# APEX K2 HIP SYSTEM (K041950)

_Apex Surgical, LLC · LPH · Dec 21, 2004 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K041950

## Device Facts

- **Applicant:** Apex Surgical, LLC
- **Product Code:** [LPH](/productcode/LPH.md)
- **Decision Date:** Dec 21, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Apex K2™ Hip Stem is intended for use as the femoral component of a primary or revision total hip replacement. This femoral hip stem is intended for uncemented fixation and single use implantation. This prosthesis may be used for the following conditions, as appropriate: . Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; . Rheumatoid arthritis; . Correction of functional deformity: . Congenital dislocation; . Revision procedures where other treatments or devices have failed; . . Femoral neck and trochanteric fractures of the proximal femur.

## Device Story

The Apex K2 Hip Stem is a femoral component for total hip arthroplasty; designed for uncemented fixation. The system comprises a rectangular tapered titanium alloy stem, modular titanium alloy necks, and modular cobalt chromium or alumina ceramic heads. The modularity allows surgeons to select component combinations to match patient anatomy and refine leg length and offset. The proximal metaphyseal region features a plasma-sprayed unalloyed titanium coating for osseointegration. Used in clinical settings by orthopedic surgeons; the device is implanted during hip replacement surgery to restore joint function and alleviate pain associated with degenerative conditions or fractures. The device may be used with the Apex Modular Acetabular Cup.

## Clinical Evidence

Bench testing only. Performance testing of modularity and plasma-sprayed unalloyed titanium coating was conducted according to relevant FDA guidance documents.

## Technological Characteristics

Materials: Titanium alloy (Ti6Al4V) stem/neck, unalloyed titanium plasma-spray coating, cobalt chromium or alumina ceramic heads. Design: Rectangular tapered stem, modular neck/head configuration. Fixation: Uncemented, porous-coated. Energy source: None (mechanical).

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- SL-Plus® ([K001942](/device/K001942.md))
- SLR-Plus® ([K021178](/device/K021178.md))
- Apex Modular™ Hip System ([K000788](/device/K000788.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the date "DEC 21 2004" in a simple, sans-serif font. The letters and numbers are uniformly sized and spaced. The text is likely part of a document or label where the date is being recorded.

DEC 21 2004

July 19, 2004

|  |  |  | Apex K2™ Hip Stem |
|--|--|--|-------------------|
|--|--|--|-------------------|

Apex Surgical, LLC

Lakeville, MA 02347

12 Harding Street

Suite 202

Contact: Edward J. Cheal, Ph.D. Managing Director (508) 947-6500 (voice) (208) 248-8227 (fax)

### 2. Device Name

1. Submitter:

Proprietary Name: Apex K2™ Hip Stem Common Name: Hip prosthesis, uncemented Classification Name: Hip joint metal/polymer/metal semi-constrained porouscoated uncemented prosthesis Requlatory Class: Class II per 21 CFR §888.3358

### 3. Intended Use

The Apex K2™ Hip Stem is intended for use as the femoral component of a primary or revision total hip replacement. This femoral hip stem is intended for uncemented fixation and single use implantation. This prosthesis may be used for the following conditions, as appropriate:

- . Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- . Rheumatoid arthritis;
- Correction of functional deformity: .
- Congenital dislocation; .
- Revision procedures where other treatments or devices have failed; .
- . Femoral neck and trochanteric fractures of the proximal femur.

#### 4. Device Description

The Apex K2™ Hip Stem consists of a rectangular tapered stem, modular necks that connect to the proximal end of the stem, and the modular heads that connect to the tapered trunion on the neck. This configuration allows the user to choose a combination of stem, neck, and head components to appropriately fit the anatomy of the patient. The various neck sizes allow for several length and lateral offset options for a given stem size. Several offset options are also available for the heads to allow further refinement of the lengths and offsets. The Apex K2 Hip Stem may be used in conjunction with the Apex Modular Acetabular Cup (K031110) for total hip arthroplasty.

The femoral stems (and modular necks) are manufactured from titanium alloy. The Apex K2 stems can be used with the cobalt chromium alloy heads and the alumina ceramic heads that are part of the Apex Modular hip system. The proximal metaphyseal region of each size femoral stem is circumferentially coated with unalloyed titanium applied by

{1}------------------------------------------------

K041950 (pg 2 of 2)

plasma spray, the same coating method used by the predicate Apex Modular hip stem. As in the predicate Apex Modular stem, the alignment pin in the stem is manufactured from wrought cobalt chromium alloy.

## 5. Predicate Device Comparison

Substantial equivalence is claimed to the SL-Plus® and SLR-Plus® hip stems distributed by Plus Orthopedics (K001942 and K021178), and the Apex Modular ™ Hip System (K000788). The table below compares the features and characteristics of the Apex Modular™ Hip Stem to these predicate devices:

|                                                               | Apex K2™ Hip<br>Stem  | Apex<br>Modular™<br>(K000788) | SL-Plus® and<br>SLR-Plus®<br>(K001942, K021178) |
|---------------------------------------------------------------|-----------------------|-------------------------------|-------------------------------------------------|
| INTENDED USE                                                  |                       |                               |                                                 |
| Primary and revision hip<br>replacement, non-<br>cemented use | Yes                   | Yes                           | Yes                                             |
| DESIGN                                                        |                       |                               |                                                 |
| Porous coated                                                 | Yes - plasma<br>spray | Yes - plasma<br>spray         | No                                              |
| Proximal coating (only)                                       | Yes                   | Yes                           | NA                                              |
| Modular head                                                  | Yes                   | Yes                           | Yes                                             |
| Surface finish                                                | Al-oxide grit blast   | Ti grit blast                 | Al-oxide grit blast                             |
| Modular neck                                                  | Yes                   | Yes                           | No                                              |
| Tapered stem                                                  | Yes                   | No                            | Yes                                             |
| Distal Cross-sectional<br>shape                               | Rectangular           | Round                         | Rectangular                                     |
| Distal slot(s)                                                | No                    | Yes                           | No                                              |
| Distal flutes                                                 | No                    | Yes - ridges                  | No                                              |
| Proximal steps                                                | No                    | Yes                           | No                                              |
| MATERIALS                                                     |                       |                               |                                                 |
| Titanium alloy (Ti6Al4V)<br>stem and neck                     | Yes                   | Yes                           | Yes                                             |
| Cobalt chromium or<br>alumina ceramic heads                   | Yes (both)            | Yes (both)                    | Yes (both)                                      |
| Titanium porous coating                                       | Yes - unalloyed       | Yes - unalloyed               | No                                              |

The Apex K2 stem geometry is similar to the SL-Plus® and SLR-Plus® distributed by Plus Orthopedics. The most significant difference between these devices is that the Apex K2 stem employs modular necks (and heads) similar to the Apex Modular™ hip system, whereas the SL-Plus and SLR-Plus stems have modular heads (only). Performance testing of the modularity was completed as per the relevant FDA guidance documents. Performance testing of the plasma sprayed unalloyed (CP) titanium coating was completed as per the relevant FDA guidance documents by APS Materials, Inc. and Bio-Coat, Inc.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract bird or human figure, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 2 1 2004

Edward J. Cheal, Ph.D. Managing Director Apex Surgical, LLC 12 Harding Street. Suite 202 Lakeville, Massachusetts 02347

Re: K041950 Trade Name: Apex K2"Hip Stem Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis. Regulatory Class: II Product Code: LPH Dated: July 19, 2004 Received: July 20, 2004

Dear Dr. Cheal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Edward J. Cheal, Ph.D.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Wlankoth

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Koria50

# Indications for Use

510(k) Number (if known): K041950

Device Name: Apex K2™ Hip Stem

Indications For Use:

The Apex K2™ Hip Stem is intended for use as the femoral component of a primary total hip replacement. This femoral hip stem is intended for uncemented fixation and single use implantation. This prosthesis may be used for the following conditions, as appropriate:

- Non-inflammatory degenerative joint disease, including osteoarthritis and . avascular necrosis;
- Rheumatoid arthritis; .
- Correction of functional deformity; .
- Congenital dislocation; .
- Revision procedures where other treatments or devices have failed; .
- Femoral neck and trochanteric fractures of the proximal femur. .

| Prescription Use | X |
|------------------|---|
|------------------|---|

AND/OR

(Per 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

> Concur ce of CDRH, Office of Device Evaluation (ODE)

ing with
ford to

**Division**

1. 1949 22 22 22 47 24

Neurological Devices

510(k) Number K041950

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda-staging.innolitics.com/device/K041950](https://fda-staging.innolitics.com/device/K041950)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).