← Product Code [GCJ](/productcode/GCJ) · K041641

# DAVOL DELIVERY SYSTEM (K041641)

_C.R. Bard, Inc. · GCJ · Jul 1, 2004 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K041641

## Device Facts

- **Applicant:** C.R. Bard, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jul 1, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The Davol Delivery System is intended to be used to facilitate the delivery of soft tissue prosthetics during the laparoscopic repair of soft tissue defects (e.g. hernia repair).

## Device Story

Davol Delivery System facilitates laparoscopic delivery of soft tissue prosthetics (e.g., hernia mesh). Device components: rolling tines/plunger assembly, hollow plastic delivery tube, 'T' handle, and optional switching stick. Operation: user rolls prosthetic using tines, inserts into delivery tube, and uses 'T' handle to plunge prosthetic into abdomen. Insertion end features flexible, three-leaf slit valve to facilitate entry into incision and allow prosthetic exit. Adjustable wing on delivery tube provides visual depth gauge and bracing point for surgeon. Used in clinical/surgical settings by surgeons. Benefits: enables uniform rolling and controlled delivery of prosthetics through laparoscopic access points.

## Clinical Evidence

Bench testing only. Testing involved rolling various large soft tissue prosthetics, inserting them into the delivery device, and plunging them onto a tabletop. Trials were repeated with saline-soaked prosthetics to simulate surgical conditions. Results demonstrated no damage to prosthetics or device, and confirmed that wetting the prosthetic facilitates delivery.

## Technological Characteristics

Hollow plastic delivery tube with rolling tines/plunger assembly and 'T' handle. Features a flexible, three-leaf slit valve at the insertion end and an adjustable external wing for depth gauging. Available in two sizes. Sterile, single-patient use. Mechanical operation; no energy source or software.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Davol Surgical Mesh Delivery System (Davol Inc.)
- MedChem Surgical Delivery System (Davol Inc.)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### SECTION 6.

## KO4 1641

## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION

### A. Submitter Information

Submitter's Name: Address:

Telephone: Fax: Contact Person: Date of Preparation: Davol, Inc. Subsidiary of C. R. Bard, Inc. 100 Sockanossett Crossroad Cranston, RI 02920 401-463-7000 ext. 2263 401-463-3845 Suzanne LaScalza May 6, 2004

Page 1 of (3

JUL 0 1 2004

#### Device Name B.

| Trade name:          | Davol Delivery System                  |
|----------------------|----------------------------------------|
| Common/Usual name:   | Soft Tissue Prosthetic Delivery System |
| Classification name: | Endoscope and Accessories              |

#### C. Predicate Device Names

Trade name: Davol Surgical Mesh Delivery System (Davol Inc.)

MedChem Surgical Delivery System (Davol Inc.) Trade name:

#### D. Device Description

The proposed device is designed to roll a soft tissue prosthetic in a tight uniform manner and provide a protective tube through which the rolled prosthetic can be delivered into the abdomen. The proposed device consists of four distinct parts, a rolling tines/plunger assembly which is used to roll the soft tissue prosthetic, a hollow plastic delivery tube to hold the rolled prosthetic, a "T" handle which is placed over the tines to facilitate the plunging motion to deliver the prosthetic, and an optional switching stick which can be used to maintain the tissue path created by a trocar sleeve. The proposed device will be provided as a sterile device for single patient use. It will be available in two sizes, a small size for the delivery of smaller soft tissue prosthetics and a large size for the delivery of larger prosthetics.

#### E. Intended Use

The Davol Delivery System is intended to be used to facilitate the delivery of soft tissue prosthetics during the laparoscopic repair of soft tissue defects (e.g. hernia repair).

{1}------------------------------------------------

### F. Technological Characteristics Summary

The Davol Delivery System and the predicate Davol device have the same intended use. Both devices are indicated to facilitate the delivery of soft tissue prosthetics, such as surgical mesh to the surgical site during laparoscopic soft tissue repair procedures (e.g. hernia repair).

The proposed device is similar to the two predicate devices in principle of operation and general design. The key differences are in the details of the device design.

In order to accommodate large soft tissue prosthetics, the large proposed device was designed with an inner diameter of 20 mm and an outer diameter of approximately 21 mm. The inner and outer diameters of the proposed device are similar to the large predicate MedChem device and fall within the size range of similar currently marketed devices.

The insertion end of the predicate devices could not be used in the proposed device since the open end of the predicate Davol device would be difficult to insert directly through an incision and even the slight taper of the predicate MedChem device would prevent the prosthetic from exiting the delivery tube. Instead, the insertion end of the proposed device is covered with a thin, flexible, rounded tip which is slit into three triangular leaves to form a valve. The rounded shape facilitates the entry of the delivery tube into the incision site while the slit valve allows the rounded tip to open wide enough to permit the prosthetic to exit. The insertion end of the proposed device does not affect the safety and effectiveness of the device since the rounded tip simply facilitates entry into the incision compared to an open insertion end.

The proposed device has an adjustable wing component on the delivery tube. which is not present in either of the two predicate devices. The wing component remains outside the body and provides a visual gauge of the insertion depth, as well as a place for the surgeon to brace against while pushing the soft tissue prosthetic through the delivery tube. This component does not adversely affect the safety and effectiveness of the device since the component remains outside the body and simply provides additional visual guide.

The switching stick of the proposed device is an optional component provided to maintain the pathway from the skin incision, through the tissue layers to the abdomen. Although neither predicate device includes a switching stick component, switching sticks, or exchange rods are sometimes packaged with currently marketed laparoscopic devices and commonly used to maintain the tissue pathway.

Preliminary bench testing was performed using several different large soft tissue prosthetics. Testing consisted of rolling the soft tissue prosthetic, inserting the

MAY 17 2004

Page 2 of '

{2}------------------------------------------------

# MAY 17 2004

Page 32

rolled prosthetic into the delivery device and plunging it out of the device onto a tabletop. Several trials were repeated after soaking the prosthetic in saline solution to simulate were ropens are sults demonstrated that neither the solution to simulate were conditions. "Thibited damage after use. Further, the trials prosmetics not the proposed des demonstrated that wetting the prosthetic facilitates ease of delivery.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling three overlapping lines or waves, often interpreted as representing the department's mission related to health and well-being.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 01 2004

C.R. Bard, Inc. c/o Mr. Robert Mosenkis President Citech 5200 Butler Pike Plymouth Meeting, Pennsylvania 19462-1298

Re: K041641

Trade/Device Name: Davol Delivery System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: June 16, 2004 Received: June 17, 2004

Dear Mr. Mosenkis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Muriam C. Provost

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## INDICATIONS FOR USE

510(k) Number (if known):

K041641

Device Name:

Davol Delivery System

Indications for Use:

The Davol Delivery System is intended to be used to facilitate the delivery of soft tissue prosthetics during the laparoscopic repair of soft tissue defects (e.g. hernia repair).

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mìriam C. Provost

(Division Sign-Off Division of General, Restorative, and Neurological Devices

**510(k) Number** K041641

---

**Source:** [https://fda-staging.innolitics.com/device/K041641](https://fda-staging.innolitics.com/device/K041641)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
