K042274 · Olympus Shirakawa Co., Ltd. · GEI · Feb 22, 2005 · General, Plastic Surgery
Device Facts
Record ID
K042274
Device Name
PSD-60/ENDOPLASMA
Applicant
Olympus Shirakawa Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Feb 22, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The PSD-60 has been designed to be used with OLYMPUS endoscope (fiberscope, video scope and rigid scope) compatible with electrosurgical accessories and other ancillary equipment for general, laparoscopic and endoscopic electrosurgery ( cutting, coagulation, including soft tissue coagulation). The argon plasma coagulation unit has been designed to be used with the electrosurgical unit, OLYMPUS PSD-60 for argon plasma coagulation (APC) of living tissues in treatment using OLYMPUS endoscope compatible with electrosurgery, argon plasma coagulation accessories and other ancillary equipment for endoscopic electrosurgery (coagulation).
Device Story
Electrosurgical unit (PSD-60) and Argon Plasma Coagulation (APC) system (ENDOPLASMA) used for tissue cutting and coagulation. PSD-60 generates high-frequency (HF) current; ENDOPLASMA ionizes Argon gas to form plasma for coagulation. System inputs include user settings via 7-segment LCD and footswitch activation; outputs include HF energy and plasma energy delivered to tissue via probes. Used in clinical settings (endoscopy/laparoscopy) by physicians. Features include automatic start/stop, visual/auditory error monitoring, and pressure regulation for Argon gas. Disposable sterile membrane filters protect the ENDOPLASMA unit from contamination. Benefits include precise tissue coagulation and cutting during endoscopic procedures.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing and comparison of technological characteristics to predicate devices.
Technological Characteristics
Electrosurgical unit with Argon plasma coagulation capability. HF energy generation; Argon gas ionization. 7-segment LCD interface. Connectivity via communication bus cable. Complies with IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-2, IEC 60601-2-18, UL2601-1, EN60529, and AAMI/ANSI HF-18. Reusable components; disposable sterile membrane filters (EO sterilized).
Indications for Use
Indicated for general, laparoscopic, and endoscopic electrosurgery (cutting and coagulation of soft tissue) and argon plasma coagulation (APC) of living tissues in patients undergoing procedures using compatible Olympus endoscopes and accessories.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
FEB 2 2 2005
Page 1 of 4
November 15, 2004
## SMDA 510(k) SUMMARY 042274 Electrosurgical unit PSD-60, ENDOPLASMA, ACCESSORIES
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR, Part 807, Subpart E, Section 807.92.
## A. GENERAL INFORMATION
- 1. Applicant
Name & Address:
Shirakawa Olympus corporation 3-1. Aza Ookamiyama Ooaza Odakura, Nishigo-mura, Nishishirakawa-gun, Fukushima, 961-8061, Japan 3002808148
Registration Number:
- 2. Initial Importer Name & Address:
OLYMPUS AMERICA INC. Two Corporate Center Drive, Melville, NY 11747-9058 2429304
Registration Number:
#### 3. Submission Correspondence Name
Address
·
- Telephone Facsimile E-mail address Registration Number
Laura Storms-Tyler Director Regulatory Affair and Quality Assurance OLYMPUS AMERICA INC. Two Corporate Center Drive, Melville, NY 11747-9058 631-844-5688 631-844-5554 Laura Storm-Tyler@olympus.com 2429304
## B. DEVICE IDENTIFICATION
#### 1. Common/Usual Name
Electrosurgical Unit, Argon Plasma Coagulation Unit and Accessories
- 2.Device Name
Electrosurgical Unit PSD-60, ENDOPLASMA, ACCESSORIES
#### 3. Classification Name
| | Classification panel | Product code | Class |
|----------|-------------------------------------------------------------------|--------------|-------|
| 878.4400 | Electrosurgical cutting and coagulation<br>device and accessories | GEI | II |
| 876.4300 | Endoscopic Electrosurgical Unit and<br>accessories | KNS | II |
xf
{1}------------------------------------------------
Page 2 of (4)
## C. IDENTIFICATION OF LEGALLY MARKETED DEVICES WHICH WE CL EQUIVALENCE
510(k)# Manufacturer Class P-Code Model ERBE Elektromedizin II GEI #k023886 ERBE VIO ESU, Model VIO300D GmbH OLYMPUS GEI #k970184 II UES-20 Electrosurgical System and associated Accessories KNS #k030194 OLYMPUS II XUES-41 Endoscopic Electrosurgical unit ERBE Elektromedizin GEI ERBE VIO APC, Model APC-2 #k024047 لل GmbH
The following listed devices are seem to be as predicate devices in consideration of its characteristic and the following table shows regulatory history.
## D. DEVICE DESCRIPTION
#### 1. Summary
- 1) PSD-60 technology
The PSD-60 is an Electrosurgery unit and is used in conjunction with a OL YMPUS ENDO-PLASMA, or can be used alone. The PSD-60 has a 7-segment LCD (Liquid Crystal Display) that provides the user with an on-screen setting and operational information.
The unit has various cutting and coagulation modes with defined effect levels to provide the physician flexibility in applications (i.e. Its ability is to generate the HF current.). The system has automatic start and stop features. When activated, the device has an alarm as well as a visual error system (i.e. malfunctions or user errors are detected with medical personnel being alerted visually and/or by sound with, in some cases, no energy being delivered.).
Upon activation, energy delivered from the PSD-60 to the tissue is displayed in watts on the 7-segment display. Also, the unit can be used in association with an OLYMPUS compatible ENDOPLASMA Argon plasma coagulator. The unit is non-sterile and is reusable.
#### 2)ENDOPLASMA technology
The OLYMPUS ENDOPLASMA is an Argon plasma coagulation system and is used in conjunction with an OLYMPUS PSD-60.
The PSD-60 provides high frequency (HF) voltage to electrically charge Argon gas from the ENDOPLASMA to form plasma in the gas stream when in close proximity to tissue. Current density concentrates at the tissue surface from ENDOPLASMA (probe) causes coagulation. The ENDOPLASMA with the PSD-60 has a 7-segment LCD (Liquid Crystal Display) that provides the user with an on-screen setting operational information.
Software in the PSD-60 controls the microprocessor chip in an ENDOPLASMA. The PSD-60/ENDOPLASMA are programmable and have error monitoring features. There is a communications bus cable for the PSD-60 to communicate with an ENDOPLASMA. A pressure reducer with sensor is provided to regulate the Argon gas going into the ENDOPLASMA. The ENDOPLASMA and its accessory are supplied non-sterile and are reusable. Cleaning/ disinfection, and sterilization (as applicable) are provided in the respective user manual .
249
{2}------------------------------------------------
Page 3 of 4
The ENDOPLASMA membrane filter is also a part of the system. The filter is disposable single-use. The filter is connected between the ENDOPLASMA (at the Argon gas port) and connector hose. The filter creates a barrier to protect the ENDOPLASMA from potential contamination. Filters are supplied sterile by means of ethylene oxide and are disposable (single use). They are sterilized by the manufacturer and the sterilization cycle has been validated
## 3) Accessories
#### 1) Footswitch
A footswitch is provided to the physician as a means to activate a mode of the PSD-60 by depressing a foot petal. A footswitch has two pedals. One is for cutting, and the other is for coagulating.
#### 2 P-cord
P-cord is provided to connect the single use of patient plate to the PSD-60.
#### 3 APC-cable
APC-cable is provided to transmit HF energy and argon gas to ENDOPLASMA probe.
- 4 APC-probe
APC-probe is provided to mix HF energy and argon gas to form ENDOPLASMA and provide plasma energy to the surface of tissue.
- 5 Irrigation adapter
Irrigation adapter is provided to clean the ENDOPLASMA probe.
- 6 Pressure reducer
Pressure reducer is provided to reduce pressure provided from the gas cylinder.
- 7 S-cord
S-cord is provided to connect the metal part of Endoscope to the PSD-60.
## 2. Design
This device has been designed to be complying with the following voluntary standards.
- · IEC60601-1-1 · IEC 60601-1-4 · UL2601-1
- ・IEC 60601-1-2 (EMC) ・EN60529 · AAMI/ANSI HF-18 · IEC 60601-2-2
- · IEC 60601-2-18
· IEC 60601-1
## 3. Materials
There aren't any patient contacting materials in PSD-60 and ENDOPLASMA. Some ancillary equipment have patient contact materials, however, there are no new patient contacting materials in those devices.
## 4. Intended Use of the device
The PSD-60 has been designed to be used with OLYMPUS endoscope (fiberscope, video scope and rigid scope) compatible with electrosurgical accessories and other ancillary equipment for general, laparoscopic and endoscopic electrosurgery ( cutting, coagulation, including soft tissue coagulation).
{3}------------------------------------------------
Page 4 of 4
The argon plasma coagulation unit has been designed to be used with the electrosurgical unit, OLYMPUS PSD-60 for argon plasma coagulation (APC) of living tissues in treatment using OLYMPUS endoscope compatible with electrosurgery, argon plasma coagulation accessories and other ancillary equipment for endoscopic electrosurgery (coagulation).
#### E. COMPARISION OF TECHNOLOGICAL CHARACTERISTICS
Theory of the operation of OLYMPUS PSD-60 and ENDOPLASMA are that the electrical energy generation in PSD-60 is supplied to human fissue via endoscopic therapy devices in order to cut and coagulate the human tissue. The system can cut and coagulate the human tissue by high frequency energy. This system is equivalent to the predicate devices, ERBE VIO300D (K023886), OLYMPUS UES-20 (K970184), ERBE VIO APC2 (K024047) and ERBE APC300 (K963189).
## F. CONCLUSION
When compared to the predicate devices, PSD-60and ENDOPLASMA do not incorporate any significant changes in the intended use, method of operation, materials, or design that could affect safety and effectiveness. Therefore, clinical data is not necessary for evaluation of safety and efficacy.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure with three arms or lines extending from the body.
FEB 2 2 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Olympus Shirakawa Company, Ltd. c/o Mr. Daniel W. Lehtonen Intertek Testing Service 70 Codman Hill Road Boxborough, Massachusetts 01779
Rc: K042274
Trade/Device Name: Electrosurgical Unit PSD-60, ENDOPLASMA, Accessorics Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: February 4, 2005 Received: February 7, 2005
Dear Mr. Lehtonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Mr. Daniel W. Lehtonen
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Miriam C Provost
Celia M. Witten, Ph.D.. M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): Device Name:
K042274 K042214
Electrosurgical Unit PSD-60, ENDOPLASMA, Accessories
Indications for Use:
The PSD-60 has been designed to be used with OLYMPUS endoscope (fiberscope, The PSD-80 has been designed to be assures and other casin also creation in culti videoscope and ligid scope) companible with Shoulded groupeny ( cutting, coagulation, including soft tissue coagulation).
The argon plasma coagulation unit has been designed to be used with the electrosurgical The argon plasma coagulation until has been doulghou (APC) of living tissues in treatment)
unit, OLYMPUS PSD-60 for argon plasma coagulation (APC) of living tissues angelophi unit, OLYMPUS PSD-0010r argon plasma obagalation (include mannus coagulation) using OLYMPUS endoscope companible than areas.sepic electrosurgery (coagulation).
Over-The-Counter-Use AND/OR × Prescription Use (Part 21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IS NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
\$\uparrow\$
(5)
510(k) Number_Ho4 2274
1 of Page
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.