K041305 · Stryker Endoscopy · MAI · Jun 14, 2004 · Orthopedic
Device Facts
Record ID
K041305
Device Name
STRYKER BIOZIP SUTURE ANCHOR SYSTEM
Applicant
Stryker Endoscopy
Product Code
MAI · Orthopedic
Decision Date
Jun 14, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Styker BioZip Suture Anchor System is intended foruse in maloutions for tissue to bone in such procedures as: Shoulder: Rotator cuff repair Bankart repair SLAP lesion repair Acromio-clavicular sepaeration repair Capsular shift/capsulolabral reconstruction Biceps tenodesis Deltoid repair Knee: Extra capsular repairs Medial collateral ligament Lateral collateral ligament Posterior oblique ligament Illiotibial band tenosis Patellar tendon repair Elbow, Wrist, Hand: Scapholunate ligament reconstruction Ulnar collateral ligament reconstruction radial Collateral ligament reconstruction Biceps tendon reattachment Foot & Ankle Medial instability repair/reconstruction Lateral instability repair/reconstruction Achilles tendon repair/reconstruction Midfoot reconstruction Hallux valgus reconstruction Pelvis: Bladder neck suspension procedures The Stryker BioZip Suture Anchor System is intended for single-use only.
Device Story
Stryker BioZip Suture Anchor System is a line extension of a previously cleared device; used for securing soft tissue to bone. Device consists of a Poly L-lactic acid (PLLA) screw-in anchor pre-threaded with non-absorbable braided polyethylene surgical sutures; pre-assembled on a disposable inserter. Used by surgeons in clinical/OR settings for orthopedic soft tissue repair. Device provides mechanical fixation of tissue to bone; benefits patient by facilitating surgical repair of ligaments and tendons. Provided sterile for single-use.
Clinical Evidence
Bench testing only. Device validated for sterility (SAL 10^-6) per ANSI/AAMI/ISO 11135 and packaging integrity per ASTM 4169. Biocompatibility confirmed per ISO-10993-1 and G95-1.
Indicated for soft tissue to bone fixation in orthopedic procedures including shoulder (rotator cuff, Bankart, SLAP, AC repair, capsular reconstruction, biceps tenodesis, deltoid repair), knee (extra capsular, MCL, LCL, POL, IT band, patellar tendon), elbow/wrist/hand (ligament reconstruction, biceps reattachment), foot/ankle (instability, Achilles, midfoot, hallux valgus), and pelvis (bladder neck suspension). Single-use only.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains a sequence of alphanumeric characters. The characters are 'K', 'O', '4', '1', '3', '0', and '5'. The characters are written in a simple, handwritten style.
Image /page/0/Picture/1 description: The image shows the word "stryke" in a bold, sans-serif font. Above the word "stryke" is a symbol that looks like a number 77. The word "stryke" is in all lowercase letters. The image is in black and white.
'JUN 1 4 2004
Endoscopy
## SUMMARY OF SAFETY AND EFFECTIVENESS
### Device Name
Classification Name:
Common and Usual Name: Proprietary Name:
Fastener, Fixation, Biodegradable, Soft Tissue 21 CFR §888.3040, Class II Bioabsorbable Suture Anchor Stryker BioZip Suture Anchor System
### Predicate Device
Stryker BioZip Suture Anchor System (#K023192), currently marketed by Stryker Endoscopy (San Jose, CA).
### Summary
This summary of Special 510(k) safety and effectiveness is being submitted in accordance with requirements of SMDA 1990.
The line extension of the Stryker BioZip Suture Anchor System is intended for use in providing a means for securing soft tissue to bone using suture. The line extension of the Stryker BioZip Suture Anchor System consists of a Poly L-lactic acid (PLLA) screw-in type anchor pre-threaded with non-absorbable braided polyethylene surgical sutures, and pre-assembled on a disposable inserter.
The line extension of the Stryker BioZip Suture Anchor System will be provided sterile for single-use (ASTM 4169). The device will be sterilized by Ethylene Oxide (ANSVAAMI/ISO 11135), including limits for Ethylene Oxide residuals and validated to a sterility assurance level (SAL) of 10 °. The device is biocompatible per ISO-10993-1 and G95-1. The line extension of the Stryker BioZip Suture Anchor System is substantially equivalent in material of construction, overall design, intended use, and safety and efficacy to the predicate device. The subject device was shown to have substantially equivalent performance when compared to the predicate device.
The line extension of the Stryker BioZip Suture Anchor System is considered substantially equivalent to the Stryker BioZip Suture Anchor System (#K023192).
Contact:
Date: May 7, 2004
Melissa Murphy Regulatory Representative Stryker Endoscopy 5900 Optical Court San Jose, CA 95138 (408) 754-2148
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three parallel lines that curve and taper, resembling a bird in flight or a flowing ribbon.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 4 2004
Ms. Melissa Murphy Regulatory Representative Stryker Endoscopy 5900 Optical Court San Jose, California 95138
Re: K041305
Trade/Device Name: Stryker BioZip Suture Anchor System Regulation Number: 21 CFR 888.3040 Regulation Name: Fastener, fixation, biodegradable, soft tissue Regulatory Class: II Product Code: MAI Dated: May 7, 2004 Received: May 17, 2004
Dear Ms. Murphy:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosure) to regary to regars and the Medical Device Amendments, or to conninered prof to May 20, 1978, the eccordance with the provisions of the Federal Food, Drug, devices that have been roomstiled mequire approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The 1 ou may, merelors, mainer of the Act include requirements for annual registration, listing of general bonavil provision practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can may be subject to sach additions, Title 21, Parts 800 to 898. In addition, FDA may ou found in the Outsting your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I Teast of advised that I Dris and that your device complies with other requirements of the Act mal I DA has made a and regulations administered by other Federal agencies. You must of any I cacal statutes and regalations and limited to: registration and listing (21 Compry with an the Act 81 e FR Part 801); good manufacturing practice requirements as set CTN in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic forth in the quality by stems (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Ms. Melissa Murphy
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to oogin manteeing your antial equivalence of your device to a legally premarket nothlication. The PDA midning of backland of .
marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your as 110 110 110 - 4659. please note the regulation entitled, province Colliati the Office of Compulance an (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general information on your responsible in the toll-free number (800) 638-2041 or Manufacturers, International and Consulter Pressonal Provincial Stations of Smamain. html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known): K041305
Device Name: Stryker BioZip Suture Anchor System
Indications For Use: The Styker BioZip Suture Anchor System is intended foruse in maloutions for tissue to bone in such procedures as:
| Shoulder: | Elbow, Wrist, Hand: |
|---------------------------------------------|-------------------------------------------------|
| Rotator cuff repair | Scapholunate ligament reconstruction |
| Bankart repair | Ulnar collateral ligament reconstruction radial |
| SLAP lesion repair | Collateral ligament reconstruction |
| Acromio-clavicular sepaeration repair | Biceps tendon reattachment |
| Capsular shift/capsulolabral reconstruction | |
| Biceps tenodesis | Foot & Ankle |
| Deltoid repair | Medial instability repair/reconstruction |
| | Lateral instability repair/reconstruction |
| Knee: | Achilles tendon repair/reconstruction |
| Extra capsular repairs | Midfoot reconstruction |
| Medial collateral ligament | Hallux valgus reconstruction |
| Lateral collateral ligament | |
| Posterior oblique ligament | Pelvis: |
| | Bladder neck suspension procedures |
| Illiotibial band tenosis | |
| Patellar tendon repair | |
The Stryker BioZip Suture Anchor System is intended for single-use only.
ర్లు Prescription Use __ (Part 21 CFR 801 Subpart D)
,
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Muriam C. Provost
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** K041305
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.