K042261 · Macropore Biosurgery, Inc. · QPB · Sep 20, 2004 · General, Plastic Surgery
Device Facts
Record ID
K042261
Device Name
MACROPORE PURICEL LIPOPLASTY SYSTEM
Applicant
Macropore Biosurgery, Inc.
Product Code
QPB · General, Plastic Surgery
Decision Date
Sep 20, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MacroPore Puricel Lipoplasty System is intended for use in the following surgical specialties when the fragmentation, emulsification, and aspiration of soft tissue is desired: - Neurosurgery -Gastrointestinal and Affiliated Organ Surgery -Urological Surgery -Plastic and Reconstructive Surgery -General Surgery -Orthopedic Surgery -Gynecological Surgery -Thoracic Surgery -Laparoscopic Surgery The MacroPore Puricel Lipoplasty System is indicated for use when the fragmentation, emulsification, and aspiration of subcutaneous fatty tissues for aesthetic body contouring is desired.
Device Story
Manual lipoaspirate system; utilizes house vacuum for energy. Components: cannula, connection tubing, waste collection container. Cannula (surgical stainless steel) features hollow tube with tip openings; communicates vacuum to tissues to fragment, emulsify, and aspirate subcutaneous fat into collection canister. Used in clinical settings by surgeons. Operator controls vacuum flow via stepped clamps on tubing. Output: collected fatty tissue/fluids in canister. Benefits: facilitates aesthetic body contouring through mechanical aspiration.
Clinical Evidence
Bench testing only. Mechanical testing performed to demonstrate substantial equivalence to predicate devices.
Technological Characteristics
Single-use, manual suction lipoplasty system. Materials: surgical stainless steel (patient contact). Energy source: external house vacuum. Dimensions: cannula 15-36cm length, 3.0-4.6mm diameter; tubing 9.6mm ID. Features: stepped clamps for flow control, integrated filter in collection canister.
Indications for Use
Indicated for patients requiring fragmentation, emulsification, and aspiration of subcutaneous fatty tissues for aesthetic body contouring across various surgical specialties including neurosurgery, GI, urological, plastic/reconstructive, general, orthopedic, gynecological, thoracic, and laparoscopic surgery.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
Predicate Devices
Byron Medical Lipoplasty / Liposuction and Tumescent Infusion Cannulae and Needles (K981172)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
Macropore Biosurgery, Inc. Kenneth Kleinhenz Director, RA 6740 Top Gun St. San Diego, California 92121
June 8, 2021
Re: K042261
Trade/Device Name: Macropore Puricel Lipoplasty System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: QPB
Dear Kenneth Kleinhenz:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated September 20, 2004. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Cindy Chowdhury, OHT4: Office of Surgical and Infection Control Devices, 240-402-6647, Cindy.Chowdhury@fda.hhs.gov.
Sincerely,
# Cindy Chowdhury -S
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized lines.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 0 2004
Mr. Kenneth K. Kleinhenz Director of Regulatory Affairs MacroPore Biosurgery, Inc. 6740 Top Gun Street San Diego, California 92121
Re: K042261
Trade/Device Name: MacroPore Puricel Lipoplasty System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: II Product Code: MUU Dated: August 20, 2004 Received: August 24, 2004
Dear Mr. Kleinhenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave teviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) to regard) the enactment date of the Medical Device American De commerce proc to thay 20, 2017 ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . devices that have been resuire approval of a premarket approval application (PMA). and Cosmetic Act (Tec) market the device, subject to the general controls provisions of the Act. The 1 ou may, therefore, manov and act include requirements for annual registration, listing of general controls provisions of ractice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elassified (600 a00 rols. Existing major regulations affecting your device can may be subject to back academant, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act that I Dr Haval statutes and regulations administered by other Federal agencies. You must or any I caeral bakates and and ing, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of it in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Mr. Kenneth K. Kleinhenz
This letter will allow you to begin marketing your device as described in your Section 510(k) rms letet notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you atten office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
KOY226/
# Indications for Use
Device Name: MacroPore Puricel Lipoplasty System
#### Indications for Use:
The MacroPore Puricel Lipoplasty System is intended for use in the following surgical specialties when the fragmentation, emulsification, and aspiration of soft tissue is desired:
- Neurosurgery
- -Gastrointestinal and Affiliated Organ Surgery
-Urological Surgery
-Plastic and Reconstructive Surgery
-General Surgery
-Orthopedic Surgery
- -Gynecological Surgery
- -Thoracic Surgery
-Laparoscopic Surgery
The MacroPore Puricel Lipoplasty System is indicated for use when the fragmentation, emulsification, and aspiration of subcutaneous fatty tissues for aesthetic body contouring is desired.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
Page 1 of 1
**510(k) Number** K042261
{4}------------------------------------------------
Koy 226/
Page 1 of 3
# SEP 2 0 2004
# ADMINISTRATIVE INFORMATION
Manufacturer Name:
Official Contact:
MacroPore Biosurgery, Inc. 6740 Top Gun Street San Diego, CA 92121
Kenneth K. Kleinhenz Director of Regulatory Affairs Telephone (858) 458-0900 Fax (858) 458-0994
#### DEVICE NAME
Classification Name: · Trade/Proprietary Name:
Suction Lipoplasty System MacroPore Puricel Lipoplasty System
### ESTABLISHMENT REGISTRATION NUMBER 2031733
#### DEVICE CLASSIFICATION AND PRODUCT CODE
As shown in 21 CFR 878.5040 Suction Lipoplasty Systems are defined as devices consisting of collection bottles, cannulas, and connecting tubing for use in aesthetic body contouring procedures. Suction Lipoplasty Systems are classified as Class II. They have been assigned Product Corde MUU.
{5}------------------------------------------------
# INTENDED USE
The MacroPore Puricel Lipoplasty System is intended for use in the following surgical specialties when the fragmentation, emulsification, and aspiration of soft tissue is desired:
- Neurosurgery -Gastrointestinal and Affiliated Organ Surgery -Urological Surgery -Plastic and Reconstructive Surgery -General Surgery -Orthopedic Surgery -Gynecological Surgery -Thoracic Surgery -Laparoscopic Surgery
The MacroPore Puricel Lipoplasty System is indicated for use when the fragmentation, emulsification, and aspiration of subcutaneous fatty tissues for aesthetic body contouring is desired.
#### DEVICE DESCRIPTION
#### Design Characteristics
The MacroPore Puricel Lipoplasty System is a single use, pre-assembled, manual lipoaspirate system that relies on house vacuum for its energy supply. The MacroPore Puricel Lipoplasty System consists of a cannula, connection tubing, and a waste collection container. The cannula handle is attached to the collection canister via connection tubing. The cannula is a hollow tube with a single opening near the tip to communicate house-vacuum to the tissues and subsequently fragment. emulsify, and aspirate subcutaneous fatty tissues from the patient into the waste collection canister for purposes of aesthetic body contouring. The collection canister contains various capped / sealed ports and a filter to trap large tissue masses
The Puricel Lipoplasty cannula is provided in various sizes ranging from 15cm - 36cm in length and 3.0 ~ 4.6mm in diameter with a single opening near the tip of the cannula. The tip region of the cannula may have a single or multiple openings that range in size from 4mm to length distributed uniformly or randomly throughout the end of the cannula. The handle of the device is 20mm in diameter and may be provided in diameters ranging from 20mm in diameter. The connecting tubing is provided with an inner diameter of 9.6mm (3/8"), an outer diameter of 14.3mm (9/16"), and a wall thickness of 2.2mm. The tubing that connects the cannula handle to the waste canister is provided in a length of 4 feet and may be provided in lengths ranging from 1 - 8 feet. The bottom of the waste collection container is also provided with the same 3/8" inner diameter connection tubing of various lengths. The proximal end of the exiting connection tubing may be provided with a barbed tubing connector to assist in the attachment of like-sized tubing for purposes of connecting the MacroPore Puricel Lipoplasty System to house vacuum and / or assorted waste traps. Connection tubing leading to and from the waste collection container is provided with a stepped clamp to allow the operator to seal the connection tubing on both sides of the waste container and prevent spillage of the collected fluids / tissues.
{6}------------------------------------------------
# Material Composition
The components of the MacroPore Puricel Lipoplasty System that have patient contact are fabricated from surgical stainless steel.
# In Vitro Testing
Mechanical testing of the MacroPore Puricel Lipoplasty System demonstrates that the device is substantially equivalent to the predicate.
# EQUIVALENCE TO MARKETED PRODUCT
MacroPore Puricel Lipoplasty System shares indications and design principles with the following predicate devices which have been determined by FDA to be substantially equivalent to preamendment devices: Byron Medical Lipoplasty / Liposuction and Tumescent Infusion Cannulae and Needles and Sound Surgical Soundvaser System, Class II medical devices that were cleared for marketing in the United States under K981172 and K022051 respectively.
#### Indications For Use
The MacroPore Puricel Lipoplasty System and the predicate devices share substantially equivalent indications for use as they are all indicated for aspiration of soft tissues in aesthetic body contouring procedures. The MacroPore Puricel Lipoplasty System shares identical indications for use language with the predicate devices.
#### Design and Materials
The design and materials of MacroPore Puricel Lipoplasty System and the predicate devices (Byron Medical Lipoplasty / Liposuction Aspiration and Tumescent Infusion Cannulae and Needles and Sound Surgical Soundvaser System) are substantially equivalent as they are all hollow tubular cannulas attached to a handle. The cannula tip of the subject device and the predicate devices contains one or several openings to allow communication between the applied vacuum and the patients tissues. The MacroPore Puricel Lipoplasty System and the Sound Surgical predicate device are substantially equivalent in design as they both consists of a cannula, connection tubing, and a waste collection container. The waste collection containers on the MacroPore Puricel Lipoplasty System and Sound Surgical predicate device are provided in substantially equivalent volume capacities.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.