CLINIQA LIQUID QC TDM CONTROLS LEVELS 1,2 & 3/CLINIQA LINICAL TDM CALIBRATION VERIFIERS LEVELS A-E FOR OLYMPUS AU SYSTEM
Applicant
Cliniqa Corporation
Product Code
JJX · Clinical Chemistry
Decision Date
Aug 16, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
Cliniqa Liquid QCᵗʰ TDM Controls Level 1, 2, & 3 are intended for use as an assayed quality control material. Three assayed levels of Acetaminophen, Amikacin, Carbamazepine, Digoxin, Disopyramide, Ethosuximide, Gentamicin, Lidocaine, Lithium, N-Acetylprocainamide, Phenobarbital, Phenytoin, Primidone, Procainamide, Quinidine, Salicylate, Theophylline, Tobramycin, Valproic Acid, and Vancomycin are provided. Cliniqa LiniCALᵗʰ TDM Calibration Verifiers Levels A – E for Olympus AU Systemsᵗʰ are assayed, liquid, quality control products which may be used to evaluate the performance of the Olympus AU Systems™ for Acetaminophen, Amikacin, Caffeine, Carbamazepine, Digoxin, Disopyramide, Gentamicin, Lidocaine, Lithium, N-Acetylprocainamide, Phenobarbital, Phenytoin, primidone, Procainamide, Quinidine, Theophylline, Tobramycin, Valproic Acid, and Vancomycin at five useful concentrations.
Device Story
Liquid quality control materials and calibration verifiers; serum-based matrix; used in clinical laboratories to monitor accuracy and precision of therapeutic drug monitoring (TDM) assays. Controls provided in three levels; calibration verifiers provided in five concentrations. Used on Olympus AU Systems to evaluate assay performance. Stability determined via Arrhenius model of accelerated elevated temperature studies; open-vial stability validated via real-time on-board testing. Provides objective verification of clinical method performance.
Clinical Evidence
No clinical data. Bench testing only. Stability characteristics determined using Arrhenius model of accelerated elevated temperature studies. Open vial stability determined via real-time on-board study on Olympus AU 400 chemistry analyzer.
Technological Characteristics
Serum-based liquid quality control material. Stability testing via Arrhenius model. Designed for use on Olympus AU Systems. No specific software algorithm or electronic connectivity described.
Indications for Use
Indicated for use as assayed quality control materials and calibration verifiers for monitoring therapeutic drug levels (TDM) on Olympus AU Systems. Intended for prescription use in clinical laboratory settings to evaluate assay performance and accuracy.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
510(k) Number: K041898
Purpose for Submission: Notification of intent to manufacture and market the device: Liquid TDM Controls (Three Levels)
Analyte: TDM – Acetaminophen, Amikacin, Carbamazepine, Digoxin, Disopyramide, Ethosuximide, Gentamicin, Lidocaine, Lithium, N-Acetylprocainamide, Phenobarbital, Phenytoin, Primidone, Procainamide, Quinidine, Salicylate, Theophylline, Tobramycin, Valproic Acid, and Vancomycin.
Calibration Verifiers – Acetaminophen, Amikacin, Caffeine, Carbamazepine, Digoxin, Disopyramide, Gentamicin, Lidocaine, Lithium, N-Acetylprocainamide, Phenobarbital, Phenytoin, Primidone, Procainamide, Quinidine, Theophylline, Tobramycin, Valproic Acid, and Vancomycin.
A. Type of Test: n/a
B. Applicant: Cliniqa Corporation
C. Proprietary and Established Names: Proprietary - Cliniqa Liquid QCᵗʰ TDM Controls Level 1, 2, & 3 – Cliniqa LiniCALᵗʰ TDM Calibration Verifiers Levels A – E for Olympus Systemsᵗʰ Established – Quality Control Material
Regulatory Information:
1. Regulation section: 21 CFR 862.1660
2. Classification: Class I
3. Product Code: JJX
4. Panel: 75
D. Intended Use:
1. Intended use(s): Cliniqa Liquid QCᵗʰ TDM Controls Level 1, 2, & 3 are intended for use as an assayed quality control material. Three assayed levels of Acetaminophen, Amikacin, Carbamazepine, Digoxin, Disopyramide, Ethosuximide, Gentamicin, Lidocaine, Lithium, N-Acetylprocainamide, Phenobarbital, Phenytoin, Primidone, Procainamide, Quinidine, Salicylate, Theophylline, Tobramycin, Valproic Acid, and Vancomycin are provided.
Cliniqa LiniCALᵗʰ TDM Calibration Verifiers Levels A – E for Olympus AU Systemsᵗʰ are assayed, liquid, quality control products which may be used to
{1}
Page 2 of 5
evaluate the performance of the Olympus AU Systems™ for Acetaminophen, Amikacin, Caffeine, Carbamazepine, Digoxin, Disopyramide, Gentamicin, Lidocaine, Lithium, N-Acetylprocainamide, Phenobarbital, Phenytoin, primidone, Procainamide, Quinidine, Theophylline, Tobramycin, Valproic Acid, and Vancomycin at five useful concentrations.
2. **Indication(s) for use**: These products can aid clinical laboratory personnel to objectively monitor the accuracy and precision of clinical methods.
3. **Special condition for use statement(s)**: For prescription use only
4. **Special instrument Requirements**: Cliniqa LiniCAL™ TDM Calibration Verifiers Levels A – E for Olympus AU Systems™ are designed for use on the Olympus AU Systems™.
E. **Device Description**: Quality Control Material (Assayed)
**Substantial Equivalence Information**:
1. **Predicate device name(s)**: Cliniqa Liquid QC™ TDM Controls Level 1, 2, & 3 are substantially equivalent to MAS PAR Brand TDM Liquid Assayed Therapeutic Drug Monitoring Controls Levels 1, 2, & 3. Cliniqa LiniCAL™ TDM Calibration Verifiers Levels A – E for Olympus AU Systems™ are substantially equivalent to LiniCAL™ Clinical Chemistry Verifiers Levels A – E for the Beckman Coulter Synchron Systems.
2. **Predicate K number(s)**: MAS PAR Brand TDM Liquid Assayed Therapeutic Drug Monitoring Controls Levels 1, 2, & 3– k032826 & LiniCAL™ TDM Calibration Verifiers Levels A – E for Olympus AU Systems™ – k031921.
3. **Comparison with predicate**:
| Cliniqa Liquid QC™ TDM Controls Level 1, 2, & 3 | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Assayed quality control material | Assayed quality control material |
| Matrix | Serum based liquid | Serum based liquid |
| Constituents | Acetaminophen, Amikacin, Carbamazepine, Digoxin, Disopyramide, Ethosuximide, Gentamicin, Lidocaine, Lithium, N-Acetylprocainamide, Phenobarbital, Phenytoin, | Acetaminophen, Amikacin, Amitriptyline, Caffeine, Carbamazepine, Chloramphenicol, Cocaine, Cyclosporine, Desipramine, Digoxin, |
{2}
Page 3 of 5
| Cliniqa Liquid QC^{tm} TDM Controls Level 1, 2, & 3 | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Target range | Primidone, Procainamide, Quinidine, Salicylate, Theophylline, Tobramycin, Valproic Acid, and Vancomycin | Disopyramide, Estriol, Ethanol, Ethosuximide, Gentamicin, Imipramine, Lidocaine, Lithium, Methotrexate, N-Acetylprocainamide, Netilmicin, Nortriptyline, Phenobarbital, Phenytoin, Primidone, Procainamide, Quinidine, Salicylate, Theophylline, Tobramycin, Total T3-Triidothyronine, Total T4, Tricyclic Antidepressants, TSH, Valporic Acid, and Vancomycin.
Assigned values differ |
| | Assigned values differ | |
| Cliniqa LiniCAL^{tm} TDM Calibration Verifiers Levels A – E for Olympus AU Systems^{tm} | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Assayed quality control material | Assayed quality control material |
| Matrix | Serum based liquid | Serum based liquid |
| Constituents | Acetaminophen, Amikacin, Carbamazepine, Digoxin, Disopyramide, Ethosuximide, Gentamicin, Lidocaine, Lithium, N-Acetylprocainamide, Phenobarbital, Phenytoin, Primidone, Procainamide, Quinidine, Salicylate, Theophylline, Tobramycin, | Albumin, BUN, calcium, creatinine, lactate, magnesium, phosphorus, total protein, triglycerides, glucose, iron, sodium, potassium, chloride. |
{3}
Page 4 of 5
| Cliniqa LiniCAL^{tm} TDM Calibration Verifiers Levels A – E for Olympus AU Systems^{tm} | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | Valproic Acid, and Vancomycin | |
F. Standard/Guidance Document Referenced (if applicable): Arrhenius model of accelerated elevated temperature studies.
G. Test Principle: N/A
Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility: N/A
b. Linearity/assay reportable range: N/A
c. Traceability (controls, calibrators, or method): “Expected Values” presented in each lot-specific insert of CLINIQA Liquid QC TDM Controls Levels 1, 2 & 3 will be generated on the Roche Integra and the Olympus AU Systems. Expected values presented in each lot-specific insert of LiniCAL TDM Calibration Verifiers Levels A – E will be generated on the Olympus AU Systems. Data points will be obtained from replicate assays obtained by multiple laboratories. A minimum of 12 data points will be used to determine the mean. Within and between assay Standard Deviations and Coefficient of Variations will be calculated for each set of data. Stability characteristics of the CLINIQA Liquid QC TDM Controls Levels 1, 2 & 3 and the of LiniCAL TDM Calibration Verifiers Levels A – E were determined using the Arrhenius model of accelerated elevated temperature studies to predict estimated storage stability at 2 – 8 °C. Open vial stability was determined in a real time on board stability study at 2-8 °C on an Olympus AU 400 chemistry analyzer.
d. Detection limit: N/A
e. Analytical specificity: N/A
f. Assay cut-off: N/A
2. Comparison studies:
a. Method comparison with predicate device: N/A
{4}
Page 5 of 5
b. Matrix comparison: N/A
3. Clinical studies:
a. Clinical sensitivity: N/A
b. Clinical specificity: N/A
c. Other clinical supportive data (when a and b are not applicable): N/A
4. Clinical cut-off: N/A
5. Expected values/Reference range: N/A
H. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.