← Product Code [FDT](/productcode/FDT) · K042076

# FUJINON G5 DUODENSCOPES; G5 DUODENOSCOPES, ED-450XT5, ED-250XT5 (K042076)

_Fujinon, Inc. · FDT · Nov 18, 2004 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K042076

## Device Facts

- **Applicant:** Fujinon, Inc.
- **Product Code:** [FDT](/productcode/FDT.md)
- **Decision Date:** Nov 18, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The device is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

## Device Story

Fujinon G5 Duodenoscopes (ED-250XT5, ED-450XT5) are flexible medical endoscopes for visualization, diagnosis, and treatment of the upper digestive tract. Devices capture optical images via distal CCD sensors and transmit video to external processors/monitors. Used in clinical settings by physicians; operators control bending, suction, and irrigation via handle-mounted valves. G5 series features design improvements over predicates: removal of L-port (lens wash), modified forceps inlet port with rubber cap for improved suction/seal, external/removable jet wash check valves for easier cleaning, and updated suction/air/water cylinders. Devices are used with existing Fujinon processors, light sources, and peripherals. Output allows real-time visualization for clinical decision-making and therapeutic intervention. Benefits include improved ergonomics, enhanced cleaning/sterilization capability, and reduced leakage during procedures.

## Clinical Evidence

Bench testing only. No clinical data provided. Device safety supported by EMC testing and comparison of patient contact materials and design specifications to predicate devices.

## Technological Characteristics

Flexible endoscope with lateral viewing (15-degree retro). Distal end diameter 13.5mm; working length 1250mm. Imaging via Interline Color CCD chip (411,988 pixels). Materials: LaSf-n17 outer lens. Connectivity: compatible with standard Fujinon processors/peripherals. Reprocessing: manual cleaning/disinfection via CA-500 adaptor. Design includes external/removable check valves and rubber forceps inlet valve cover.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Fujinon ED-400XL ([K944620](/device/K944620.md))
- Fujinon ED-200XU ([K944759](/device/K944759.md))

## Reference Devices

- Fujinon Double Balloon Enteroscopy System ([K040048](/device/K040048.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

NOV 18 2004

K042076

# 510 (k) Summary

#### Date Prepared (21 CFR 807.92(a)(1){

July 29, 2004 [Amended 10/27/04]

#### Submitter's Information [21 CFR 807.92(a)(1)|

Joseph M. Azary C/o Fujinon Inc. 543 Long Hill Avenue Shelton, CT. 06484

Azary Technologies has received authorization to submit this 510(k) on behalf of the sponsor Fujinon Inc., 10 High Point Drive, Wayne, NJ 07470, Establishment Registration# 2431293.

### Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)|

The device trade names are: Fujinon G5 Duodenoscopes Common Name: Duodenoscopes Classification: Class II, 21 CFR 876.1500, FAK

#### Predicate Device | 21 CFR 807.92(a)(3)|

- . Fujinon ED-400XL - K944620
- . Fujinon ED-200XU - K944759

The G5 changes were described and cleared by FDA in the 510(k) submission for the Fujinon Double Balloon Enteroscopy System - K040048.

This 510(k) captures some minor design changes that have occurred during the evolution of the product line resulting in the G5 family of scopes. Although the changes are believed to be minor, the 510(k) is being submitted to account for "design creep" and to ensure that FDA has the most current information concerning the Fujinon Duodenoscopes.

The subject devices have the same indications for use, composition of patient contact materials, viewing direction, length, and reprocessing/sterilization method as the predicates. The subject devices use the same processors and peripherals as the predicate device.

The main differences between the subject devices and predicate devices are as follows:

- Minor differences with observation range, field of view, diameter, and bending capacity. ◆
- . The subject device includes the G5 upgrade, which is characterized by the following minor differences:
	- The L-Port has been eliminated. The L-port functioned as a lens wash port . Doctors had ্ the option to take a syringe to inject a fluid to use it as a high pressure wash for the lens.

{1}------------------------------------------------

K. L. L. L.
13. 1. 1.

This function was eliminated because demand was low and it was rarely used by the surgeons.

- A G5 forceps inlet port was modified. The new port is smaller and comes with a rubber ೧ A OF foreeps mice port and rubber cap help increase suction and reduce leakage.
- The jet wash line check valve was removed. Internal check valves were removed to elimi-਼ I he jet wasn'tine cheek farro has research and to facilitate cleaning, disinfection, and sterilization. The valves are now external and removable.
- The suction and air/water cylinders and valves were upgraded. They were updated to ac-0 commodate the new valves. The function of the valves is the same.
- of the FOV, which is the rubber forceps inlet valve cover. This helps create a wa-0 tertight seal when the endoscope is used.
- terright sour when the eleaning adaptor. The cleaning adaptor allows the scope to be con-ಂ nected to tubes for cleaning.

# Description of the Device |21 CFR 807.92(a)(4)|

The Fujinon G5 Duodenoscopes are medical endoscopes used for the visualization of the duodenum and upper I he f unitor US Duodenosoppe are mosevation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

The G5 Duodenoscopes are offered in a 200 and 400 series. The ED-250XT5 represents the G5 Duodenoscope 200 series and the ED-450XT5 represents the G5 Duodenoscope 400 series.

The G5 Duodenoscopes include minor changes that improve the useability, ergonomics, and cleaning of the devices. The G5 scopes do not have an L-Port. The L-port functioned as a lens wash port. Doctors had the option to take a syringe to inject a fluid to use it as a high pressure wash for the lens. This function was eliminated because demand was low and it was rarely used by the surgeons.

A G5 forceps inlet port was modified. The new port is smaller and comes with a rubber cap. The new design helps increase suction and reduces leakage.

The jet wash line check valve was removed. Internal check valves were removed to eliminate the potential for clogging and to facilitate cleaning, disinfection. The valves are now external and removable.

The suction and air/water cylinders and valves were upgraded. They were updated to accommodate the new valves. The function of the valves is the same.

The G5 Duodenoscopes also feature the addition of the FOV, which is the rubber forceps inlet valve cover. This helps create a watertight seal when the endoscope is used. Upgrade to CA-500 cleaning adaptor. The cleaning adaptor allows the scope to be connected to tubes for cleaning.

The G5 Duodenoscopes are used with the same processors, monitors, hard copy units, and carts as the predicate devices. Each Duodenoscope is packaged in a protective carrying case with lens cleaner, silicon oil, forceps valve, waterproof cap, S connector cap, protective cap, adapters, valves, and the Operation Manual.

The Fujinon G5 Duodenoscopes are used in conjunction with other peripherals specified in the Operation Manual such as:

- . Light Source

{2}------------------------------------------------

- Processor .
- Cart .
- . Data Keyboard
- Foot Switch .
- . Monitor
- Video Printer .
- Camera and Hard Copy Unit .
- VCR .
- ElectroSurgical Instruments .

# Specifications for ED-450XT5 and ED-250XT5

|                           | ED-450XT5                  | ED-250XT5                  |
|---------------------------|----------------------------|----------------------------|
| Viewing Direction         | Lateral (15 degrees retro) | Lateral (15 degrees retro) |
| Observation Range         | 3-60mm                     | 5-60mm                     |
| Field of View             | 100 degrees                | 100 degrees                |
| Image Size                | Super Image                | Super Image                |
| Distal End Diameter       | 13.5mm                     | 13.5mm                     |
| Flexible Portion Diameter | 12.6mm                     | 12.6mm                     |
| Bending Capacity UP       | 130 degrees                | 130 degrees                |
| Bending Capacity DOWN     | 90 degrees                 | 90 degrees                 |
| Bending Capacity LEFT     | 90 degrees                 | 90 degrees                 |
| Bending Capacity RIGHT    | 110 degrees                | 110 degrees                |
| Forceps Channel Diameter  | 4.2mm                      | 4.2mm                      |
| Working Length            | 1250mm                     | 1250mm                     |
| Total Length              | 1550mm                     | 1550mm                     |

A 8 4 2 16

{3}------------------------------------------------

K-mmer
77 444

|                                                                           |                                  | ED-450XT5                     | ED-250XT5                     |
|---------------------------------------------------------------------------|----------------------------------|-------------------------------|-------------------------------|
| Material of Outer Lines (GI)                                              |                                  | LaSf-n17                      | LaSf-n17                      |
| Lens Diameter (mm)                                                        |                                  | 2.2                           | 2.2                           |
| Electronic Video Imaging System                                           | Total Pixels                     | 411,988                       | 411,988                       |
|                                                                           | Pixels per square millimeter     | 83,892                        | 83,892                        |
|                                                                           | Size of Pixel (H*V mm)           | 0.0032*0.003725               | 0.0032*0.003725               |
|                                                                           | Active area of CCD chip (H*V mm) | 2.458*1.840                   | 2.458*1.840                   |
|                                                                           | Type of CCD chip                 | Interline Color Chip          | Interline Color Chip          |
| Image Quality Resolution (lines / mm) - Near Point                        |                                  | 15.9 lines/mm<br>(6mm)        | 14.3 lines/mm<br>(7mm)        |
| Image Quality Resolution (lines/mm) - Far Point                           |                                  | 2.00 lines/mm<br>(60mm)       | 1.59 lines/mm<br>(60mm)       |
| Magnification (specify monitor size and type used for this specification) |                                  | 26 (5m) at 14<br>inch monitor | 31 (5m) at 14 inch<br>monitor |
| Focal Length                                                              |                                  | 1.459mm                       | 1.459mm                       |

## Intended Use [21 CFR 807.92(a)(5)]

The device is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

# Technological Characteristics [21 CFR 807.92(a)(6)]

Fujinon, Inc. believes that the subject device is substantially equivalent to the predicate device. The subject device have the same indications for use, composition of patient contact material, viewing direction, length, device nave ne same indications for abs, bettredicate. The subject devices use the same processors and peripherals as the predicate devices.

The main differences are the minor changes associated with the G5 upgrade. The G5 changes were previ-The nam offerences are the 510(k) submission for the Fujinon Double Balloon Enteroscopy System -K040048.

# Performance Data [21 CFR 807.92(b)(1)]

The subject device has been subjected to and passed EMC testing requirements. The patient contact materials in the Duodenoscopes are identical to the materials used in the predicate device.

#### Conclusion [21 CFR 807.92(b)(3)]

We believe the changes are minor and conclude that the subject devices are as safe and effective as the predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 1 8 2004

Fujinon, Inc. c/o Mr. Joseph M. Azary Azary Technologies™ LLC 543 Long Hill Avenue SHELTON CT 06484

Re: K042076

Trade/Device Name: G5 Duodenoscopes Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 FDT Dated: October 27, 2004 Received: October 28, 2004

Dear Mr. Azary:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmadsmam.html

Sincerely yours,

Nancy C. brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

K.4 2076

510(k) Number (if known):

Device Name: Fujinon Inc. G5 Duodenoscopes

The device is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

David A. Bergman

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 042016 510(k) Number.

Over-The-Counter Use ____________

(Optional Format 1-2-96)

Page 4

---

**Source:** [https://fda-staging.innolitics.com/device/K042076](https://fda-staging.innolitics.com/device/K042076)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
