The MirroScope™ System is indicated for use during dental examinations and procedures when a video image is desired.
Device Story
MirroScope™ is an intraoral camera system consisting of a handpiece with a mirror, prisms, lenses, and a 1/4" CCD camera. Six LEDs provide illumination. A disposable protective sheath is used for hygiene. The system captures video signals from the handpiece, which are transmitted via cable to a control box containing a Camera Control Unit (CCU) and Video Processing Unit (VPU). The system is available in two configurations: PRO-200™ (external display) and VIEW-300™ (includes 12.1" TFT screen). Operated by dental professionals in a clinical setting, the device provides real-time video visualization of the oral cavity to assist in dental examinations and procedures. The output is displayed on a monitor, allowing the clinician to view areas of the mouth that are difficult to see directly, thereby facilitating diagnosis and treatment.
Clinical Evidence
No clinical trial data provided. Evidence consists of bench testing, including electrical safety, electromagnetic compatibility, software verification and validation, and performance testing. Clinical images were evaluated to confirm system performance.
Technological Characteristics
Intraoral camera with 1/4" CCD sensor, LED illumination, and optical prism/lens assembly. System includes CCU, VPU, and optional TFT display. Complies with IEC 60601-1, IEC/EN 60601-1-2, IEC 60601-1-4, EN 1441, and ISO 14971. Connectivity via cable; optional RF transmitter. Disposable protective sheaths used for infection control.
Indications for Use
Indicated for use during dental examinations and procedures requiring video imaging. Intended for use by dental professionals.
Regulatory Classification
Identification
A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.
Predicate Devices
AcuCam Concept IV Intraoral Camera System (K000112)
{0}------------------------------------------------
## SEP 2 1 2004 510(K) SUMMARY
510(k) Number K_______________________________________________________________________________________________________________________________________________________________
## Applicant's Name:
Miras Mirror Imaging Solutions Ltd. 31 Habarzel St. Tel Aviv, 69710 Israel Tel.: 972-3-6442595 Fax: 972-3-6442597
## Contact Person:
Dorit Winitz Ph.D Biomedical Strategy Ltd. 11 Menachem Begin Street. Ramat Gan 52521, Israel Tel: 972-3- 6123281 Fax: 972-3-6123282
## Date Prepared:
June 2004
#### Trade Name:
MirroScope™ System
## Classification Name:
UNIT, OPERATIVE DENTAL Product Code: Regulation No: Class: I Panel: Dental
#### Classification:
Product Code: EIA Regulation No: 872.6640 Class: I Panel: Dental
{1}------------------------------------------------
### Predicate Device:
- The AcuCam Concept IV Intraoral Camera System (Dentsply 물 International, Inc.), cleared under K000112
- The Cdr-Cam System (Schick Technologies, Inc.), cleared 비 under K963778
- The FlexiScope Optomodul (Schoelly Fiberoptic, Gmbh), cleared under K984641
- Class I Exempt Dental Mirror (product code EAX) I
## Performance Standards:
No performance standards have been established for such device under Section 514 of the Federal Food, Drug, and Cosmetic Act. However, the MirroScope™ System complies with the following voluntary standards:
- IEC 60601-1(1988)+A1(1991)+A2(1995) .
- IEC/EN 60601-1-2 (2001) .
- CAN/CSA-C22.2 No. 601.1.2-94(R1999); CISPR 11(2003) . group 1 class B
- . IEC 60601-1-4 (1999)
- . EN 1441 (1997)
- ISO 14971 (2000) .
## Intended Use / Indication for Use:
The MirroScope™ System is indicated for use during dental examinations and procedures when a video image is desired.
## Device Description:
The MirroScope™ System is an intraoral camera that is mounted on a customary mirror and complimentary imaging accessories. The System consists of the following main components:
#### 1. The MirroScope™
The MirroScope™ is a handpiece comprised of a mirror (MirroLens), alignment of prisms and lenses, a miniature 1/4" CCD camera Remote Head Unit (RHU), located in the handle. The video signal from the RHU is transferred to the Camera Control Unit (CCU) located in a control box, via a lightweight cable. A set of six LED's illuminates the working area around the mirror. A disposable, single use protective sheath is provided to be mounted over the handpiece prior to each use, to ensure proper hygiene and to serve as a barrier cover.
{2}------------------------------------------------
2. Two complementary packages, the PRO-200™ and the VIEW-300™, to provide either an overall solution or an add-on solution to the existing infrastructure:
The PRO-200™ includes the following:
- Control Box: .
- o Camera Control Unit (CCU)
- o Video Processing Unit (VPU)
- Power Supply Unit (PSU) o
- RF Transmitter (Optional) o
- O Infrared Receiver (IRR)
- Docking station for MirroScope™ o
- R-400 Infrared remote control .
- External AC/DC adaptor .
- Foot pedal (Optional) .
- RF Receiver (Optional) .
The VIEW-300™ includes the following:
- Control box: .
- o Camera Control Unit (CCU)
- O Video Processing Unit (VPU)
- o Power Supply Unit (PSU)
- o 12.1" TFT screen and Converter
- O Infrared Receiver (IRR)
- o Docking station for MirroScope™
- R-400 Infrared remote control .
- External AC/DC adaptor .
- Accessories: .
- Foot pedal (Optional) .
## Substantial Equivalence:
The MirroScope™ System and its predicate devices share the same intended use and similar materials and design. A comprehensive testing program was developed and performed in order to verify that the MirroScop™ System does not raise any new safety and effectiveness issues in comparison to its predicate device. This includes the following testing and activities:
- Electrical safety and electromagnetic compatibility testing . according to IEC 60601-1 (and amendments), and IEC 60601-2 standards.
- Software verification and validation testing .
- Performance testing .
- Effective clinical images and use evaluation .
{3}------------------------------------------------
- Hazard analysis including risk level and solutions performed . in compliance with AAMI/ISO 14971-1, 2000 is attached
Altogether, Miras Mirror Imaging Solutions Ltd. believes that the MirroScope™ System is safe and effective for its intended use and is substantially equivalent to its predicate devices cited above without raising any new safety and/or effectiveness issue.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the agency's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
SEP 2 1 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Miras Mirror Imaging Solutions, Limited C/O Dr. Dorit Winitz Company Consultant Biomedical Strategy Limited 11 Menachem Begin Street Ramat Gan 52521, ISRAEL
Re: K041805
Trade/Device Name: The MirroScope™ System Regulation Number: 872.6640 Regulation Name: Dental Operative Unit and Accessories Regulatory Class: I Product Code: EIA Dated: June 20, 2004 Received: July 6, 2004
Dear Dr. Winitz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 wr ), irence of be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{5}------------------------------------------------
## Page 2 - Dr. Winitz
Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advisou that 1 DTC issuaires on that your device complies with other requirements Incall that I DA mas made a actessand regulations administered by other Federal agencies. of the Act of ally t oderal states the requirements, including, but not limited to: registration Tourinust comply with an the Hosever (21 CFR Part 801); good manufacturing practice and listing (21 CFR Part 807), labeling systems (QS) regulation (21 CFR Part 820); and if requirents as set form in the quality ijon control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) Fins letter will anow you to began finding of substantial equivalence of your device to a premarket notification. - The Fire results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specific at not to pliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may overn outer generalant and Consumer Assistance at its toll-free Dribber (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Siva Runno
Chiu-Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
K641805
Device Name:
The MirroScope™ System
Indications for Use:
The MirroScope™ System is indicated for use during dental examinations and procedures when a video image is desired.
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
510(k) Number_
Prescription Use _ 1 (Per 21 CFR 801.109)
Over the Use
Counter 1
Susan Knapp
(Division Sign-Off) (Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: K041605
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.