K041682 · Pie Medical Imaging BV · LLZ · Jul 27, 2004 · Radiology
Device Facts
Record ID
K041682
Device Name
CAAS MRV
Applicant
Pie Medical Imaging BV
Product Code
LLZ · Radiology
Decision Date
Jul 27, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The CAAS MRV has been designed for quantitative functional analysis of the left and right ventricle of the heart based on multi-slice, multi-phase cine MRI images. Specifically, it enables the user to: 1. Delineate automatically or semi-automatically the inner and outer wall of the ventricles, as well as the papillary muscles, on the MRI images 2. Derive from these contours the global and regional functional parameters like Ejection Fraction, Stroke Volume, Wall Movement etc. The CAAS MRV has been designed to be used in everyday clinical practice, as well as for research purposes like clinical research trials.
Device Story
CAAS MRV is a software tool for cardiac functional analysis; inputs multi-slice, multi-phase cine MRI datasets from various vendors via CD-ROM, hard disk, or PACS. Software enables automatic, semi-automatic, or manual delineation of inner/outer ventricular walls and papillary muscles. Transforms contours into quantitative outputs: ventricular volumes, ejection fraction, stroke volume, and regional wall motion. Used in clinical practice and research settings by clinicians. Results displayed on-screen or as hardcopy for diagnostic support and clinical decision-making; benefits include standardized, efficient quantification of cardiac function.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Software-based image processing system. Operates on multi-slice, multi-phase cine MRI datasets. Features automated, semi-automated, and manual contouring tools for ventricular wall and papillary muscle segmentation. Outputs include volumetric and motion parameters. Standalone software package.
Indications for Use
Indicated for quantitative functional analysis of left and right heart ventricles in patients undergoing multi-slice, multi-phase cine MRI. Used for clinical diagnosis and research trials.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
MRI-Magnetic resonance Analytical Software System (K994283)
Submission Summary (Full Text)
{0}------------------------------------------------
r
K041-682
This summary statement complies with 21CFR, section 807.92(c). Date summary prepared: 19 March 2004
This premarket notification has been submitted by Pie Medical Imaging BV and covers the CAAS MRV software package. Pie Medical Imaging is located at:
JUL 27 2004
| Pie Medical Imaging BV |
|-----------------------------|
| Becanusstraat 13 D 01 |
| 6216 BX Maastricht |
| The Netherlands |
| Phone +31.43.3281328 |
| Fax +31.42.3281329 |
| Email: carla.devries@pie.nl |
| The contact person is: | Ms. Carla de Vries, Quality Assurance Officer |
|-----------------------------|--------------------------------------------------|
| The trade name is: | CAAS MRV |
| The common name is: | Magnetic Resonance Ventricular analysis software |
| The classification name is: | Image Processing System (LLZ), CFR 892.2050. |
The above as stated in 21 CFR, part 892.1570, has been classified as regulatory Class II.
The CAAS MRV software package is substantially equivalent to the quantitative analysis software package MRI-Magnetic resonance Analytical Software System, K994283.
The CAAS MRV is a software tool designed for the functional analysis of the heart based on multi-slice, multi-phase MR images of the heart. Therefore it provides functionality to import and view cine MR datasets of the heart from several vendors from CDROM, hard disk or (optionally) a PACS system. Next, the inner and outer wall of the ventricles can be determined either automatically, semi-automatically or manually. From these contours the ventricular volume, the ejection fraction and other related parameters are determined. Next to the quantification of the ventricular volumes, also the motion of the ventricular wall is quantified. All results of the analysis are available on screen as well as hardcopy, and can be saved.
The intended use of CAAS MRV is to enable the user to:
1. Delineate the inner and outer wall of the ventricles automatically or semi-automatically, as well as the papillary muscles, on MRI images
2. Derive from these contours the global and regional parameters like Ejection Fraction, Stroke Volume, Wall Movement etc.
The CAAS MRV has been designed to be used in everyday clinical practice to support clinical diagnoses, as well as for research purposes like clinical research trials.
The CAAS MRV is substantially equivalent to the predicate device mentioned in this summary by using the same technological characteristics and intended use.
The CAAS MRV is produced under the same Quality Assurance system applicable to the development and production of products currently marketed by Pie Medical Imaging.
SoftSol Doc. 12556
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## JUL 2 7 2004
Ms. Carla de Vries Quality Assurance Officer Pie Medical Imaging bv Becanusstraat 13D 6216 BX Maastricht THE NETHERLANDS
Re: K041682 Trade/Device Name: CAAS MRV Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: June 14, 2004 Received: June 21, 2004
Dear Ms. de Vries:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmaldsmamain.html.
Sincerely yours,
Nancy C. Brygdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Notification - CAAS MRV
| INDICATION FOR USE STATEMENT | | page 1 of 1 |
|------------------------------|----------|-------------|
| 510(k) number (if known): | K041682 | |
| Device Name: | CAAS MRV | |
| Indications For Use: | | |
The CAAS MRV has been designed for quantitative functional analysis of the left and right ventricle of the heart based on multi-slice, multi-phase cine MRI images. Specifically, it enables the user to: 1. Delineate automatically or semi-automatically the inner and outer wall of the ventricles, as well as the papillary muscles, on the MRI images
2. Derive from these contours the global and regional functional parameters like Ejection Fraction, Stroke Volume, Wall Movement etc.
The CAAS MRV has been designed to be used in everyday clinical practice, as well as for research purposes like clinical research trials.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel A. Sypm
(Division Sign-Off) Division of Reproductive, Abde and Radiological Devic 510(k) Number -
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use
12552
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.