K041675 · Pie Medical · IYN · Jul 23, 2004 · Radiology
Device Facts
Record ID
K041675
Device Name
PICUS, PARUS, FALCO AND AQUILA ULTRASOUND SYSTEMS
Applicant
Pie Medical
Product Code
IYN · Radiology
Decision Date
Jul 23, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Pie Medical's Picus, Parus, Falco & Aquila Ultrasound Imaging Systems are to be used by or under the direction of a physician to perform general noninvasive and invasive diagnostic high resolution ultrasound imaging studies, to include: fetal, cardiac, transrectal, transvaginal, abdominal, neonatal cephalic, pediatric, peripheral vascular, small organs, intraoperative abdominal, intraoperative vascular, musculoskeletal (conventional and superficial) and wounds.
Device Story
Wound imaging method utilizes high-resolution ultrasound to evaluate wound extent. Procedure involves covering wound with sterile coupling gel and commercially available ultrasound couplant sheet (e.g., Civco HydroScan); ultrasound probe covered with sterile sheath (e.g., Civco). Linear or convex array transducer scanned over wound to acquire images. Clinician uses online measurement package to determine linear parameters, circumference, or depth for volume calculation. Output allows medical specialists to evaluate treatment efficacy by monitoring changes in wound area. Used in clinical settings by physicians or under their direction. Avoids adhesive edges on couplant sheets to prevent damage to sensitive periwound skin.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is based on technological similarities to predicate devices and the established use of ultrasound for wound imaging as demonstrated by the predicate Hudson system.
Technological Characteristics
Ultrasound imaging systems (Picus, Parus, Falco, Aquila) using linear/convex array transducers. Imaging modes include 2D, M-mode, and 3D (Picus). Electrical safety per EN60601-1; ultrasound safety per Track 1/3. Connectivity and software details not specified beyond standard measurement packages.
Indications for Use
Indicated for diagnostic ultrasound imaging of fetal, cardiac, transrectal, transvaginal, abdominal, neonatal cephalic, pediatric, peripheral vascular, small organs (thyroid, breast, testicles), intraoperative (abdominal/vascular), musculoskeletal (conventional/superficial), and wound applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K041675
510(k) Summary Picus. Parus. Falco & Aquila Ultrasound stems Pie Medical
# 510(k) Summary
The following safety and effectiveness summary has been prepared pursuant to requirement for 510(k) summaries specified in 21CFR¶807.92(a).
### 807.92(a)(1)
## Submitter Information
| Carri Graham, Official Correspondent | |
|--------------------------------------|-------------------------------|
| 7992 Castleway Drive | |
| Indianapolis, IN 46250 | |
| Phone: | (317) 849-1916, extension 103 |
| Facsimile: | (317) 577-9070 |
Contact Person: Carri Graham
Date: June 18, 2004
807.92(a)(2)
| Trade Name: | Ultrasound Imaging Systems | |
|---------------------------|---------------------------------------------------------------------------|----------------------|
| Common Name: | Ultrasound Imaging System | |
| Classification Name(s): | Ultrasonic pulsed echo imaging system<br>Diagnostic Ultrasonic Transducer | 892.1560<br>892.1570 |
| Classification Number: | 90IYO/90ITX | |
| 807.92(a)(3) | Predicate Device(s) | |
| Pie Medical | Picus | K023512 / K00288 |
| Pie Medical | Parus | K003725 |
| Pie Medical | 100LC/100S/485 | K002357 |
| Hudson Diagnostic Imaging | Hudson 2020/2040/2060<br>Ultrasound Scanners | K022928 |
Additional Substantial Equivalence Information is provided in the following substantial Equivalence Comparison Table.
{1}------------------------------------------------
510(k) Summarv Picus, Parus. Falco & Aquila Ultrasound Imaging Systems Pie Medical
### 807.92(a)(4)
### Device Description
Wound imaging is a method for evaluating the extent of wounds, using high resolution ultrasound for image acquisition and applying the online measurement package of the ultrasound system to measure either the wounds' linear parameters, or the circumference and depth for volume determination. This information is used for the medical specialist to evaluate the efficacy of a particular treatment in the reduction of the wound area.
The proposed wound imaging method uses a commercially available ultrasound couplant sheet, such as the Civco's HydroScan Couplant Sheet (Class 1), to cover a wound filled with a sterile coupling gel, such as Curasol (K953450). A linear or convex array ultrasound transducer is scanned over the covered wound to image the extent of the wound. Note that it is important that the cover not be sealed with adhesive edges area, as there is a potential to cause damage to the sensitive skin area adjacent to the wound. We believe that this is a significant advantage over previously described methods which use an adhesive material on the edge of the couplant sheet, and risk damaging the sensitive skin area proximal to the wound. To prevent contamination of the wound during the scan the ultrasound probe is covered with a commercially available sterile sheath, such as those available from Civco (K970515).
The clinician can choose to perform area length or circumference measurements to evaluate the dimensions of the wound, the circumference yielding a better determination of the wounds extent in situations where the wound has an irregular geometry.
#### Intended Use(s) 807.92(a)(5)
Pie Medical's Picus, Parus, Falco & Aquila Ultrasound Imaging Systems are to be used by or under the direction of a physician to perform general noninvasive and invasive diagnostic high resolution ultrasound imaging studies, to include: fetal, cardiac, transrectal, transvaginal, abdominal, neonatal cephalic, pediatric, peripheral vascular, small organs, intraoperative abdominal, intraoperative vascular, musculoskeletal (conventional and superficial) and wounds.
{2}------------------------------------------------
| General<br>characteristics | ' Pie Medical<br>Picus, Parus, Falco &<br>Aquila Systems<br>This Submission | Pie Medical<br>Picus, Parus, Falco &<br>Aquila Systems<br>K002880/023513/<br>003725 | Hudson Diagnostic<br>Imaging<br>Hudson<br>2020/2040/2060<br>Systems<br>K022928 |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Intended Use | Fetal, Abdominal, Pediatric,<br>Small Organ, Intraoperative<br>abdominal, Peripheral<br>vascular, Cardiac,<br>Transrectal, Transvaginal,<br>Musculoskeletal-<br>conventional & superficial,<br>Abdominal, Neonatal<br>Cephalic & Wounds | Fetal, Abdominal,<br>Pediatric, Small Organ,<br>Intraoperative<br>abdominal, Peripheral<br>vascular, Cardiac,<br>Transrectal,<br>Transvaginal,<br>Musculoskeletal-<br>conventional &<br>superficial, Abdominal,<br>Neonatal Cephalic &<br>Wounds | Wounds |
| Transducer Type | Linear | 8.0 Mhz 30C/40mm<br>(#410647)<br>7.5 Mhz 40mm<br>(#410503) | 8.0 Mhz 40mm<br>(#402198)<br>6-8 Mhz 60mm<br>(#410054)<br>5.0 Mhz 40mm<br>(#410503) |
| | | 8.0 Mhz 30C/40mm<br>(#410647)<br>7.5 Mhz 40mm<br>(#410503) | |
| General<br>characteristics | ¹ Pie Medical<br>Picus, Parus, Falco &<br>Aquila Systems<br>This Submission | Pie Medical<br>Picus, Parus, Falco &<br>Aquila Systems<br>K002880/023513/<br>003725 | Hudson Diagnostic<br>Imaging<br>Hudson<br>2020/2040/2060<br>Systems<br>K022928 |
| Imaging Modes | 2D / M-mode | 2D / M-mode<br>3D (Picus) | 2D / M-mode |
| Safety | | | |
| Electrical Safety | EN60601-1 | EN60601-1 | EN60601-1 |
| Ultrasound Safety | Track 1/3 | Track 1/3 | Track 1/3 |
{3}------------------------------------------------
# 8007.92(a)(6)
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a circular seal with an emblem in the center. The emblem consists of three parallel, curved lines that resemble a stylized bird or wing design. Encircling the emblem is text that reads "U.S. DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged along the perimeter of the circle, with the emblem positioned in the center.
Public Health Service
JUL 2 3 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Pie Medical % Ms. Carrie Graham Consultant The Anson Group 7992 Castleway Drive INDIANAPOLIS IN 46250
Re: K041675
Trade Name: Picus, Parus, Falco, and Aquila Ultrasound Imaging System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Name: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN, IYO, and ITX Dated: June 18, 2004 Received: June 21, 2004
Dear Ms. Graham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Picus, Parus, Falco, and Aquila Ultrasound Imaging System, as described in your premarket notification:
### Transducer Model Number
### 7.5 MHz L40, Picus 8MHz LA DF 30C/40mm, Parus, Falco, & Aquila
{5}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device it may or secreer to basis of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please 11 you the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
David A. Lyman
Image /page/5/Picture/8 description: The image contains a handwritten word, "for", in cursive script. The word is written in black ink on a white background. The letter "f" has a long, curved descender, and the "o" and "r" are connected. The writing appears to be somewhat stylized.
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{6}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form - Appendix F
#### Parus System
| | | Mode of Operation | | | | | | | | |
|----------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | | | | | | P [2] | |
| Abdominal | | P | P | | | | | | P [2] | |
| Intraoperative (Abdominal) | | P | P | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | P | P | | | | | | P [2] | |
| Small Organ (specify) [1] | | P | P | | | | | | P [2] | |
| Neonatal Cephalic | | P | P | | | | | | P [2] | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | P | P | | | | | | P [2] | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | P | P | | | | | | P [2] | |
| Transvaginal | | P | P | | | | | | P [2] | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | P | P | | | | | | P (2) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | | | | | | P [2] | |
| Musculo-skeletal Superficial | | P | P | | | | | | P [2] | |
| Other - Wounds | | N | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N=new indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
[1] Small organs include Thyroid, Breast and Testicles.
[2] Applicable combined modes: B+B; B+M
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel A. Leyman
(Division Sign-Off) Division of Reproductive, and Radiological Devic 510(k) Number
{7}------------------------------------------------
Falco System
| Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows. | | |
|--------------------------------------------------------------------------------------------------|--|--|
|--------------------------------------------------------------------------------------------------|--|--|
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | | | | | | P [2] | |
| Abdominal | | P | P | | | | | | | |
| Intraoperative (Abdominal) | | P | P | | | | | | P [2] | - |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | P | P | | | | | | P [2] | |
| Small Organ (specify) [1] | | P | P | | | | | | P [2] | |
| Neonatal Cephalic | | P | P | | | | | | P [2] | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | P | P | | | | | | P[2] | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | P | P | | | | | | | |
| Transvaginal | | P | P | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | P | P | | | | | | P [2] | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | | | | | | P [2] | |
| Musculo-skeletal Superficial | | P | P | | | | | | P [2] | |
| Other - Wounds | | N | | | | | | | | |
N=new indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
[1] Small organs include Thyroid, Breast and Testicles.
[2[ Applicable combined modes: B+B; B+M
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH. Office of Device Evaluation (ODE)
Daniel C. Boyann
(Division Sign-Off) Division of Reproductive, I and Radiological Devices 510(k) Number _
{8}------------------------------------------------
Aquila System
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | | | | | | P [2] | |
| Abdominal | | P | P | | | | | | P [2] | - |
| Intraoperative (Abdominal) | | P | P | | | | | | P [2] | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | P | P | | | | | | P [2] | |
| Small Organ (specify) [1] | | P | P | | | | | | P [2] | |
| Neonatal Cephalic | | P | P | | | | | | P [2] | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | P | P | | | | | | P [2] | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | P | P | | | | | | P [2] | |
| Transvaginal | | P | P | | | | | | P [2] | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | P | P | | | | | | P [2] | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | | | | | | P [2] | |
| Musculo-skeletal Superficial | | P | P | | | | | | P [2] | |
| Other - Wounds | | N | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N=new indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
[1] Small organs include Thyroid, Breast and Testicles.
[2] Applicable Combined Modes: B+B; B+M
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Damil A. Sypm
(Division Sign-Off) Division of Reproductive, Al and Radiological Devices 1 ( ( ) Number __
{9}------------------------------------------------
Picus System
| | Mode of Operation | | | | | | | | | | |
|----------------------------------|-------------------|---|---|------------|-----|--------------------------|------------------------------|------------------------------|-----------------------|--------------------------|--|
| Clinical Application | A | B | M | PWD<br>(D) | CWD | Color<br>Doppler<br>(CD) | Amplitude<br>Doppler<br>(AD) | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify)<br>3D | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | P[2] | P | |
| Abdominal | | P | P | P | | P | P | | P[2] | P | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) [1] | | P | P | P | | P | P | | P[2] | P | |
| Neonatal Cephalic | | P | P | P | | P | P | | P[2] | P | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | P | P | P | | P | P | | P[2] | P | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | P | P | P | | P | P | | P[2] | P | |
| Transvaginal | | P | P | P | | P | P | | P[2] | P | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | P | P | P | | P | P | | P[2] | P | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | P[2] | | |
| Musculo-skeletal Superficial | | P | P | P | | P | P | | P[2] | | |
| Other - Wounds | | N | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N=new indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
[1] Small organs include Thyroid, Breast and Testicles.
[2] Applicable combined modes: B+D; B+CD+D; B+AD+D
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Severson
(Division Sign-Division of Reprodu and Radiological De 510(k) Numbe
{10}------------------------------------------------
Probe #410503 - Picus 7.5 MHz L40
| Clinical Application | A | B | M | PWD<br>(D) | CWD | Color<br>Doppler<br>(CD) | Amplitude<br>Doppler<br>(AD) | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify)<br>3D |
|----------------------------------|---|---|---|------------|-----|--------------------------|------------------------------|------------------------------|-----------------------|--------------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | - | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) [1] | | P | P | P | | P | P | | P [2] | P |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | P | P | P | | P | P | | P [2] | P |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | P[2] | |
| Musculo-skeletal Superficial | | P | P | P | | P | P | | P[2] | |
| Other - Wounds | | N | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
N=new indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
[1] Small organs include Thyroid, Breast and Testicles.
[2] Applicable combined modes: B+D; B+CD+D; B+AD+D
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Otfice of Device Evaluation (ODE)
Daniel G. Ingram
(Division Sign-Off Division of and Radiolog 510(k) Number
{11}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form - Appendix F
Probe# 410647 - Parus, Falco & Aquila 8MHz LA DF 30C/40mm
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify)<br>3D |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | P [2] | |
| Intraoperative (Abdominal) | P | P | P | | | | | | P [2] | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | P | P | | | | | | P [2] | |
| Small Organ (specify) [1] | | P | P | | | | | | P [2] | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | P | P | | | | | | P [2] | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | | | | | | P [2] | |
| Musculo-skeletal Superficial | | P | P | | | | | | P [2] | |
| Other - Wounds | N | N | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N=new indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
[1] Small organs include Thyroid, Breast and Testicles.
[2] Applicable combined modes: B+B; B+M
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel A. Seymour
(Division Sign-O nroductive and Radi 510(k) Num
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.