LIGASURE PRECISE MODEL #LS1200/LIGASURE 5MM LAPAROSCOPIC SEALER DIVIDER MODEL # LS1500
K041587 · Valleylab · GEI · Sep 28, 2004 · General, Plastic Surgery
Device Facts
Record ID
K041587
Device Name
LIGASURE PRECISE MODEL #LS1200/LIGASURE 5MM LAPAROSCOPIC SEALER DIVIDER MODEL # LS1500
Applicant
Valleylab
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 28, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LigaSure™ Open Dissector Divider is a bipolar electrosurgical instrument intended for use with the LigaSure Generator in open, general and gynecologic surgical procedures where ligation and division of vessels is desired. This device creates ligation by application of bipolar electrosurgical RF energy and pressure to vessels/tissue interposed between the jaws of the device. The ligation can then be transected using the built in cutting mechanism. The device can be used for vessels up to and including 7 mm in diameter and tissue bundles as large as will fit into the jaws of the instrument. Indications for use include general, open procedures including urologic, vascular, thoracic, and plastic and reconstructive, procedures where dissection, ligation and division of vessels are performed such as: spleenectomies, thryoidectomies, nephrectomies, prostatectomies, bowel resection and colectorny, The LigaSure Vessel Sealing System has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.
Device Story
Bipolar electrosurgical instrument for open surgery; used with LigaSure Vessel Sealing Generator. Device clamps tissue/vessels up to 7mm diameter between jaws; applies RF energy and pressure to create hemostatic seal; mechanical cutting mechanism divides tissue. Operated by surgeon via device button or footswitch. Connects to generator via 10-foot cable with 'smart' connector for automatic instrument identification. Single-use, sterile device. Provides surgeons with integrated dissection, sealing, and division capabilities; reduces need for multiple instruments; improves procedural efficiency.
Clinical Evidence
Bench testing and preclinical study performed to verify design specifications and functional performance. No clinical trial data provided.
Technological Characteristics
Bipolar electrosurgical instrument; 10-foot cable with 'smart' connector for generator identification; mechanical cutting mechanism; single-use; sterile. Operates via RF energy delivery.
Indications for Use
Indicated for patients undergoing open general, urologic, vascular, thoracic, and plastic/reconstructive surgical procedures requiring vessel/tissue ligation and division. Contraindicated for tubal sterilization or tubal coagulation procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
Valleylab LigaSure™ Open Dissector Divider
K041587
### Submitter Information 1.
Valleylab A division of Tyco Healthcare Group LP 5920 Longbow Drive Boulder, CO 80301 Contact: Charles M. Copperberg Manager, Regulatory Telephone: 303-530-6247
Page ① of 2
Date summary prepared: September 20, 2004
### Name of Device 2.
Trade or Proprietary Name: LigaSure™ Open Dissector Divider, Catalog Number LS4100
Common Name: Bipolar, Open Electrosurgical Instrument
Classification Name:
- Electrosurgical Cutting and Coagulation Device and Accessories, and .
- Gynecologic Electrocautery and Accessories .
### Predicate Devices 3.
The LigaSure™ Open Dissector Divider (catalog number LS4100) is substantially equivalent to the following legally marketed devices:
- For configuration, sealing and dissection capabilities the LS4100 is . substantially equivalent to the Valleylab LS1200 LigaSure™ Precise Instrument (K010010)
- For cutting capabilities, the LS4100 is substantially equivalent to the Valleylab . LS1500 LigaSure™ 5mm Laparoscopic Sealer-Divider Instrument (K031011)
## 4. Device Description
The LigaSure Open Dissector Divider is a multi-functional, electrosurgical instrument intended for use with the LigaSure Vessel Sealing Generator (K981916) when performing open surgery. The instrument is capable of sealing and dividing vessels and tissue clamped between its jaws, grasping tissue, and dissection. Sealing of vessels and tissue containing vessels can be activated
{1}------------------------------------------------
Page ② of 2
using the activation button on the device or via a footswitch. The instru incorporates a mechanical cutting mechanism that divides tissue.
The instrument attaches to the LigaSure™ generator with a ten (10) foot cable and a "smart" connector that identifies the instrument type to the generator. The instrument is supplied sterile for single-use.
#### 5. Intended Use
The LigaSure™ Open Dissector Divider is a bipolar electrosurgical instrument intended for use with the LigaSure Generator in open, general and gynecologic surgical procedures where ligation and division of vessels is desired. This device creates ligation by application of bipolar electrosurgical RF energy and pressure to vessels/tissue interposed between the jaws of the device. The ligation can then be transected using the built in cutting mechanism. The device can be used for vessels up to and including 7 mm in diameter and tissue bundles as large as will fit into the jaws of the instrument.
Indications for use include general, open procedures including urologic, vascular, thoracic, and plastic and reconstructive, procedures where dissection, ligation and division of vessels are performed such as: spleenectomies, thryoidectomies, nephrectomies, prostatectomies, bowel resection and colectorny,
The LigaSure Vessel Sealing System has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.
### 6. Summary of Technological Characteristics
The LigaSure™ Open Dissector Divider has the same basic technological characteristics as the predicate devices noted above.
### 7. Performance Data
Performance (bench) testing and a preclinical study were done to ensure that the LigaSure™ Open Dissector Divider functions as intended, and meets design specifications. Sufficient data were obtained to show that the device is substantially equivalent to the predicate devices, and meets safety and effectiveness criteria.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes extending from its head, enclosed within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the border of the circle.
Public Health Service
SEP 2 8 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Charles M. Copperberg Manager, Regulatory Valleylab 5920 Longbow Drive Boulder, Colorado 80301
Re: K041587
K041387
Trade/Device Name: LigaSure™ Open Dissector Divider Regulation Number: 21 CFR 878.4400 Regulation Name: 21 CFR 078. Proo
Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: September 14, 2004 Received: September 15, 2004
Dear Mr. Copperberg:
We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 510(x) premainer is substantially equivalent (for the indications
referenced above and have determined the device is substantial in interestate referenced above and have determined the arrest and devices marketed in interstate for use stated in the encrosure) to regally manated provice Amendments, or to commerce prior in May 28, 1970, the classified in accordance with the Federal Food, Drug, devices that have been icclassified in accerative who a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a previous actuals groups and Cosment Act (Act) that do not require subject to the general controls provisions of the Act. The
You may, therefore, market the device, subject to the general magatesti You may, merceiole, market the devices, boojes requirements for annual registration, listing of general controls provisions of the Her Herman somments of and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), in If your device is classified (soc aboro) ins. Existing major regulations affecting your device can may be subject to such additional controller is notions.
be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may be found in the Ood of Peachas oncerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not meant Please be advised that I DA 3 issuance or our device complies with other requirements of the Act that TDA has made a decormination administered by other Federal agencies. You must of any Federal statutes and regulations annualing, but not limited to: registration and listing (21 comply with an the Act 3 requirements, morealing practice requirements as set CFK Fart 807), fabeling (21 CFR Part 820); and 1f applicable, the electronic form in the quality by sections (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Charles M. Copperberg
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally premaintented predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Usc
510(k) Number (if known): K041587
Device Name: LigaSure™ Open Dissector Divider
Indications for Use:
The LigaSure™ Open Dissector Divider is a bipolar electrosurgical instrument intended for use with the LigaSure™ Generator in open, general, surgical procedures where dissection, ligation and division of vessels is desired. This device creates ligation by application of bipolar electrosurgical RF anerizy and pressure to vessels/tissue interposed between the jaws of the device. The ligation can then be transected using the built in cutting mechanism. The device can be used for vessels up to and including 7 mm in diameter and tissue bundles as large as will fit into the jaws of the instrument.
Indications for use include general, open procedures including urologic, vascular, thoracic, and plastic and reconstructive, procedures where dissection, ligation and division of vessels are performed such as: spleenectomies, thryoidectomies, nephrectomies, prostatectomies, bowel resection are clobe vomy.
The LigaSure Vessel Sealing System has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
Page | of | |
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number_Ko 41587
(Posted November 13, 2003)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.