K041499 · Toshibamedical Systems Corporation · IYN · Jun 10, 2004 · Radiology
Device Facts
Record ID
K041499
Device Name
SSA-770A, APLIO VERSION 5.5
Applicant
Toshibamedical Systems Corporation
Product Code
IYN · Radiology
Decision Date
Jun 10, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The APLIO is intended to be used for the following type of studies; fetal, abdominal, intraoperative, pediatric, small organs, neonatal cephalic, adult cephalic, cardiac, transreseophageal, peripheral vascular, musculo-skeletal (both conventional and superficial) and laparoscopic.
Device Story
APLIO (SSA-770A) is a mobile diagnostic ultrasound system; utilizes wide array of probes (2-12 MHz) including linear, convex, and sector arrays. System captures acoustic signals; processes data to generate B-mode, M-mode, Pulsed Wave Doppler, Color Doppler, Amplitude Doppler, Color Velocity Imaging, Harmonic Imaging, and Tissue Doppler Imaging (TDI-Q). Used in clinical settings (clinics, hospitals) by trained healthcare professionals. Output displayed on monitor for real-time visualization; assists clinicians in diagnostic assessment of anatomy and blood flow. Benefits include non-invasive diagnostic imaging across diverse clinical applications.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing and adherence to recognized standards, including IEC 60601, IEC 60601-2-37, and AIUM-NEMA UD2/UD3 output measurement and display standards.
Technological Characteristics
Mobile ultrasound system; frequency range 2-12 MHz; probes include flat linear, convex, and sector arrays. Complies with IEC 60601, IEC 60601-2-37, and AIUM-NEMA UD2/UD3 standards. Supports multiple imaging modes (B, M, Doppler, Harmonic, TDI).
Indications for Use
Indicated for fetal, abdominal, intraoperative, pediatric, small organ (thyroid, parathyroid, breast, scrotum, penis), neonatal/adult cephalic, cardiac, transesophageal, transrectal, transvaginal, peripheral vascular, laparoscopic, and musculoskeletal (conventional/superficial) imaging.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Toshiba SSA-770A, Aplio Version 4.0 Diagnostic Ultrasound (k032281)
Siemens Medical Solutions Sequoia 8.0 Diagnostic Ultrasound System (k032281)
Reference Devices
Toshiba SSA-700A (k022400)
Previous 510(k) for this device (k013633)
Submission Summary (Full Text)
{0}------------------------------------------------
K041499
### (JUN 1 0 2004
#### 510(k) Summary of Safety and Effectiveness: 21 CFR 807.92
| Submitter's Name: | Toshiba America Medical Systems, Inc. |
|--------------------------|--------------------------------------------------------------------------------------------|
| Address: | PO Box 2068,2441 Michelle Drive Tustin, CA 92781-2068 |
| Contact: | Paul Biggins, Sr. Manager of Regulatory Affairs |
| Telephone No.: | (714) 730-5000 |
| Device Proprietary Name: | SSA-770A, APLIO Version 5.5 |
| Common Name: | Diagnostic Ultrasound System |
| Classification: | |
| Regulatory Class: | II |
| Review Category: | Tier II |
| | Ultrasonic Pulsed Doppler Imaging System -- Product Code: 90-IYN<br>[Fed.Reg.No.:892.1550] |
| | Ultrasonic Pulsed Echo Imaging System - Product Code: 90-IYO<br>[Fed.Reg.No.:892.1560] |
| | Diagnostic Ultrasonic Transducer – Product Code: 90-ITX<br>[Fed.Reg.No.:892.1570] |
#### Identification of Predicate Devices:
Toshiba America Medical Systems believes that this device is substantially equivalent to:
- 1) Toshiba SSA-770A, Aplio Version 4.0 Diagnostic Ultrasound; 510(k) control number k032281
- 2) Siemens Medical Solutions Sequoia 8.0 Diagnostic Ultrasound System; 510(k) control number k032281
#### Device Description:
The APLIO Ultrasound System is a mobile system. This system is a Track 3 device that employs a wide array of probes that include flat linear array, convex linear array, and sector array with a frequency range of approximately 2 MHz to 12 MHz.
#### Intended Use:
The APLIO is intended to be used for the following type of studies; fetal, abdominal, intraoperative, pediatric, small organs, neonatal cephalic, adult cephalic, cardiac, transrectal, transvaginal, transesophageal, peripheral vascular, musculo-skeletal (both conventional and superficial) and laparoscopic.
#### Safety Considerations:
This device is designed and manufactured in conjunction with the Quality System Regulation, IEC 60601 (applicable portions), IEC60601-2-37 (applicable portions), and the AIUM-NEMA UD2 Output Measurement Standard as applied to Track 3 Ultrasound systems and the AIUM-NEMA UD3 Output Display Standard.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, facing to the right. The eagle is composed of three curved lines, giving it a modern and abstract appearance.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### JUN 1 0 2004
Toshiba America Medical Systems, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K041499
K041477
Trade Name: APLIO Diagnostic Ultrasound System, Model SSA-770A Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulatory Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN, IYO, and ITX Dated: June 4, 2004 Received: June 7, 2004
#### Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Section 310(K) prematerially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosale) to regars atment date of the Medical Device Amendments, or to commerce prior to May 20, 1970, the enability with the provisions of the Federal Food, Drug, and devices that nave been receasined in are reader we device, subject to the general controls Cosment Act (Act). Tou may, mercrors, in the Act include requirements for annual provisions of the Act. "The general vehiclering practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use I ins determination of savstanial oqur allers upposes in the many of the many of the many of notification:
#### Transducer Model Number
| PST-25AT | PLT-1204AX | PST-37CT |
|-----------|------------|-----------|
| PVT-375AT | PC-20M | PST-30BT |
| PVT-661VT | PET-510MB | PLT-704AT |
| PLT-805AT | PLT-1202S | PLT-1204A |
| PST-20CT | PET-704LA | PVT-375AX |
{2}------------------------------------------------
| PST-65AT | PLT-308P | PVT-375BT |
|-----------|-----------|-----------|
| PLT-604AT | PET-508MA | PVT-382BT |
| PST-50AT | PVT-770RT | |
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (Scc above) into elabor regulations affecting your device can be may be subject to such additions, Title 2, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register
furnier announcenting concerning your as not the the the the mination does not mean that Flease be advised mat 1 271 's issualled of a eveemonomies with other requirements of the Act or any FDA has made a decemination administered by other Federal agencies. You must comply with all r cocration and regulations administed to: registration and listing (21 CFR Part 807); the Act s requirements, merading, but notacturing practice requirements as set forth in the quality labeling (21 CFR Part 007), good manazive; and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device The I DA midnig of substantial equivice and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please If you desire specific ad roo for your and additionally, for questions on the promotion comact the Office of Compines at (301) 594-4639 Compliance at (301) 594-4639. Also, and advertising of your deviles, prease vehicles by reference to premarket notification" (21 CFR picase note in regulation on the responsibilities under the Act may be obtained from I at 801:27). Other general Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
David An. Lynn
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{3}------------------------------------------------
System _X__ Transducer____ SSA-770A______________________________________________________________________________________________________________________________________________________________________ Model_ 510(k) Number(s)
| | | | | | | | | Mode of Operation | | | | |
|----------------------------------|---|---|-------------|-------------|------------------|----------------------|------------------------------|-----------------------|---------------------|-----------------|------------|-----------|
| Clinical Application | B | M | P<br>W<br>D | C<br>W<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging | 1.5<br>Harmonic | 1.5<br>RSI | TDI-<br>Q |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | P | P | P | P | P | P | P | P | P | | | |
| Abdominal | P | P | P | P | P | P | P | P | P | | | |
| Intraoperative (Specify)** | P | P | P | | P | P | P | P | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | | |
| Pediatric | P | P | P | P | P | P | P | P | P | | | |
| Small Organ (Specify)*** | P | P | P | | P | P | P | P | P | | | |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | | | |
| Adult Cephalic | P | P | P | P | P | P | P | P | P | | | |
| Cardiac | P | P | P | P | P | P | P | P | P | E' | P | P |
| Transesophageal | P | P | P | P | P | P | P | P | | E' | | P |
| Transrectal | P | P | P | | P | P | P | P | P | | | |
| Transvaginal | P | P | P | | P | P | P | P | P | | | |
| Transurethral | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | | | |
| Laparoscopic | P | P | P | | P | P | P | P | | | | |
| Musculo-skeletal | P | P | P | | P | P | P | P | P | | | |
| Superficial | | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | P | P | P | | P | P | P | P | P | | | |
entralia.
N= new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Combined Modes: B/M; B/PWD; Additional Comments: Additional Collanceral Conancils: BDF/MDF/PWD;B-TDI; M-TDI; 2D/CWD; CHI/2D; FEI/2D; CHI/BDF; FEI/BDF
| E'-- Added via LTF against SSA-700A 510(k) control number K022400 |
|-------------------------------------------------------------------|
| Previous 510(k) for this device k013633 |
| ** Abdominal |
| *** For example: thyroid, parathyroid, breast, scrotum and penis |
(P).EASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON OTHER PAGES IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Jamilah Begum
(Division Sign Division of Re and Radio 510(k) Numbe
{4}------------------------------------------------
Transducer_X__ System __ PST-25AT_ Model_ 510(k) Number(s)
| Clinical Application | B | M | P<br>W<br>D | C<br>W<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging | 1.5<br>Harmonic | 1.5<br>RSI | TDI-Q |
|----------------------------------|---|---|-------------|-------------|------------------|----------------------|------------------------------|-----------------------|---------------------|-----------------|------------|-------|
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Intraoperative (Specify) | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Neurological | | | | | | | | | | | | |
| Pediatric | P | P | P | P | P | P | P | P | P | | | |
| Small Organ (Specify) | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | P | E¹ | | P |
| Cardiac | P | P | P | P | P | P | P | P | P | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | | |
N= new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Combined Modes: B/M; B/PWD; Additional Comments: Additional Continution ContinuedPWD;B-TDI; M-TDI; 2D/CWD; BDF/CWD; CHI/2D; FEI/2D; CHI/BDF; FEVBDF
E' - Added via LTF against SSA-700A 510(k) control number K022400 Previous 510(k) for this device k013633
> (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON OTHER PAGES IF NEVEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Ch. Sayre
(Division Sign-Off) Division of Reproductive, Abdo and Radiological Devic 510(k) Number
{5}------------------------------------------------
Transducer X System __ PVT-375AT Model_ 510(k) Number(s)
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------|-------------------|---|-------------|-------------|------------------|----------------------|------------------------------|-----------------------|---------------------|-----------------|-------------|-----------|
| Clinical Application | B | M | P<br>W<br>D | C<br>W<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging | 1.5<br>Harmonic | ાં ર<br>RSI | TDI-<br>Q |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | P | P | P | | P | P | P | P | P | | | |
| Abdominal | P | P | P | | P | P | P | P | P | | | |
| Intraoperative (Specify) | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Pediatric | P | P | P | | P | P | P | P | P | | | |
| Small Organ (Specify) | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | | |
N= new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Combined Modes: B/M; B/PWD; Additional Comments: Additional Comments: BDF/MDF/PWD;B-TDI; M-TDI; 2D/CWD; BDF/CWD; CHI/2D; FEI/2D; CHI/BDF; FEI/BDF;
Previous 510(k) for this device k013633
(PI.E.A.SE DO NOT WRITE BELOW THIS LINE · CONTINUE ON OTHER PAGES IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David h. Seryson
(Division Sign-Off) Division of Reproductive, Abdon and Radiological Devices 510(k) Number _
{6}------------------------------------------------
Transducer X System _______________________________________________________________________________________________________________________________________________________________________ PVT-661 VT Model_ 510(k) Number(s)
| | Mode of Operation | | | | | | | | | | | |
|--------------------------------|-------------------|---|-------------|-------------|------------------|----------------------|------------------------------|-----------------------|---------------------|-----------------|------------|-------|
| Clinical Application | B | M | P<br>W<br>D | C<br>W<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging | 1.5<br>Harmonic | 1.5<br>RSI | TDI-Q |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Intraoperative (Specify) | | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ (Specify) | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | P | P | P | | P | P | P | P | P | | | |
| Transvaginal | P | P | P | | P | P | P | P | P | | | |
| Transurethral | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | | | |
| Superficial | | | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | | | |
| Conventional | | | | | | | | | | | | |
N= new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Combined Modes: B/M; B/PWD; Additional Comments: Auditional Comments:
BDF/PWD; BDF/MDF; BDF/MDF/PWD;B-TDI; M-TDI; 2D/CWD; BDF/CWD; CHI/2D; FEI/2D; CHI/BDF; FEI/BDF
Previous 510(k) for this device k013633
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON OTHER PAGES IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David R. Ingram
(Division Sign-Off) Division of Reproductive, Abdo and Radiological Device 510(k) Number _
{7}------------------------------------------------
Transducer _X System _ PLT-805AT Model_ 510(k) Number(s)
| | | | | | | | | Mode of Operation | | | | | |
|-------------------------------|---|---|-------|-------|---------------|-------------------|------------------------|--------------------|------------------|--------------|---------|--------|--|
| Clinical Application | B | M | P W D | C W D | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (Specify) | Harmonic Imaging | 1.5 Harmonic | 1.5 RSI | TDI- Q | |
| Ophthalmic | | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | | |
| Intraoperative (Specify) | | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | | |
| Neurological | | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | | |
| Small Organ (Specify) | P | P | P | P | P | P | P | P | P | | | | |
| Neonatal Cephalic | | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | | | | |
| Laparoscopic | | | | | | | | | | | | | |
| Musculo-skeletal | P | P | P | | P | P | P | P | P | | | | |
| Superficial | | | | | | | | | | | | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | | | | |
N= new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Combined Modes: B/M; B/PWD; Additional Comments: Additional Commonts: BDF/MDF/PWD;B-TDI; M-TDI; 2D/CWD; BDF/CWD; CHI/2D; FEI/2D; CHI/BDF; FEI/BDF
Previous 510(k) for this device k013633
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON OTHER PAGES IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Ch. Johnson
(Division Sign-Off) Division of Reproductive, Abdo and Radiological Devices : 10(k) Number _______________________________________________________________________________________________________________________________________________________________
{8}------------------------------------------------
Transducer _X System _ PST-20CT______________________________________________________________________________________________________________________________________________________________________ Model_ 510(k) Number(s)
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------|-------------------|---|-------------|-------------|------------------|----------------------|------------------------------|-----------------------|---------------------|-----------------|------------|-----------|
| Clinical Application | B | M | P<br>W<br>D | C<br>W<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging | 1.5<br>Harmonic | 1.5<br>RSI | TDI-<br>Q |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | P | P | P | P | P | P | P | P | P | | | |
| Abdominal | | | | | | | | | | | | |
| Intraoperative (Specify) | | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ (Specify) | | | | | | | | | | | | |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac | P | P | P | P | P | P | P | P | P | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | | |
N= new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Combined Modes: B/M; B/PWD; Additional Comments: BDF/PWD; BDF/MDF; BDF/MDF/PWD;B-TDI; M-TDI; 2D/CWD; BDF/CWD; CHI/2D; FEV2D; CHI/BDF; FEVBDF
Previous 510(k) for this device k013633
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON OTHER PAGES IF NEEDEDI Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel k. Leyman
(Division Sign-Off) Division of Reproductive, and Radiological Device 510(k) Number
{9}------------------------------------------------
Transducer_X System __ PLT-1204AX____________________________________________________________________________________________________________________________________________________________________ Model_ 510(k) Number(s)
| Mode of Operation | | | | | | | | | | | | |
|----------------------------------|---|---|-------------|-------------|------------------|----------------------|------------------------------|-----------------------|---------------------|-----------------|------------|-------|
| Clinical Application | B | M | P<br>W<br>D | C<br>W<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging | 1.5<br>Harmonic | 1.5<br>RSI | TDI-Q |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Intraoperative (Specify) | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Neurological | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ (Specify) | P | P | P | | P | P | P | P | P | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | | | |
| Laparoscopic | | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | P | P | P | | P | P | P | P | P | | | |
| Musculo-skeletal<br>Conventional | P | P | P | | P | P | P | P | P | | | |
N= new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Combined Modes: B/M; B/PWD; Additional Comments: Additional Continents: BDF/MDF/PWD;B-TDI; M-TDI; 2D/CWD; BDF/CWD; CHI/2D; FEI/2D; CHI/BDF; FEI/BDF
Previous 510(k) for this device k013633
(PLEASE DO NOT WRITE BELOW THIS LINE / CONTINUE ON OTHER PAGES IF NEEDED) Concurrence of CDRII, Office of Device Eyaluation (ODE)
David A. Syverson
(Division Sign-Off) (Division of Reproductive, Abdo and Radiological Devices 51()(k) Number _______________________________________________________________________________________________________________________________________________________________
{10}------------------------------------------------
Transducer X System __ PC-20M Model_ 510(k) Number(s)
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------|-------------------|---|-------------|-------------|------------------|----------------------|------------------------------|-----------------------|---------------------|-----------------|------------|-----------|
| Clinical Application | B | M | P<br>W<br>D | C<br>W<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging | 1.5<br>Harmonic | 1.5<br>RSI | TDI-<br>Q |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Intraoperative (Specify) | | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | | |
| Pediatric | | | | P | | | | | | | | |
| Small Organ (Specify) | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac | | | | P | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Musculo-skelctal | | | | | | | | | | | | |
| Superficial | | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | | |
N= new indication; P = Previously Cleared by i DA; E = Added under Appendix E (LTF)
Combined Modes: B/M; B/PWD; Additional Comments: Additional Comments: BDF/MDF/PWD;B-TDI; M-TDI; 2D/CWD; BDF/CWD; CHI/2D; FEI/2D; CHI/BDF; FEVBDF
Previous 510(k) for this device k013633
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON OTHER PAGES IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Lyman
(Division Sign-Off) Division of Reproductive, Abdom and Radiological Devices 510(k) Number __
{11}------------------------------------------------
Transducer X System __ PET-510MB Model_ 510(k) Number(s)
| Mode of Operation | | | | | | | | | | | | |
|----------------------------------|---|---|-------------|-------------|------------------|----------------------|------------------------------|-----------------------|---------------------|-----------------|------------|-----------|
| Clinical Application | B | M | P<br>W<br>D | C<br>W<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging | 1.5<br>Harmonic | 1.5<br>RSI | TDI-<br>Q |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Intraoperative (Specify) | | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ (Specify) | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac | | | | | | | | | | E | | P |
| Transesophageal | P | P | P | P | P | P | P | P | P | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | | | |
| Superficial | | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | | |
entonal
N= new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Combined Modes: B/M; B/PWD; Additional Comments: Auditional Continents: - BDF/MDF/PWD:B-TDI; M-TDI; 2D/CWD; BDF/CWD; CHI/2D; FEI/2D; CHI/BDF; FEVBDF
Previous 510(k) for this device k013633
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON OTHER PAGES IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David M. Segeen
(Division Sign-Off) Division of Reproductive, Abdom and Radiological Devices 510(k) Number ___
{12}------------------------------------------------
Transducer_X System __ PLT-1202S Model_ 510(k) Number(s)
| Mode of Operation | | | | | | | | | | | | |
|-------------------------------|---|---|-------------|-------------|------------------|----------------------|------------------------------|-----------------------|---------------------|-----------------|------------|-------|
| Clinical Application | B | M | P<br>W<br>D | C<br>W<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging | 1.5<br>Harmonic | 1.5<br>RSI | TDI-Q |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Intraoperative (Specify) | P | P | P | | P | P | P | P | P | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ (Specify) | P | P | P | | P | P | P | P | P | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | | | |
| Laparoscopic | | | | | | | | | | | | |
| Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | | | |
N= new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Combined Modes: B/M; B/PWD; Additional Comments: Additional Comments:
BDF/PWD; BDF/MDF; BDF/MDF/PWD;B-TDI; M-TDI; 2D/CWD; BDF/CWD; CHV2D; FEI/2D; CHI/BDF; FEI/BDF
Previous 510(k) for this device k013633
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON OTHER PAGES IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Egerom
(Division Sign-Off) Division of Reproductive, Ab and Radiological Devices 510(k) Number --
{13}------------------------------------------------
Transducer _X System _ PET-704LA_ Model_
510(k) Number(s)
| Clinical Application | B | M | P W D | C W D | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (Specify) | Harmonic Imaging | 1.5 Harmonic | 1.5 RSI | TDI-Q |
|-------------------------------|---|---|-------|-------|---------------|-------------------|------------------------|--------------------|------------------|--------------|---------|-------|
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Intraoperative (Specify) | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Neurological | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ (Specify) | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Laparoscopic | P | P | P | | P | | P | P | | | | |
| Musculo-skeletal | | | | | | | | | | | | |
| Superficial | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
N= new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Combined Modes: B/M; B/PWD; Additional Comments: Additional Continuer; BDF/MDF/PWD;B-TDI; M-TDI; 2D/CWD; BDF/CWD; CHI/2D; FEI/2D; CHI/BDF; FEI/BDF
Previous 510(k) for this device k013633
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON OTHER PAGES IF NEFDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Legerm
(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number _
{14}------------------------------------------------
Transducer _X System _______________________________________________________________________________________________________________________________________________________________________ PST-37CT_ Model_ 510(k) Number(s)
| | Mode of Operation | | | | | | | | | | | |
|--------------------------------|-------------------|---|-------------|-------------|------------------|----------------------|------------------------------|-----------------------|---------------------|------------|------------|-------|
| Clinical Application | B | M | P<br>W<br>D | C<br>W<br>D | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging | 1.5<br>THI | 1.5<br>RSI | TDI-Q |
| Ophthalmic | | | | | | | | | E | | | |
| Fetal | E | E | E | E | E | E | E | E | E | | | |
| Abdominal | E | E | E | E | E | E | E | E | E | | | |
| Intraoperative (Specify) | | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | | |
| Pediatric | E | E | E | E | E | E | E | E | E | | | |
| Small Organ (Specify) | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | | | |
| Superficial | | | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.