K041882 · Pointcare Technologies, Inc. · GKZ · Nov 23, 2004 · Hematology
Device Facts
Record ID
K041882
Device Name
FLOWCARE SYSTEM
Applicant
Pointcare Technologies, Inc.
Product Code
GKZ · Hematology
Decision Date
Nov 23, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
The FlowCare™ System is an automated immune hematology system intended for in vitro diagnostic use in the direct enumeration of White Blood Cell populations and certain T-lymphocyte subsets from human whole blood. The FlowCare™ CD4 Reagent Kit using the FlowCare™ System is intended for in vitro diagnostic use in clinical laboratory settings.
Device Story
FlowCare™ System is an automated immune hematology analyzer for clinical laboratory use. It processes anticoagulated whole blood samples using proprietary non-fluorescent reagents containing anti-CD4 monoclonal antibodies conjugated to colloidal gold. The system performs automated aspiration, mixing, and selective erythrocyte lysing. The instrument uses light scatter (forward scatter and time-of-flight) to differentiate leukocyte populations. Software performs automated cluster analysis to gate lymphocytes and quantify CD4+ T-lymphocytes based on scatter signatures resulting from the gold-conjugate binding. The system includes a touch screen computer for operation and data analysis, and an inkjet printer for results. It provides quantitative results for WBC count, lymphocyte subsets, and CD4 counts, aiding in the assessment of immune status, particularly in immunocompromised populations. The system is factory-calibrated and utilizes barcode reading for sample identification.
Clinical Evidence
Multi-site prospective study at four investigational sites (n=414 samples). Compared FlowCare System results to commercially available reference methods. Pooled data analysis demonstrated comparable means and ranges for all measured parameters (WBC count, lymphocyte percentage/number, CD4 count/percentage). Bench testing included electrical safety (UL 61010A-1, CSA C22.2 1010-1, EN 61326-1, EN 55011), precision (within-run, total, and within-day), linearity across cell concentration ranges, and carryover assessment (negligible).
Technological Characteristics
Compact benchtop flow-based analyzer. Uses non-fluorescent reagents and light scatter measurements for cell analysis. Automated sample processing. Electrical safety compliance: UL 61010A-1, CSA C22.2 1010-1, EN 61326-1, EN 55011. Software-based automated cell population cluster analysis.
Indications for Use
Indicated for the analysis of whole blood samples to determine White Blood Cell Count, Lymphocyte Percentage, Lymphocyte Number, CD4 T-Lymphocyte Count, and CD4 Percentage of total Lymphocytes.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Coulter® tetraONE™ System for EPICS XL Flow Cytometry Systems (K990172)
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE AND INSTRUMENT TEMPLATE
A. 510(k) Number:
K041882
B. Purpose for Submission:
To seek clearance for the FlowCare™ CD4 Reagent and the FlowCare™ System, are components of an automated immune hematology system intended for in vitro diagnostic use in the direct enumeration of White Blood Cell populations and certain T-lymphocyte subsets from human whole blood.
C. Measurand:
White Blood Cell Count, Lymphocyte Percentage (of WBCs), Lymphocyte Number, CD4 T-Lymphocyte Count, CD4 Percentage (of total Lymphocytes)
D. Type of Test:
Quantitative, automated flow-based
E. Applicant:
PointCare Technologies, Inc.
F. Proprietary and Established Names:
FlowCare™ System and FlowCare™ CD4 Reagent Kit
G. Regulatory Information:
1. Regulation section:
21 CFR 864.5220, Automated Differential Cell Counter
2. Classification:
Class II
3. Product Code:
GKZ
4. Panel:
81 Hematology
H. Intended Use:
1. Intended use(s):
The FlowCare™ System is an automated immune hematology system intended for in vitro diagnostic use in the direct enumeration of White Blood Cell populations and certain T-lymphocyte subsets from human whole blood.
The FlowCare™ CD4 Reagent Kit using the FlowCare™ System is intended for in vitro diagnostic use in clinical laboratory settings.
{1}
Page 2 of 11
2. **Indication(s) for use:**
Whole blood samples can be analyzed with the FlowCare™ System for the following parameters:
White Blood Cell Count
Lymphocyte Percentage (of White Blood Cells)
Lymphocyte Number
CD4 T-Lymphocyte Count
CD4 Percentage (of total Lymphocytes)
3. **Special condition for use statement(s):**
N/A
4. **Special instrument Requirements:**
N/A
I. **Device Description:**
The FlowCare™ System consists of an analysis instrument, a touch screen computer and a standard inkjet printer.
The touch screen computer is the central processing unit for system operation and data analysis. The Windows-based software analyzed the patient sample for immune hematology parameters with optimized cluster finding algorithms.
The FlowCare™ CD4 Reagent Kit has been formulated specifically for use with the FlowCare™ System to provide optimal assay performance for White Blood Cell differential parameter analysis and CD4 counting.
J. **Substantial Equivalence Information:**
1. **Predicate device name(s):**
a. Coulter® Gen-S Hematology Analyzer
b. Coulter® LH 750 Hematology Analyzer
c. Coulter® tetraONE™ System for EPICS XL Flow Cytometry Systems
2. **Predicate K number(s):**
a. K962988
b. K032342 & K011342
c. K990172
{2}
Page 3 of 11
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Sample | Same as predicate | Coulter LH 750 & Gen-S: Anticoagulated whole blood |
| Intended Use | Determination of WBC count, Lymphocyte count, Lymphocyte percentage, CD4 T-Lymphocyte count and CD4 percentage | Coulter LH 750 & Gen-S: Determination of WBC count, Lymphocyte count, Lymphocyte percentage plus other hematologic parameter |
| Principle of Operation | Light scatter | Coulter tetraONE: Determination of CD4 percentage plus other lymphocyte populations |
| | | Coulter tetraONE: Light Scatter |
| Differences | | |
| Item | Device | Predicate |
| CD4 Reagent | Anti-CD4 monoclonal antibody conjugated to colloidal gold substrate | Coulter tetraONE: Anti-CD4 monoclonal antibody labeled with fluorochrome |
| Instrumentation | Compact flow-based immune hematology analyzer with computer and software | Coulter LH 750 & Gen-S: Standard hematology analyzer with computer and software |
| Principle of Operation | Light scatter | Coulter tetraONE: Standard flow cytometer with computer and software |
| | | Coulter LH 750 & Gen-S: Impedance |
K. Standard/Guidance Document Referenced (if applicable):
Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA (December 4, 2001)
L. Test Principle:
The FlowCare™ System reports CD4 T-Lymphocyte counts in conjunction with White Blood Cell (WBC) count and total Lymphocyte count information, on the basis
{3}
Page 4 of 11
of software analysis of established light scatter measurements with the use of proprietary non-fluorescent reagents.
All assay steps are performed on whole blood with capped bar-coded reagent tubes designed for use on the FlowCare™ System. All assay aspiration, dispensing and mixing steps are automated. The whole blood specimen is introduced into the CD4 Reagent Tube, where an anti-CD4 monoclonal antibody, conjugated to a colloidal gold substrate, recognizes and binds to the surface of leukocytes with CD4 receptors, such as CD4 positive T-Lymphocytes and monocytes.
Following a binding step incubation, the reaction mixture is subjected to selective lysing of erythrocytes. The lysing preserves the white blood cells in their near-native state, for subsequent flow-based leukocyte differential analysis. The instrument software analyzes, under flow, different leukocyte cell populations based on their light scatter properties. Automated cell population cluster analysis is performed by the software and results provided with no operator interpretation.
WBC populations are differentiated by the software on the basis of their forward scatter and time-of-flight signatures. The software positively gates on the lymphocyte cell cluster, excluding in the process, monocytes. CD4 positive T-lymphocytes are distinguished from CD4 negative T-lymphocytes, based on their greater degree of scatter on extinction versus right angle scatter, due to the bound CD4-colloidald gold particle conjugate bound onto their cell surfaces.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
A precision study was performed to assess the within-run precision and total precision of the FlowCare™ System by replicate measurements of control materials. Commercially available hematology (STRECK STaK-Chex®, tri-level – low, normal high) controls were used for WBC count, Lymphocyte count and Lymphocyte percentage of WBC count. CD4 quality controls (STRECK CD- Chex® Plus, bi-level – low, normal) quality controls were analyzed on two instruments, in triplicate runs, three times per day over three days. A high level is not available with the STRECK CD-CHEX controls. The testing was performed at PointCare Technologies, Inc., in Ashland, MA
The statistical measures of precision include the grand means along with standard deviations and coefficients of variation for both within-run precision and total precision.
{4}
Page 5 of 11
The % CV for within-run precision and total precision demonstrate acceptable performance.
All % CVs for within-run and within-day precision demonstrated acceptable performance. Results of the precision studies met acceptance criteria.
In addition, a precision study was performed to assess the within-day precision of the FlowCare™ System by replicate measurements of whole blood samples at four concentration levels. The samples were analyzed on two instruments with three separate runs. In each run, three replicate measures were performed for each level of blood sample. Testing was performed at PointCare Technologies, Inc., in Ashland, MA.
All whole blood samples were analyzed for WBC count, Lymphocyte count, Lymphocyte percentage of WBC count, CD4 count and CD4 percentage of Lymphocyte count.
The results demonstrated acceptable performance for within-day precision.
b. Linearity/assay reportable range:
A linearity study was performed to assess the performance of the FlowCare™ System over a wide range of cell concentrations for the measured parameters of WBC count, Lymphocyte count and CD4 count. Additionally, CD4 count was further evaluated in the low range. For both the full range and low range study, a minimum of six concentration levels were analyzed in quadruplicate on two instruments.
Samples for the full range study were prepared using a concentrated whole blood sample diluted with autologous platelet poor plasma to achieve the desired concentration levels. The 50% (normal range) sample was used to determine the expected values at the other concentration levels.
For the CD4 count low range stuffy, a whole blood sample was obtained from a donor known to have a CD4 count of approximately 400 cells/μL. The low range concentration levels were prepared by diluting the whole blood sample with autologous platelet poor plasma. The undiluted whole blood sample was used to determine the expected values at the other concentration levels.
{5}
Page 6 of 11
The sample for each concentration was analyzed four times in random sequence on each of the two FlowCare instruments. Testing was performed at PointCare Technologies, Inc., in Ashland, MA.
The results for each parameter were evaluated for linearity by the method of Snedecor and Cochran (Statistical Methods, Iowa State University Press, Sixth Edition, 1967, Chapter 15, section 15.4). The method involves fitting a linear regression, a quadratic regression and a simple one-way analysis of variance to the replicate results of the assayed parameter, at each of the expected concentration levels. The test for departure from linearity is an F-test which compares the mean square error from the quadratic regression against the pooled mean square error. Large values of the F-test lead to rejection of a linear fit. Acceptable linearity was demonstrated in all cases.
Least squares regressions analysis was also performed for each parameter. In all cases, linearity was demonstrated, and the expected and actual FlowCare values were shown to be highly correlated.
Full Range Linearity:
| Regression Statistics: | S/N 1816 | | | S/N 2185 | | |
| --- | --- | --- | --- | --- | --- | --- |
| N | WBC# | Lymph# | CD4# | WBC# | Lymph# | CD4# |
| | 21 | 21 | 21 | 22 | 19 | 20 |
| Correlation | 0.998 | 0.967 | 0.981 | 0.984 | 0.994 | 0.992 |
| Slope | 1.018 | 0.960 | 1.002 | 0.957 | 0.969 | 0.950 |
| Intercept | -131.732 | -29.898 | -85.463 | 204.348 | -9.586 | -7.461 |
Low Range Linearity:
| Regressions Statistics: | CD4 Count | |
| --- | --- | --- |
| | S/N 1816 | S/N 1858 |
| N | 27 | 27 |
| Correlation | 0.994 | 0.994 |
| Slope | 1.042 | 0.975 |
| Intercept | -12.932 | 13.295 |
{6}
Page 7 of 11
FlowCare System Reportable Ranges:
| Parameter | Reportable Range | Units |
| --- | --- | --- |
| WBC Count | 1.0 - 23.0 | 10³/μL |
| Lymphocyte% | 10 - 75 | % |
| Lymphocyte Count | 0.3 - 6.0 | 10³/μL |
| CD4 Count | 50 - 3000 | μL |
| CD4% | 0 – 80 | % |
c. Traceability (controls, calibrators, or method):
The CD4 monoclonal antibody used in the FlowCare CD4 Reagent Kit is derived from clone number RPA-T4. The monoclonal antibody recognized the human CD4 cell surface antigen, a 55 kD glycoprotein expressed by the helper-inducer subset of T-Lymphocytes and more weakly by monocytes.
The antigen specificity of the CD4 monoclonal antibody has been previously established by the Fourth and Fifth International Workshops for Leukocyte Typing.
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
A multi-site prospective study was conducted to evaluate the performance of the FlowCare System and reagents for the determination of White Blood Cell count, Lymphocyte count, Lymphocyte percentage of White Blood Cells, CD4 T-Lymphocyte count, and CD4 percentage of total Lymphocytes. Results from the FlowCare method were compared to those from currently available, FDA-cleared reference methods.
The study was conducted across four investigational sites in the U.S. and Sub-Saharan Africa using normal and abnormal whole blood samples tested in parallel.
{7}
Page 8 of 11
Reference methods for WBC counts, Lymphocyte counts and Lymphocyte %: Coulter® Gen-S, Coulter® LH750, Sysmex® E2500
Reference methods for CD4%: Coulter® EPICS XL tetraONE™ SYSTEM, and tetraCHROME™ 4-color reagents, BDIS FacsCalibur™ and MultiTest™ 4-color reagents
Reference method for CD4 count: Dual Platform (Reference Lymphocyte counts x Reference CD4%)
## Study Procedure
For each investigational site, blood specimens were analyzed by the reference methods in accordance with the procedures recommended by the respective instrument manufacturers and those further defined by the standard operating procedures and quality control programs of the site. Analyses by the FlowCare method were performed following training provided by PointCare Technologies, Inc. and in accordance with procedures described in the preliminary instrument operator manual and reagent kit package insert.
Samples analyzed in the clinical study were obtained from the routine adult populations available in each investigational site. Blood samples were collected into EDTA whole blood specimen collection tubes. Prior to analysis, the samples were kept at ambient room temperature for analysis within 10 hours of the subject blood draw.
The whole blood samples tested at the two sites in the United States were obtained primarily from a healthy donor population. In contrast, the samples tested at the two sites in Africa were from a population predominantly composed of immunocompromised individuals including those known or suspected of being infected with HIV or with other clinical conditions that may be associated with altered CD4 T-Lymphocyte counts and other hematological parameters.
Analyses of the pooled data for all sites show comparable means and ranges for the FlowCare and reference method parameter. (See table below)
{8}
Page 9 of 11
| Parameter | N | Corr. Coeff. | Slope | Intercept | Mean | | Range | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | | FlowCare | Reference | FlowCare | Reference |
| All Sites | | | | | | | | |
| WBC# | 403 | 0.9623 | 1.116 | -335.13 | 6194.5 | 5851.9 | 1900-19700 | 1820-18100 |
| LYM# | 403 | 0.9145 | 1.081 | -44.62 | 1956.8 | 1850.1 | 400-6500 | 500-5320 |
| LYM% | 403 | 0.9487 | 0.955 | 1.405 | 33.1 | 33.2 | 4.5-80.2 | 6.2-78.8 |
| CD4# | 425 | 0.9121 | 1.034 | 9.289 | 678.5 | 647.1 | 0-2318 | 7-2090 |
| CD4% | 401 | 0.9507 | 0.950 | 1.406 | 35.9 | 36.3 | 0.4-73.9 | 1.0-68.0 |
b. Matrix comparison: Not applicable
3. Clinical studies:
a. Clinical sensitivity: Not applicable
b. Clinical specificity: The antigen specificity of the CD4 monoclonal antibody has been previously established by the Fourth and Fifth International Workshops for Leukocyte Typing.
c. Other clinical supportive data (when a and b are not applicable):
4. Clinical cut-off: Not applicable
5. Expected values/Reference range: Whole blood specimens were collected form apparently healthy males and females in the Northeastern United States, without selection on the basis of age or race. Expected results for FlowCare parameter are presented based on a 95% normal distribution and compare closely with expected results observed with reference methods.
| FlowCare Parameter | N | Range | | Mean ± SD |
| --- | --- | --- | --- | --- |
| | | Minimum | Maximum | |
| WBC Count | 206 | 4400 | 12100 | 6882.5 ± 1618.76 |
| Lymph Count | 206 | 1100 | 3300 | 1963.6 ± 525.62 |
| Lymph % | 206 | 16.4 | 45.3 | 28.8 ± 6.21 |
| CD4 Count | 207 | 468 | 1702 | 928.5 ± 284.53 |
| CD4% | 205 | 31.3 | 65.8 | 47.7 ± 7.46 |
{9}
Page 10 of 11
N. Instrument Name:
FlowCare™ System
O. System Descriptions:
1. Modes of Operation:
Fully-automated cap piercing, closed tube system with position recognition
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐
3. Sample Identification:
Barcode reading and tube position verification
4. Specimen Sampling and Handling:
Fully-automated closed tube piercing, automated mixing and control reaction temperature
5. Calibration:
The system is factory calibrated for Total WBC count, Lymphocyte count and CD4 count.
6. Quality Control:
Commercial Hematology quality control (QC) material should be analyzed for WBC count and Lymphocyte count parameter, and a commercial CD4 QC material for assessing CD4 Reagent Kit performance.
Whole blood specimens may be used instead of commercial controls, provided there are corresponding values available from other instruments for reference purposes.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The “L. Performance Characteristics” Section Of The SE Determination Decision Summary:
A carryover study was performed to assess the effect of a whole blood sample on background counts in subsequent analyses for the measured parameters. Negligible carryover was observed.
Q. Proposed Labeling:
The labeling is sufficient and satisfies the requirements of 21 CFR section 809.10.
{10}
Page 11 of 11
**R. Conclusion:**
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.