← Product Code [LON](/productcode/LON) · K041572

# BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-PIPERCILLIN (GN) 0.5-128 UG/ML (K041572)

_Becton, Dickinson and Co. · LON · Aug 12, 2004 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K041572

## Device Facts

- **Applicant:** Becton, Dickinson and Co.
- **Product Code:** [LON](/productcode/LON.md)
- **Decision Date:** Aug 12, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1645
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic gram-negative and gram-positive bacteria of human origin.

## Device Story

The BD Phoenix System is an automated microbiology platform for rapid identification and antimicrobial susceptibility testing (AST) of bacterial isolates. The system utilizes sealed, self-inoculating polystyrene panels containing biochemicals and antimicrobial agents. Inputs consist of pure bacterial cultures inoculated into Phoenix ID or AST broth. The system performs broth microdilution testing by monitoring redox indicator changes and bacterial turbidity via continuous incubation at 35°C; readings are taken every 20 minutes. The system outputs MIC values and categorical interpretations (S, I, or R). Used in clinical microbiology laboratories by trained technicians; results assist clinicians in selecting appropriate antibiotic therapy for patients with bacterial infections.

## Clinical Evidence

Performance evaluated via method comparison against NCCLS broth dilution reference panels using 1,781 clinical and challenge isolates. Combined results showed 94.3% Essential Agreement (EA) and 93.8% Category Agreement (CA). Reproducibility demonstrated >95% across three sites. Quality control testing using NCCLS strains confirmed system accuracy. No clinical sensitivity/specificity studies performed as this is an in vitro diagnostic AST system.

## Technological Characteristics

Automated colorimetric oxidation-reduction growth-based system. Materials: molded polystyrene trays with 136 micro-wells. Energy: 35°C incubation. Connectivity: barcode scanning, automated software-driven analysis. Software: Expert system using NCCLS-derived rules. Dimensions: multi-well tray format. Sterilization: not specified.

## Regulatory Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

## Predicate Devices

- VITEK® System (PMA No. N50510)
- BD Phoenix™ Automated Microbiology System with Gatifloxacin ([K020321](/device/K020321.md))
- BD Phoenix™ Automated Microbiology System with Ofloxacin ([K020323](/device/K020323.md))
- BD Phoenix™ Automated Microbiology System with Levofloxacin ([K020322](/device/K020322.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Page 1 of 7

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE

A. 510(k) Number: k041572

B. Purpose for Submission:
To add piperacillin at 0.5 – 128 ug/mL to the Gram-Negative AST panels

C. Analyte:
Piperacillin at 0.5-128 ug/mL

D. Type of Test:
Antimicrobial Susceptibility Test (Quantitative) colorimetric oxidation-reduction, growth-based

E. Applicant:
Becton, Dickinson &amp; Company

F. Proprietary and Established Names:
BD Phoenix™ Automated Microbiology System – Piperacillin Gram Negative

G. Regulatory Information:

1. Regulation section:
21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle
Antimicrobial

2. Classification:
Class II

3. Product Code:
LON

4. Panel:
83 Microbiology

H. Intended Use:

1. Intended use(s):
BD Phoenix™ Automated Microbiology System:
The BD Phoenix™ Automated Microbiology System is intended for
in vitro quantitative determination of antimicrobial susceptibility by minimal
inhibitory concentration of gram-negative aerobic and facultative anaerobic
bacteria belonging to the family Enterobacteriaceae and
non – Enterobacteriaceae and gram-positive bacteria belonging to the genera
Staphylococcus and Enterococcus.

{1}

Page 2 of 7

The BD Phoenix™ GN Panel: The BD Phoenix™ Automated Microbiology System is intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of gram-negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacteriaceae and non-Enterobacteriaceae.

2. Indication(s) for use:
This submission is for the addition of the antibiotic piperacillin at concentrations of 0.5 – 128 µg/mL to the gram negative susceptibility panel.

3. Special condition for use statement(s):
Prescription Use Only
Results for Proteus spp., Providencia spp., and Stenotrophomonas maltophilia have been excluded in the BD Phoenix™ therefore no results will be reported. An alternate method should be performed with these combinations.

4. Special instrument Requirements:
Not Applicable

I. Device Description:

The BD Phoenix™ Automated Microbiology System includes instrumentation and software, sealed and self-inoculating molded polystyrene trays with 136 micro-wells containing dried reagents, and specific inoculum broth formulations for ID and AST Indicator. A growth control well is included. The organism to be tested must be a pure culture and be preliminarily identified as gram positive or gram negative. Colonies are then suspended in broth, and equated to a 0.5 McFarland with the recommendation to use the BD CrystalSpec™ Nephelometer. A further dilution is made into an AST broth, which contains an AST indicator, prior to inoculating the panel. The AST broth is a cation-adjusted formulation of Mueller-Hinton broth containing 0.01% Tween 80. After adding the indicator solution to the AST inoculum the color is blue and after inoculation and incubation goes to pink to colorless as reduction in the panel well proceeds. Inoculated panels are barcode scanned and loaded into the BD Phoenix™ Automated Microbiology System instrument where the panels are continuously incubated at 35°C. The AST has a final inoculum of 5 × 10⁵ CFU/ml. The instrument incubates, reads and records the results of the biochemical substrates and antimicrobial agents and interprets the reactions to give an ID of the isolate and MIC value and category interpretation of the antimicrobial agents. Organisms growing in the presence of a given antimicrobic agent reduce the indicator, signaling organism growth and resistance to the antimicrobic agent. Organisms killed or inhibited by a given antimicrobic agent do not cause reduction of the indicator and therefore do not produce a color change. Additional interpretation is done using software driven "EXPERT" System using rules derived from the NCCLS standards.

{2}

Readings are taken every 20 minutes with an ID result available between 2-12 hours and an AST result available between 4-16 hours. This is only an autoread result; there are no manual readings possible.

# J. Substantial Equivalence Information:

1. Predicate device name(s): VITEK® System
2. Predicate K number(s): N50510
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended use | Intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of gram-negative aerobic and facultative anaerobic bacteria. | same  |
|  Isolates | Isolated colonies from culture used | Isolated colonies from culture used  |
|  Results | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR)  |
|  Incubation conditions | <16 hours | <16 hours  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  Inoculum preparation | Inoculum density equated to 0.5 McFarland standard | Inoculum density equated to 1.0 McFarland standard  |
|  Reading algorithm | Results are determined from serial twofold dilutions of antimicrobial agents | Results are determined from extrapolation of doubling dilutions  |
|  Technology | Automated growth based enhanced by use of a redox indicator (colorimetric oxidation-reduction) to detect organism growth. | Automated growth based with detection using an attenuation of light measured by an optical scanner.  |

{3}

Page 4 of 7

## K. Standard/Guidance Document Referenced (if applicable):

“Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA”; NCCLS M7 (M100-S14)
“Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard.”

## L. Test Principle:

The system employs conventional, colorimetric, fluorogenic and chromogenic substrates to identify the genus and species of the isolate. The AST portion of the BD Phoenix™ Automated Microbiology System is a broth based microdilution method that utilizes a redox indicator (colorimetric oxidation-reduction) to enhance detection of organism growth. The MIC is determined by comparing growth in wells containing serial two-fold dilutions of an antibiotic to the growth in “growth control wells” which contain no antibiotic.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

**a. Precision/Reproducibility:**
Thirteen isolates were evaluated for site to site and inter site reproducibility demonstrating &gt;95% reproducibility. The ten isolate study described in the guidance document was used (10 organisms tested 3 times on 3 days at 3 sites).

**b. Linearity/assay reportable range:**
Not applicable

**c. Traceability, Stability, Expected values (controls, calibrators, or method):**
NCCLS recommended Quality Control strains were tested (see table below). The Phoenix results demonstrate that the system can produce QC results in the recommended range.

{4}

Page 5 of 7

|  Organism | Concentration | Reference results | Phoenix™ results  |
| --- | --- | --- | --- |
|  |   |   |   |
|  E. coli
ATCC 25922
(range 1-4 ug/ml ) | 1 | 1 |   |
|   |  2 | 119 | 54  |
|   |  4 | 88 | 505  |
|   |  8 |  | 4  |
|   |  16 |  | 11  |
|   |  32 |  | 1  |
|  |   |   |   |
|  P aeruginosa
ATCC 27853
range 1-8 ug/mL | ≤ 0.5 |  | 1  |
|   |  1 |  | 6  |
|   |  2 | 46 | 13  |
|   |  4 | 147 | 434  |
|   |  8 | 15 | 105  |
|   |  16 |  | 11  |
|   |  32 |  | 2  |
|   |  16 |  |   |
|   |  |   |   |
|   |  |   |   |

Inoculum density control: The organism suspension density of the ID broth was equivalent to a 0.5 McFarland standard using the BBL™ CrystalSpec™ Nephelometer which was verified each day of testing. Internal data was used to demonstrate that the use of the BBL™ CrystalSpec™ Nephelometer would produce reproducible results. Five different instruments were used.

d. Detection limit:
Not applicable

e. Analytical specificity:
Not applicable

f. Assay cut-off:
Not applicable

2. Comparison studies:

a. Method comparison with predicate device:
The NCCLS recommended broth dilution reference panel was prepared according to the NCCLS recommendation and used to compare with the Phoenix™ results. Clinical testing was performed at four sites. The testing included both fresh clinical isolates and stock isolates along with a challenge set with known results. A

{5}

Page 6 of 7

comparison was provided to the reference method with the following agreement.

|   | total | EA | % EA | Eval Total | Eval EA | Eval EA % | CA total | CA % | #R | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Clinical | 1515 | 1422 | 93.9 | 905 | 819 | 90.5 | 1406 | 92.8 | 598 | 82 | 23 | 4  |
|  Challenge | 266 | 257 | 96.6 | 208 | 199 | 95.7 | 264 | 99.2 | 58 | 2 | 0 | 0  |
|  Combined | 1781 | 1679 | 94.3 | 1113 | 1018 | 91.5 | 1670 | 93.8 | 656 | 84 | 23 | 4  |

EA-Essential Agreement
CA-Category Agreement
Eval- evaluable
R-resistant isolates

maj-major discrepancies
vmj-very major discrepancies
min- minor discrepancies

Essential agreement (EA) is when the BD Phoenix™ panels agree with the reference test panel results exactly or within one doubling dilution of the reference method. Category agreement (CA) is when the BD Phoenix™ panel result interpretation agrees exactly with the reference panel result interpretation.

The test device had a growth rate of &gt;95%.

b. Matrix comparison:
Not applicable

3. Clinical studies:

a. Clinical sensitivity:
Not applicable

b. Clinical specificity:
Not applicable

c. Other clinical supportive data (when a and b are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
Enterobacteriaceae and Non-Enterobacteriaceae excluding Pseudomonas aeruginosa; ≤16 (S), 32-64(I), ≥128 (R)
Pseudomonas aeruginosa ≤ 64 (S), ≥ 128 (R)
The expected value range, interpretative criteria and QC are the same as recommended by the FDA and NCCLS. All values will be included in the package insert.

{6}

Page 7 of 7

N. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K041572](https://fda-staging.innolitics.com/device/K041572)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com