BONE-LOK MVP CORTICAL-CANCELLOUS COMPRESSION DEVICE

K042244 · Triage Medical, Inc. · HWC · Sep 16, 2004 · Orthopedic

Device Facts

Record IDK042244
Device NameBONE-LOK MVP CORTICAL-CANCELLOUS COMPRESSION DEVICE
ApplicantTriage Medical, Inc.
Product CodeHWC · Orthopedic
Decision DateSep 16, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The BONE-LOK® MVP Cortical-Cancellous Compression Device is indicated for use in the general management of fractures and reconstructive surgery.

Device Story

BONE-LOK® MVP is a double-helix screw with a compression-locking collar; used for bone fixation in fracture management and reconstructive surgery. Device is cannulated; available in 3.5mm or 4.5mm diameters. Operated by surgeons in clinical/OR settings. Provides mechanical compression and stabilization of bone segments. Single-use only; supplied sterile via gamma radiation.

Clinical Evidence

Bench testing only. No clinical data provided. Biocompatibility established via material standards (ASTM F-136 and F-138).

Technological Characteristics

Double-helix screw with compression-locking collar; cannulated design. Materials: Titanium 6Al-4V (ASTM F-136) or stainless steel (ASTM F-138). Diameters: 3.5mm and 4.5mm. Sterilization: Gamma radiation. Single-use.

Indications for Use

Indicated for general management of fractures and reconstructive surgery in patients requiring bone fixation.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP 1 6 2004 t Notification Special 5 BONE-LOK® MVP Cortical-Cancellous Compression Device (Helical Compression Anchor System, Cannulated, Version 2) # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS ## SUBMITTER INFORMATION | A. Company Name: | Triage Medical, Inc | |-----------------------|--------------------------------------------------------------------| | B. Company Address: | 13700 Alton Parkway<br>Suite 160<br>Irvine, CA 92618 | | C. Company Phone: | (949) 472-0006 | | D. Company Facsimile: | (949) 472-0016 | | E. Contact Person: | Gayle Hirota<br>Manager, Quality Assurance & Regulatory<br>Affairs | ## DEVICE IDENTIFICATION | A. | Trade Name: | BONE-LOK® MVP Cortical-Cancellous<br>Compression Device | |----|----------------------|------------------------------------------------------------------| | B. | Catalog Number: | TMCD-35-1520S to TMCD-35-4050S<br>TMCD-45-3040S to TMCD-45-6070S | | C. | Common Name: | Bone Fixation Screw | | D. | Classification Name: | Smooth or Threaded Metallic Bone Fixation<br>Fastener | | E. | Product Code: | HWC | | F. | Device Panel: | Orthopaedic | | G. | Device Class: | Class II (per 21 CFR 888.3040) | ### IDENTIFICATION OF MODIFIED DEVICE The BONE-LOK® MVP Cortical-Cancellous Compression Device is similar in basic design and intended use to the Triage Medical Helical Compression Anchor System, Cannulated, Version 2 cleared under 510(k) K013903. The device modification includes devices fabricated from Titanium 6Al-4V alloy, which meets the requirements of ASTM F-136. {1}------------------------------------------------ ### DEVICE DESCRIPTION The BONE-LOK® MVP Cortical-Cancellous Compression Device is a double-helix screw with a compression-locking collar. It is available as 3.5mm or 4.5mm diameter device with a compression looking ochan 1978 - The BONE-LOK® MVP Cortical-Cancellous Compression Device is intended for single use only. Devices provided "STERILE" are double-pouched in Tyvek®ffilm pouches and sterilized by gamma radiation. #### INTENDED USE The BONE-LOK® MVP Cortical-Cancellous Compression Device is indicated for use in the general management of fractures and reconstructive surgery. #### TECHNOLOGICAL CHARACTERISTICS The BONE-LOK® MVP Cortical-Cancellous Compression Device is similar in basic design, construction and mechanical performance to the previously cleared Helical Compression Anchor System, Cannulated, Version 2. The device modification includes devices fabricated from titanium alloy in addition to devices fabricated from stainless steel alloy. ## BIOCOMPATIBILITY AND PERFORMANCE DATA The materials the BONE-LOK® MVP Cortical-Cancellous Compression Device are made from meet the requirements of ASTM F-138 (stainless steel) or ASTM F-136 (titanium). These materials are currently being utilized in a myriad of legally marketed orthopedic devices. Because the inertness and biocompatibility of these materials have already been established, additional testing is not required. Performance test results indicate that the device is safe and satisfies functional performance requirements when used as indicated. ## CONCLUSIONS DRAWN FROM STUDIES The test results demonstrate that the modified BONE-LOK® MVP Cortical-Cancellous Compression Device is substantially equivalent to the predicate device and is capable of safely and accurately performing the stated intended use. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the department's name encircling an abstract symbol. The symbol resembles an eagle or bird in flight, composed of three curved lines. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 1 6 2004 Ms. Gayle Hirota Manager, Quality Assurance and Regulatory Affairs Triage Medical, Inc. 13700 Alton Parkway Irvine, California 92618 Re: K042244 Trade/Device Name: BONE-LOK® MVP Cortical-Cancellous Compression Device Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC Dated: August 16, 2004 Received: August 19, 2004 Dear Ms. Hirota: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Gayle Hirota This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Miriam C. Provost Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### SECTION 1.6 INDICATIONS FOR USE FORM # Indications for Use 510(k) Number (if known): BONE-LOK® MVP Cortical-Cancellous Compression Device Name: Device The BONE-LOK® MVP Cortical-Cancellous Compression Indications For Use: Device is indicated for use in the general management of fractures and reconstructive surgery. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Miriam C. Provost (Division Sign-Off) Division of General, Restorative, and Neurological Devices **510(k) Number** k042244
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...