ALLOMATRIX TCP, ALLOMATRIX W/CASO4 FILLED TCP, AND ALLOMATRIX PLUS CBM
Applicant
Wrightmedicaltechnologyinc
Product Code
MQV · Orthopedic
Decision Date
Sep 16, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ALLOMATRIX® TCP, ALLOMATRIX® w/ CaSO4 Filled TCP, and ALLOMATRIX® PLUS CBM are indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. ALLOMATRIX® TCP, ALLOMATRIX® w/ CaSO4 Filled TCP and ALLOMATRIX® PLUS CBM are intended to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.
Device Story
ALLOMATRIX® products (TCP, CaSO4 filled TCP, and PLUS CBM) are bone void fillers supplied as kits containing premeasured powder, TCP/CBM chips, and mixing solution. Clinicians mix components to create a putty, which is then manually packed into non-structural bony voids or gaps in the extremities or pelvis. The device acts as a scaffold for bone repair. It is intended for single-patient use in a surgical setting. The product incorporates demineralized bone matrix (DBM) processed for viral inactivation and assayed for osteoinductive potential.
Clinical Evidence
No human clinical data provided. Performance evaluated via bench testing and a canine model using radiographic and histological methods. Viral inactivation validation performed using a panel of model human viruses.
Technological Characteristics
Bone void filler kit containing TCP, CaSO4, and DBM/CBM components. Supplied sterile for single-patient use. Materials include TCP, CaSO4, and DBM. No electronic or software components.
Indications for Use
Indicated for patients with surgically created or traumatic osseous defects in the skeletal system (extremities and pelvis) that are not intrinsic to the stability of the bony structure.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Submission Summary (Full Text)
{0}------------------------------------------------
K041663 1/2
# SEP 1 6 2004
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe in accordance with the Food and Drag Formance with 21 CRF 807, this information serves as a
Medical Devices Act of 1990 and in conformance with 21 CRF 807, this information o Medical Devices Act of 1970 and in ven the use of ALLOMATRIX® TCP, ALLOMATRIX® w/ CaSO4 Filled TCP, and ALLOMATRIX® PLUS CBM.
| Submitted By: | Wright Medical Technology, Inc. |
|-------------------------------------|----------------------------------------------------------------------------------|
| Date: | September 15, 2004 |
| Contact Person: | Brian Young<br>Director, Regulatory Affairs |
| Proprietary Name: | ALLOMATRIX® TCP, ALLOMATRIX® w/<br>CaSO4 Filled TCP, and ALLOMATRIX® PLUS<br>CBM |
| Common Name: | Bone Void Filler |
| Classification Name and Reference: | Filler, Calcium Sulfate Preformed Pellets - Class II<br>888.3045 |
| Device Product Code and Panel Code: | Orthopedics/MQV |
### DEVICE INFORMATION
#### INTENDED USE A.
INTENDED USE ADDOMIXTRIA - 15-15-14-11-11-11-11-11-11-11-11-11-11-11-11-11-11-11-2017-11-11 of the bony structure. ALLOMATRIX® TCP, ALLOMATRIX® w/ CaSO4 Filled TCP and ALLOMATRIX® PLUS CBM are intended to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.
#### DEVICE DESCRIPTION B.
DEVICE DESCREAT TOT!
ALLOMATRIX® TCP, ALLOMATRIX® w/ CaSO4 Filled TCP and ALLOMATRIX® PLUS CBM come in the form of a kit with a premeasured powder, TCP and/or CBM chips, premeasured mixing solution, and the tools necessary to mix the components. emps, premeasured himmy all the mixing solution supplied in the kit, the resultant putty can then be handled and placed in the appropriate bone voids. This product is supplied sterile for single patient use.
{1}------------------------------------------------
K041663
### C. SUBSTANTIAL EQUIVALENCE INFORMATION
ALLOMATRIX® TCP, ALLOMATRIX® w/ CaSO4 Filled TCP and ALLOMATRIX® PLUS CBM were found to be substantially equivalent to the predicate devices. The safety and effectiveness of ALLOMATRIX® TCP, ALLOMATRIX® w/ CaSO4 Filled TCP and ALLOMATRIX® PLUS CBM is adequately supported by the substantial equivalence information, materials data, and testing results provided within this Premarket Notification.
#### Osteoinductivity Potential
Each lot of DBM incorporated into ALLOMATRIX® TCP, ALLOMATRIX® with CaSO4 Filled TCP and ALLOMATRIX® Plus CBM (hereafter ALLOMATRIX® Putty Products) is assayed in vitro for a native protein as a surrogate test marker for osteoinductive potential. Results from this immunoassay were correlated to the athymic rat model for the DBM alone and the ALLOMATRIX Putty Products.2 Testing each lot of DBM with this immunoassay assures that only DBM with osteoinductivity potential is used in the ALLOMATRIX® Putty Products. Although only one native protein is used as the test marker, it is the combination of various proteins in the DBM that is responsible for its osteoinductivity potential. The combination of DBM, and binding medium has not been evaluated for osteoinductivity; therefore, it is unknown to what extent the formulation components may alter the osteoinductive character of the DBM. Additionally, it is unknown how osteoinductivity potential. measured by this surrogate immunoassay, will correlate with human clinical performance of the Allomatrix Putty Products.
- l Data on file at Wright Medical Technology, Inc.
- 2 Lindholm TS, Urist MR. A quantitative analysis of new bone formation by induction in compositive grafts of bone marrow and bone matrix, Clin Orthop 1980 Jul-Aug;(150):288-300.
#### Viral Inactivation Validation
The method for processing the DBM and CBM contained in ALLOMATRIX® TCP, ALLOMATRIX® w/ CaSO4 Filled TCP and ALLOMATRIX® PLUS CBM was evaluated for its viral inactivation potential. A panel of model potential human viruses representing various virus types, sizes, shapes, and genomes were evaluated. The viral inactivation testing demonstrated suitable viral inactivation potential of the processing method for a wide spectrum of potential human viruses.
#### Product Performance Testing
Performance of ALLOMATRIX® TCP, ALLOMATRIX® w/ CaSO4 Filled TCP and ALLOMATRIX® PLUS CBM were evaluated in a canine model by radiographic and histological methods.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is an emblem of an eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 6 2004
Mr. Brian J. Young Director, Regulatory Affairs Wright Medical Technology 5677 Airline Road Arlington, Tennessee 38002
Re: K041663
ALLOMATRIX® TCP, ALLOMATRIX® w/ CaSO4 Filled TCP, and ALLOMATRIX® PLUS CBM Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler devices Regulatory Class: Class II Product Code: MQV Dated: June 14, 2004 Received: June 18, 2004
Dear Mr. Young:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices macketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal aqencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing pregetive
{3}------------------------------------------------
Page 2 - Mr. Brian J. Young
requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE
510(K) Number (if known):_K04 (663
ALLOMATRIX® TCP, ALLOMATRIX® w/ CaSO4 Filled TCP and Device Name: ALLOMATRIX® PLUS CBM
Indications for Use:
ALLOMATRIX® TCP, ALLOMATRIX® w/ CaSO4 Filled TCP, and ALLOMATRIX® PLUS CBM are indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. ALLOMATRIX® TCP, ALLOMATRIX® w/ CaSO4 Filled TCP and ALLOMATRIX® PLUS CBM are intended to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.
Prescription Use X (Per21 CFR 801.109)
OR
Over-The Counter Use (Optional Format 1-2-96)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Muriam C. Provost
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
K041663 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.