BUTLER NUCARE PROPHYLAXIS PASTE WITH NOVAMIN

K041371 · Novamin Technology, Inc. · EJR · Aug 23, 2004 · Dental

Device Facts

Record IDK041371
Device NameBUTLER NUCARE PROPHYLAXIS PASTE WITH NOVAMIN
ApplicantNovamin Technology, Inc.
Product CodeEJR · Dental
Decision DateAug 23, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6030
Device ClassClass 1
AttributesTherapeutic

Indications for Use

Butler NuCare® Prophylaxis Paste with NovaMin® is intended for cleaning and polishing procedures as a part of a professionally administered dental prophylaxis treatment. Secondarily, Butler NuCare™ Prophylaxis Paste with NovaMin® can be used for the immediate relief of tooth sensitivity, post-scaling and root planing.

Device Story

Butler NuCare™ Prophylaxis Paste with NovaMin® is a dental prophylaxis paste used by dental professionals during cleaning and polishing procedures. The paste contains NovaMin® (calcium sodium phosphosilicate), which reacts upon exposure to an aqueous environment (saliva/water) to physically occlude exposed dentinal tubules. This occlusion blocks hydrodynamic flow, providing immediate relief from tooth sensitivity following scaling and root planing. The material eventually reacts to form hydroxycarbonate apatite (HCA), which is structurally similar to natural tooth mineral. The device is applied topically by a dental professional in a clinical setting.

Clinical Evidence

Bench testing only. Tubule occlusion efficacy was evaluated using a dentin disk model, demonstrating significant tubule occlusion compared to controls. Biocompatibility testing confirmed no hazardous effects.

Technological Characteristics

Active ingredient: NovaMin® (calcium sodium phosphosilicate). Composition includes Ca, Na, Si, P, and O. Mechanism: surface reaction in aqueous environment to form hydroxycarbonate apatite (HCA) for physical tubule occlusion. Form factor: prophylaxis paste.

Indications for Use

Indicated for dental patients requiring professional prophylaxis cleaning and polishing, and for patients experiencing tooth sensitivity following scaling and root planing procedures.

Regulatory Classification

Identification

An oral cavity abrasive polishing agent is a device in paste or powder form that contains an abrasive material, such as silica pumice, intended to remove debris from the teeth. The abrasive polish is applied to the teeth by a handpiece attachment (prophylaxis cup).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) Premarket Notification NovaMin Technology, Inc. Butler NuCare™ Prophylaxis Paste with NovaMin® K041371 CONFIDENTIAL ## AUG 2 3 701 #### 1. SUBMITTER INFORMATION: | Name: | NovaMin Technology, Inc. | |------------|----------------------------| | Address: | 13709 Progress Blvd., #23 | | | Alachua, Florida 32615 USA | | Phone: | (386) 418-1551 | | Facsimile: | (386) 418-1465 | | Contact: | David C. Greenspan, Ph.D. | May 19, 2004 Preparation Date: #### 2. DEVICE NOMENCLATURE: | Trade Name: | Butler NuCare™ Prophylaxis Paste with NovaMin® | |----------------------|------------------------------------------------| | Common Name: | Dental Prophylaxis Paste | | Classification Name: | Prophylaxis Paste | #### 3. LEGALLY MARKETED PREDICATE DEVICE: | Device Name: | Butler GUM® Prophylaxis Paste with NovaMin® | |----------------|---------------------------------------------| | 510(k) Number: | K024343 | | Applicant: | NovaMin Technology, Inc. | #### 4. DEVICE DESCRIPTION: Butler NuCare™ Prophylaxis Paste with NovaMin® is a product that is intended for cleaning and polishing procedures as a part of a professionally administered dental prophylaxis treatment. Butler NuCare™ Prophylaxis Paste with NovaMin® is also designed to physically occlude dentinal tubules for the immediate relief of tooth sensitivity, post-scaling and root planing. NovaMin® (calcium sodium phosphosilicate) is composed of elements that occur naturally in the body (Ca, Na, Si, P, and O . When exposed to an aqueous environment, NovaMin® undergoes a rapid surface reaction, allowing it to physically occlude tubules. Within a short period of time, essentially all of the NovaMin® reacts to form hydroxycarbonate apatite (HCA), which is chemically and structurally similar to natural tooth mineral. #### 5. INTENDED USE: Butler NuCare® Prophylaxis Paste with NovaMin® is intended for cleaning and polishing procedures as a part of a professionally administered dental prophylaxis treatment. Secondarily, Butler NuCare™ Prophylaxis Paste with NovaMin® can be used for the immediate relief of tooth sensitivity, post-scaling and root planing. {1}------------------------------------------------ ### 6. TECHNOLOGICAL CHARACTERISTICS: The technological characteristics of Butler NuCare™ Prophylaxis Paste with NovaMin® and Butler The technological characteries of Dain® are the same. Both devices are designed to polish and Clean the tooth surface as part of a professionally administered dental prophylaxis treatment. Both cican the tooth surface as part of a pross hypersensitivity associated with exposed dentin by the devices are also tesigned to renove resources poth surface. Both devices use NovaMin" to deposition of a calcium phosphate layer that occludes and blocks hydrodynamic flow. produce a calcially phosphate in the ingredients used to make the base of the products, and .in the material used as the abrasive agent. ### 7. SAFETY AND PERFORMANCE DATA: Many different biocompatibility tests have been performed on the NovaMin® particulate, the active ingredient in Butler NuCare™ Prophylaxis Paste with NovaMin®. The results of these tests indicate there is no evidence of any hazardous effects to the patient if the product is used as directed. The tubule occlusion efficacy of Butler NuCare™ Prophylaxis Paste with NovaMin® was evaluated rice table overalian block model. The results indicate Butler NuCare™ Prophylaxis Paste with NovaMin® occludes a significant number of tubules when compared with controls. #### 8. CONCLUSIONS: Butler NuCare™ Prophylaxis Paste with NovaMin® is considered to be substantially couvvalent to the Dutter Nacked predicate device, Butler GUM® Prophylaxis Paste with NovaMin® (K024343). The legariy marketed prodicate and biocompatibility data demonstrate the safety and efficacy of Butler NuCarc™ Prophylaxis Paste with NovaMin® for the intended uses. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines forming its body and wings. The symbol is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 2 3 2004 Dr. David C. Greenspan Vice President & Chief Technology Officer NovaMin Technology, Incorporated 13709 Progress Boulevard #23 Alachua, Florida 32615 Re: K041371 Ko+1371 Trade/Device Name: Butler NuCare™ Prophylaxis Paste with NovaMin** Regulation Number: 872.6030 Regulation Name: Oral Cavity Abrasive Polishing Agent Regulatory Class: 1 Product Code: EJR Dated: May 21, 2004 Received: May 25, 2004 Dear Dr. Greenspan: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your sectetermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device interstate conninetee prior to may 20, 1970, consided in accordance with the provisions of Amendinents, or to do roob and Cosmetic Act (Act) that do not require approval of a premarket the rederal Pood, Drag, and Ooou may, therefore, market the device, subject to the general approvin upprovisions of the Act. The general controls provisions of the Act include controls provisions of the ristration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting (FMA). it may be subject to meet adam federal Regulations, Title 21, Parts 800 to 898. In your device can be found in ther announcements concerning your device in the Eederal Register. {3}------------------------------------------------ Page 2 - Dr. Greenspan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. or the Act of any vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and instillents as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if requirence as between as product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a premaired predicated. - > > device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), n you donto the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilitics under the Act from the may obtain of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Qels Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ # OF INDICATIONS FOR U , 413711 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: _________________________________________________________________________________________________________________________________________________________________ INDICATIONS FOR USE: Butler NuCare™ Prophylaxis Paste with NovaMin® is intended for cleaning and polishing. Secondation of the comment of the red dental monbulaxis treatment. Secondary Butler NuCare™ Prophylaxis Pasce with Novahin Tismatores was is treatment. Secondarily: procedures as part of a professionally administered dental for a immediate relief of procedures as part of a professionally administered demails teamers. In the inimediate relief of Buttor into post-scaling and root planing. (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) (Division Sign-On) Division of Anesthesiology, General Hospital, Division Control, Dental Devices 510(k) Number k(1137) Prescription Use __X __ OR (Per 21 CFR 801.109) Over-The-Counter Use ___ SECTION D – STATEMENT OF INDICATIONS FOR USE
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...