K042002 · Ela Medical, Inc. · DQK · Oct 20, 2004 · Cardiovascular
Device Facts
Record ID
K042002
Device Name
SYNESCOPE, EASYSCOPE, SYNECOM, HOLTERREADER
Applicant
Ela Medical, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Oct 20, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
SyneScope™ is intended to analyze, edit, review, report, store and transfer multi-channel ECG recording on pediatric or adult patients (for periods up to 96 hours). These recordings are provided by: SpiderView™ Holter ECG recorder (K032466), SyneFlash™ Holter ECG (K990727), or other compatible cassette tape recorder. In addition, any ISHNE compatible recording can be read by SyneScope™ EasyScope™ is intended to analyze, edit, review, report, store and transfer 2-3 channel ECG recording on pediatric or adult patients (for periods up to 96 hours). These recordings are provided by: SpiderView™ Holter ECG recorder (K032466), SyneFlash™ Holter ECG (K990727), or other compatible cassette tape recorder. SyneCom™ is intended to download, transfer for analysis, print and store multi-channel ECG recordings on pediatric or adult patients (for periods up to 96 hours). These recordings are provided by: SpiderView™ Holter ECG recorder (K032466) or SyneFlash™ Holter ECG (K990727). HolterReader is intended to view and print Holter report from SyneScope™, EasyScope™ and SyneCom™. SyneScope™, EasyScope™, SyneCom™ and HolterReader are intended to be used under the supervision of licensed and trained practitioners, in a hospital or clinic setting. all superville for Holter monitoring include, but are not limited to, evaluation of the following: Patient symptoms such as syncope, dizziness or palpitations. Ischemia, especially in patients who cannot exercise or in patients with variant angina. Function of an implanted pacemaker or defibrillator.
Device Story
Software suite (SyneScope, EasyScope, SyneCom, HolterReader) for processing Holter ECG data; inputs include digital recordings from SpiderView/SyneFlash recorders, cassette tapes, or ISHNE-compatible files. System performs multi-channel arrhythmia analysis, ST-segment deviation analysis, pacemaker activity detection, and HRV analysis (time/frequency domain). Operates on Windows-based PCs with flash/tape readers. Clinicians use software to review, edit, and report ECG findings; supports trend review, beat-to-beat editing, and full disclosure. Output facilitates clinical diagnosis of cardiac symptoms and ischemia; aids in monitoring implanted device function. Used in hospitals/clinics by trained staff.
Clinical Evidence
Bench testing only. Functional performance and safety testing conducted per AAMI/ANSI EC38:1998 and IEC 60601-2-47. Software verification and validation testing performed at module and functional levels. Field validation protocols completed to confirm performance in clinical environments.
Technological Characteristics
Windows-based software applications; PC-based architecture; supports digital and cassette tape inputs; network compatible; connectivity via flash-card/tape readers and internet; supports ISHNE file formats; multi-channel arrhythmia analysis; ST-segment analysis; pacemaker detection; HRV analysis; 96-hour recording capacity.
Indications for Use
Indicated for pediatric and adult patients requiring Holter ECG monitoring for up to 96 hours to evaluate symptoms (syncope, dizziness, palpitations), ischemia, or pacemaker/defibrillator function. Used under supervision of licensed practitioners in clinical settings.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
042002
page 1 of 5
# 510 (k) Summary of Safety and Effectiveness
### Submitter:
ELA Medical, Inc. 2950 Xenium Lane North, Suite 120 Plymouth, MN 55441 Phone: (763) 519-9400 Fax: (763) 519-9440
## Contact Person:
Susan Olive, Regulatory Affairs Manager
Devices trade names / Common names:
- Trade Name: SyneScope™ 道
- Common name: Holter ECG analysis software (or Holter ECG analyzer, or Holter O ECG scanner).
- Trade Name: EasyScope™ 지
- Common name: Holter ECG analysis software (or Holter ECG analyzer, or Holter o ECG scanner).
- Trade Name: SyneCom™ 해
- o Common name: Holter ECG data transfer software or system.
- Trade Name: HolterReader 특
- Common name: Holter ECG report viewer and printing software or tool.
Note: SyneCom™ and HolterReader are considered accessories of Holter ECG analyzers (they can be distributed separately to analyzers but they are intended to be used in conjunction with them).
Classification name:
Computer, diagnostic, programmable DQK (21CFR 870.1425)
## Predicate devices:
The legally marketed devices to which ELA Medical is claiming equivalence are:
- SyneTec (ELA Medical): ELA Medical "SyneTec™" (510(1) K002817, ELA -Medical Inc.).
- SyneView (ELA Medical): ELA Medical "SyneView™" (510(k) K990727, ELA ﺖ Medical Inc.) including SyneView2™
- H-Scribe (Mortara): Mortara Instrument, Inc. "H-Scribe" (510(k) K004017, Mortara -Instrument, Inc.).
{1}------------------------------------------------
K042002
page 2 of 5
- Holter For Windows + (Rozinn): Rozinn Electronics, Inc. "Holter For Windows +" (based on 510(k) K930564, Northeast Monitoring, Inc.) including also Rozinn Electronics, Inc. "Netlink/NetScan" (510(k) K020213, Rozinn Electronics, Inc.).
- Mida (Ortivus sold by Philips Medical Systems): based on "Mida System Models -1000/1100" (510(k) K896396, Medical Graphics Corp.)
## Device description:
SyneScope™, EasyScope™, SyneCom™ and HolterReader are software applications that Syncocope - ; Lasy Seoper recordings obtained with an ELA Medical Holter ECG digital and & U [SuneFlash™( K990727) and SpiderView™(K032466)], a standard cassette tape recorder, or an ISHNE compatible recording (for SyneScope™ only).
SyneScope™, EasyScope™, SyneCom™ and HolterReader are Microsoft Windows-based Dynoodoop ; Lasy oopersonal computers equipped with flash-card readers and/or tape cassette readers.
## Intended use:
SyneScope™ is intended to analyze, edit, review, report, store and transfer multi-channel ECG recording on pediatric or adult patients (for periods up to 96 hours). These recordings are provided by:
- SpiderView™ Holter ECG recorder (K032466), -
- SyneFlash™ Holter ECG (K990727), or -
- other compatible cassette tape recorder. -
In addition, any ISHNE compatible recording can be read by SyneScope™
EasyScope™ is intended to analyze, edit, review, report, store and transfer 2-3 channel ECG recording on pediatric or adult patients (for periods up to 96 hours). These recordings are provided by:
- SpiderView™ Holter ECG recorder (K032466), -
- SyneFlash™ Holter ECG (K990727), or -
- other compatible cassette tape recorder. ﺳ
SyneCom™ is intended to download, transfer for analysis, print and store multi-channel ECG recordings on pediatric or adult patients (for periods up to 96 hours). These recordings are provided by:
- SpiderView™ Holter ECG recorder (K032466) or -
- SyneFlash™ Holter ECG (K990727). -
{2}------------------------------------------------
042002
page 3 of 5
HolterReader is intended to view and print Holter report from SyneScope™, EasyScope™ and SyneCom™.
SyneScope™, EasyScope™, SyneCom™ and HolterReader are intended to be used under the supervision of licensed and trained practitioners, in a hospital or clinic setting. all superville for Holter monitoring include, but are not limited to, evaluation of the following:
- Patient symptoms such as syncope, dizziness or palpitations. .
- Ischemia, especially in patients who cannot exercise or in patients with variant . angina.
- Function of an implanted pacemaker or defibrillator. .
| Model | SyneScope | EasyScope | SyneTec | SyneView/<br>SyneView 2 | H-Scribe | Holter for<br>Windows |
|-----------------------------------------------------------|-----------------------------------------------------|------------------------|----------------------|--------------------------|----------------------|-------------------------------|
| Company | ELA Medical | ELA Medical | ELA Medical | ELA Medical | Mortara | Rozinn |
| 510(k) Number | | | K002817 | K990727 | K004017 | K930564 |
| CE / MDD<br>certification | Yes | Yes | Yes | Yes | Yes | Yes |
| Type | Holter Analyzer | Holter Analyzer | Holter Analyzer | Holter Analyzer | Holter Analyzer | Holter Analyzer |
| PC based | Yes | Yes | Yes | Yes | Yes | Yes |
| Operating<br>System | Microsoft<br>Windows | Microsoft<br>Windows | Microsoft<br>Windows | Microsoft<br>Windows | Microsoft<br>Windows | Microsoft<br>Windows |
| Network<br>Compatible | Yes | Yes | Yes | Yes | Yes | Yes |
| Read digital | Yes | Yes | Yes | Yes | Yes | Yes |
| Cassette Tape | Yes | Yes | Yes | Yes | Yes | Yes |
| ECG Channels | Basic: 2/3 | Basic: 2/3 | Basic: 2/3 | Basic: 2/3 | Basic: 2/3 | Basic: 2/3 |
| | Max: 12 | Max: 3 | Max: 3 | Max: 3 | Max: 12 | Max: 12 |
| Multi-Channel<br>Arrhythmia<br>Analysis | Yes | Yes | Yes | Yes | Yes | Yes |
| 24H Full<br>Disclosure | Yes | Yes | Yes | Yes | Yes | Yes |
| ST deviation<br>analysis | Yes | Yes | Yes | Yes | Yes | Yes |
| Pacemaker<br>Activity | Yes | Yes | Yes | Yes | Yes | Yes |
| Trend Review | Yes | Yes | Yes | Yes | Yes | Yes |
| Templates<br>editing | Yes | Yes | Yes | Yes | Yes | Yes |
| Beat to Beat<br>editing | Yes | Yes | Yes | Yes | Yes | Yes |
| Strip editing | Yes | Yes | Yes | Yes | Yes | Yes |
| ECG<br>Superimposition | Yes | Yes | Yes | Yes | Yes | Yes |
| Page Mode | Yes | Yes | Yes | Yes | Yes | Yes |
| Editing report | Yes | Yes | Yes | Yes | Yes | Yes |
| Model | SyneScope | EasyScope | SyneTec | Syne View/<br>SyneView 2 | H-Scribe | Holter for<br>Windows+ |
| Company | ELA Medical | ELA Medical | ELA Medical | ELA Medical | Mortara | Rozinn |
| 510(k) Number | | | K002817 | K990727 | K004017 | K930564 |
| Customized<br>report | Yes | Yes | Yes | Yes | Yes | Yes |
| Archiving | Yes | Yes | Yes | Yes | Yes | Yes |
| Printing ECG<br>strip | Yes | Yes | Yes | Yes | Yes | Yes |
| Printing Full<br>Disclosure | Yes | Yes | Yes | Yes | Yes | Yes |
| Import/Export<br>capabilities | Yes | No | Yes | No | Yes | Yes |
| Import ASCII<br>RR Files | Yes | No | Yes | No | | |
| Export Results in<br>ASCII format | Yes | No | Yes | No | | |
| Export RR or<br>HR files in ASCII<br>format | Yes | No | Yes | No | | |
| Export ECG files<br>in ISHNE format | Yes | No | Yes | No | | |
| Read High<br>Resolution ECG<br>file | Yes | No | Yes | No | | |
| Export High<br>Resolution ECG<br>files in ISHNE<br>format | Yes | No | Yes | No | | |
| Heart Rate<br>Variability<br>(HRV) | Yes | Partial | Yes | Partial | Yes | Yes |
| Time Domain<br>(TD) | Yes | No | Yes | No | Yes | Yes |
| Frequency<br>Domain (FD) | Yes | No | Yes | No | No | Yes |
| Multiday | Yes (96H) | Yes (96H) | No | No | Yes (48H) | Yes (48H) |
| True 12-Lead<br>ECG | Yes | No | No | No | Yes | Yes |
| 12-Lead ECG<br>Derived | Yes (Dower<br>method - from<br>XYZ-Lead<br>system)* | No | No | No | No | No |
| 12-Lead ST<br>segment analysis | Yes | No | No | No | Yes | Yes |
| Holter Data<br>Transfer | Yes<br>(like SYNECOM) | Yes<br>(like SYNECOM) | No | No | Yes | Yes |
| Holter ECG<br>Report<br>Viewer/Editor | Yes (Holter<br>Reader) | Yes (Holter<br>Reader) | No | No | Yes | Yes |
| Import ISHNE<br>ECG file | Yes | No | No | No | No | No |
| Internet | Yes | Yes | No | No | Yes | Yes (with<br>Netlink/Netscan) |
## Comparison of technology characteristics to predicate devices:
{3}------------------------------------------------
042002
page 4 of 5
* see Attachment 10.8 for details.
·
{4}------------------------------------------------
y2002
page 5 of 5
#### Summary of Studies:
The following functional testing was performed on the SyneScope™, EasyScope™ SyneCom™ and HolterReader:
| Test group | Tests | Report of results |
|-------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|----------------------|
| SyneScopeTM, EasyScopeTM,<br>SyneComTM and HolterReader<br>safety and performance<br>testing. | Safety and performance tests<br>according to the AAMI/ANSI<br>EC38:1998 & IEC 60601-2-47<br>standards. | See Attachment 10.9. |
| SyneScopeTM, EasyScopeTM,<br>SyneComTM and HolterReader<br>software verification and<br>validation testing. | Module and functional testing<br>for SyneScopeTM,<br>EasyScopeTM, SyneComTM<br>and HolterReader software<br>applications | See Attachment 10.6. |
| SyneScopeTM, EasyScopeTM,<br>SyneComTM and HolterReader<br>field testing | Field validation protocol and<br>evaluation form completion. | See Attachment 10.4. |
All test systems were representative of final production devices unless otherwise specified.
Compliance with European Medical Device Directive 93/42/CEE and European standards IEC-60601-2-47 was verified.
See Attachment 10.7 for copies of the SyneScope™, EasyScope™, SyneCom™ and HolterReader declarations of conformity and CE certificates.
#### Conclusion:
The information presented in this submission provides reasonable assurance that the SyneScope™, EasyScope™, SyneCom™ and HolterReader will perform in a safe and effective manner.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## OCT 2 0 2004
ELA Medical, Inc. c/o Ms. Susan Olive Regulatory Affairs Manager 2950 Xenium Lane North, Suite 120 Plymouth, MN 55441
Re: K042002
K042002
Trade Name: SyneScope™, EasyScope™, SyneCom™, and HolterReader Holter Analyzers
11:11 PM EasyScope™, EasyScope 1125 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: DQK Dated: July 23, 2004 Received: July 26, 2004
Dear Ms. Olive:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section > IQ(x) premainer is substantially equivalent (for the indications
referenced above and have determined the devices marketed in interstate referenced above and nave determined the devices marketed in interstate
for use stated in the enclosure) to legally marketed predicate devices Amendments. for use stated in the enclosure) to regary manced produce aredical Device Ameralments, or to
commerce prior to May 28, 1976, the enactment of the wassision of the Federal Foo commerce prior to May 28, 1976, the enaculient and of the Federal Food. Drug, devices that have been reclassified in accordance who are approval application (PMA).
and Cosmetic Act (Act) that do not require approval assesses approvisions of the Act. T and Cosmetic Act (Act) that do not require approval or a pal controls provisions of the Act. The
You may, therefore, market the device, subject to the general conversior at You may, therefore, market the device, subject to annual registration, listing of
general controls provisions of the Act include requirements misbranding and general controls provisions of the Act mende requirements as a seasonst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (secaded) in clicier case in personations affecting your device can
may be subject to such additional controls. Existing major regulations allec may be subject to such additional controlist Extrong may on to 898. In addition, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 to Register be found in the Code of Federal Regaralier, our device in the Federal Register.
{6}------------------------------------------------
Page 2 - Ms. Susan Olive
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA's issuance of a substance oquires with other requirements of the Act
that FDA has made a determination that your device adders Reguirements of the that FDA has made a determination in a your develop other Federal agencies. You must
or any Federal statutes and regulations administed to: registration and listing or any Federal statutes and regulations and limited to registration and listing (21
comply with all the Act's requirements, including, but not limited to registration as set comply with all the Act s requirements, including practice requirements as set
CFR Part 807); labeling (21 CFR Part 801); god manufacturing practice requirements as set CFR Part 807); labeling (21 CFR Part 801); good manufacturing nd if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and 1000-1050. forth in the quality systems (QS) regulanton (2 - CF (CFR) 0f the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 53 -542 of the Act); 21 CFR 1000-1050 product radiation control provisions (Sections 351-37), and recorp), as in the stock of S (VC)
This letter will allow you to TB (Secures restants againstical nyour device to This letter will allow you to begin marketing your active of your device of your device to a legally
premarket notification. The FDA finding of substantial equivales on thus premarket notification. The PDA inding of subsamial equively .
marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
rstiled. If you desire specific advice for your device of bar noomig the regulation entitled,
contact the Office of Compliance at (301) 594-4646. Also, please not of the regulation contact the Office of Compliatics at (301) 57 1 107.07) you may obtain.
"Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtained from the "Misbranding by reference to premarket nothlivanom (210) Act may be obtained from the Other general information on your responsionial and Consumer Assistance at its toll-free number
Division of Small Manufacturers, International and Consumer Assistance at its Division of Simall Mail Mail 2011 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
B. Hummer for
MD
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number: K042002
Device Name:
Boves Hames
SyneScope™ (Holter ECG analysis software or Holter ECG analyzer, or Holter ECG scanner).
ecanner)
EasyScope™ (Holter ECG analysis software or Holter ECG analyzer, or Holter ECG scanner)
SyneCom™ (Holter ECG data transfer software or system)
HolterReader (Holter ECG report viewer and printing software or tool)
Indications For Use:
SyneScope™ is intended to analyze, edit, review, report, store and transfer multi-Synescope - Is Intended to anialyze, radult patients (for periods up to 96 hours). These recordings are provided by:
- SpiderView™ Holter ECG recorder (K032466), -
- SyneFlash™ Holter ECG (K990727), or ﺴ
- other compatible cassette tape recorder. -
In addition, any ISHNE compatible recording can be read by SyneScope™
EasyScope™ is intended to analyze, edit, review, report, store and transfer 2-3 channel Easyocope - 13 intended to analyal etients (for periods up to 96 hours). These recordings are provided by:
- SpiderView™ Holter ECG recorder (K032466), -
- SyneFlash™ Holter ECG (K990727), or -
- other compatible cassette tape recorder. -
SyneCom™ is intended to download, transfer for analysis, print and store multi-channel Synecom "Is Intended to download intinuts (for periods up to 96 hours). These recordings are provided by:
- SpiderView™ Holter ECG recorder (K032466), -
- SyneFlash™ Holter ECG (K990727) -
HolterReader is intended to view and print Holter reports from SyneScope™ p
orts of TM and Comment The TW EasyScope™ and SyneCom™ .
page 1 of 2
{8}------------------------------------------------
SyneScope™, EasyScope™, SyneCom™ and HolterReader are intended to be used under the supervision of licensed and trained practitioners, in a hospital or clinic setting. Applications for Holter monitoring include, but are not limited to, evaluation of the following:
- Patient symptoms such as syncope, dizziness or palpitations. .
- Ischemia, especially in patients who cannot exercise or in patients with variant . angina.
- Function of an implanted pacemaker or defibrillator. .
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
B.B. Simmons
(Division Sign-Off) Division of Cardiovascular Devices 510(k) Number
Page 2
1 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.