# NEXFLEX TOTAL HIP SYSTEM (K042252)

_Alphatec/Nexmed · LPH · Dec 6, 2004 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K042252

## Device Facts

- **Applicant:** Alphatec/Nexmed
- **Product Code:** [LPH](/productcode/LPH.md)
- **Decision Date:** Dec 6, 2004
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

1) When used as a hemi-hip replacement system, it is intended for osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal involvement of the corresponding acetabulum, femoral head or neck fractures, aseptic necrosis of the femoral head, previous failed hip arthroplasty where there is evidence of sufficient bone quality to adequately set the implant. 2) When used as a total hip replacement system, it is intended for osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal involvement of the corresponding acetabulum. femoral head or neck fractures, aseptic necrosis of the femoral head, previous failed hip arthroplasty where there is evidence of sufficient bone quality to adequately set the implant. 3) In addition, the SIERRA™ hip stem is intended for cases where alternative modes of trealment appear less preferable and the associated risks of a total hip replacement are thought to be acceptable. It is intended for severely disabled joints, which could result from arthritis or late stages of avascular necrosis and revisions of unsuccessful acetabular cup arthroplasty and/or femoral procedure.

## Device Story

SIERRA™ Femoral Hip Stem; sterile orthopedic implant for hip arthroplasty. Tapered distal stem design; available in cemented (CoCrMo) or uncemented (Ti 6Al-4V, porous/HA coated) configurations. Options include standard/high offset necks and collared/non-collared designs. Used by orthopedic surgeons in clinical/OR settings to restore range of motion and treat joint deformities. Implanted via surgical procedure; provides structural support for hip joint. Benefits include pain relief and improved mobility for patients with severe joint degeneration or failed prior implants.

## Clinical Evidence

Bench testing only. Mechanical and dynamic testing performed to verify structural integrity and performance characteristics.

## Technological Characteristics

Materials: Titanium alloy (Ti 6Al-4V, ASTM F136) or Cobalt chrome (CoCrMo, ASTM F75). Design: Tapered distal stem, standard/high offset necks, collared/non-collared options. Coatings: Porous or porous with hydroxylapatite (HA). Energy source: None (mechanical implant). Connectivity: None. Sterilization: Sterile.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- NexFlex™ Total Hip System ([K033580](/device/K033580.md))
- Synergy HA Coated Porous Femoral Stems ([K002996](/device/K002996.md))
- Synergy Cemented Hip Stems ([K990369](/device/K990369.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K042252

# DEC - € 2004

### SIERRA™ Femoral Hip Stem 510(k) SUMMARY August 2004

- l . Company: Nexmed, Inc. 6110 Corte Del Cedro Carlsbad, CA 92009, USA (760) 431-9286
#### II Contact Person: Ellen Yarnall, Director of Regulatory Affairs

#### 111 Trade/Proprietary Name: SIERRA™ Femoral Hip Stem

#### IV Product Description:

The SIERRA™ Femoral Hip Stem is a sterile component used to help restore patient range of motion and aid in the treatment of other deformities as listed in the Indications for Use. The SIERRA™ Femoral Hip Stem has a tapered distal stem design and is available in a variety of options including cemented, porous coated or porous coated with hydroxylapatite (HA) coating. They are also available with standard or high offset necks, as well as collared and non-collared designs. This submission seeks clearance for the SIERRA™ Femoral Hip Stem, which is a line extension to the NexFlex™ Total Hip System. The SIERRA™ Femoral Hip Stem is manufactured from titanium alloy (Ti 6Al-4V, ASTM F136) or cobalt chrome (CoCrMo, ASTM F75). Cobalt chrome stems are intended for cemented use. The porous coated and HA/porous coated stems are intended for uncemented use.

#### V. Classification

| LPH (21 CFR 888.3358) | Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis          |
|-----------------------|---------------------------------------------------------------------------------------------|
| JDI (21 CFR 888.3350) | Hip joint metal/polymer semi-constrained cemented prosthesis                                |
| KWY (21 CFR 888.3390) | Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented or uncemented prosthesis. |

#### VI Indications for Use

- 1) When used as a hemi-hip replacement system, it is intended for osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal involvement of the corresponding acetabulum, femoral head or neck fractures, aseptic necrosis of the femoral head, previous failed hip arthroplasty where there is evidence of sufficient bone quality to adequately set the implant.
- 2) When used as a total hip replacement system, it is intended for osteo-, theumatoid, and posttraumatic arthritis of the hip with minimal involvement of the corresponding acetabulum. femoral head or neck fractures, aseptic necrosis of the femoral head, previous failed hip arthroplasty where there is evidence of sufficient bone quality to adequately set the implant.
- 3) In addition, the SIERRA™ hip stem is intended for cases where alternative modes of trealment appear less preferable and the associated risks of a total hip replacement are thought to be acceptable. It is intended for severely disabled joints, which could result from arthritis or late stages of avascular necrosis and revisions of unsuccessful acetabular cup arthroplasty and/or femoral procedure.

{1}------------------------------------------------

## SIERRA™ Femoral Hip Stem 510(k) SUMMARY August 2004

### VII Substantial Equivalence:

The SIERRA™ Femoral Hip Stem is substantially equivalent to femoral hip stems from the following predicate devices:

Trade/Proprietary Name NexFlex™ Total Hip System Synergy HA Coated Porous Femoral Stems Synergy Cemented Hip Stems

Manufacturer Nexmed, Inc. Smith & Nephew, Inc.

Clearance K033580 K002996

Smith & Nephew, Inc.

Kago369

### VIII Performance Data:

Mechanical and dynamic testing of the SIERRA™ Femoral Hip Stem was performed.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle with three lines representing its wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC - 6 2004

Ms. Ellen A. Yarnall Director of Regulatory Affairs Nexmed, Inc. 6110 Corte Del Cedro Carlsbad, California 92009

Re: K042252

Tro 12/Device Name: Nexflex Total Hip System, Sierra Femoral Hip Stem Regulation Number: 21 CFR 888.3358; 21 CFR 888.3350; 21 CFR 888.3390 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis; Hip joint metal/polymer semi-constrained cemented prosthesis; Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis

Regulatory Class: II Product Code: LPH, JDI, KWY Dated: November 15, 2004 Received: November 16, 2004

Dear Ms. Yarnall:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Ms. Ellen A. Yarnall

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Marlie N. Millbern

Celia Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): K042252

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications for Use:

- 1) When used as a hemi-hip replacement system, it is intended for osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal involvement of the corresponding acetabulum, femoral head or neck fractures, aseptic necrosis of the femoral head, previous failed hip arthroplasty where there is evidence of sufficient bone quality to adequately set the implant.
- 2) When used as a total hip replacement system, it is intended for osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal involvement of the corresponding acetabulum, femoral head or neck fractures, aseptic necrosis of the femoral head, previous failed hip arthroplasty where there is evidence of sufficient bone quality to adequately set the implant.
- 3) In addition, the NexFlex™ Total Hip System is intended for cases where alternative modes of treatment appear less preferable and the associated risks of a total hip replacement are thought to be acceptable. It is intended for severely disabled joints, which could result from arthritis or late stages of avascular necrosis and revisions of unsuccessful acetabular cup arthroplasty and/or femoral procedure.

Mule N Millause

and Neurological Devices

510(k) Number K042252

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

---

**Source:** [https://fda-staging.innolitics.com/device/K042252](https://fda-staging.innolitics.com/device/K042252)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).