Use (intended use): For use as an aid in the detection and localization of 3. subgingival dental calculus.
- 4. Description of the Device: This submission is for a modification of a device system cleared under K983658, the DIAGNOdent®. The modification is in the form of a new accessory probe tip which can be connected to the unmodified DIAGNOdent® electronics system. The probe tip is longer and more slender to allow for the periodontal use. In-vitro and preclinical studies have shown that that laser fluorescence is correlated with materials in calculi and concrements 46. The researchers identified the method of subgingival calculi detection to be a reproducible, objective method for assessing the root surface". Although the intensity of the detection was influenced by surrounding media, the differences between cementum and calculus was highly significant in media like air, electrolytic solution and blood".
a. M. Folwaczny, R. Heym, A. Mehl and R. Hickel, "Subgingival Calculus detection with Fluorescence induced by 655 nm InGaAsP Diode Laser Radiation", J. Periodontol, June 2002, p 597-601
b. F. Krause, A. Braun, M. Frentzen, "The possibility of detecting subgingival calculus by laserfluorescence in vitro", Lasers Med. Sci. (2003) 18: 32-35
{1}------------------------------------------------
\$\beta.2\$
K042394
- 5. Safety and Effectiveness, comparison to predicate device:
| Comparison<br>Areas | DIAGNOdent® K983658 | DETECTAR, K023367 | DIAGNOdent® Perio Tip |
|------------------------|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Indications<br>for use | Detection of Dental Caries | Detection of subgingival<br>dental calculus | Detection of subgingival<br>dental calculus |
| Probe<br>technology | Fiber Optic | SAME | SAME |
| Light Source | 655 nm <1 mw Laser | Two LEDs. The first is in<br>the red area of the visible<br>spectrum (RED-LED) and<br>shows a wavelength peak<br>at 635 in a band between<br>600 and 670nm. The<br>second LED is in the near<br>infrared spectrum (NIR-<br>LED) and shows a<br>wavelength peak at 880<br>nanometers in a band<br>between 800 and 900nm. | 655 nm <1 mw Laser |
| Returned<br>light | Fluorescence | Not specified.<br>(proprietary) | Fluorescence |
| Sterilization | Probe tip only, autoclave | SAME | SAME |
| Target<br>population | Dentists' offices | SAME | SAME |
- 6. Conclusion: In all important respects, the DIAGNOdent® Perio Tip is substantially equivalent to This conclusion is based on the DIAGNOdent® K983658 and DETECTAR, K023367. indications for use, in-vitro, and clinical studies.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three lines representing its wings and body. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The seal is black and white.
AUG 2 - 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Kavo America Corporation C/O Mr. Daniel Kamm Kamm & Associates P.O. Box 7007 Deerfield, Illinois 60015
Re: K042394
Trade/Device Name: DIANGOdent® Perio Tip Regulation Number: 21 CFR 872.1745 Regulation Name: Laser Fluorescence Caries Detection Device Regulatory Class: II Product Code: NBL Dated: July 27, 2005 Received: July 28, 2005
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicated in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Kamm
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Ours
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K042394
## Indications for Use
510(k) Number (if known):
Device Name:__________________________________________________________________________________________________________________________________________________________________
Indications For Use:
For use as an aid in the detection and localization of subgingival dental calculus.
Prescription Use _X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Suoer Rurse
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devic
510(k) Number: K02344
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.