FLEXISKIN POWDER FREE NITRILE EXAMINATION GLOVES, NON STERILE
K042239 · Pt. Mandiri Inti Buana · LZA · Dec 16, 2004 · General Hospital
Device Facts
Record ID
K042239
Device Name
FLEXISKIN POWDER FREE NITRILE EXAMINATION GLOVES, NON STERILE
Applicant
Pt. Mandiri Inti Buana
Product Code
LZA · General Hospital
Decision Date
Dec 16, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The Powder Free Nitrile Examination Gloves, Blue, Non Sterile is a disposable device intended for medical purposes that is worn on the examiner's to prevent contamination between patient and examiner.
Device Story
Device is a disposable, non-sterile, powder-free nitrile examination glove. Worn by healthcare professionals during patient examinations to provide a protective barrier against cross-contamination. Manufactured from synthetic rubber. Device is provided in various sizes (XS to XL) with specified dimensions and physical property requirements. Performance is verified through water leak testing for pinholes and physical property testing (tensile strength, elongation) per ASTM standards. Used in clinical settings to protect both patient and examiner.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including ASTM D 6319-00aE3 compliance for physical properties, dimensions, and pinhole AQL, as well as biocompatibility testing (primary skin irritation and dermal sensitization).
Indicated for use as a disposable, non-sterile, synthetic rubber examination glove worn on the hands of healthcare personnel to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Submission Summary (Full Text)
{0}------------------------------------------------
K 042239
Image /page/0/Picture/1 description: The image shows a stylized, gothic-style monogram consisting of the letters 'M', 'I', and 'B'. The letters are rendered in a bold, black color, with sharp, angular edges that give them a medieval or fantasy-like appearance. The design is compact and the letters are closely intertwined, creating a unified and distinctive logo or emblem.
1.0
PT MANDIRI INTI BUANA
DEC 1 6 2004
Jalan Sei Belumai Desa Dalu 10 A Dusun I Tanjung Morawa - 20362 SUMUT - INDONESIA
+62-61-7944880 Tel Fax +62-61-7944882
## 510 (K) SUMMARY
| Submitter: | |
|------------|-----------------------------------------|
| Name | PT MANDIRI INTI BUANA |
| Address | Jl. Sei Belumai, Desa Dalu 10 A Dusun I |
| | Tanjung Morawa - 20362 |
| | SUMUT – INDONESIA |
| Phone No. | +62-61-7944880 |
| Fax No. | +62-61-7944882 |
Date of Summary Prepared:
### 2.0 Contact Person:
| Name | : | Mr. Sasitharan Nair |
|---------|---|---------------------|
| Phone | : | +62-61-7944880 |
| Fax No. | : | +62-61-7944882 |
### 3.0 Name or the device:
| Trade Name | : | 1) Flexiskin and |
|---------------------|---|----------------------------------------------------------------------|
| | | 2) Multiple or Customers' Trade Name |
| Device Name | : | Powder Free Nitrile Blue Examination Gloves,<br>Blue, Non<br>Sterile |
| Common Name | : | Examination Gloves |
| Classification Name | : | Nitrile Blue Examination Gloves (Class I) |
### 4.0 Identification of The Legally Marketed Device:
Class I Nitrile Examination Gloves, 80LZA, powder free, that meets all the Test.
### 5.0 Description of The Device
The Powder Free Nitrile Examination Gloves, Blue, Non Sterile meets all the requirements of ASTM standard D 6319-00a53 and FDA 1000 ml Water Leak Test.
### 6.0 Intended Use of The Device
The Powder Free Nitrile Examination Gloves, Blue, Non Sterile is a disposable device intended for medical purposes that is worn on the examiner's to prevent contamination between patient and examiner.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a stylized logo with the letters "MB" in a gothic-style font. The letters are bold and black, with intricate details and sharp edges. The overall design has a vintage or medieval feel, with the letters slightly overlapping each other.
Jalan Sei Belumai Desa Dalu 10 A Dusun 1 Tanjung Morawa - 20362 SUMUT - INDONESIA
+62-61-7944880 Tel +62-61-7944882 Fax
### Summary of The Technological Characteristics of The Device 7.0
The Powder Free Latex Examination Gloves, Blue, Non Sterile (Contains 50 micrograms or less of Total Water Extractable Protein per gram) are summarized with the following technological characteristics compared to ASTM or equivalent standards.
| CHARACTERISTICS | STANDARDS | DEVICE<br>PERFORMANCE |
|-----------------------|---------------------------------------------------------------|-----------------------------------------------------------------------------|
| Dimension | D 6319-00aE3 | Meets |
| Physical Properties | D 6319-00aE3 | Meets |
| Freedom from Pinholes | D 6319-00aE3<br>FDA 21 CFR 800.20 | Meets |
| Powder Residue | D 6319-00aE3<br>D6124 - 01 | < 2 mg/glove |
| Biocompatability | Primary Skin Irritation in<br>Rabbits<br>Dermal Sensitization | Passes<br>(No primary skin irritation)<br>Passes<br>(No contact sensitizer) |
### 8.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data
The performance test data of the non-clinical tests that support a determination of substantial equivalent is the same as mentioned immediately above.
### 9.0 Substantial Equivalent Based on Assessment of Clinical Performance Data Clinical data is not needed for gloves or for most devices cleared by the 510 (k) process.
### Conclusion 10.0
It can be concluded that The Powder Free Nitrile Blue Examination Gloves, Blue, Non Sterile will perform according to the gloves performance standards referenced in Section (7) above and meet ASTM standards, and FDA requirements for water leak test on pinhole AQL. Consequently, this device is substantially equivalent to currently marketed device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a stylized, black and white logo or emblem. It features two large, bold letters, 'M' and 'B', interlocked or placed closely together. The font style is somewhat gothic or medieval, with thick strokes and sharp angles, giving the letters a strong, imposing appearance. The overall design is simple yet striking, making it easily recognizable.
Jalan Sei Belumai Desa Dalu 10 A Dusun 1 Tanjung Morawa - 20362 SUMUT – INDONESIA
+62-61-7944880 Tel +62-61-7944882 Fax
### Premarket Notification 510 (k) Submission Applicant: 1.0
| Name | : PT MANDIRI INTI BUANA |
|-----------|------------------------------------------------------------------------------------------|
| Address | : Jl. Sei Belumai, Desa Dalu 10 A Dusun I<br>Tanjung Morawa - 20362<br>SUMUT - INDONESIA |
| Country | : INDONESIA |
| Phone No. | : +62-61-7944880 |
| Fax No. | : +62-61-7944882 |
#### 1.1 Activity of Applicant: [ ] Importer [V] Manufacturer | Repacker [ ] Other [V] Contact Manufacturing | Consultant
### 1.2 Contact Person:
Proprietary or Trade Name:
| Name | : | Mr. Sasitharan Nair |
|---------|---|---------------------|
| Phone | : | +62-61-7944880 |
| Fax No. | : | +62-61-7944882 |
### Truthful and Accurate Statement: As shown in Attachment A 2.0
- Indication for Use Statement: As shown in Attachment B 3.0
- Trade Name: 4.0
1) Glovetex and
2) Multiple Free Nitrile Examination Gloves, Blue, Non Sterile.
### 5.0 Name and Location of ACTUAL Manufacturer:
| Name | PT MANDIRI INTI BUANA |
|-----------|-----------------------------------------|
| Address | Jl. Sei Belumai, Desa Dalu 10 A Dusun I |
| | Tanjung Morawa - 20362 |
| | SUMUT – INDONESIA |
| Country | INDONESIA |
| Phone No. | +62-61-7944880 |
| Fax No. | +62-61-7944882 |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a stylized logo with the letters 'MB' in a bold, gothic-style font. The letters are rendered in solid black, creating a strong contrast against the white background. The font choice gives the logo a vintage or medieval feel. The letters are slightly distorted, adding to the unique character of the design.
Jalan Sci Bclumai Desa Dalu 10 A Dusun 1 Tanjung Morawa - 20362 SUMUT - INDONESIA
Tel +62-61-7944880 +62-61-7944882 Fax
### 6.0 Labels, Labeling and Advertisings:
Copy of dispenser box labeling is provided in Attachment C
### 7.0 Classification Information:
- 7.1 Device Class: I
- 7.2 Substantial Equivalent Device Description: Nitrile Examination Gloves
### 7.3 Product Code
- [] Vinyl 80 LYZ
- [ ] Polymer 80 LZA
- [] Specialty 80 LZC
- [] Finger Cot 80 LZB
- [ ] Other 80 FMC = =
- [V] Nitrile 80 LZA
- [ ] Latex 80 LYY
- [] Latex (Powdered) LYY
- [] Latex (Powder-free) 80 LYY
- [ ] Latex (hypoallergenic) 80 LYY
- [ ] Latex ( powder-free/hypoallergenic- 80LYY
- ] Latex (protein label claim) 80 L Y Y
### 8.0 Specifications:
| Size | 230 | WIDTH, mm |
|------|-----|--------------|
| XS | 230 | $70 \pm 10$ |
| S | 230 | $80 \pm 10$ |
| M | 230 | $95 \pm 10$ |
| L | 230 | $111 \pm 10$ |
| XL | 230 | $115 \pm 10$ |
| SINGLE WALL THICKNESS, mm (minimum) | |
|-------------------------------------|------|
| Finger | 0.08 |
| Palm | 0.08 |
| | BEFORE AGING | AFTER AGING<br>At @ 100°C for 22 hr |
|---------------------|--------------|-------------------------------------|
| Tensile Strength | 14 MPa, min | 14 MPa, min |
| Ultimate Elongation | 500 % min | 400 % min |
ASTM Pinhole Requirements: Single, General Inspection Level G-1, AQL 2.5
### Reference Performance Standards: 8.1
ASTM D 6319 - 00a53 Adherence: [V] FULL [ ] PATRIAL
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a stylized, black and white graphic of the letters "MB" in a gothic-style font. The letters are bold and textured, giving them a rough, almost medieval appearance. The overall design is simple yet striking, with the letters dominating the composition.
Jalan Sei Belumai Desa Dalu 10 A Dusun 1 Tanjung Morawa - 20362 SUMUT - INDONESIA
+62-61-7944880 Tel 462-61-7944882 Fax
### Quality Assurance Testing (of Finished Gloves): 9.0
Does quality assurance conform to ALL ASTM D 6319-00a53 procedures and FDA water leak test?
[V] YES [ ] NO
A summary of Quality Assurance Testing Procedure, Sampling Plan and Acceptance Criteria for Gloves Inspection and Testing, and Device Test Report of Compliance of finished gloves are provided in Attachment D, E and F
### 10.0 Dusting Powder
Information of Powder used is provided in Attachment
#### Finished Powder Free Gloves: 11.0
Information for Finished Powder Free Gloves is provided in Attachment G
#### 12.0 Colour Additive:
A material safety data sheet for the colour additive used and colour fastness test report are provided in Attachment I and J
### 13.0 Glove Biocompatibility:
Powder Free Nitrile Examination Glove, Blue, Non Sterile has been sent for biocompatibility test i.e. FDA Primary Skin Irritation Test and FDA Dermal Sensitization Test were performed on the finished gloves and the test result are provided in Attachment K and L respectively.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# DEC 1 6 2004
Mr. P. Sasitharan Nair Quality Regulatory Manager PT Mandiri Inti Buana Jalan Sei Belumai, Desa Dalu 10 A Dusun 1 Tanjung Morawa-20362 SUMUT-INDONESIA
Re: K042239
Trade/Device Name: Powder Free Nitril Examination Gloves, Blue, Non Sterile Regulation Number: 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LZA Dated: November 23, 2004 Received: November 26, 2004
Dear Mr. Nair:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Nair
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K 04 2 2 3 9
Device Name: POWDER TREE NITRILE EXAMINATION GLOVES, BLUE NON STERILE
Indications For Ilse: pouder Eroe Nitrile Examination Gloves, Non Sterile is a disposable device and make of Synthetic Is a Gropodosatibits rubber like Characteristics Intended For Medical Purpose that is worn on the Incended roz finger to prevent contamination between patient and examiner -
Prescription Use CERBO1 . 109 (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONNIINUE ON ANOTHER PAGE IN NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sonette M. Chins Doos
(Division Sign-Cip Division of Anesths stology Infection Control, Dental Dental Di
510(k) Number. K042239
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.