DISCOVERY ST PET/CT IMAGING SYSTEM WITH AN ALTERNATE DETECTOR OPTION
K042257 · General Electric Co. · KPS · Sep 8, 2004 · Radiology
Device Facts
Record ID
K042257
Device Name
DISCOVERY ST PET/CT IMAGING SYSTEM WITH AN ALTERNATE DETECTOR OPTION
Applicant
General Electric Co.
Product Code
KPS · Radiology
Decision Date
Sep 8, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1200
Device Class
Class 2
Indications for Use
The GE Discovery ST System with the alternate detector option is intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery ST with alternate detector option is to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. This device can also assist in radiotherapy planning. The Discovery ST with alternate detector option can also be used as a stand-alone head and whole body multislice computed tomography (CT) diagnostic imaging system.
Device Story
Integrated PET/CT imaging system; combines GE Positron Emission Tomography and GE LightSpeed CT. Inputs: PET radiopharmaceutical distribution data and CT X-ray attenuation data. Operation: CT images used to correct PET non-uniform attenuation and facilitate anatomical localization of emission activity. Output: Integrated PET/CT images for metabolic/physiologic assessment. Used in clinical settings by trained healthcare professionals (physicians, technologists). Output enables visualization of lesions, disease, and organ function; assists in diagnosis, staging, and radiotherapy planning. Benefits: improved anatomical localization of metabolic activity and attenuation correction.
Clinical Evidence
No clinical data. Device evaluated for electrical and radiation safety; conforms to applicable medical device safety standards.
Technological Characteristics
Integrated PET/CT system. PET component for metabolic imaging; CT component for attenuation correction and anatomical localization. Multislice CT capability. Conforms to 21 CFR 820, ISO 9001, and ISO 13485 quality systems. Electrical and radiation safety verified via independent evaluation.
Indications for Use
Indicated for head and whole-body PET imaging and CT diagnostic imaging in patients requiring assessment of metabolic and physiologic functions, including evaluation, diagnosis, staging, and follow-up of lesions, cancer, cardiovascular disease, and brain dysfunction; also for radiotherapy planning. Used by trained healthcare professionals.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
Predicate Devices
GE Discovery ST Positron Emission Tomography System (K010641)
Submission Summary (Full Text)
{0}------------------------------------------------
K042251
SEP = 8 2004
Special 510(k) Premarket Notification Discovery ST with alternate detector option August 18, 2004
#### Attachment B:
Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92(c).
# GE Healthcare
| Submitter: | GE Medical Systems<br>PO Box 414<br>Milwaukee, WI 53201 |
|------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person. | D. Duersteler<br>Safety and Regulatory Engineering<br>Telephone: 262-312-7029; Fax: 262-312-7144 |
| Date Prepared: | August 18, 2004 |
| Device Name: | Discovery ST PET/CT Imaging System an with alternate detector option.<br>Emission Computed Tomography System, 21 CFR 892.1200, 90-KPS |
| Marketed Device: | GE Medical Systems GE Discovery ST Positron Emission Tomography System, 510(k)<br>Number K010641, currently in commercial distribution. |
Device Description: The Discovery ST PET/CT Imaging System an with alternate detector option integrates a GE Positron Emission Tomography system and a GE LightSpeed CT System. In addition to providing CT and PET stand-alone capabilities, it uses the CT images to correct for non-uniform attenuation of the PET images and to facilitate localization of the emission activity in the patient anatomy.
Indications for Use: The GE Discovery ST System with the alternate detector option is intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images.
The Discovery ST with alternate detector option is to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. This device can also assist in radiotherapy planning.
The Discovery ST with alternate detector option can also be used as a stand-alone head and whole body multislice computed tomography (CT) diagnostic imaging system.
Comparison with Predicate Device: The Discovery ST with alternate detector option is of a comparable type and substantially equivalent to the currently marketed GE Discovery ST PET/CT System. It has the same technological characteristics, is comparable in key safety and effectiveness features, uses the same basic design, construction, and materials, and has the same intended uses as the predicate device.
Summary of Studies: The device has been evaluated for electrical, and radiation safety, and conforms with applicable medical device safety standards.
#### Clinical Tests: None required.
Conclusion: Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines, and established methods of patient examination. Intended uses and fundamental scientific technology are the same as the legally marketed GE Discovery ST System. The design and development process of the manufacturer conforms with 21 CFR 820, ISO 9001 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards and compliance is verified through independent evaluation with factory surveillance. Therefore, it is the opinion of GE Medical Systems that the Discovery ST with alternate detector option PET/CT Imaging System is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized representation of a human figure, with three overlapping profiles suggesting a family or community. The profiles are rendered in a simple, flowing line. Encircling the figure is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", arranged in a circular fashion to follow the contour of the logo.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD. 20850
### SEP = 8 2004
Mr. D. Duersteler Safety and Regulatory Engineering General Electric Company GE Medical Systems W-1250 P.O. Box 414 MILWAUKEE WI 53201Re: K042257
Trade/Device Name: Discovery ST PET/CT Imaging System with Alternate Detector Option Regulation Number: 21 CFR 882.1200 Regulation Name: Emission computed tomography system Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 90 KPS and JAK Dated: August 18, 2004 Received: August 20, 2004
Dear Mr. Duersteler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to ought maining of substantial equivalence of your device to a legally premarket notheation: "The PDF intrally stimally for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire spoolite da roof the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, 101 queenoms on me f-4639. Also, please note the regulation entitled, "Misbranding Other of Somphalles as (s notification" (21CFR Part 807.97) you may obtain. Other general by receited to premarked nonitodnen. (er the Act may be obtained from the Division of Small miorination on 30ar respensional Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brighton
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Special 510(k) Premarket Notification Discovery ST with alternate detector option August 18, 2004
#### STATEMENT OF INDICATIONS FOR USE
510(k) Number (if known):
KD42257
Device Name: Discovery ST PET/CT Imaging System with alternate detector option
## Indications for Use
The GE Discovery ST System with the alternate detector option is intended for head and whole body The GE Discovery ST Oystem with the altomate ooky (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images.
The Discovery ST with alternate detector option is to be used by trained health care professionals The Discovery of With anomate dologicals in the body for the assessment of metabolic for lineging the distinution of fudiophamaceaes in the evaluation, diagnosis, staging, (Invectural) and follow up of lesions, This sam assist function such as (but not limited to) cancer, restaging, and follow up on looked and this device can also assist in radiotherapy planning.
The Discovery ST with alternate detector option can also be used as a stand-alone head and whole body multislice computed tomography (CT) diagnostic imaging system.
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801-109)
✓
OR Over-The-Counter Use
Daniel R. Lyman
(Division Division of 510(k) Num
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.